manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Renishaw
  6. •
  7. Measuring Instrument
  8. •
  9. Renishaw P-GA01-0001 User manual

Renishaw P-GA01-0001 User manual

Other Renishaw Measuring Instrument manuals

Renishaw RENGAGE OMP600 User manual

Renishaw

Renishaw RENGAGE OMP600 User manual

Renishaw RMP600 User manual

Renishaw

Renishaw RMP600 User manual

Renishaw VM11 User manual

Renishaw

Renishaw VM11 User manual

Renishaw RMP40 User manual

Renishaw

Renishaw RMP40 User manual

Renishaw MIH User manual

Renishaw

Renishaw MIH User manual

Renishaw XK10 User manual

Renishaw

Renishaw XK10 User manual

Renishaw RMP40 User manual

Renishaw

Renishaw RMP40 User manual

Renishaw Primo Radio Part Setter User manual

Renishaw

Renishaw Primo Radio Part Setter User manual

Renishaw Primo User manual

Renishaw

Renishaw Primo User manual

Renishaw RTS User manual

Renishaw

Renishaw RTS User manual

Renishaw RMP40 User manual

Renishaw

Renishaw RMP40 User manual

Renishaw REVO User manual

Renishaw

Renishaw REVO User manual

Renishaw OMM-S SPRINT User manual

Renishaw

Renishaw OMM-S SPRINT User manual

Renishaw OMP400 User manual

Renishaw

Renishaw OMP400 User manual

Renishaw PH10M-iQ PLUS User manual

Renishaw

Renishaw PH10M-iQ PLUS User manual

Renishaw LP2 User manual

Renishaw

Renishaw LP2 User manual

Renishaw TP7 Series User manual

Renishaw

Renishaw TP7 Series User manual

Renishaw MP7 User manual

Renishaw

Renishaw MP7 User manual

Renishaw TP1 User manual

Renishaw

Renishaw TP1 User manual

Renishaw PH10 PLUS Series User manual

Renishaw

Renishaw PH10 PLUS Series User manual

Renishaw LP2 User manual

Renishaw

Renishaw LP2 User manual

Renishaw RMP600 User manual

Renishaw

Renishaw RMP600 User manual

Renishaw SCP600 User manual

Renishaw

Renishaw SCP600 User manual

Renishaw RVP User manual

Renishaw

Renishaw RVP User manual

Popular Measuring Instrument manuals by other brands

Powerfix Profi 278296 Operation and safety notes

Powerfix Profi

Powerfix Profi 278296 Operation and safety notes

Test Equipment Depot GVT-427B user manual

Test Equipment Depot

Test Equipment Depot GVT-427B user manual

Fieldpiece ACH Operator's manual

Fieldpiece

Fieldpiece ACH Operator's manual

FLYSURFER VIRON3 user manual

FLYSURFER

FLYSURFER VIRON3 user manual

GMW TG uni 1 operating manual

GMW

GMW TG uni 1 operating manual

Downeaster Wind & Weather Medallion Series instruction manual

Downeaster

Downeaster Wind & Weather Medallion Series instruction manual

Hanna Instruments HI96725C instruction manual

Hanna Instruments

Hanna Instruments HI96725C instruction manual

Nokeval KMR260 quick guide

Nokeval

Nokeval KMR260 quick guide

HOKUYO AUTOMATIC UBG-05LN instruction manual

HOKUYO AUTOMATIC

HOKUYO AUTOMATIC UBG-05LN instruction manual

Fluke 96000 Series Operator's manual

Fluke

Fluke 96000 Series Operator's manual

Test Products International SP565 user manual

Test Products International

Test Products International SP565 user manual

General Sleep Zmachine Insight+ DT-200 Service manual

General Sleep

General Sleep Zmachine Insight+ DT-200 Service manual

Sensa Core Lacto Spark user manual

Sensa Core

Sensa Core Lacto Spark user manual

VOLTCRAFT UTS-1980 operating instructions

VOLTCRAFT

VOLTCRAFT UTS-1980 operating instructions

EBCHQ 94915 Operation manual

EBCHQ

EBCHQ 94915 Operation manual

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Balluff

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Pulsar IMP Lite Series instruction manual

Pulsar

Pulsar IMP Lite Series instruction manual

ETCR 6800 user manual

ETCR

ETCR 6800 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Leaflet
H-1000-4054-02-B
Trigger force gauge guide
Specification
Overall length 95 mm
Needle length 41 mm
Depth 27 mm
Width 43 mm
Range 4 g to 35 g range with 1 g graduations
Renishaw part no. P-GA01-0001
All Renishaw’s touch-trigger probes have an optimum trigger force
setting for general purpose applications as detailed in table below:
Trigger force
CMM
probes
Stylus length
(typical)
Optimum trigger force
(preset by Renishaw)
Trigger force
range
TP1 (S) 31 mm
(PS1-1R) 0.15 N (15 gf) 0.1 N to 0.5 N
(10 gf to 50 gf)
TP2-5 way 10 mm
(PS1-2R)
0.07 N to 0.08 N
(7 gf to 8 gf)
0.07 N to 0.15 N
(7 gf to 15 gf)
TP6 / TP6A 21 mm
(PS1-12R)
0.11 N to 0.13 N
(11 gf to 13 gf)
0.11 N to 0.3 N
(11 gf to 30 gf)
Trigger force
Trigger force is the amount of pressure applied by the helical
compression spring onto the pivotal plate and bearing points to hold
the stylus mount in place.
The trigger force is preset by Renishaw but can be altered for any of
the following reasons:
• To permit the use of a longer styli on the probe.
• To permit the use of heavier styli on the probe.
• If the preset trigger force has decreased due to probe use.
• If the acceleration of the CMM is causing illegal triggers.
NOTE:
Changing the probe’s trigger force will affect it’s measurement
performance. It is important to requalify the probe configuration and
check the measurement performance of the probe system after any
adjustment has been made to the trigger force.
Checking trigger force with a Renishaw gram
gauge
1. Ensure that the probe is held firmly in position (preferably on a
CMM) and connected to an interface to detect a probe trigger.
2. Establish the direction of trigger which gives the maximum
resistance. The most practical way of doing this is to gently
deflect the stylus with a finger, trying different directions (see
sketch below). There are three lobes which produce three
maximum and three minimum force directions. These can be
easily found with a minimum of practice.
Australia
T+61 3 9521 0922
Eaustralia@renishaw.com
Austria
T+43 (0) 2236 379790
Eaustria@renishaw.com
Brazil
T+55 11 4195 2866
Ebrazil@renishaw.com
The People’s Republic of China
T+86 21 6353 4897
Eshanghai@renishaw.com
Canada
T+1 905 828 0104
Ecanada@renishaw.com
Czech Republic
T+420 548 216 553
Eczech@renishaw.com
France
T+33 1 64 61 84 84
Efrance@renishaw.com
Germany
T+49 (0) 7127 981-0
Egermany@renishaw.com
Hong Kong
T+852 2753 0638
Ehongkong@renishaw.com
Hungary
T+36 23 502 183
Ehungary@renishaw.com
India
T+91 80 25320 144
Eindia@renishaw.com
Israel
T+972 4 953 6595
Eisrael@renishaw.com
Italy
T+39 011 966 10 52
Eitaly@renishaw.com
Japan
T+81 3 5366 5315 (or 5324)
Ejapan@renishaw.com
The Netherlands
T+31 76 543 11 00
Ebenelux@renishaw.com
Poland
T+48 22 577 1180
Epoland@renishaw.com
Russia
T+7 095 231 1677
Erussia@renishaw.com
Singapore
T+65 6897 5466
Esingapore@renishaw.com
Slovenia
T+386 (1) 527 2100
E[email protected]
South Korea
T+82 2 2108 2830
Esouthkorea@renishaw.com
Spain
T+34 93 663 3420
Espain@renishaw.com
Sweden
T+46 (0)8 584 90 880
Esweden@renishaw.com
Switzerland
T+41 55 415 50 60
Eswitzerland@renishaw.com
Taiwan
T+886 4 2251 3665
Etaiwan@renishaw.com
UK (Head Office)
T+44 (0)1453 524524
Euk@renishaw.com
USA
T+1 847 286 9953
Eusa@renishaw.com
For all other countries
T+44 1453 524524
Einternational@renishaw.com
Renishaw worldwide
© 2001 - 2006 Renishaw Issued 0106 Part no. H-1000-4054-02-B
3. Set the gram gauge maximum force indicator to the zero point
and place flat on the CMM table. Move the stylus tip to the same
height above the surface as the gram gauge lever.
4. Slide the gram gauge slowly sideways so that the flat point on
the end of the lever touches the probe stylus ball (ensuring that
the probe is deflected in the low trigger force direction). Very
slowly continue to move the gram gauge until the probe triggers,
at which point stop immediately, back off the gram gauge and
read the maximum force indicator.
5. Repeat this procedure three or four times to ensure consistent
results.
Measuring
trigger force
NOTE:
Use the following formula to
convert gf to Newtons:
Newtons = gf / 100
Trigger force adjustment
The trigger force of a TP1 (S) / TP2-5W / TP6 / TP6A probe is preset
by Renishaw at an optimal performance setting, but can be altered if
necessary as follows:
1. Remove the probe from the quill of your CMM.
2. Insert the correct size hexagonal key (supplied with every probe)
into the centre of the shank until you locate a grub screw.
3. Adjust this grub screw to alter the trigger force of the probe:
- Clockwise increases the trigger force.
- Anticlockwise decreases the trigger force.
Establishing direction
of minimum resistance
A = High trigger force direction
B = Low trigger force direction
A
B