
19
22
Battery Replacement
1. Replace battery when " appears in the upper right corner "
of LCD display. (See Figure 9)
2. Slide battery cover down as shown in Figure 10.
3. Remove battery and install 2 new AAA alkaline batteries
as shown . in Figure 11
4. Slide battery cover back on.
Figure 9
Figure 10
Figure 11
20
Care And Cleaning
1. The probe window must be kept clean, dry, and undamaged at all
times to ensure accurate readings. The accuracy of temperature
readings can be affected by damage to the probe window, or the
presence of dirt and ear wax on the probe window.
2. Fingerprints, earwax, dust and other soiling compounds reduce
transparency of the window and result in lower temperature
readings.
3. The probe window is the most delicate part of the thermometer.
To safely clean the window, gently wipe its surface with a cotton
swab slightly moistened with isopropyl alcohol and immediately
wipe dry with a clean cotton swab. After cleaning, allow at least
5 minutes drying time before taking temperatures.
Note: Do not use any chemical other than isopropyl alcohol to
clean the probe window.
4. Use a soft, dry cloth to clean the thermometer display and
exterior.
5. The thermometer is not waterproof. Do not submerge the unit in
water when cleaning.
6. Store the thermometer in a dry location, free from dust and
contamination and away from direct sunlight.
7. Periodic cleaning and disinfection of the device following in
order to prevent patient cross infection. Please clean and
disinfect the thermometer immediately after each use.
21
Care And Cleaning
8. Ensure that children do not use the instrument unsupervised;
some parts are small enough to be swallowed.
9. Do not remove or modify the equipment without permission.
10. Strong electromagnetic fields may interfere with the proper
operation of the thermometer. The device needs special
pre-cautions regarding EMC according to the EMC information.
11
12. Put the thermometer back to the original packaging after using.
For cleaning:
1)Soak a clean soft cloth in drinking water, wring it out and then
wipe the thermometer(including probe) no less than 3 times;
2)Then use another clean soft cloth to wipe the residue water
on thermometer;
3)Put the thermometer in the original packaging.
For disinfection:
1)Soak a clean soft cloth in drinking water, wring it out and then
wipe the thermometer(including probe) no less than 3 times;
2)Then use another clean soft cloth to wipe the residue water
on thermometer;
3)Using a clean soft cloth dipped in 70% medical alcohol,
wipe the probe 3 times, at least one minute for each time.
4)Using a clean cotton swab dipped in 70% medical alcohol,
wipe the sensor window 3 times;
5)Wait at least 10 minutes to let the alcohol volatilization and
then put the thermometer in the original packaging.
. It is not intended for use in the oxygen rich environment and
presence of flammable anesthetic mixture with air, oxygen or
nitrous oxide.
Specifications
Ear/Forehead mode: 34.0℃~43.0℃(93.2℉~109.4℉)
Object mode: 0℃~100℃(32℉~212℉)
Ear/Forehead mode:
±0.2℃ (0.4℉) during 35.5℃~42.0℃ (95.9℉~107.6℉)
at 15℃~35℃ (59.0℉~95.0℉) operating temperature range
±0.3℃ (0.5℉) for other measuring and operating
temperature range
Object mode: ±4% or ±2℃(4℉) whichever is greater
Measuring range
Laboratory accuracy
Display resolution 0.1℃ or 0.1℉
Measure time
Ear mode :
Forehead: Approx. three seconds
/Object mode Approx. one second
Measuring site Ear canal(Ear Mode)、Forehead(Forehead Mode)
Oral (This thermometer converts the ear temperature to
display its “oral equivalent.”)
Reference body site
Operation mode Ear/Forehead mode(Adjust mode)
10℃~40℃(50℉~104℉), 15%~85%RH, non-condensing
00hPa ~ 1060hPaAtmospheric Pressure : 7
Approx.85grams(with batteries)
DC3V(2×AAA battery)
Operating temperature
range:
Storage and transport
temperature range
-25℃~ 55℃ (-13℉~131℉), 15%~95%RH, non-condensing
700hPa ~ 1060hPaAtmospheric Pressure :
Clinical accuracy
Dimension
Weight
Battery
Battery life Approx. 1 year/6000 readings
Ingress protecting rating
IP22
Clinical bias:0.05℃(0.09℉)
Clinical repeatability:0.11℃(0.20℉)
Limits of agreement:1.04℃( 1.88℉)
Shock withstands drop of 3 feet
Expected service life Three years
155 * 39* 50mm