precisionplus L30 User manual

P. 1
P. 2
P. 3
3. Contraindications
Should be used with caution on infants, pregnant women, hemophiliacs and
patients with heart problems.
4. Adverse Reactions
●Patients may feel tense, uncomfortable or experience pain during use of the device.
● Patients hearing may be affected during use of the device.
1. Indication for Use
● This air motors is intended to operate/drive dental handpiece attachments
for dental applications.
● To be used for dental treatments by trained dental professionals.
2. Precautions for Handling and Operation
● Devices are only to be used for the purpose specified in Intended for Use.
● All precautions should be read and understood before use.
● Instructions are provided in order to prevent the risk of personal injury or damage
to the dental handpiece.
WARNING: Indicates a hazard that may result in serious injury/device
damage if instructions are not correctly followed.
CAUTION: Indicates a hazard that may result in mild to moderate
injury / device damage if instructions are not correctly followed.
CAUTION
● Users are required to protect themselves, the patient and third parties from danger.
● Users are responsible for correct and intended usage, regular maintenance and
continual inspection of the air motor.
● Avoid impact or contamination of air motor.
● Do not disassemble or alter the air motor except as recommended in our user
instructions.
● Correct eye, face and personal protection must be used by operator, assisting
staff and patient whilst air motor is being used.
● If the air motor begins operating abnormally, immediately stop use and contact
your authorized Precision Plus dealer.
● Air motor is supplied in a non-sterile state and must be autoclaved before use.
● Air motor should be carefully tested for vibration, noise or overheating prior to use,
if it has not been used for a long time.
USER MANUAL
DENTAL AIR MOTOR
For your safety and the safety of your patients please read this user manual
carefully before use and file for future reference.
This manual is published by Precision Plus.
Precision Plus does not guarantee its contents and reserves the right to amend
without prior notice. Amendments will be published in new editions of this manual.
L30 Air Motor
L30 Rev. V2.3 © Precision Plus
L30-M4

7.
Connection & Disconnection of Handpiece to Air Motor
1. Connect handpiece to air motor until it
locks into place. (Fig.1)
2. Pull on handpiece to make sure it is
securely attached to air motor.
1. Hold air motor and handpiece attachment separately and pull the
handpiece attachment off in an axial direction.(Fig.1)
7.1 Connection
7.2 Disconnection
P. 4
P. 6
P. 5
P. 7
6. Air Supply Pressure Setting
Air supply pressure must be tested and set to comply with specified values in
the table 5. Specifications.
WARNING: Do not exceed the specified air supply pressure values in table
5. Specifications.
CAUTION: Air supply must be free of dust, moisture and oil.
8. Test Prior to Use
Once handpiece is correctly connected to air motor, check for unusual
vibration, noise or overheating. If there are any abnormalities, see troubleshooting
chart or contact your authorized Precision Plus dealers. if problem persists.
Model L30-M4 (4 holes)
Hose Connection Type ISO 9168 Type 2
(Midwest 4 holes)
Handpiece Connection Type ISO 3964
Air Supply Pressure 3 kg/cm² (42 p.s.i.)
Negative Rotating Speed of Motor (3 kg/cm²)
Positive Rotating Speed of Motor (3 kg/cm²)
Maximum Speed 20,000 r.p.m. ±10%
20,000 r.p.m. ±10%
20,000 r.p.m. ±10%
Air Consumption (3 kg/cm²) <66NL/min
Rated Torque >0.0005N.m (at 3 kg/cm²)
5. Specifications
Gear Ratio 1:1
10. Post-Use Maintenance
CAUTION: Do not use strong/highly acidic water, strong acid/alkaline
chemicals, chlorine containing solutions or solvents such as
benzine or thinner to wipe, immerse or clean the device.
Do not use ultrasonic bath during the cleaning process.
CAUTION: Motor must have completely stopped rotating before
connecting / disconnecting.
9. Forward/Reverse Rotation
Fig. 2
Fig. 3
1. Forward Rotation:Turn FWD/REV
regulating sleeve to “F”.(Fig 2)
F: Forward Rotation.
2. Reverse Rotation:Turn FWD/REV
regulating sleeve to “R”. (Fig 3)
R: Reverse Rotation.
Fig. 1

Lubricant must be applied before autoclaving.
10.3 Lubricating
P.10
P. 11
P. 8
P. 9
1. Use a high quality spray lubricant intended for use with
dental handpieces / air motors (Not included with device).
2. Insert spray nozzle into drive air tube of air motor (shown
in Fig 4). Hold the air motor firmly and spray for
approximately 2 to 3 seconds.
3. Lubricant should be seen to be expelled from exhaust air
tube. (Fig 5) for Midwest 4 holes.
Fig. 5
Fig. 4
Spray Lubricant
Midwest 4 holes
Midwest 4 holes
1. Use correct eye, face and personal protection to avoid possible infection.
2. Disconnect handpiece from air motor.
3. Remove air motor from air supply hose.
10.1 Preparation
Disconnect the air motor from the tube. Remove gross soiling of the instrument with
cold water (<40°C) immediately after use. Don’t use a fixating detergent or hot water
(>40°C) as this can cause the fixation of residuals which may influence the result of
the reprocessing process. Store the instruments in a humid surrounding.
Safe storage and transportation to the reprocessing area to avoid any damage and
contamination to the environment.
10.2 Cleaning
10.2.1 Preparation at the Point of Use:
Do a manual pre-cleaning, until the air motor is visually clean. Submerge the air
motor in a cleaning solution and flush the lumens with a water jet pistol with cold tap
water for at least 10 seconds. Clean the surfaces with a soft bristle brush.
10.2.2 Pre-cleaning
Use a washer-disinfector meeting the requirements of the ISO 15883 series.
Put the instrument into the machine on a tray. Connect the instrument with the WD
by using suitable adapter and start the program:
• 4 min pre-washing with cold water (<40°C);
• emptying
• 5 min washing with a mild alkaline cleaner at 55°C
• emptying
• 3 min neutralising with warm water (>40°C);
• emptying
• 5 min intermediate rinsing with warm water (>40°C)
• Emptying
The automated cleaning processes have been validated by using 0.5% neodisher
MediClean forte (Dr. Weigert).
Note: Acc. to EN ISO 17664 no manual reprocessing methods are required for these
devices. If a manual reprocessing method has to be used, please validate it prior to use.
10.2.3 Automated Cleaning:
Automated Thermal Disinfection in washer/disinfector under consideration of
national requirements in regards to A0 value (see EN 15883).
A disinfection cycle of 5 min disinfection at 93°C has been validated for this device
to achieve an A0 value of 3000.
10.2.4 Disinfection:
Drying of outside of instrument through drying cycle of washer/disinfector. If needed,
additional manual drying can be performed through lint free towel. Insufflate cavities
of instruments by using sterile compressed air.
10.2.5 Drying:
CAUTION: Visual inspection for cleanliness of the air motor. If necessary, p
erform reprocessing process again until the air motor is visibly clean. Before
packaging and autoclaving, make sure that the air motor has been maintained
acc. to this instruction.

Air motor should be sterilized after each and every treatment.
1. Air motor should be lubricated as per user manual.
2. Insert the air motor into an autoclave pouch and seal pouch correctly. The
autoclave pouch should comply with the requirements of EN ISO 11607.
3. Sterilize at 134 °C for 5 mins. Sterilization of instruments by applying a fractionated
pre-vacuum steam sterilization process (according to EN 285/EN 13060/EN ISO
17665) under consideration of the respective country requirements. Flash
sterilization is not allowed on lumen instruments!
4. Air motor should be stored in a sealed pouch in a dry, clean and dust free place
out of direct sunlight until required for use.
10.5 Sterilization
10.4 Oil Lubrication
1. Apply 1-2 drops of high quality lubricant oil intended for use with dental
handpiece / air motor (not included with device) into drive air tube (Fig. 1)
2.
Connect air motor to air supply hose. Operate the motor for approximately 10 seconds.
P.15
CAUTION
● Always follow the instructions provided by lubricant manufacturer. Be aware of all
precautions and warnings.
● Always hold lubricant can in an upright position.
● Hold air motor firmly to avoid it being dislodged due to lubricant spray pressure.
CAUTION
● Air motor must be cleaned and lubricated before sterilization.
● Any remaining blood on outer surface may become clotted and cause operational
failure.
● Do not use autoclaves exceeding 137°C during sterilization.
●Sterilize according to EN13060, class B. Always follow autoclave manufacturer's
instructions for use.
● Reprocessed products should be stored, protected from dust in a dry, dark and
cool place with minimum exposure to germs.
● Autoclave sterilization is the only agreed method to correctly sterilize this air motor.
The validity of other sterilization methods is not confirmed or guaranteed.
● Resistance to reprocessing: 250 cycles (reprocessing cycles include cleaning and
sterilization
The device has not been subjected to a sterilization process
Serial number
Conforms to CE European Directive of “Medical device directive 93/42/EEC”
Authorized representative in the European Community.
Batch code
Manufacturer
P.12
P.13
P.14
0197
NON
STERILE
Medical Device
Consult instructions for use
Sterilizable in a steam sterilizer (autoclave) at the temperature 134°C(273°F)
Washer-disinfector for thermal disinfection
12. Symbols
Points to check
Rotation
Process
‘O’ rings Make sure all O’ rings are in correct place and are undamaged.
Connect and start Air Motor. Check for abnormal rotation, noise, overheating
and vibration.
11. Regular Maintenance Checks
Regular maintenance should be performed every 3 months as per the below chart.
If any abnormalities are found, immediately contact your authorized Precision Plus dealer.
* See trouble shooting chart if problems are discovered.

If the device will not work correctly after following above troubleshooting solutions,
please contact Precision Plus or an authorized maintenance centre.
Malfunction Cause Solution
High noise, weak torque,
stops rotating Bearing damaged Replace the bearing
Normal sound but weak torque Air pressure too low Check and correct working air pressure
13. Troubleshooting
made in China
Precision Plus products are warranted against manufacturing and material defects.
Precision Plus reserves the right to analyze and determine the cause of any problem.
The Warranty is voided should the product not be used in accordance to the user manual,
not for its intended purpose, has not been repaired by Precision Plus or an authorized
Precision Plus repairer, or has non genuine Precision Plus parts installed. Replacement
parts will be made available for seven years after the date of discontinuation of the model.
In order to avoid the health risks to operators handling the disposal of medical devices,
as well as the risks of environment contamination caused thereof, a surgeon or a dentist
is required to confirm the device is sterilized prior to disposal. Ask a specialized firm
who is licensed to dispose specially controlled medical wastes, to dispose this product for
you. Always follow your country specific laws, directives, standards and guidelines for
disposal of medical devices.
14. Warranty
15. Disposal of Product
P.16
P.17
Precision Plus China Limited
Flat 103, No. 122, Tianqiao Road,
Tutang, Changping, Dongguan,
Guangdong, China
Tel: +86-769-82306233
Fax: +86-769-82301566
L30-M4 12-2021
Premium Plus Poland sp. z o.o.
ul. Bukowska 27
62-081 Wysogotowo
Poland
Tel: 48-61-880-10-94
Email: [email protected]
imported and distributed by:
manufactured by:
This manual suits for next models
1
Table of contents
Other precisionplus Dental Equipment manuals
Popular Dental Equipment manuals by other brands

BRASSELER USA
BRASSELER USA EndoSync TR-CM BSL Operation instructions

KDF
KDF Cascom SE instruction manual

Lares Research
Lares Research 557 Instructions for use

DentalEZ
DentalEZ Star TITAN Indications for Use

Acteon
Acteon SATELEC Newtron Booster user manual

Morita
Morita TorqTech ST-DH Series operating instructions