Incufridge RS-IF-233
Troubleshooting
Symptom Fault Action
Temperature on the display
is higher than the actual
Incufridge may be in need
of calibration
See calibration instructions
in the manual
Temperature on the display
is lower than the actual
Incufridge may be in need
of calibration
See calibration instructions
in the manual
Power cord may be dis-
connected
Inspect all power supply
conditions
Display screen does not
turn on
Power switch may be off Inspect power switch in
back
Fan buzzes during
operation
Something may be caught
in the fan
Turn off Incufridge and
inspect internal and external
fans. Carefully remove
foreign objects if any are
found
Temperature stays at
ambient and does not heat
or cool
Possible defective
components
Contact Revolutionary
Science
Fan is not operational Possible defective
components
Contact Revolutionary
Science
If symptoms persist, please contact Revolutionary Science Support Center at
Table of Contents
Introduction 2
Recommended Safeguards 3
Operating Instructions 4
Care and Maintenance 5
Calibration 5
Specifications 6
Contact Us 6
Registration 7
Warranty 7
Warranty Registration Card 8
Checklist 9
Notes and Service Record 9
Troubleshooting 10
NOTE: This manual contains important information, safeguards and operating
Instructions for the operation of the REVOLUTIONARY SCIENCE model RS-IF-233
Incufridge by Alternative Pioneering Research & Development, Inc. (APR&D, Inc.)
© Alternative Pioneering Research and Development, Inc.
Instruction Manual for the
Revolutionary Science RS-IF-233
Incufridge
R E VO L U TI O N A RY S C IE N C E
Man ufa cturer of Preci sio n Labora tor y Equ ipmen t
17319 Lake Blvd.
Shafer, MN 55074
Toll Free: (800) 775-7942
Fax: (775) 218-7015
R E V O LU T I O N A R Y S C I E N C E
E-mail: [email protected]
Website: www.revsci.com
1

R E V O L U T I O N A R Y S C I E N C E
Incufridge RS-IF-233
Introduction
Thank you for purchasing the Revolutionary Science RS-IF-233 Incufridge. We are
confident that when used correctly, our Incufridge will provide years of reliable
service in your lab, classroom, or fieldwork. We have made every effort to use high
quality, durable parts to deliver the performance expected of instruments costing 3 to
4 times as much.
Please let us know what you think of our products, positive or negative, so we can
continue to enhance and improve our products in the future.
Sincerely yours,
REVOLUTIONARY SCIENCE
2
R E V O L U T I O N A R Y S C I E N C E
Incufridge RS-IF-233
Checklist
Before packaging material is discarded please make sure all parts are accounted for:
__ 2 Shelves
__ Incufridge
Notes, Service Record, etc.
______________________________________________________
______________________________________________________
______________________________________________________
______________________________________________________
______________________________________________________
______________________________________________________
______________________________________________________
______________________________________________________
______________________________________________________
______________________________________________________
______________________________________________________
______________________________________________________
______________________________________________________
______________________________________________________
______________________________________________________
______________________________________________________
______________________________________________________
______________________________________________________
______________________________________________________
______________________________________________________
______________________________________________________
______________________________________________________
______________________________________________________
______________________________________________________
9

R E V O L U T I O N A R Y S C I E N C E
Incufridge RS-IF-233
8
Please fill out this page, or a copy of it, and send it to the address below to validate
your warranty.
--------------------------------------------------------------------------------
Warranty Registration Card for the
REVOLUTIONARY SCIENCE RS-IF-233 Incufridge
(Please Print)
Date Purchased:_________________ Serial Number:______________________
Purchased From:___________________________________________________
Name of Purchaser:________________________________________________
• Company • School • Individual (Please check one)
Name of institution:________________________________________________
Department:______________________________________________________
Address:_________________________________________________________
_______________________________________________________________
Intended use:_____________________________________________________
Where did you hear of our product?:____________________________________
Comments:______________________________________________________
_______________________________________________________________
Mail to: Revolutionary Science
17319 Lake Blvd., Shafer, MN 55074
Incufridge RS-IF-233
R E V O L U T I O N A R Y S C I E N C E
3
To maximize the use of this Incufridge, please complete this short list before using the
RS-IF-233:
1. Examine carton and packaging for possible damage caused during shipping. If
damage is present contact carrier at once. Failure to do so immediately may release
the carrier from any liability of damage.
2. Fill out the checklist (page 9 of instruction manual). If something is missing please
contact Revolutionary Science.
3. Carefully read the instruction manual.
4. Fill out warranty registration card and mail it to Revolutionary Science.
Recommended Safeguards
1. Never attempt to disassemble or modify the RS-IF-233.
2. Do not clean any parts with solvents.
3. Do not drop.
4. Keep door closed for effective temperature stabilization.
5. Do not operate on wet surfaces.
6. Make sure the power cord is clear of hot surfaces and other hazards. We
recommend any slack in the cord to be wrapped up and held together with a twisty
tie.
7. Avoid spills of any kind on or in the unit. This may contaminate or corrode the
surfaces.
8. Do not use around flammables, combustible materials, or explosives.
9. Keep back vents clear (at least two inches in every direction) from other
equipment and avoid setting unit on unstable surfaces. This will allow for effective air
convection.
10. Do not heat any substances that may release fumes considered to be toxic.
11. Avoid placing samples directly on floor of chamber, instead use shelf to insure
even distribution of temperature.
12. Like all heating and cooling equipment their efficiency depends upon the external
or ambient temperature. This product has a maximum efficiency at an ambient
temperature of 22°C, provided that all other safeguards are met. If the ambient
temperature increases or decreases, the maximum efficiency will decrease.

R E V O L U T I O N A R Y S C I E N C E
Incufridge RS-IF-233
Recommended Room and Storage Conditions:
· Do not store your REVOLUTIONARY SCIENCE RS-IF-233 in direct sunlight, or in
dusty or poorly ventilated areas.
· Do not store in a cold or moist room for extended periods of time.
· Never immerse your REVOLUTIONARY SCIENCE RS-IF-233 in water.
Operating Instructions
(1) PLACEMENT:
Place unit on a stable, hard surface away from the edge of the counter and other
moving equipment.
(2) LOADING:
Open the Incufridge by gently pulling outward (away from the unit) on the door
handle. With the door open and shelves properly inserted, load your samples leaving
adequate space between samples for proper air circulation. Avoid placing samples
directly on floor of chamber, instead use shelf to insure even distribution of
temperature.
(3) CONNECTING THE POWER:
120-240V use: Insert power cord into an outlet designed for 120-240v use. Make
sure the power cord is clear of hot surfaces and other hazards. Any slack in the cord
should be wrapped up and held together with a twisty tie.
(4) SET THE TEMPERATURE:
After power is connected, flip power switch on backside of unit to On position.
Default display is actual chamber temperature. To increase chamber temperature,
press and hold the arrow button pointed up for 2+ seconds. To decrease chamber
temperature, press and hold the arrow button pointed down for 2+ seconds. To
display current set temperature, press either button.
4
R E V O L U T I O N A R Y S C I E N C E
Incufridge RS-IF-233
Registration
For your future reference and to expedite answers to your questions, please
complete the following information:
Date Purchased:___________________________________________________
Serial Number:____________________________________________________
Model Number:___________________________________________________
Purchased From:___________________________________________________
Full Two Year Warranty
Your REVOLUTIONARY SCIENCE RS-IF-233 is warranted to be free from defects of
material or workmanship under normal use for a period of two years from date of
purchase. If your RS-IF-233 proves to be defective within this warranty period, Rev-
olutionary Science will repair or replace your unit at our option. This warranty will
be void if malfunction is caused by accident, misuse, or negligence, including tamper-
ing, abuse, or damage in transit. This warranty excludes any relief for incidental
or consequential damages. Some states do not allow limitations on how long an im-
plied warranty lasts or the exclusion or limitation of incidental or consequential
damages, so the above limitation may not apply to you. This warranty gives you spe-
cific legal rights, you may also have other rights, which may vary from state to state.
Your RS-IF-233 comes with several built in safety features. Any attempt to inter-
fere with the operation of these features makes this warranty null and void.
To validate this warranty, be sure to complete and return the Warranty Registration
Card, found on page 8.
7

R E V O L U T I O N A R Y S C I E N C E
Incufridge RS-IF-233
6
Incufridge Specifications
Model No.: RS-IF-233basic
Precision: ±0.7°C (at 20°C ambient)
Uniformity: Internal temperature uniformity ±1.0°C (at 20°
C ambient)
Display: 3 digits, 0.1°C resolution
Controller: Digital microcomputer
Power Requirements: 120-240VAC, 50-60 Hz, 150 W, 12 VDC
Net Weight: 34 lb.
Outside Dimensions: 15" wide x 23” deep x 15.5" tall
Inside Dimensions: 12" wide x 16.5" deep x 10.25” tall
Capacity: 33L (1.2 cu. ft.)
Temperature Range: Accurately holds from 7°C to 60°C (at 20°C
ambient)
Warranty: 2 year, repair or replacement
Contact Us
Mailing address: Toll Free: (800) 775-7942
Revolutionary Science Fax: (775) 218-7015
Shafer, MN 55074 USA Web site: www.revsci.com
R E V O L U T I O N A R Y S C I E N C E
Incufridge RS-IF-233
5
Care and Maintenance
Your Incufridge should be cleaned periodically to ensure optimal performance, clean-
liness, and to eliminate contamination. Clean your REVOLUTIONARY SCIENCE
RS-IF-233 only when it is not plugged in. Using a soft cloth soaked in water and
bleach or mild detergent, wipe the inside and outside of the unit. Then let air dry or
wipe with a dry cloth. Be careful not to scratch surfaces with abrasive cleaners or
dirty cloths.
User Calibration Instructions
Note: To set temperature, follow instructions on page 4.
Setup:
Set desired temperature on Incufridge and insert calibrated thermocouple or
thermometer closest to the geometric center of the chamber. Close door and check
accuracy of temperature 1 to 2 hours later. If temperature is off by more than 0.5°C,
record Incufridge display reading and thermometer reading and calibrate accordingly,
following these three simple steps.
1. Press up and down buttons simultaneously. Display should read “CAL.”
2. After "CAL" times out a new number will display. This is the calibration off-set
number. Press up arrow if you desire to increase the temperature on the display
reading or down arrow if you wish to decrease the temperature on the display.
3. Wait for three seconds for the display to return to the temperature reading.
NOTE: Incufridge will maintain adjusted calibration if it is turned off or restarted.
Calibration may be adjusted anytime.
Temperature Uniformity:
Geometric extremes (i.e. within an inch or two from the inside of the door, ceiling,
fan, walls etc) may be less precise.
For best results, it is recommended to place specimens and samples on the trays, and
if possible, close to the geometric center.
Other Revolutionary Science Laboratory Equipment manuals

Revolutionary Science
Revolutionary Science incufridge professional series User manual

Revolutionary Science
Revolutionary Science SANICLAVE FRONT-LOADER Series User manual

Revolutionary Science
Revolutionary Science Saniclave RS-SC-102 User manual

Revolutionary Science
Revolutionary Science saniclave 200 RS-SC-200 User manual
Popular Laboratory Equipment manuals by other brands

Applied Photonics
Applied Photonics LIBSCAN 25+ user manual

Showa Denko
Showa Denko Shodex ORpak CDBS-453 Operation manual

Kinematics
Kinematics 4400/PVC operating instructions

Stereotaxis
Stereotaxis Niobe PM3.1 Installation Verification and Testing

PHYWE
PHYWE 11207-30 quick start guide

Agilent Technologies
Agilent Technologies XY Operator's manual