Rion UV-15 User manual

3-20-41 Higashimotomachi, Kokubunji, Tokyo 185-8533, Japan
http://www.rion.co.jp/english/
INSTRUCTION MANUAL
Vibration Meter Unit
UV-15


i
Organization of this manual
This manual describes the features, operation, and other aspects of the Vibration
Meter Unit UV-15. If the unit is used together with other equipment to congure a
measurement system, consult the documentation of all other components as well.
The section starting on page iii contains important information about safety. Be
sure to read and observe these precautions in full.
This manual contains the following sections.
Outline
Gives basic information about the unit and contains a block diagram.
Names of Parts and Functions
Briey identies and explains the front panel controls, rear panel connec-
tors and all other parts of the unit.
Linking
Explains how to congure multiple units as a system.
Power Supply Connection
Explains how to make connections for powering the unit.
Accelerometer Selection and Connection
Explains the types of accelerometers that can be used, how to connect them,
and how to set them up for measurement.
Output Connections
Explains how to use the output connectors on the rear panel.
Operation Modes
Explains the operation modes necessary for measurement, including input
mode, measurement mode, and calibration mode (Output CAL).

ii
Performance Characteristics
Contains charts that show the characteristics of the high-pass lter and low-
pass lter as well as velocity (VEL) and displacement (DISP) frequency
response characteristics.
Specications
Lists the technical specications of the unit.
Reference Material
Provides an explanation of output signal delay. Maintenance parts are also
listed in this section.
* Company names and product names mentioned in this manual are
usually trademarks or registered trademarks of their respective owners.

iii
FOR SAFETY
In this manual, important safety instructions are specially marked as shown
below. To prevent the risk of death or injury to persons and severe damage
to the unit or peripheral equipment, make sure that all instructions are fully
understood and observed.
Caution
WARNING
Important
Disregarding instructions
printed here incurs the risk of
death or severe injury to per-
sons.
Disregarding instructions
printed here incurs the risk of
injury to persons and/or dam-
age to peripheral equipment.
Disregarding instructions
printed here incurs the risk of
damage to the product.
Mentioned about the tips to
use this unit properly. (This
messages do not have to do
with safety.)
Note

iv
To conform to the EU requirement of the Directive
on Waste Electrical and Electronic Equipment,
the symbol mark on the right is shown on the
instrument.
The product described in this manual is in conformity with the
following standards;
Electrical equipment for measurement control and laboratory use.
RION Co., Ltd.
3-20-41 Higashimotomachi, Kokubunji, Tokyo 185-8533, Japan
AB
AC adapter
NC-99 series
Core
Core A: ZCAT2749-0430 (TDK)
Core B~H: ZCAT2436-1330 (TDK)
Equivalent product
C
EG
H
Core
DF
This product may cause interference if used in residential areas.
Such use must be avoided unless the user takes special measures
to reduce electromagnetic emissions to prevent interference to
the reception of radio and television broadcasts.

v
Precautions
•Operate the unit only as described in this manual.
• Do not disassemble the unit or attempt internal alterations.
• Observe the following precautions before using the unit:
- Make sure that all connections are properly and safely established.
- Make sure that the unit is operating normally.
• The permissible ambient temperature range for operation of the unit is
‑10°C to +50°C. Relative humidity must be 90% or below.
• Do not use or store the unit in locations which
- may be subject to strong magnetic elds or strong radiation, or
- may be subject to high levels of dust or splashes of water, or
- may be subject to gases or air with high salt or sulphur content, or
are in the vicinity of stored chemicals, or
- may be subject to high temperature, humidity, or to direct sunlight,
or
- may be subject vibrations or shock.
• Always switch off the power after using the unit.
• When disconnecting cables, always hold the plug or connector and do not
pull the cable.
• Use only the specied AC adapter or other specied power source.
• This is a precision device. Take care not to drop the unit and protect it
from shocks.
• The LCD panel of the unit can easily become scratched. Do not tap the
panel with a pointed object such as a pencil, screwdriver, etc.
• In case of malfunction, do not attempt any repairs. Note the condition of
the unit clearly and contact the supplier.
• When disposing of the unit, follow national and local regulations regard-
ing waste disposal.

vi
Contents
FOR SAFETY............................................................................... iii
Precautions......................................................................................v
Outline............................................................................................1
Names of Parts and Functions.........................................................4
Front panel.................................................................................4
Rear panel..................................................................................6
LCD panel .................................................................................7
Operation panel .........................................................................9
Top panel .................................................................................11
Bottom panel ...........................................................................11
Linking .........................................................................................12
Linking procedure ...................................................................12
Removing the cable from the link connector............................ 17
Linking with Battery Unit BP-17 .............................................18
Linking with Interface Unit UV-22..........................................19
Using a single UV-15 unit ........................................................20
Power Supply Connection..............................................................21
Using an AC adapter................................................................22
Using the Battery Unit BP-17...................................................23
Accelerometer Selection and Connection ......................................25
Accelerometer connection........................................................26
INPUT (PREAMP) connector wiring diagram...................27
Accelerometer mounting..........................................................28
Output Connections.......................................................................30

vii
Operation Modes...........................................................................31
Input setup mode .....................................................................31
TEDS communication mode....................................................34
Setting the ID number .............................................................35
Measurement setup mode.........................................................36
Calibration mode (OUTPUT CAL) ..........................................38
Measurement mode..................................................................39
Measurement range ............................................................41
Frequency range .................................................................41
Measurement upper limit.................................................... 41
Measurement lower limit....................................................42
Check mode.............................................................................45
Using the Interface Unit UV-22 (option) ..................................46
Restoring the factory default settings.......................................48
Performance Characteristics..........................................................49
Specications ................................................................................51
Reference Material ........................................................................58
Delay of output signal..............................................................58
User Filter................................................................................59
Maintenance parts ...................................................................61

viii

1
Outline
The UV-15 is a single-channel vibration meter unit with a wide measurement
range. It supports connection of piezoelectric accelerometers, accelerom-
eters with integrated preamplier (24 V DC, 4 mA power supplied), TEDS
compliant accelerometers with integrated preamplier, and accelerometers
connected via the Preamplier VP-26A. The chassis of up to 16 units can be
linked to create a multi-channel conguration.
Two types of input connectors are provided (Microdot and 7-pin), and the
unit is equipped with a backlight LCD panel, alarm LEDs, operation keys,
AC and DC output connectors (BNC), DC IN jack, and link connector.
The following accessories are available as options, allowing the conguration
of advanced measurement systems suited to various requirements.
Sound level Meter Unit UN-14
Single-channel sound level meter unit designed for connection of a
measurement preamplier microphone.
In combination with the UV-15, a multi-channel sound and vibration
level measurement system can be congured.
AC adaptor NC-99 series (CE mark), NC-97 series
AC adaptor supplies the UV-15 with the DC power.
It also can supply linked multiple unit with the DC power.
Battery Unit BP-17
Designed to hold eight IEC R14 (size C) batteries, this unit can be
linked to the UN-14 and/or UV-15 to provide power.
The power switch on the front panel of the BP-17 can be used to switch
the entire system on and off.

2
Outline
Interface Unit UV-22
Linking the UV-22 allows connection to a computer for controlling
the sound and vibration level measurement system and for transfer
of measurement data.
Both USB and Ethernet connections are supported.

3
Outline
UV-15 Block Diagram
DC
9 V to 15
V
±
12
V
+3.3
V
+5
V
+1.8
V
+24
V
CCLD
24
V 4
mA
TEDS
AMP
-20
dB
AMP
0
dB
SENSE AMP
1
SENSE AMP
2
SENSE AMP
3
SENSE AMP
4
AMP
1
AMP
2
AMP
3
AMP
4
AMP
5
DSP (high-pass lter, low-pass lter,
effective value processing etc.)
A/D D/A
AMP
6
AMP
7
EEPROM
AMP
-20
dB
AMP
0
dB
CPU
AC
OUT
DC
OUT
OVER
OVER
ACC (m/s2)
VEL (mm/s)
DISP (mm)
Power
supply
Microdot
connector Charge amplier 0 dB
Charge amplier -20 dB Integrator
Integrator
7-pin connector
Preamplier power supply ±12 V
LCD
Battery unit
OVERLED I/O

4
Names of Parts and Functions
Front panel
Sub LED
Lights up in green when a control command from the Interface Unit
UV-22 has been received normally and when the Master/Slave function
is used.
Overload indicator LED
Lights up in red when saturation occurs in the internal signal processing
circuitry. In such a case, correct measurement is not possible.
PICKUP connector
Serves for connection of the accelerometer, accelerometer with integrated
preamplier, or TEDS compliant accelerometer with integrated preamplier.
PICKUP
VIBRATION METER UNIT
UV-15
CAL
DISPLAY
INPUT
MEAS
RANGE
RMS
EQ
PEAK
EQ P-P
SETUP
PICKUP connector
Sub LED Overload indicator LED
LCD panel
Operation panel

5
Names of Parts and Functions
Operation panel
Control keys for measurement mode (range selection, display switching
etc.) are located here.
LCD panel
Shows setup information, a bar graph display, numeric display etc.
When the unit is powered from batteries using the Battery Unit BP-17,
the display backlight is activated for 10 seconds when any key is pressed.
When the unit is powered from an AC adapter, the display backlight is
always on.

6
Names of Parts and Functions
Rear panel
INPUT (PREAMP) connector
This connector is designed for connection of a piezoelectric accelerometer
via the Preamplier VP-26A.
AC OUT connector
Supplies an AC output signal equivalent to the input signal.
DC OUT connector
Supplies a DC output signal equivalent to the input signal.
Name plate
Shows the serial number and other information about the unit.
DC IN jack
Serves for connection of the optional AC adapter or an external DC
source.
DC IN
9-15V
+
15
No. 12345678
RION CO.,LTD.
MADE IN JAPAN
AC OUT
DC OUT
INPUT
(PREAMP)
DC IN jack
INPUT
(PREAMP) connector
AC OUT connector
DC OUT connector
Name plate

7
Names of Parts and Functions
LCD panel
ID number indication
Serves to identify the unit in a multi-channel conguration. The number
display range is 1 to 16.
Input indicators
The input selection is shown here.
• FRONT CHARGE: Piezoelectric accelerometer
• FRONT CCLD:
Accelerometer with integrated preamplier
• FRONT CCLD TEDS: TEDS compliant accelerometer with inte-
grated preamplier
• REAR: Preamplier VP-26A connected
Calibration (CAL) indicator
This indicator blinks when you press the CAL key. The AC OUT and DC
OUT connectors on the rear panel supply a calibration signal for calibra-
tion of external equipment.
Range/sensitivity (RANGE/SENS) indicators
When the numeric display shows the current range, the RANGE indicator is
on. When the readout shows the sensitivity, the SENS indicator is on. When
the numeric display shows the measurement value, both indicators are off.
REMOTE
FRONT
REAR
ACOUST
CAL
m/s
2
mm/s
mm
F
S
10ms
CHARGE
CCLD
T
EDS
No.
RANGE SENS
dB
UN
FS
50%
10%
HPF
LPF
Hz
kHz
Numeric display
ID number indication
Input indicators
Calibration indicator
Range/sensitivity
indicators
Unit indicators
Arrow indicators
LCD panel
LPF indicator
HPF indicator
Bar graph

8
Names of Parts and Functions
Unit indicators
Show the measurement unit for the currently selected measurement
mode.
• Acceleration (ACC): m/s2
• Velocity (VEL): mm/s
• Displacement (DISP): mm
During sensitivity indication, no unit is shown.
Arrow indicators
Below these arrows are labels for RMS, EQ PEAK, and EQ P-P indica-
tion characteristics. The arrow for the currently selected characteristic
is visible.
LPF indicator
Shows the selected low-pass lter setting.
HPF indicator
Shows the selected high-pass lter setting.
Bar graph
Shows the level of the signal as a bar graph indication.
Numeric display
Shows the measurement value, range, sensitivity, and error information.
When the range is changed during measurement value display, the new
range setting is shown for about 10 seconds.

9
Names of Parts and Functions
Operation panel
MEAS key
Serves to select the following modes.
Measurement mode setting
Acceleration (ACC): m/s2
Velocit y ( VEL): m m /s
Displacement (DISP): mm
Display characteristics setting
RMS
EQ PEAK
EQ P-P
HPF (high-pass lter) setting
LPF (low-pass lter) setting
RANGE keys
During measurement, these keys serve to select the range. In the setup
mode, the keys serve to change a setting.
PICKUP
VIBRATION METER UNIT
UV-15
CAL
DISPLAY
INPUT
MEAS
RANGE
RMS
EQ
PEAK
EQ P-P
SETUP
RMS, EQ PEAK,
EQ P-P labels
MEAS key
RANGE keys
PICKUP connector
(See page 4)
DISPLAY key
CAL key
INPUT key

10
Names of Parts and Functions
DISPLAY key
Serves to switch the numeric display between range indication and mea-
surement value indication.
CAL key
Pressing this key causes the AC OUT and DC OUT connectors on the
rear panel to supply a calibration signal for calibration of external equip-
ment.
INPUT key
Serves to make the following settings.
Input selection
: FRONT CHARGE/FRONT CCLD/FRONT CCLD TEDS/
REAR
Sensitivity selection:
0.100 to 0.999, 1.00 to 9.99, 10.0 to 99.9
ID number: 1 to 16
RMS, EQ PEAK, EQ P-P labels
The arrow indication above the display characteristics selected with the
MEAS key is visible.
Table of contents
Other Rion Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Leister
Leister EXAMO 300F USB operating instructions

WEL
WEL Data Scribe quick start guide

Katronic Technologies
Katronic Technologies KATflow 200 operating instructions

Onicon
Onicon System-10 BTU Meter Installation and operation guide

ENDA
ENDA EI1410 NTC quick start guide

Omega
Omega DPG1000ADA-30V100 instruction sheet