ROHO Smart Check User manual

Supplier: This manual must be given to the user of this product.
Operator (Individual or Caregiver): Before using this product, read instructions and save for future reference.
shape fitting technology
®
Operation Manual
DRAFT

Table of Contents
Intended Use
Important Safety Information
2-3
Legend 4
Specifications
Parts Detail
5
Overview 6
Smart Setup Instructions 7-8
Check Instructions 9
Optional Setup Instructions
Hand Check Guide
10
Hard Reset Instructions
Disconnection and Storage
Troubleshooting
Limited Warranty
11
Cleaning and Disinfecting
Battery Replacement
Disposal
12
Español 13
Français 25
Deutsch 37
ROHO, Inc.
100 North Florida Avenue, Belleville, Illinois 62221-5429
ROHO.com
© 2015 ROHO, Inc.
The following are trademarks and registered trademarks of ROHO, Inc.: ROHO®, Sensor Ready™,shape fitting technology®and Smart Check™.
Products in this manual may be covered by one or more U.S. and foreign patents and trademarks. Smart Check™by ROHO®patent pending.
ROHO, Inc. has a policy of continual product improvement and reserves the right to amend this document.
DRAFT

2
Smart Check by ROHO Operation Manual
Intended Use
Smart Check™by ROHO®(Smart Check) is a cushion feedback system that is intended to assist the user by saving an
inflation range during cushion setup. During subsequent inflation level checks, Smart Check is intended to indicate
whether adjustments are needed to remain in the saved range.
Smart Check must be connected to a ROHO®Single Compartment Cushion with Sensor Ready™Technology (Sensor
Ready Cushion).
Important Safety Information
• Read this instruction manual carefully and completely before using the product.
• ROHO medical products are intended to be part of an overall care regimen. ROHO, Inc. recommends that a clinician
make product recommendations based upon an evaluation of the individual’s medical and therapeutic needs and
overall condition (for example, potential allergic reactions). Use of this device does not eliminate routine inspection
of the product and all normal activities associated with skin and soft tissue protection (weight shifts, skin
inspection, overall skin care, etc.)
Warnings
• Smart Check by ROHO has been designed for use only with the ROHO Sensor Ready Cushion. Using a cushion
that is not a ROHO Sensor Ready Cushion will affect Smart Check functions, reduce or eliminate the benefits of
the cushion, and increase risk to skin and other soft tissue. Using a cushion that is not a ROHO Sensor Ready
Cushion may also result in damage to Smart Check and will void the warranty.
• DO NOT use an under-inflated cushion or an over-inflated cushion because the product benefits will be reduced
or eliminated, resulting in an increased risk to skin and other soft tissue. If the cushion does not appear to be
holding air, or if you are not able to inflate or deflate the cushion, see the TroubleshooTing section of the manual
that was provided with the cushion. Immediately contact your clinical caregiver, distributor, or ROHO, Inc. if the
problem persists.
• Check inflation frequently, at least once a day, using the Smart Check or a hand check.
• Perform Smart Setup or Optional Setup again if the following changes occur: 1) a different individual is
using Smart Check; 2) a different chair or a different cushion is being used; or 3) the individual experiences
physiological changes (for example, weight loss or weight gain). Before performing a new Setup, you must clear
your Smart Check setting; see hard reseT insTrucTions.
•KEEP BATTERIES OUT OF REACH OF CHILDREN AND ANY INDIVIDUALS WHO HAVE A TENDENCY TO PLACE INEDIBLE
OBJECTS IN THEIR MOUTHS. Swallowing or inhalation may lead to serious injury or death. IMMEDIATELY seek
emergency medical assistance.
• USE CAUTION when in presence of children to avoid a strangulation hazard. Detach Smart Check when not in
use and store in a safe and accessible location.
DRAFT

3
Smart Check by ROHO Operation Manual
Important Safety Information, Continued
Warnings, Continued
• Using an accessory (for example, a pump or a cushion cover) other than those provided by ROHO, Inc., may be
unsafe and could void the product warranty.
• DO NOT modify Smart Check or any components. Doing so may result in damage to the product and will void the
warranty.
•DO NOT use in the presence of flammable anesthetics and/or in an oxygen-rich environment. Doing so will
increase the risk of fire or explosion.
• If your cushion feels unstable during Smart Setup, perform a Hard Reset and then repeat Smart Setup. If you
still feel unstable, perform an Optional Setup and consult with a healthcare provider.
• The cushion MUST be properly sized to the user. IF IT IS NOT, Smart Check may not find the individual’s inflation
range during Smart Setup, reducing or eliminating the benefits of the cushion and increasing the risk to skin
and other soft tissue.
• All aspects of Smart Check use described in this manual can be performed by the individual or by a caregiver. If
you are unable to perform any tasks described in this manual, seek assistance.
• BURN HAZARD: If ambient temperature exceeds 47°C (117°F), DO NOT maintain contact with product surface for
more than ten (10) minutes.
• Smart Check MUST be used ONLY with a compatible ROHO cushion cover. Use of a different cover may interfere
with the accuracy of Smart Check functions.
• DO NOT perform maintenance on Smart Check while it is in use. Doing so may result in improper functioning.
• Set up your cushion in the usual environmental conditions. Perform Check when environmental conditions,
including temperature and altitude, change significantly.
Cautions
• Daily: Inspect Smart Check quick disconnect, tubing, and device housing for wear or breakage, and to ensure inter-
component attachment. DO NOT continue to use Smart Check if any part is damaged or cannot be attached securely.
ROHO, Inc. recommends that Smart Check be inspected periodically by a qualified technician.
• DO NOT submerge Smart Check in liquid, as it will damage the device.
• Protect Smart Check from blunt force shocks that may cause damage and/or breakage.
• DO NOT expose the product to high heat, open flames, or hot ashes.
• DO NOT use the inflation valve or any Smart Check components as a handle for carrying or for pulling the cushion.
• Prolonged exposure to ozone may degrade some of the materials used in the manufacture of this product, affect the
performance of this product, and void the product warranty.
• Prolonged exposure to UV light / sunlight may cause product labels to fade.
DRAFT

4
Smart Check by ROHO Operation Manual
Legend
Warning
Refer to instructions for use.
Manufacturing location
Keep dry.
Wipe by hand.
Air dry.
Do not use bleach.
Low battery: amber light
Temperature Limit
Humidity Limitation
(relative, non-condensing)
Atmospheric Pressure Limitation
Choking hazard
Do not discard in household trash.
User identification
IEC 60601-1: Type BF Applied Part
0 - 6
DRAFT

5
Smart Check by ROHO Operation Manual
Specifications
Storage Conditions
-25°C (-13°F)
+70°C (158°F) 93%
0-5%
Operating Conditions
93%
0-5%
0°C (32°F)
48°C (118°F)
1060 hPa
700 hPa
Altitude Limitation (above sea level):
Lower -410 m ( -1,350 ft.)
Upper 3185 m (10,450 ft)
- Smart Check: polycarbonate with polyester
faceplate
- Storage Case: polystyrene
- Tubing: polyurethane
- Quick Disconnect: acetal
- This product complies with RoHS2.
- Smart Check is a type BF Applied Part.
Smart Check has been tested and complies
with IEC 60601-1, IEC 60601-1-2, and IEC
60601-1-11; certified to CAN/CSA Standard
C22.2 No. 60601-1. Electromagnetic
interference is not caused by Smart Check
and does not interfere with Smart Check.
- Expected Service Life: 5 years
- Batteries: one (1) year typical operating time
- Smart Check is an IP54-rated device
(protected from dust and splashing water).
Parts Detail
Package Contents: Smart Check™ by ROHO®, Smart Check instructions, registration card, storage case
Product Registration: Mail the enclosed product registration or register your product at ROHO.com.
For information about your cushion, refer to the ROHO®Single Compartment Cushion with Sensor Ready™Technology
Operation Manual.
Smart Check Faceplate
Refer to instructions for
faceplate details.
Tubing
Quick
Disconnect
Operation Manual
Product Registration Card Storage Case
DRAFT

6
Smart Check by ROHO Operation Manual
Overview
This page explains how to use Smart Check to find,
save, and check your cushion inflation setting.
Please read before using Smart Check.
Smart Setup and Check:
1. Follow the Smart Setup InStructIonS. Smart Check will
find and save your cushion inflation setting for you.
2. Follow the check InStructIonS. When you press the Check
Button, Smart Check will tell you if you need to adjust
your cushion.
IMPORTANT!
• Perform Setup (either Smart Setup or Optional Setup) when using Smart Check for the first time.
• Perform Smart Setup or Optional Setup again if the following changes occur: 1) a different individual is
using Smart Check; 2) a different chair or a different cushion is being used; or 3) the individual experiences
physiological changes (for example, weight loss or weight gain). Before performing a new Setup, you must
clear your Smart Check setting; see hard reseT insTrucTions.
• Perform a Check frequently, at least once a day!
Optional Setup is an alternative to Smart Setup.
Optional Setup may be preferred in the following situations:
- A qualified seating specialist has determined an alternate immersion and envelopment level for the
individual.
- The individual must accommodate needs (such as severe asymmetrical posture) that are beyond the
immersion and envelopment setting found by Smart Check during Smart Setup.
- The cushion is being used on an upholstered surface (such as a Captain’s Seat).
Warnings:
• It is recommended that a qualified seating specialist be consulted when saving a preferred cushion
inflation setting with Optional Setup. (added)
• Optional Setup should not be used if the individual weighs less than 90 lbs. (40 kg).
Optional Setup (and Check)
1. Find your preferred cushion inflation setting without using Smart Check. (Follow the ROHO®Single
Compartment Cushion with Sensor Ready™Technology Operation Manual.)
2. Follow the optIonal Setup InStructIonS,using Smart Check to save your inflation setting.
3. Follow the check InStructIonS. When you press the Check Button, Smart Check will tell you if you need to adjust
your cushion.
Proper Cushion Inflation Setting: A) sink down into the
cushion and B) “float” above the cushion base.
B
A
DRAFT

7
Smart Check by ROHO Operation Manual
Smart Setup Instructions
Smart Check will find and save the inflation setting for you, on your cushion, in your
chair. NOTE: Results may vary based on body type.
• Perform the following steps in the usual environmental conditions.
• The cushion MUST be overinflated to start Smart Setup. You may
temporarily feel unstable; if so, proceed with caution.
• The individual MUST be seated on the cushion in the chair before using
Smart Check.
Symbols and Lights for Smart Setup
Red Light
“Remove air.”
Green Light
“Setup is complete.”
Red Light
“Add air.”
Setup
Button
Check
Button
3. Sit in the chair in your usual position.
Pull to remove the tab (first-time use only).
4. Make sure the inflation valve is closed.
Attach Smart Check to the cushion. You will
hear a “click” when the connection is secure.
1. Place the cushion in the chair. Place the
nozzle of the pump over the inflation valve.
If a cushion cover is used, place the cover on
the cushion before inflating.
2. Open the inflation valve. Overinflate the
cushion until it bulges.
Close the inflation valve.
Remove the pump.
3
2Open
Close
Cushion Cover
Quick Disconnect Inflation Valve
1
Quick Disconnect
4
DRAFT

8
Smart Check by ROHO Operation Manual
Store Smart Check in a safe and accessible location. See disconnecTion and sTorage instructions.
Smart Setup Instructions, Continued
Your cushion is now set up.
(Next: See CheCk InstruCtIons.)
When you see the flashing green
light, close the inflation valve. The
green light means that Setup is
complete.
5b. Press (but do not hold) the Setup Button.
The red down arrow light will now flash.
Open the inflation valve, release air (approximately 30
to 60 seconds)*, and wait for the flashing green light.
If the red down arrow light stops flashing before you
have opened the valve, go back to 5a.
*To ensure an accurate reading: Remain still and avoid
restricting air flow while you release air from the cushion.
5a. Press the Check Button.
When you see the red down arrow light,
proceed to step 5b.
NOTE: If you see the red up arrow light
when you press the Check Button:
1) Make sure the quick disconnect is securely
attached, and 2) Open the inflation valve,
add more air, close the inflation valve, and
press the Check Button again. Repeat until
you see the red down arrow light.
5a.
5b.
DRAFT

9
Smart Check by ROHO Operation Manual
How to Perform a Check:
1. You should be in your usual sitting position.
2. Press the Check Button.
3. Refer to the table below.
• Perform a Check frequently, at least once a day!
• If a cover was used while setting up the cushion, the same cover must be used
during Check.
• Perform a Check when there is a significant change in altitude.
Check Instructions
Perform a Check after completing Setup (Smart Setup or Optional Setup).
Check Table
NOTE: After you perform a Check, you may choose to perform a Hand Check to become familiar with your inflation range. See the hand check GuIde.
NOTE: You may be able to add or remove a small amount of air and still be at your cushion inflation setting. Smart Check saves your cushion
inflation setting with some room for adjustment.
Light Status Action
The cushion is within the saved
inflation setting. No further action is required.
Light Possible Causes Action
Your weight may be unevenly
distributed.
Perform a weight shift to circulate air in the cushion, then press
the Check Button to get a new reading.
The quick disconnect is not securely
attached to the cushion.
Make sure the connection is secure, then press the Check Button
to get a new reading. See troubleShootInG in this manual.
The cushion needs more air. Open the valve and add more air to the cushion (6 to 10 pumps).
Close the valve. Press the Check Button to get a new reading.
Repeat as needed.
If you are unable to make adjustments on your own,
discontinue use until assistance is available.
The cushion or the quick disconnect
on the cushion has a slow leak.
See TROUBLESHOOTING in the ROHO®Single Compartment
Cushion with Sensor Ready™Technology Operation Manual or
contact Customer Care.
Light Cause Action
The cushion has too much air. Open the valve and remove air from the cushion until you see the
green light. Close the valve.
If you are unable to perform this task, seek assistance.
Symbols and Lights for Check
Red Light
“Remove Air.
See the Check
Table.”
Green Light
“Within the saved
inflation setting.”
Setup
Button
Check
Button
Red Light
“Action required.
See the Check
Table.”
DRAFT

10
Smart Check by ROHO Operation Manual
Optional Setup Instructions
Optional Setup is an alternative to Smart Setup that records your preferred cushion inflation setting. See overvIew for more information.
Part One: Adjust Your Cushion Without Smart Check
1. Perform a Hard Reset. See hard reSet InStructIonS.
NOTE: For first-time use, remove the battery pull tab.
2. Inflate the cushion and place it in the chair. Sit on the cushion
in your usual position. Adjust the cushion by following the cuShIon adjuStment
instructions in the ROHO®Single Compartment Cushion with Sensor Ready™
Technology Operation Manual.
3. Perform a hand check to confirm immersion and envelopment. (See the hand
check GuIde in this manual.)
Symbols and Lights for Optional Setup
Green Light
“Setup is complete.”
Setup
Button
Check
Button
Part Two: Save your Setting with Smart Check
• It is recommended that a qualified seating specialist be consulted when saving a preferred cushion inflation setting
with Optional Setup.
• Perform the following steps in the usual environmental conditions.
4. Make sure Smart Check
is attached to your
cushion. You will hear
a “click” when the
connection is secure.
5. Make sure the inflation valve is
closed. Press the Check Button
to activate Smart Check. You will
see either a or a red light.
DO NOT add or remove air.
Store Smart Check in a safe and accessible location.
See disconnecTion and sTorage instructions.
4s
6. Press and hold the Setup Button for four (4)
seconds until you see a green light. The green
light means that your preferred cushion inflation
setting has been saved. The green light DOES
NOT indicate a Proper Cushion Inflation Setting.
See overvIew for more information.
Your cushion is now set up. (Next: See CheCk InstruCtIons.)
Hand Check Guide
Use this guide to adjust your cushion during Optional Setup.
A hand check is also recommended after performing Smart Setup so you will have a
reference point if you need to adjust your cushion without Smart Check (for example, if
the device has been misplaced).
1. Place your hand between yourself and the cushion. Feel for the lowest bony
prominence. You should be able to slightly move your fingertips. The recommended
distance between the individual and the seating surface is ½ - 1 inch (1.5 cm to 2.5 cm).
2. If it feels like the cushion has too much air or not enough air, adjust as needed and see check InStructIonS and overvIew.
½" - 1" (1.5 - 2.5 cm)
Before you perform a hand check, refer to imporTanT safeTy informaTion and read
the warning about using an under-inflated cushion or an over-inflated cushion.
Quick
Disconnect
DRAFT

11
Smart Check by ROHO Operation Manual
Hard Reset Instructions
Before you perform a new Setup, you must clear the setting on Smart Check by performing a Hard Reset.
NOTE: Smart Check may be attached or detached when performing the Hard Reset.
1. Make sure there are no steady or flashing lights.
2. Simultaneously press and hold the Check Button AND the
Setup Button for five (5) seconds. Each of the lights will
flash one time, indicating that your previous setting has
been removed.
Disconnection and Storage
To Disconnect Smart Check from the Cushion: Press down on the gray latch, then separate.
To Store Smart Check: You may detach and store Smart Check in a safe, accessible location
when not in use (for example, the storage case provided with Smart Check).
Confirm that Smart Check is not pressing against your body.
Troubleshooting
Problem Action
No lights are displayed on Smart Check after you press
the Check Button.
Make sure the pull tab has been removed from the battery
compartment. Confirm the battery type and battery orientation.
Replace the batteries if needed. Contact Customer Care if the
issue continues.
Smart Setup: After you press the Check Button, Smart
Check still displays the red “add air”light, but you
have confirmed that the cushion is overinflated and
that the quick disconnect is securely attached.
Inspect the O-ring on the cushion’s quick disconnect for wear or
damage. If damage is found, contact Customer Care.
Inspect the cushion for a slow leak. See troubleShootInG in the
ROHO®Single Compartment Cushion with Sensor Ready™
Technology Operation Manual.
Optional Setup: You do not see a green light after
pressing and holding the Setup Button for four (4)
seconds.
Repeat Part 2 of optIonal Setup InStructIonS. If you still do not see a
green light, note that Optional Setup may not be compatible with
your physical needs or preferences. Refer to the overvIew and Smart
Setup InStructIonS, or contact your clinician or Customer Care.
You require assistance with the use or maintenance of
Smart Check or you encounter unexpected problems.
Contact Customer Care.
Limited Warranty
Limited Warranty Term: 24 months from the date the product was originally purchased. See also the ROHO Limited
Warranty supplement provided with your product, or contact Customer Care.
+
5s
DRAFT

12
Smart Check by ROHO Operation Manual
Cleaning and Disinfecting
• Cleaning and disinfecting are separate processes. Cleaning must precede disinfection.
• When used by the same individual, clean and disinfect product as often as needed. Before use by a
different individual, clean, disinfect, and check product for proper functioning.
• Follow safety guidelines on cleaning and disinfecting product labels.
• Smart Check should not be cleaned with acetone, MEK, toluene, or adhesive removers.
• DO NOT immerse Smart Check in liquid.
To Clean: Wipe Smart Check with mild detergent and a clean, soft cloth or a sponge. Air dry.
To Disinfect: Wipe Smart Check with bleach-free disinfecting wipe. Air dry.
Sterilization: Smart Check is not packaged in a sterile condition, nor is it intended or required to be sterilized
prior to use. High temperatures accelerate aging and cause damage to the product assembly.
Disposal
Dispose of the product and/or components in accordance with the applicable regulations in your
jurisdiction, (including electronic waste disposal). DO NOT INCINERATE.
Battery Replacement
The low-battery light means the batteries
should be replaced.
-Smart Check will remember your saved setting after battery
replacement.
-To preserve battery life, Smart Check automatically goes into
“sleep” mode when not in use.
-Expected service life of batteries: one (1) year.
1. Detach Smart Check from the cushion at the quick disconnect.
2. Use a Phillips head screwdriver with a non-conductive handle to loosen the screw and open the battery door.
3. Use the screwdriver to release the batteries.
4. Refer to "+" and insert two (2) new 1.5 volt alkaline button cell batteries (LR44, AG13 or equivalent).
5. Close the battery door.
6. Tighten the screw. NOTE: If the screw is not tightened completely, the batteries may be exposed to moisture,
resulting in product damage.
KEEP BATTERIES OUT OF REACH OF CHILDREN AND ANY INDIVIDUALS WHO HAVE A TENDENCY
TO PLACE INEDIBLE OBJECTS IN THEIR MOUTHS. Swallowing or inhalation may lead to serious
injury or death. IMMEDIATELY seek emergency medical assistance.
+
0 - 6
DRAFT

T20185
Revision: 2/22/16
Print Date: Qty:
Mailing/Shipping Address:
1501 South 74th Street, Belleville, IL 62223-5900 USA
Administrative Offices:
100 N. Florida Avenue, Belleville, IL 62221-5429 USA
U.S.: 800-851-3449 • Fax 888-551-3449
Outside the U.S.: 618-277-9150 • Fax 618-277-6518
ROHO.com
DRAFT
Table of contents
Popular Security System manuals by other brands

Konig
Konig SEC-ALARM200 user manual

Texecom
Texecom Premier Elite iProx installation manual

Honeywell Home
Honeywell Home Ademco Vista-V21SIALTE Installation and setup guide

Clas Ohlson
Clas Ohlson SST-14 instruction manual

Sunsky
Sunsky YA-700-GSM-2 user manual

Cooper Security
Cooper Security i-On 40 installation guide