Quidel Sofia Influenza A+B User manual

120 µL Pipette
Test Procedure
NOTE: The procedures for testing swab specimens versus liquid specimens in viral transport media are different. Read carefully.
All specimens must be at room temperature before testing.
Expiration date: Check expiration date on each individual test package or outer box before using. Do not use any test past the expiration date on the label.
Verify that Sofia or
Sofia 2 is set to the
desired Mode:
WALK AWAY or
READ NOW. See
the “Using Sofia”or
“Using Sofia 2”
section for more
information.
Dispense all of the Reagent
Solution into the Reagent
Tube. Swirl the Reagent Tube
to dissolve its contents.
Place the patient swab
sample into the Reagent
Tube. Roll the swab at
least 3 times while
pressing the head
against the bottom and
side of the Reagent
Tube.
Roll the Swab head
against the inside of the
Reagent Tube as you
remove it. Dispose of
the used Swab in your
biohazard waste.
Fill the provided Small, Clear
120 µL Fixed Volume Pipette with
patient sample from the Reagent
Tube.
To fill the Fixed Volume Pipette
with the patient sample:
a) FIRMLY squeeze the top bulb.
b) Still squeezing, place the Pipette
tip into the sample.
c) With the Pipette tip still in the
sample, release pressure on bulb
to fill the Pipette.
Firmly squeeze the top bulb to
empty the contents of the Small,
Clear 120 µL Fixed Volume Pipette
into the Test Cassette sample well.
Extra liquid in the overflow bulb is
OK.
NOTE: The Fixed Volume Pipette is
designed to collect and dispense the
correct amount of patient sample.
Discard the Pipette in your
biohazard waste.
NOTE: Do not pour sample from the
Reagent Tube. Use the provided
Small, Clear 120 µL Fixed Volume
Pipette.
Proceed to the
“Using Sofia”or
“Using Sofia 2”
section to complete
the test.
Swab Test Procedure (Nasal/Nasopharyngeal)
Study the Package Insert and User Manual
thoroughly before using Quick Reference
Instructions. This is not a complete
Package Insert.
IMPORTANT! Read instructions carefully before
beginning. The test procedure below is unique to the
Sofia Influenza A+B FIA and may differ from other Sofia
FIA procedures.
CLIA Complexity: Waived for direct nasal swab, nasopharyngeal
swab and nasopharyngeal aspirate/wash specimens
QUICK REFERENCE
INSTRUCTIONS
For use with Sofia and Sofia 2.
Rx only
CLIA Complexity: Moderate for nasopharyngeal swab and
nasopharyngeal aspirate/wash specimens eluted in
transport media

120 µL
Pipette
Verify that Sofia or
Sofia 2 is set to the
desired Mode:
WALK AWAY or
READ NOW. See
the “Using Sofia”or
“Using Sofia 2”
section for more
information.
Dispense all of the Reagent
Solution into the Reagent
Tube. Swirl the Reagent Tube
to dissolve its contents.
Fill the provided Large, Pink
250 µL Fixed Volume Pipette
with patient sample from the
collection cup.
To fill the Fixed Volume
Pipette with the sample:
a) FIRMLY squeeze the top
bulb.
b) Still squeezing, place the
Pipette tip into the patient
sample.
c) With the Pipette tip still in
the patient sample, release
pressure on bulb to fill the
Pipette.
NOTE: To obtain accurate
results, avoid mucoid
substances when filling the
Large, Pink 250 µL Fixed
Volume Pipette with patient
sample from the collection cup.
Firmly squeeze the top bulb to
empty the contents of the
Large, Pink 250 µL Fixed
Volume Pipette into the
Reagent Tube. Extra liquid in
the overflow bulb is OK.
NOTE: Once the sample is
added to the Reagent Tube,
vigorously mix prior to
adding the sample to the Test
Cassette.
NOTE: The Fixed Volume
Pipette is designed to collect
and dispense the correct
amount of patient sample.
Discard the Pipette in your
biohazard waste.
Fill the provided Small,
Clear 120 µL Fixed
Volume Pipette with
patient sample from the
Reagent Tube.
Firmly squeeze the top bulb
to empty the contents of
the Small, Clear 120 µL
Fixed Volume Pipette into
the Test Cassette sample
well. Extra liquid in the
overflow bulb is OK.
NOTE: Do not pour sample
from the Reagent Tube.
Use the provided Small,
Clear 120 µL Fixed Volume
Pipette.
Proceed to the
“Using Sofia”or
“Using Sofia 2”
section to complete
the test.
250 µL
Pipette
Liquid Specimens Test Procedure
250 µL
Pipette

WALK AWAY/READ NOW Modes
Refer to the Sofia User Manual.
Sofia may be set to two different modes (WALK AWAY and READ NOW). The procedures for each mode are described below.
In WALK AWAY Mode, the user immediately inserts the Test Cassette into Sofia.
Sofia will automatically time the test development, and the results will be
displayed in 15 minutes.
The user places the Test Cassette onto the counter or bench top for 15 minutes (outside of
Sofia). The user manually times this development step. The Test Cassette MUST remain on
the bench for 15 minutes to get an accurate result. Then, the user inserts the Test Cassette
into Sofia. In READ NOW Mode, Sofia will scan and display the test result in 1 minute. Note:
Results will remain stable for an additional 15 minutes after the recommended
development time of 15 minutes.
RUN TEST
1. Input the User ID with the barcode scanner or enter
the data using the key pad.
NOTE: If you scan the wrong barcode, use the Arrow
Keys on Sofia to re-highlight the field and simply
rescan using the correct barcode. The previous one will
be overwritten with the right barcode.
3. Press Start Test and the Sofia drawer will
open.
4. Verify that the correct mode (WALK
AWAY or READ NOW) has been selected.
Insert the Test Cassette into the drawer.
Then gently close the drawer.
2. Input the Patient ID or Order # with the barcode
scanner or enter the data using the key pad.
5. Sofia will start automatically and display the progress. In WALK AWAY
Mode, the test results will be displayed in
approximately 15 minutes. In READ NOW
Mode, the test results will be displayed in
1 minute. See Sofia Interpretation of
Results section.
Allow the test to develop for the FULL 15 minutes BEFORE placing it into Sofia.
WALK AWAY Mode
READ NOW Mode
Using Sofia

When the test is complete, the results will be displayed on the Sofia screen. The results can be
automatically printed on the integrated printer if this option is selected. Test Lines will not be
visible to the naked eye.
Results: The Sofia screen will display results for the procedural control as being “valid or invalid.” If
the control is “invalid,” retest with a new patient sample and new Test Cassette.
Reader Display
Interpretation
Flu A: Positive
Flu B: Negative
Procedural Control: Valid
Positive Test for Flu A
(influenza A antigen present)
Flu A: Negative
Flu B: Positive
Procedural Control: Valid
Positive Test for Flu B
(influenza B antigen present)
Flu A: Positive
Flu B: Positive
Procedural Control: Valid
Positive Test for both Flu A and Flu B
(influenza A and B antigen present)
Flu A: Negative
Flu B: Negative
Procedural Control: Valid
Negative Test for Flu A and Flu B
(no antigen detected)
Flu A: Invalid
Flu B: Invalid
Procedural Control: Invalid
Result Invalid
From the main menu, select
Run QC.
Follow the prompts on the screen. Scan the QC Card (located on the kit box).
Sofia prompts the user to select the desired mode (WALK AWAY or READ NOW).
Run the External Control Swabs.
Follow the Swab Test Procedure of this Quick Reference Instructions. First test the
Positive Swab, followed by the Negative Swab.
After both the Positive and Negative Swabs have been run, the results will be
displayed as “Passed” or “Failed.”
Sofia Interpretation of Results
Sofia External Quality Control (External Positive and Negative Swabs are supplied in the kit)

WALK AWAY/READ NOW Modes
Refer to the Sofia 2 User Manual.
Sofia 2 may be set to two different modes (WALK AWAY and READ NOW). The procedures for each mode are described below.
In WALK AWAY Mode, the user immediately inserts the Test Cassette into Sofia
2. Sofia 2 scans the Test Cassette periodically during the test development time.
Positive test results will be displayed between 3 and 15 minutes. Negative test
results will be displayed at 15 minutes.
The user first places the Test Cassette onto the counter or bench top for 15 minutes
(outside of Sofia 2). The user manually times this development step. The Test Cassette
MUST remain on the bench for 15 minutes to get an accurate result. Then, the user inserts
the Test Cassette into Sofia 2. In READ NOW Mode, Sofia 2 will scan and display the test
result within 1 minute. Note: Results will remain stable for an additional 15 minutes after
the recommended development time of 15 minutes.
RUN TEST
1. Input the User ID with the integrated barcode scanner
or manually enter the data using the key pad.
NOTE: If you scan the wrong barcode, re-highlight the
field using the touchscreen on Sofia 2. The previous
one will be overwritten with the right barcode.
3. Verify that the correct mode (WALK
AWAY or READ NOW) has been
selected. Press and open the Sofia
2 drawer.
4. Insert the Test Cassette into the
drawer. Then gently close the drawer.
2. Input the Patient ID and Order # (if applicable) with
the integrated barcode scanner or manually enter the
data using the key pad.
5. Sofia 2 will start automatically and display the progress. In WALK AWAY
Mode, the test results will be displayed in approximately 15 minutes. In
READ NOW Mode, the test results will be displayed within 1 minute. See
Sofia 2 Interpretation of Results section.
Allow the test to develop for the FULL 15 minutes BEFORE placing it into Sofia 2.
Using Sofia 2
WALK AWAY Mode
READ NOW Mode

Reader Display
Interpretation
Flu A:
Flu B:
Procedural Control:
Positive Test for Flu A
(influenza A antigen present)
Flu A:
Flu B:
Procedural Control:
Positive Test for Flu B
(influenza B antigen present)
Flu A:
Flu B:
Procedural Control:
Positive Test for both Flu A and Flu B
(influenza A and B antigen present)
Flu A:
Flu B:
Procedural Control:
Negative Test for Flu A and Flu B
(no antigen detected)
Flu A:
Flu B:
Procedural Control:
Result Invalid
From the main menu,
select Run QC.
Follow the prompts on the screen. Scan the QC Card (located on the kit box).
Sofia 2 prompts the user to select the desired mode (WALK AWAY or READ NOW).
Run the External Control Swabs.
Follow the Swab Test procedure of this Quick Reference Instructions. First test the
Positive Swab, followed by the Negative Swab.
After both the Positive and Negative Swabs have been run, the results will be displayed as
or .
INTENDED USE
The Sofia Influenza A+B FIA employs immunofluorescence to detect influenza A and influenza B viral nucleoprotein antigens in direct nasal swab, nasopharyngeal swab, and nasopharyngeal
aspirate/wash specimens and nasopharyngeal swab and nasopharyngeal aspirate/wash specimen in transport media from symptomatic patients. This qualitative test is intended for use
as an aid in the rapid differential diagnosis of acute influenza A and influenza B viral infections. The test is not intended to detect influenza C antigens. A negative test is presumptive and it
is recommended these results be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay. Negative results do not preclude influenza virus infections and should not
be used as the sole basis for treatment or other patient management decisions. The Sofia Influenza A+B FIA may be used with Sofia or Sofia 2.
ASSISTANCE
If Sofia, Sofia 2 or the assay do not perform as expected, contact Quidel Technical Support 800.874.1517 (in the U.S.), 858.552.1100 (outside the U.S.), technicalsupport@quidel.com, or
your local distributor.
Sofia 2 Interpretation of Results
Sofia 2 External Quality Control (External Positive and Negative Swabs are supplied in the kit)
Quidel Corporation
San Diego, CA 92121 USA
quidel.com
1219205EN00 (06/17)
When the test is complete, the results will be displayed on the Sofia 2 screen. Test Lines will not
be visible to the naked eye.
Results: The Sofia 2 screen will display results for the procedural control as being or . If the
control is , retest with a new patient sample and new Test Cassette.
Reference the Package Insert for Warnings and Precautions, Specimen Collection and Handling, and Quality Control.
Study the Package Insert and User Manual thoroughly
before using Quick Reference Instructions. This is not a
complete Package Insert.
Table of contents