1.H. Precautions
• Important Safety Instructions. Save these instructions.
• This IVD instrument complies with the EMC emission and immunity requirements described in
IEC-61326-2-6 and the safety requirements of IEC 61010-1 and IEC 61010-2-101.
• Changes or modifications to this unit not expressly approved by the party responsible for compliance
could void the user’s authority to operate the equipment.
• This equipment has been designed and tested to CISPR 11 Class A. In a domestic environment it may
cause radio interference, in which case you may need to take measures to mitigate the interference.
• Do not use this device in proximity to sources of strong electromagnetic radiation (e.g., unshielded
intentional RF sources), as these may interfere with the proper operation.
• Do not use this instrument for anything other than its intended use.
• Always disconnect the power before cleaning or performing routine maintenance.
• Do not disassemble unit.
• Ensure cartridges, elution tubes and plungers have been securely inserted in their correct positions and
orientation. Failure to do so may result in damage to the instrument.
• Use only Promega-supplied cartridges, plungers and elution tubes. Do not reuse cartridges, plungers or
elution tubes.
• If the equipment is used in a manner other than that specified by Promega, the protection provided by
the equipment may be impaired.
• Keep hands clear of instrument platform as it moves in and out of the instrument.
• During elution, the heated elution block at the front of the platform becomes very hot. Do not touch.
• To avoid muscle strain or back injury, use lifting aids and proper lifting techniques when removing or
replacing the instrument.
• Equipment can be hazardous due to the use of chemical and biohazardous substances.
• Instrument door should be opened and closed only by using the Maxwell®CSC software. Do not
manually pry the door open or override the door sensor during protocol operation, as it will result in the
protocol being aborted.
• This instrument is intended for use with human samples. Use appropriate personal protective equip-
ment (gloves, safety goggles, lab coat, etc.) for handling and disposing of biohazardous materials.
• Do not attempt to replace the UV lamp. Replacing the UV lamp requires special tools. Contact Promega
Technical Services if the UV lamp needs replacement.
• Do not load any additional software programs on the Tablet PC supplied with Maxwell®CSC Instru-
ment. Additional programs may cause the application to slow down.