Realtech REAL CLEAN SYSTEM I User manual


Real Clean System I v 2.1
Real Tech Inc. 2
1 Contents
2Overview_____________________________________________________________________________4
3Safety and Warranty Registration__________________________________________________________5
3.1 Safety Instructions _________________________________________________________________5
3.2 Warranty Registration_______________________________________________________________5
4Technical Specifications_________________________________________________________________6
5Installation____________________________________________________________________________6
5.1 Unpacking and Inspection ___________________________________________________________6
5.2 Assembly ________________________________________________________________________7
5.3 Mounting_________________________________________________________________________7
5.4 Plumbing_________________________________________________________________________8
5.4.1 Connecting an M UV254 Analyzer _________________________________________________9
5.5 Electrical and Communication _______________________________________________________10
6Start Up and Operation_________________________________________________________________11
6.1 Start Up_________________________________________________________________________11
6.2 Operation _______________________________________________________________________12
7Cleaning ____________________________________________________________________________13
8Cleaning Configuration_________________________________________________________________14
8.1 M UV254 Analyzer Menu ___________________________________________________________15
9Maintenance _________________________________________________________________________16
10 Parts and Accessories _________________________________________________________________16
11 Cabinet Dimensions and Clearance_______________________________________________________17
12 Warranty ____________________________________________________________________________18
List of Tables
Table 1: Real Clean System I Specifications _____________________________________________________6
Table 2: Items and quantity list for unpacking ____________________________________________________6
Table 3: Commercial cleaning solutions________________________________________________________13
Table 4: Menu functions and configuration for Real Clean System ___________________________________14
Table 5: UV254 M Series menu selection of cleaning configuration __________________________________15
Table 6: Maintenance tasks and frequency for Real Clean system___________________________________16
Table 7: Spare parts for Real Clean system_____________________________________________________16

Real Clean System I v 2.1
Real Tech Inc. 3
List of Illustrations
Figure 5.1: Wash bottle holder ________________________________________________________________7
Figure 5.2: Mounting cabinet _________________________________________________________________7
Figure 5.3: Sample bolt and washer for attachment________________________________________________7
Figure 5.4: Real Clean system ports ___________________________________________________________8
Figure 5.5: Real Clean plumbing with Dual Feed__________________________________________________8
Figure 5.6: M UV254 analyzer and Real Clean plumbing connections _________________________________9
Figure 5.7: M Series analyzer connections _____________________________________________________10
Figure 5.8: Real Clean system communication and power connections _______________________________10
Figure 7.1: Effective cleaning ________________________________________________________________13
Figure 7.2: Inadequate cleaning______________________________________________________________13

Real Clean System I v 2.1
Real Tech Inc. 4
2 Overview
Thank you for purchasing from Real Tech Inc.
The Real Clean System I is an automatic chemical cleaning system for use with
any Real Tech Inc. M2000 or M3000 series analyzers. The purpose of the
system is to reduce fouling of the flow cell within the analyzer.
The Real Clean and M series Analyzer are plumbed so that process water goes through the Real Clean to the
Analyzer. When a cleaning cycle starts, the Real Clean system halts input of process water and pumps
cleaning solution from its wash bottle to the flow cell of the M series analyzer on a timed schedule, or when
needed, immediately through manual intervention.
A portion of cleaning fluid is recaptured and directed back into the wash bottle after a cleaning cycle is
completed. Analyzer measurement resumes after the tubing is cleared of residual cleaning fluid by influx of
process water.
The cleaning system is controlled through the Operator Interface of the Real Tech M series analyzer. From
here, the system is enabled, and a cleaning frequency is selected.
Cleaning of the flow cell is best achieved when cleaning frequency and cleaning fluid work in unison to prevent
the accumulation of fouling agents inside the flow cell.
*Your Serial Number is located on the inside of the cabinet door.
Model Number
UVT-078130
Serial Number*
On schedule,
cleaning fluid drawn
from bottle by
pump.
Cleaning fluid to
flow cell for a
10-minute hold
time.

Real Clean System I v 2.1
Real Tech Inc. 5
3 Safety and Warranty Registration
3.1 Safety Instructions
Please read and follow all safety instructions outlined in this owner’s manual prior to
installation and/ or operation.
1. Visually inspect the device prior to operation. If it seems broken or damaged in any way, do not use.
Contact Real Tech Inc. for a replacement.
2. Ensure that all responsible personnel carefully read this manual before installing or servicing the device.
3. Failure to properly install and maintain this device may impact its effectiveness and warranty.
4. Improper use of this device may cause injury.
3.2 Warranty Registration
Please register within 30 days
To activate your standard two-year warranty for your product, the end user of the product is responsible for
filling out the warranty registration information for their company. Failure to do so could void the warranty. See
the warranty section for full details on Real Tech’s two-year warranty program.
Please register online at: www.realtechwater.com/product-registration.
Thank you for choosing Real Tech!

Real Clean System I v 2.1
Real Tech Inc. 6
4 Technical Specifications
Table 1: Real Clean System I Specifications
CHARACTERISTIC
TECHNICAL DATA
CLEANING FREQUENCY
User configurable cleaning cycle
CLEANING CAPACITY
1L storage container
DIMENSIONS
10”high x 18” wide x 6” deep (25 cm x 46 cm x 15 cm)
ENCLOSURE
Wall mountable
FLUID CONNECTIONS
¼” (6 mm) push-to-connect fittings and flex tubing
ELECTRICAL
Operates from 24 VDC
STORAGE TEMP.
-20°C to +60°C (-4°F to +140°F)
OPERATING TEMP.
0°C to 45°C (32°F to 113°F)
WEIGHT
10 lb (4.5 kg)
WETTED MATERIAL
Polypropylene, Marprene
WARRANTY
2-year limited warranty
5 Installation
5.1 Unpacking and Inspection
To begin, remove the Real Clean system from the packaging and
carefully inspect the product to ensure that no visible damage has
occurred during shipping. The following items will be packed inside
the box:
Table 2: Items and quantity list for unpacking
ITEM
QUANTITY
Real Clean System
1
1L Wash Bottle
1
Wash Bottle Holder
1
Power and Communication Cable
1
Owner’s Manual
1
Please ensure all items are unpacked and accounted for before moving on to assembly.

Real Clean System I v 2.1
Real Tech Inc. 7
5.2 Assembly
The wash bottle holder must be mounted onto the Real Clean I
system from the underside of the enclosure as shown in Figure 5.1.
Fittings are supplied with the holder.
After mounting the enclosure, fill the 1L wash bottle supplied with
cleaning fluid. Recommended fluids are outlined in Section 7 –
Cleaning. Connect the tubing from the system to the spout on the
1L bottle supplied.
5.3 Mounting
The system is enclosed in a wall mountable cabinet.
The cabinet should be attached to the wall at all four
of the mounting holes provided (top left and right;
bottom left and right in Figure 5.2).
Use screws and washers that are sturdy enough to
support the cabinet weight of approximately 10 lbs.
Cabinet dimensions are to the nearest 1/8 inch. Refer
to Section 11, Cabinet Dimensions and Clearance for
drawings of dimensions and clearance.
The expanded view at right (Figure 5.3) shows
an example of a ¼” bolt and washer inserted in
the front side of the cabinet.
The system should be mounted in a convenient
location near the M series instrument to be
cleaned (less than 2 ft.).
Figure 5.1: Wash bottle holder
Figure 5.2: Mounting cabinet
Figure 5.3: Sample bolt and washer
for attachment

Real Clean System I v 2.1
Real Tech Inc. 8
5.4 Plumbing
Process water from a pressurized source flows through the Real Clean system and into the connected
instrument for measurement.
Piping is required to be ¼” OD flex tubing of whatever pressure rating is needed for the particular site. Figure
5.4 displays the two push-to-connect inlet and outlet ports on the side of the Real Clean system.
The Real Clean system is connected in series with the instrument to be cleaned. It is preferred to keep the
tubing between the Real Clean and instrument less than 2 feet. Configurations for the M series instruments are
outlined in Figure 5.6.
Note: If installing with a Dual Feed system, the Real Clean will be installed prior to the Dual feed on the Stream
1 port (See Figure 5.5). Refer also to the Dual Feed manual for further instructions.
Figure 5.4: Real Clean system ports
From Process
To Analyzer
To Stream 1
From Process
To Analyzer
Figure 5.5: Real Clean plumbing with Dual Feed

Real Clean System I v 2.1
Real Tech Inc. 9
5.4.1 Connecting an M UV254 Analyzer
1. Connect ¼” OD flex tubing from a pressurized source to
the IN port on the Real Clean system.
2. Connect ¼” tubing from the OUT port on the Real Clean
to the IN port on the side of the analyzer.
3. Connect ¼” tubing to the OUT port on the analyzer and
divert to drain.
IMPORTANT: The test water must be allowed to flow unrestricted to
drain so that no significant pressure will occur inside the flow cell.
Figure 5.6: M UV254 Analyzer and Real Clean plumbing connections
Water In
Water Out
(To Drain)
Real Clean System

Real Clean System I v 2.1
Real Tech Inc. 10
5.5 Electrical and Communication
The power and communication line from the M Series
analyzer feeds into the Real Clean system via an
electrical conduit fitting at the bottom of the cabinet. The
system is powered and controlled through the
connected M series analyzer.
Connections at the M Series Analyzer are shown in
Figure 5.7.
•White and green wires to terminal 7 and 8 (Clean
Out). Polarity is not important.
•Black wire to terminal 9 (24 VDC Out -).
•Red wire to terminal 10 (24 VCD Out +).
The terminal block connections at the Real Clean
system control board are shown in Figure 5.8.
1. Connect the white and green wires leading from
the M Series analyzer to the right terminal
block. Polarity is not important.
2. Connect the power wires leading from the M
Series analyzer to the left terminal block. The
black power wire connected to the left (-)
terminal and red wire to the right (+) terminal.
Figure 5.8: Real Clean system communication and power
connections
Figure 5.7: M Series Analyzer connections

Real Clean System I v 2.1
Real Tech Inc. 11
6 Start Up and Operation
6.1 Start Up
Once plumbing and electrical connections have been made, you need to enable the cleaning system and enter
settings for cleaning configuration. This is accomplished through the M Series Analyzer’s Operator Interface.
Section 8 gives more detail on menu selections.
From the main menu, press the down arrow button to scroll to the +Cleaning system menu. Access the Main
Menu page by pressing the Menu button. You may need to press the button more than once to have the screen
displayed.
1. To enable the cleaning system, scroll down to ‘Enable Clean’ and press Enter to enable the cleaning
system. The 0 turns to 1, which signifies the item has been selected and therefore, enabled.
Scroll down by
pressing the down
arrow repeatedly.
Press Enter to
expand the
+Cleaning System
menu.

Real Clean System I v 2.1
Real Tech Inc. 12
2. Scroll down from the last entry until the +Clean Frequency menu is displayed. Expand the menu by
pressing the Enter button. Seven selections are available. The default is 4: 3 days. De-select a
frequency by scrolling to it, and then pressing Enter to remove the checkmark. Conversely, select an
option by scrolling the arrow to it, and then press Enter to place a checkmark beside it.
Implementing these two settings is the minimum you need for set up. The system is ready for use. For all other
functions, refer to Section 8, Cleaning Configuration. If the system encounters any problems, associated alarms
or warnings are displayed on the Operator Interface of the M Series Analyzer.
To turn the Real Clean System off, disable the device by removing the checkbox opposite Enable Clean
(reverse of step 1).
6.2 Operation
The system consists of a 3-way stainless solenoid valve, a peristaltic pump, and a 1L chemical cleaner storage
container. The cleaning process is initiated automatically on a preselected time interval or can be manually
triggered.
The cleaning process consists of the following steps:
1. Solenoid shuts off the process water flow to the analyzer flow cell (LED turns on).
2. Solenoid opens flow from peristaltic pump to the analyzer flow cell.
3. Peristaltic pump turns on and pumps cleaning solution into the analyzer flow cell.
4. The pump turns off and the system waits to allow the chemical cleaning solution to remove fouling and
staining present on the walls of the quartz flow cell.
5. The pump turns on in the opposite direction to re-capture the cleaning chemical.
6. Solenoid valve opens the water flow from the process to purge out the cleaning solution (LED turns off).
The process is now complete.
During the cleaning cycle, the measured UVA/UVT value shown on the M series instrument is held to the last
value measured before the cycle started and released after the purge time is complete.

Real Clean System I v 2.1
Real Tech Inc. 13
7 Cleaning
The Real Clean system comes with a 1L chemical storage bottle. After
a cleaning cycle, approximately 85% of the used cleaning fluid is
pumped back into the bottle reducing the need to replenish the fluid
on a regular basis.
Two important factors that affect the sensor performance are the
cleaning fluid used to clean the flow cell, and how often the flow cell
is cleaned (cleaning frequency).
Cleaning Frequency
The optimum frequency of cleaning is quite variable depending on
the amount and type of fouling agents in the water. This will be
dependent on the application and installation point.
For wastewater, a higher frequency will be required (ex. 1-24hrs)
whereas high purity water will require a lower frequency (ex. once a
month). Each installation will be different and require configuration
onsite.
Once in operation, observe the flow cell for signs of fouling prior to
the first cleaning cycle. If fouling is present, increase the cleaning
frequency accordingly. The measured data can also be observed for
signs of a saw tooth effect as shown in Figure 7.1. If the saw tooth
shows an upward trend, as shown in Figure 7.2, the sensor is drifting
and cleaning frequency must be increased.
Cleaning Fluid
Selecting an effective cleaning solution is essential for optimal
performance. The following commercial cleaning solutions and acids
are readily available to remove common fouling agents.
Table 3: Commercial cleaning solutions
SOLUTION
FOULING AGENTS
Lactic Acid <20% solution (CLR®)
Removes lime, calcium, rust, magnesium, and other dissolved minerals
Sulfamic Acid <10% solution (Lime Away®)
Removes lime, calcium, rust, magnesium, and other dissolved minerals
Phosphoric Acid <30% solution (Rust Off,
Hagasen Blue)
Removes lime, calcium, rust and color staining
Citric Acid <20% solution
Removes mineral scaling
Sodium Hypochlorite <6% solution (Not bleach)
Removes oil, grease and biofilm
Acetic Acid <20% solution
Removes oil and grease
Sulfuric Acid <10% solution
Removes oil and grease
Figure 7.1: Effective cleaning
Figure 7.2: Inadequate cleaning

Real Clean System I v 2.1
Real Tech Inc. 14
8 Cleaning Configuration
The Real Clean system is controlled and configured through the M series analyzer. Below is a general
summary of the functions and configurations for the system.
Table 4: Menu functions and configuration for Real Clean System
MENU
DESCRIPTION
-Cleaning System
Enable Clean
Cleaning must be enabled for the Real Clean system to operate.
Reset Clean
When the cleaning fluid is changed or refilled, the user must do a Reset to reset the counter on
the cleaning bottle.
Force Clean
At any time, the user can trigger a cleaning cycle by selecting force clean. This does not
interfere with the normal cleaning schedule.
Pumping Time
Pumping Time allows for configuration of time (in seconds) the cleaning system pumps
cleaning fluid to the sensor. The pumping time should be configured to ensure the cleaning
fluid reaches and fills the quartz flow cell prior to the 10-minute holding period. Longer
distances between the cleaning system and sensor will require a longer pumping time. Default
setting is 120 seconds, suitable for a tube length of 24-36 inches.
-Clean Frequency
Clean Frequency allows the frequency of cleaning to be configured i.e. hour, day, week, or
month. Default is 3 days.
-Alarms
If enabled, alarm conditions are displayed on the main menu.

Real Clean System I v 2.1
Real Tech Inc. 15
8.1 M UV254 Analyzer Menu
Table 5: UV254 M Series menu selection of cleaning configuration
MENU ITEM
DESCRIPTION
CLEANING SYSTEM
Enable Clean
Press Enter then up/down to select 0 or 1 to turn on automatic cleaning system
Reset Clean
Press Enter then up/down to select 0 or 1 to reset fluid alarms after refilling fluid
Force Clean
Press Enter then up/down to select 0 or 1 to force a cleaning cycle
Pumping Time
Press Enter then up/down to select 60, 90 or 120 seconds
CLEAN FREQUENCY
0: No Cleaning
1: 4 hours
2: 12 hours
3: 24 hours
4: 3 days
5: 1 week
6: 2 weeks
7: 1 month
Select
Press Enter then up/down to select cleaning frequency (value 0-7)
ALARMS
Cleaning
Alarm status (0= no alarm 1=alarm)
Clean Fluid Empty
Alarm status (0= no alarm 1=alarm)
Clean Fluid Low
Alarm status (0= no alarm 1=alarm)

Real Clean System I v 2.1
Real Tech Inc. 16
9 Maintenance
Following the manufacturer’s recommended maintenance schedule will ensure optimal performance of the
product. Parts and accessories are outlined in Section 10.
A standard recommended maintenance schedule for the Real Clean system is as follows:
Table 6: Maintenance tasks and frequency for Real Clean system
FREQUENCY
TASK
Site Specific
•Automatic cleaning cycle configuration to ensure proper sensor cleaning
Weekly
•Visual inspection for leaks, corrosion or other issues.
•Inspect cleaning fluid levels. Replenish if necessary.
•View connected analyzer or controller display for configured cleaning alarms.
10 Parts and Accessories
Table 7: Spare parts for Real Clean system
PRODUCT NAME
DESCRIPTION
PART NUMBER
Cleaning Fluid Bottle
1 Litre wash bottle for chemical fluid storage
S-166011
Chemical Pump Tubing
Replacement internal pump tubing for chemical pump
UVT-139013

Real Clean System I v 2.1
Real Tech Inc. 17
11 Cabinet Dimensions and Clearance
The front, side, and back views of the cabinet are shown below to the nearest 1/8 inch. The back view gives the
dimensions relevant for spacing of screws to mount the cabinet. Allow a clearance of 6 inches around the
cabinet for unit connection.

Real Clean System I v 2.1
Real Tech Inc. 18
12 Warranty
Real Tech Inc. (Real Tech) warrants to the original
purchaser of this instrument that it will be free of
defects in material and workmanship, in normal use
and service, for a period of two years from date of
delivery to the original purchaser. Real Tech's
obligation under this warranty is limited to replacing
or repairing the instrument, at Real Tech’s option, at
Real Tech's factory. Excluded from this limited
warranty are accessories and supplies.
The original purchaser is responsible for return of
the instruments, or parts thereof, to Real Tech's
factory. This includes all freight and insurance
charges incurred in shipping to and from Real
Tech's factory.
Real Tech is not responsible for any personal injury
or damage to the instrument, or parts thereof,
resulting from any accident, improper use of the
instrument, negligence or willful neglect on the part
of the user of the instrument. Real Tech is not
responsible for any personal injury or defects to the
instrument resulting from repairs or alterations
made by any person not authorized by Real Tech.
Real Tech assumes no liability for indirect or
consequential damage, losses, injury or costs of
any kind, and the original purchaser, by placement
of any order for the instrument, or parts thereof,
shall be deemed liable for any and all damages
incurred by the use or misuse of the instruments, or
parts thereof, by the purchaser, its employees, or
others, following receipt thereof.
Carefully inspect this product for shipping damage, if
damaged, immediately notify the shipping company
and arrange an on-site inspection. Real Tech cannot
be responsible for damage in shipment and cannot
assist with claims without an on-site inspection of the
damage.
This warranty is given in lieu of and excludes all other
warranties, conditions and guarantees, expressed or
implied, statutory or otherwise including those of
merchantability and fitness for use. No agent or
representative is authorized to assume any liability
for Real Tech.
For further assistance or service please contact:
Real Tech Inc.
1150 Champlain Court
Whitby ON L1N 6K9 Canada
Tel: 1.877.779-2888
Tel: 905.665.6888
Fax: 905.665.7025
Email: service@realtechwater.com
Website: www.realtechwater.com
Table of contents
Popular Cleaning Equipment manuals by other brands

Dibo
Dibo JMB-M 500/30 manual

Super Products
Super Products CAMEL MAX Series Operator's manual

Menzer
Menzer ESM 406 Floor Tec Translation of the original operating instructions

Kärcher
Kärcher K 7 Premium operating instructions

Nilfisk-Advance
Nilfisk-Advance SC3000 Instructions for use

Zoef Robot
Zoef Robot Evert RB80Z user manual