PAGE 4 M-UM016-05
Non-Volatile Memory
The RP 460 uses a non-volatile memory system. Data is not lost if the unit is turned o while
testing, if the battery is removed, or if the unit is stored for long periods of time.
USB Download
Before transferring data from the RP 460 to the computer, download and install the ODM®
brand soware, available online at www.ripley-tools.com/downloads.
1. Plug the RP 460 into the computer using the included USB cable. If this is the first
time it has been plugged in, wait at least ten seconds for the drivers to install.
2. Place the ODM® brand soware into Dump/Sync mode by navigating to Settings >
Data Mode > Dump Mode.
3. Press the USB button on the RP 460 to dump/sync all data points. The data from the
RP 460 appears in the box on the le side of the window.
4. Fill out the fields on the right side of the window (Customer, User, Test Site, etc).
5. Navigate to Settings > Report Settings to identify a minimum and maximum value
for the PASS/FAIL functionality. Choose a color code scheme from this window, if
necessary.
6. To create the final report, click File > Export to Excel.
Data Streaming
The RP 460 can be used in a live streaming mode with the ODM® brand soware.
1. With the RP 460 plugged in via USB, place the ODM® brand soware into Active Mode
by navigating to Settings > Data Mode > Active Mode.
2. Click the Save button on-screen to save the live reading displayed in the window.
Save as many readings as needed.
3. To create the final report, click File > Export to Excel.
Maintenance
Unpacking & Inspection
The RP 460 is supplied with a so carry case, 2.5 mm adapter installed on the unit, 3.0V CR2
battery, USB cable, and an instruction card. Contact Ripley immediately if any part of the
unit or packaging is damaged, or otherwise unsatisfactory.
Low Battery
The RP 460 provides over 1000 operations under normal conditions. When the BATT
indicator is shown on the device screen, the CR2 battery should be recharged or replaced,
depending on the battery type.
Optical Connector Interface
The RP 460 is equipped with a universal connector adapter that is compatible with many
popular style connectors. The connector adapter interface should be kept covered and
protected from contamination. Care must be taken to avoid objects that may damage the
glass surface of the detector mount. If scratches or breaks occur on the surface, please
contact Ripley for proper service.
Calibration
All Ripley ODM® brand products include a 2-year warranty. Ripley recommends
recalibration every two years to ensure adherence to NIST measurement standards.
The first recalibration is free within the warranty period. Contact Ripley Tools for return
information.