IMPORTANT:
READ MANUAL COMPLETELY BEFORE OPERATING THIS DEVICE
Basic delivery technique is described. Also, this manual contains instructions on periodically
required checks to be performed by the user. These checks are necessary to insure the proper
performance of this device and its safety features. Retain this manual for future reference.
WARNINGS AND PRECAUTIONS
These warnings and precautions are to help you
to understand how to safely operate the
Nitronox device. A WARNING alerts you to a
possible hazard to people. A CAUTION alerts
you to the possibility of equipment damage.
WARNING: New or modified installations -
properly connected gas pipelines are
absolutely essential to patient safety. The
ultimate responsibility of assuring that lines
are not crossed rests with the user. See next
page for details.
WARNING: An oxygen enriched environment
can accelerate the spread of ignited materials.
Therefore, when used in conjunction with
energy producing devices such as lasers, RF
sources, or other heat sources the user must
adhere to the instructions for use of those
devices to avoid ignition of combustible
materials.
WARNING: Nitronox Inhalation Analgesia
systems are intended to be used by medical
professionals trained in its use, and are
responsible for administering nitrous oxide and
oxygen for medical applications with
consideration of contraindications for use as
well as for other risks that could cause harm to
the healthcare professional, staff, patient, or
facility
WARNING: Do not use this device for the
administration of general anesthesiaor as a part
of, or in conjunction with, a general anesthesia
administration system.
WARNING: Inspect and maintain the
analgesia delivery system to prevent N2O leaks
in all hoses, connections and fittings. Have all
leaks repaired immediately.
WARNING: Use Scavenging: Medical workers
are exposed to N2O during administration of
N2O/O2 conscious sedation analgesia. Controls
are effective in the patient treatment area to
achieve low levels of ppm (parts per million)
exposure. Controls include System
Maintenance, Ventilation and Work Practices.
Your accessory Scavenger System is an
important part of the system of controls in
medical settings.
WARNING: During any power outage,
remember to turn OFF the tank valves (or
external supplier). With centralized, electrically
powered gas systems, if gas was flowing when
the power went out and the Nitronox is left ON,
gas will be capable of flowing when the power is
restored.
WARNING: When using a single use full face
mask (not supplied sterile), dispose of mask
after use to prevent patient cross-
contamination. When using multiple use mask,
follow manufacturer’s sterilization instructions.
WARNING: Always use clean, dry medical
gradegases. Donotintroducemoistureorother
contaminants into the system and flowmeter
device. Never oil or grease any part of this
system (minimize fire or explosion
potential). Cylinder mount: Do not clean
check valve entrance or sealing washer
(replace as needed) with disinfectant. User
should ensure that the cylinder valve is clear of
dust or dirt which may be carried to regulator
and cause damage or accident. One clearing
technique is to “crack” [secure] cylinder valve
before installing cylinder to E-Block. Open valve
slightly and then close. Do not discharge flow of
gas at any person or flammable material.
CAUTION: Do not attempt to repair, alter or
calibrate this device. Unauthorized repair,
alteration or misuse of this device is likely to
adversely affect the performance and will void
the warranty.