Raytech SynchroKing 4S Mk2 User manual

RayTech Automotive
SynchroKing 4 Mk2 - Instruction Manual
www.raytechautomotive.com 1 of 6 Rev 2
Aug 2020
Application
The SynchroKing 4S Mk2is a universal vacuum synchronization/balancing tool for both
carburetors and throttle valve bodies. It is suitable for any combustion engine with four intake
ports from cars, motorcycles, mopeds, scooters, carts, mowers and outboard engines. The
SynchroKing 4S Mk2is designed to provide a compact, easy to use and economical
synchronisation tool for home and workshop use. The SynchroKing 4Sis primarily designed for
use with engines with four inlet manifolds but is in principle suitable for any engine with more than
one intake. If not all four ports are used, the device will automatically ignore the sensor input from
the disconnected ports, making it possible to use the device on 2 and 3 manifold engines. The
device comes with a set of 4 of each M5 and M6 nipples, enabling easy connection to the
vacuum test ports on most vehicles
This Owner's manual must be read in conjunction with the Workshop manual of the vehicle
before using the SynchroKing 4S. Before proceeding the user should as a minimum be familiar
with the test ports in the manifolds, the idle adjuster screws and the balancing adjuster screws.
Features:
1. Australian designed and manufactured.
2. High quality, robust housing.
3. Digital display for vacuum for four vacuum channels
4. Bar displays for differential vacuum display
5. Push-button operated dampening control
6. No battery required.
7. Reverse polarity protection.
8. Auto calibration function at startup.
9. 1-year limited warranty for the original Purchaser.
The package includes:
•1 x SynchroKing 4S Mk2device, c/w integrated power lead
•4 x vacuum hoses
•4 x M5 brass nipples
•4 x M6 brass nipples
•Detailed and illustrated user manual

RayTech Automotive
SynchroKing 4 Mk2 - Instruction Manual
www.raytechautomotive.com 2 of 6 Rev 2
Aug 2020
Device Description:
1. Main displays - The two main displays indicate the vacuum measured in the inlet
manifolds in mm of mercury [mmHg].
2. Bar displays - Each bar display indicates the level of vacuum relative to the average
vacuum in the inlet manifolds.
3. Pushbuttons ' + ' and ' - ' Used to set the desired dampening factor and initiate a manual
re-calibration
4. Power cable - Connects to the vehicle's battery to power the device.
5. Hoses - Connect to the nipples in the inlet manifolds.
6.
4 x M6 and 4 x M5 nipples - screw into the test ports of the vehicle's intake manifolds.
Figure 1 - Device Overview

RayTech Automotive
SynchroKing 4 Mk2 - Instruction Manual
www.raytechautomotive.com 3 of 6 Rev 2
Aug 2020
Connections
Diagram 2 shows the SynchroKing 4S Mk2connected to the battery and inlet manifolds
Figure 2 - Connection Diagram
Function description
Vacuum measurement
The SynchroKing 4S Mk2displays the vacuum of each individual channel in mmHg with a
resolution of 2mmHg. Values under 20mmHg are supressed and the device will display ' - - - ' for
any suppressed channels. If the channel is suppressed then it will be excluded from the
calculation of the average value of all active channels.
Vacuum differential
The SynchroKing 4S Mk2calculates the average value of all active channels and compares the
individual value of each channel to this average. The deviation from average is displayed on the
bar graph of the channel. Each bar represents 20 mmHg vacuum differential, whereas the two
middle being lit indicate a pressure differential of less than 20mmHg (i.e the vacuum being
practically the same in both manifolds). See Figure 3 for examples.
Dampening
The SynchroKing 4S Mk2is equipped with a dampening function to produce a meaningful and
readable display of the vacuum level. The vacuum in the inlet manifold of a running engine will
fluctuate which will result in a rapidly updating readout which makes it virtually impossible to read
the vacuum level. (Equivalent to a rapidly bouncing needle on a conventional gauge).
To overcome this, the dampening function will store the average value of many measurements
into the device memory and produce a display value based on these, resulting in a more stable
readout (equivalent to a choke valve on a conventional gauge). The dampening factor is

RayTech Automotive
SynchroKing 4 Mk2 - Instruction Manual
www.raytechautomotive.com 4 of 6 Rev 2
Aug 2020
selectable in 100 increments. A higher dampening factor will result in a more stable display but
reduced sensitivity and vice versa. The user should set the dampening factor such that a
satisfactory balance is struck between readability and sensitivity.
At startup the device will be set at the default dampening factor of 30.
The dampening factor can be adjusted in increments of 1 by pressing the '+' or '-' button
momentarily. The dampening factor can be increased or decreased in increments of 10 by
pressing and holding the '+' or '-' button respectively.
Once one of the buttons is pressed, the device will display 'dp' in the No.2 display and the current
dampening level in the No.3 display. After setting the desired dampening level and releasing the
button, the device will revert to normal operation after 1 second.
Calibration
At startup the SynchroKing 4S Mk2will perform an automatic calibration, during which the
output signal of each sensor unit is baselined at zero. (Note that values below 20 are suppressed
thus not showing on the display). This function ensures that each of the four sensors are
referencing the same start value.
The device is also equipped with a manual calibration option. In normal operation it is not
necessary to execute a manual calibration. Only in the event that the previous (auto) calibration
was not successful, manual calibration is needed. Incorrect calibration occurs for instance when
the pressure on one or more channels was not atmospheric at the calibration.
The manual calibration function is provided in order not to initiate a reset of the dampening level
as is the case with turning the device off and on.
To execute a manual calibration, press both the '+' and '-' buttons at the same time. The No.2
display will show 'CAL'. The No.3 display will show a count down by adding a ' - ' to the display
every one second. Calibration is complete when the No.3 display goes blank. Releasing either
button before the cycle is complete will abort the calibration.
Ensure that at the time of calibration, all channels are at atmospheric pressure. Vacuum or
pressure being present at the time of calibration will result in incorrect measurements. It is
suggested to turn the engine of and have the throttle wide open when calibrating.

RayTech Automotive
SynchroKing 4 Mk2 - Instruction Manual
www.raytechautomotive.com 5 of 6 Rev 2
Aug 2020
Operation
The following paragraphs describe the generic synchronization procedure which will be applicable
to most vehicles. The user should read these in conjunction with the vehicle's service manual.
Some manufacturers specify specific vacuum levels.
1. Find the test port plugs in the inlet manifold test ports and remove them.
2. Select the correct size of nipple (either M5 or M6) and install them into the manifold's test
ports. Apply 'snub tight' torque only, enough to compress the O-rings. Some vehicles are
fitted with vacuum test port that accept the 5mm hose to be connected directly; if this is
the case the nipples are unnecessary.
3. Push the hoses onto nipples, take note which manifold connects to which port on the
SynchroKing 4S Mk2
4. Connect the alligator clamps to the vehicle's battery. The device will now come on-line
and execute to automatic calibration. - CAUTION: Observe correct polarity and take care
not to cause a short circuit.
5. The display should now light up and indicate "- - -" on all displays.
6. Start the vehicle and allow the engine to warm up to normal operating temperature.
7. With the throttle fully closed, adjust, as necessary, the stationary adjuster screw on the
carburettor or throttle valve body, until the bar displays equal vacuum in both inlet
manifolds. (See Figure 3).
8. The dampening factor may be used to reduce the 'bounce' on the display that the engine
may produce. Bear in mind that increasing dampening will reduce bounce but reduces
sensitivity and vice versa.Most engines will not achieve a perfect and stable readout, the
user should aim for the best practically achievable setting.
9. After the idle balancing has been completed, proceed by slowly opening the throttle and
allow the engine speed to increase to roughly 1/3 of maximum RPM. Repeat the
procedure above for idling only using the balancing adjuster screw(s) - refer to the
manual.
10. Re-check and re-adjust the idle and balancing screws as required and stop the engine.
11. Remove the hoses.
12. Allow the nipples to cool of and remove them from the manifolds.
13. Reinstall the test port plugs (do not forget any seals or gaskets, observe the instructions
in the workshop manual).

RayTech Automotive
SynchroKing 4 Mk2 - Instruction Manual
www.raytechautomotive.com 6 of 6 Rev 2
Aug 2020
Figure 3 - Examples of Bar Display indications
Specification
Dimensions 160 x 120 x 32 mm
Nippple M5 dimensions M5, d_barb=6, L=25 mm
Nippple M5 dimensions M6, d_barb=6, L=25 mm
Supply Voltage Max 15 V
Supply Voltage Min 6 V
Current draw, typical 30 mA
Current draw, max 60 mA
Operating Temp, Max 50 deg C
Operating Temp, Min -5 deg C
Measurement resolution 2 mmHg
Measurement range 0 - 400 mmHg
Typical maximum deviation between channels at 200mmHg
<2mm mmHg
Limited Warranty-
The SynchroKing 4S Mk2
comes with a limited 1-year warranty for the original purchaser. RayTech will repair, replace
or refund any unit found to have defects in material workmanship during this period. Any warranty claims must be
submitted to raytechcontact@yahoo.com. Opening of the device will void the Warranty.
The customer is liable for return shipping. RayTech will pay for shipping back to the customer, only if the device has been
determined to be defective by RayTech.
RayTech does not guarantee the fitness of its products for any purpose other than described in this manual. RayTech
shall not be liable for any loss from use of the product or accidental or consequential damages.
Table of contents
Other Raytech Measuring Instrument manuals

Raytech
Raytech SynchroKing 2 Mk2 User manual

Raytech
Raytech Micro Junior 2 User manual

Raytech
Raytech WR14-R User manual

Raytech
Raytech WR50-12 User manual

Raytech
Raytech WR14 User manual

Raytech
Raytech TraceTek TTDM-128 User manual

Raytech
Raytech TR-Mark III 250V User manual

Raytech
Raytech MC2 User manual

Raytech
Raytech WR14 User manual

Raytech
Raytech TR-1P User manual
Popular Measuring Instrument manuals by other brands

Cal Test Electronics
Cal Test Electronics CT4197 user manual

HP
HP 415E Operating and service manual

MONARCH INSTRUMENT
MONARCH INSTRUMENT ACT-3X User manual and reference guide

Teledyne
Teledyne 100E instruction manual

IDEAL Networks
IDEAL Networks LanXPLORER Quick reference guide

CHAMELEON
CHAMELEON IHD6-CAD-PPMID manual