oxycure MD300 C63 User manual

D.SPEC.05.10 Pulse oxymètre MD300-C63 - Mode d'emploi_EN
v. 28/06/2019 1
Fingertip Puls Oximeter
MD300 C63
User manual
Start-up: summarised instructions
1. Before use, disinfect the device and the finger used for measured.
2. Insert your finger into the device until the end and nail towards the screen!
3. Press the start / stop button
4. Remove the finger. The device turns off automatically.
Periodic maintenance.
1. Regular disinfection.

D.SPEC.05.10 Pulse oxymètre MD300-C63 - Mode d'emploi_EN
v. 28/06/2019 2
Thank you for choosing an Oxycure concentrator.
The Oxycure range, from left to right
Oxycure portable concentrators
Oxycure portable concentrator – Inogen G4
Oxycure portable concentrator – Inogen G3
Oxycure portable concentrator – Zen-O lite
Oxycure portable concentrator – Simply Go Mini
Oxycure portable concentrator – SimplyGo
Oxycure portable concentrator – eQuinox
Oxycure concentrators
Oxycure – Kröber concentrator
Oxycure cylinders (gaseous oxygen)
Oxycure B2 cylinder – 0.4 m³
Oxycure B5 cylinder – 1.0 m³
Oxycure tanks (liquid oxygen)
Oxycure liquid oxygen tank – 25.6 m³
Oxycure portable liquid oxygen tank (1.0 m³)
More information at www.oxycure.be

D.SPEC.05.10 Pulse oxymètre MD300-C63 - Mode d'emploi_EN
v. 28/06/2019 3
OXYCURE : ORDERS 081 22 15 90
Oxycure Belgium plc.
Business Park Fernelmont
L. Génicot, 9
B-5380 Fernelmont
Belgium
Tel + 32 (0)81 22 15 90
Fax + 32 (0)81 22 15 99
http://www.oxycure.be
e-mail : oxycure@oxycure.be
OPENING HOURS
From Monday to Friday 09.00 – 12.30 and 13.30 – 18.00
CUSTOMER SERVICE 0800 98 0 68
Our customer service operates 24/7.
It is exclusively to urgent installations or repairs on our medical devices.
Call to guard service
Leave a message on the answering machine with your name and phone number and
give a brief reason of your call ;
The customer service will call you within 20 minutes ;
If the customer service did not call you within 20 minutes, call again.
Notes
1. Technical interventions (repairs) are included in the price of the assistance of the
leasing.
2. Deliveries of goods, outside the opening hours, will be charged.

D.SPEC.05.10 Pulse oxymètre MD300-C63 - Mode d'emploi_EN
v. 28/06/2019 4
OPERATOR’S
MANUAL
V1.0C63
Fingertip Pulse Oximeter
General Description
Oxygen Saturation is a percentage of Oxyhemoglobin (HbO2) capacity, compounded with
oxygen, by all combinative hemoglobin (Hb) capacity in blood In other words, it is
consistency of Oxyhemoglobin in blood. It is a very important parameter for the Respiratory
circulation System. Many respiratory diseases can result in oxygen saturation being lowered
in human blood. Additionally, the following factors can reduce oxygen saturation: Automatic
regulation of organ dysfunction caused by Anesthesia, Intensive Postoperative Trauma,
injuries caused by some medical examinations. That situation might result in light-
headedness, asthenia, and vomiting. Therefore, it is very important to know the oxygen
saturation of a patient so that doctors can find problems in a timely manner.
The fingertip pulse Oximeter features small size, low power consumption, convenient
operation and portability. It is only necessary for a patient to put one of his fingers into the
fingertip photoelectric sensor for diagnosis, and a display screen will show oxygen
saturation. It has been proven in clinical experiments that it also features high precision and
repeatability.
Measurement principle
Principle of the Oximeter is as follows: A mathematical formula is established making use of
Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive
hemoglobin (RHb) and Oxyhemoglobin (HbO2) in glow and near-infrared zones. Operation
principle of the instrument: Photoelectric Oxyhemoglobin Inspection Technology is adopted
in accordance with Capacity Pulse Scanning and Recording Technology, so that two beams
of different wavelength of lights (660nm glow and 940nm near infrared light) can be focused
onto a human nail tip through a clamping finger-type sensor. A measured signal obtained
by a photosensitive element, will be shown on the Oximeter’s display through process in
electronic circuits and microprocessor shown on the Oximeter’s display through electronic
circuits and a microprocessor.
Diagram of Operation Principle
1. Red and Infrared-ray Emission Tube
2. Red and Infrared-ray Receipt Tube
Precautions for use
1Do not use the pulse oximeter in an MRI or CT environment
2Do not use the pulse oximeter in situations where alarms are required.
3Explosion hazard: Do not use the pulse oximeter in an explosive atmosphere.
4The pulse oximeter is intended only as an adjunct in patient assessment. It must be
used in conjunction with other methods of assessing clinical signs and symptoms.
5Check the pulse oximeter sensor application site frequently to determine the positioning
of the sensor and circulation and skin sensitivity of the patient.
6Do not stretch the adhesive tape while applying the pulse oximeter sensor. This may
cause inaccurate readings or skin blisters.
7Before use, carefully read the manual.
8The pulse oximeter has no SpO2alarms; it is not for continuous monitoring, as indicated
MD300C63

D.SPEC.05.10 Pulse oxymètre MD300-C63 - Mode d'emploi_EN
v. 28/06/2019 5
by the symbol.
9Prolonged use or the patient’s condition may require changing the sensor site
periodically. Change sensor site and check skin integrity, circulatory status, and correct
alignment at least every 4 hours.
10 Inaccurate measurements may be caused by autoclaving, ethylene oxide sterilizing, or
immersing the sensors in liquid may cause inaccurate readings.
11 Significant levels of dysfunctional hemoglobins (such as carbonxy- hemoglobin or
methemoglobin)
12 Intravascular dyes such as indocyanine green or methylene blue
13 SpO2measurements may be adversely affected in the presence of high ambient light.
Shield the sensor area (with a surgical towel, or direct sunlight, for example) if
necessary.
14 Excessive patient movement
15 Venous pulsations
16 Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or
intravascular line.
17 The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia
18 The patient is in cardiac arrest or is in shock.
19 Fingernail polish or false fingernails may cause inaccurate SpO2readings.
Follow local ordinances and recycling instructions regarding disposal or recycling
of the device and device components, including batteries.
Product Properties
1Operation of the product is simple and convenient.
2The product is small in volume, light in weight and convenient in carrying.
3Power consumption of the product is low.
4Low voltage warning will be indicated in visual window when battery voltage is so low
that normal operation of the Oximeter might be influenced.
5The product will automatically be powered off when no signal is in the product for longer
than 8 seconds.
Product Operation Scope
Fingertip PULSE OXIMETER is a portable non-invasive, spot-check, oxygen saturation of
arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patient at home, and
hospital (including clinical use in internist/surgery, Anesthesia, intensive care and etc). Not
for continuously monitoring.
The PULSE OXIMETER requires no routine calibration or maintenance other than
replacement of batteries.
Operation Instructions
1Installing two AAA batteries correctly.
2Nip the clamp as diagram.
3Insert one finger into rubber hole of the Oximeter fully.
4Press the switch button once on front panel.
5Your finger and body do not tremble during the Oximeter is working.
6Read corresponding data from display screen.
7Six display modes
Table of contents
Other oxycure Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Getinge
Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Mettler Electronics
Mettler Electronics Sonicator 730 Maintenance manual

Pressalit Care
Pressalit Care R1100 Mounting instruction

Denas MS
Denas MS DENAS-T operating manual

bort medical
bort medical ActiveColor quick guide

AccuVein
AccuVein AV400 user manual