Rosenberger 19CK010-150 User manual

User Manual
SMP Calibration Kit 19CK010-150
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.com
Tel. : +49 8684 18-0
Email : [email protected]
Page
1 / 16
Dieses Dokument ist urheberrechtlich geschützt
● This document is protected by copyright ● Rosenberger Hochfrequenztechnik GmbH & Co. KG
19CK010-150

User Manual
SMP Calibration Kit 19CK010-150
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.com
Tel. : +49 8684 18-0
Email : [email protected]
Page
2 / 16
Dieses Dokument ist urheberrechtlich geschützt
● This document is protected by copyright ● Rosenberger Hochfrequenztechnik GmbH & Co. KG
Contents
1. General Information
Calibration Kit Normal Use 1.1
Warnings 1.2
Kit Contents 1.3
Kit Documentation 1.4
Calibration Certificate 1.5
2. Specifications
Environmental Requirements 2.1
Electrical and Mechanical Specifications 2.2
3. Standard Definitions
Standard Definitions 3.1
Installation of Standard Definitions 3.2
Standard Definitions file nomenclature 3.3
4. VNA Calibration
VNA Calibration 4.1
Calibration Check 4.2
5. Calibration Kit Service
Calibration Interval 5.1
Spare Parts 5.2
Where to send a Kit for Re-Calibration / Repair 5.3
Extent of Rosenberger Re-Calibration Services 5.4
6. Use, Maintenance and Care
Connector Care 6.1
Interface Dimensions and Gauging 6.2
Handling and Storage 6.3
7. Connections

User Manual
SMP Calibration Kit 19CK010-150
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.com
Tel. : +49 8684 18-0
Email : [email protected]
Page
3 / 16
Dieses Dokument ist urheberrechtlich geschützt
● This document is protected by copyright ● Rosenberger Hochfrequenztechnik GmbH & Co. KG
1. General Information
This User Manual is valid for the Calibration Kit
19CK010-150 (Industrial Version)
1.1 Calibration Kit Normal Use
Calibration Kits are used as an accessory for Vector Network Analyzers (VNA) to perform
vector-error correction. Vector-error correction is the process of characterizing systematic errors
of the measurement system by measuring known calibration standards, and then
mathematically removing the effects of these errors from subsequent measurements. The
measurement accuracy of the calibrated VNA depends mainly on the quality of the calibration kit
used. Contributing factors are noise, linearity and temperature stability of the VNA, transmission
and reflection stability of a test port cable as well as operator experience and care.
VNAs usually offer several techniques for correcting systematic measurement errors. The
standards in this calibration kit allow you to perform one-port OSL (OSM) or two-port OSLT
(TOSM) and TRM (TMR) calibrations from DC up to 40 GHz for measurements of components
with SMP connectors.
RPC-2.92 to SMP adapters are included in this calibration kit to interface with standard
2.92 mm, K or 3.5 mm test port cables and VNAs.
1.2 Warnings
The calibration standards of this calibration kit are mechanically extremely sensitive devices.
High accelerations e.g. when falling down or high forces to the connector fingers will damage
them.
When connecting and disconnecting male and female connectors keep them in a straight line as
much as possible. Radial misalignment (more than permitted) stresses the fingers of the female
outer conductor significantly and leads to bent or broken fingers. This results in a very unstable
or bad RF performance.
Connect the calibration standards only to test ports and other devices that have been cleaned
and checked for gauge, concentricity and integrity of the contact fingers. Detailed procedures
are described in Chapter 6 and 7.
Tighten the connector coupling nuts of the RPC-2.92 interfaces to the recommended torque with
a torque wrench (e.g. Rosenberger 03W021-000) while holding the RPC-2.92 female counter-
part with a wrench (e.g. Rosenberger Combi Wrench 03W021-000). Do not over torque!
Consider the power applied to the Load calibration standards. Maximum power limits for the
sum of RF and DC power are defined in the specific data sheets.
It is good lab practice to use at least an ESD wrist strap when handling calibration standards or
DUTs on a VNA although the calibration standards themselves are not ESD sensitive. Please
consult your VNA manual.

User Manual
SMP Calibration Kit 19CK010-150
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.com
Tel. : +49 8684 18-0
Email : [email protected]
Page
4 / 16
Dieses Dokument ist urheberrechtlich geschützt
● This document is protected by copyright ● Rosenberger Hochfrequenztechnik GmbH & Co. KG
Follow further limitations given in the specific data sheets.
1.3 Kit Contents
Table 1-1 Parts of the 19CK010-150 Calibration Kit
Device Part number Remarks Quantity
Open circuit plug 19S12L-000D3 1
Open circuit jack 19K12L-000D3 1
Short circuit plug 19S12S-000D3 1
Short circuit jack 19K12S-000D3 1
Calibration load plug 19S150-C10D3 1
Calibration load jack 19K150-C10D3 1
Calibration adaptor plug/plug 19S121-S20D3 1
Calibration adaptor jack/jack 19K121-K20D3 1
Calibration adaptor plug/jack 19S121-K20D3 1
Calibration adaptor plug/plug 02S119-S20D3 1
Calibration adaptor jack/jack 02K119-K20D3 1
1.4 Kit Documentation
USB-Stick
Standard Definitions as data files for Vector Network Anaylzer Families PNA
(Keysight/Agilent) and ZVA (Rohde&Schwarz). Calibration Certificate as PDF-file.
Standard Definitions Cards
Printed Standard Definitions that can be used on nearly all Vector Network Analyzers.
Kit Info Card
Handling precautions and information for installing Standard Definitions on a Vector Net-
work Analyzer.
Calibration Certificate
See chapter Calibration Certificate.
User Manual
(This document)
1.5 Calibration Certificate
Each calibration kit comes with a calibration certificate. Measurement results and uncertainties
for all calibration standards are included.
Standard scope of delivery is the Factory Calibration.

User Manual
SMP Calibration Kit 19CK010-150
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.com
Tel. : +49 8684 18-0
Email : [email protected]
Page
5 / 16
Dieses Dokument ist urheberrechtlich geschützt
● This document is protected by copyright ● Rosenberger Hochfrequenztechnik GmbH & Co. KG
An Accredited Calibration is not available.
2. Specifications
2.1 Environmental Requirements
The electrical performance of vector network analyzers and test port cables are sensitive to am-
bient temperature drift. Most manufacturers limit the allowable temperature drift to ± 1 K during
measurement calibration and during measurements when the network analyser error correction
is turned on. For further information please refer to the VNA specifications.
NOTE: With your fingers being a heat source, avoid unnecessarily handling the devices
during calibration.
The calibration standards are less sensitive to temperature changes. The operating temperature
range for which the specifications are valid is defined in the specific data sheets.
2.2 Electrical and mechanical Specifications
Electrical and mechanical specifications are defined for all calibration standards in the specific
data sheets.
3. Standard Definitions
3.1 Standard Definitions
Standard definitions provide the data needed to mathematically describe the electrical charac-
teristics of each calibration standard. For this calibration kit standard definitions are available:
Model based (standard delivery): These are the traditional standard definitions and can be
used on both old and modern VNAs. The values of these constants are theoretically derived
from the physical dimensions and material of each calibration standard, or from actual meas-
ured response. All necessary values are printed on the Standard Definition Cards and stored
as data files on an USB stick. Both are included with this calibration kit.
These standard definitions are used in the calibration of the VNA to determine the systematic
errors of the measurement system. Please always use the most recent source of information
(Standard Definitions Cards or VNA specific files).

User Manual
SMP Calibration Kit 19CK010-150
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.com
Tel. : +49 8684 18-0
Email : [email protected]
Page
6 / 16
Dieses Dokument ist urheberrechtlich geschützt
● This document is protected by copyright ● Rosenberger Hochfrequenztechnik GmbH & Co. KG
3.2 Installation of Standard Definitions
The standard procedure to install Standard Definitions is to import them from the USB stick onto
your VNA. As standard delivery data files for Vector Network Anaylzer Families PNA
(Keysight/Agilent) and ZVA (Rohde&Schwarz) are included. Data files for other VNA families
are available on request. If the specific data file is not available, the standard definitions have to
be installed manually with the provided information from the Standard Definitions Cards.
USB-stick
The Kit Info Card gives a brief description for the VNAs described above.
For all others please consult the VNA operating manual.
Manual installation
The manual installation is a more complex process. Please consult the VNA operating man-
ual.
Some network analyzers allow a step by step input of the model based standard definitions
during the VNA calibration without previously loading them.
Updating
After a re-calibration the standard definitions are optimized on the basis of the actual meas-
urements.
Data files have to be imported like described before.
For a manual update only the values that have changed have to be modified.
3.3 Standard Definitions file nomenclature
The file name includes the calibration kit type, the serial number, the type of standard definitions
and the date of calibration. The file extension defines the VNA types the file can be used on. For
further Information see the Application Note AN001.
KitType (SN)_StdDef_CalDate.Ext
KitType 19CK010-150 or customized type
(SN) Serial number (5 character) enclosed by brackets
StdDef ‘mb’ stands for model based and ‘db’ for data based, refer to 3.1
CalDate Date of kit calibration, these standard definition relate to
Ext e.g. ‘ckt’ file can be used on Keysight/Agilent VNAs
PNA family and compatible ones
‘calkit’ file can be used on Rohde&Schwarz VNAs
ZVA, ZVB, ZVBT, ZNB families and compatible ones
Example file name:
19CK010-150 (BH003)_mb_2020-05-30.calkit

User Manual
SMP Calibration Kit 19CK010-150
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.com
Tel. : +49 8684 18-0
Email : [email protected]
Page
7 / 16
Dieses Dokument ist urheberrechtlich geschützt
● This document is protected by copyright ● Rosenberger Hochfrequenztechnik GmbH & Co. KG
4. VNA Calibration
4.1 VNA Calibration
A VNA is only as useful as the accuracy of the measurements it makes, and this requires the in-
strument to be calibrated. The calibration process employs a technique called vector error cor-
rection, in which error terms are characterized using known standards so that errors can be re-
moved from actual measurements. The process of removing these errors requires the errors
and measured quantities to be measured vectorially (thus the need for a vector network analyz-
er).
Main setting parameters:
It is good practice to start your measurement by selecting the main setting parameters.
Sweep type (linear / logarithmic)
Sweep range (start frequency / stop frequency)
Number of points (frequency step size)
Source power
IF bandwidth (small IF bandwidth basically increases the dynamic range)
Sweep time
Connect test port cables and adapters to your VNA as needed for your specific measurement
task.
Calibration techniques
A very large number of calibration techniques have been developed to calibrate a VNA. The fol-
lowing list describes calibration techniques you can apply with this calibration kit along with their
features and applicability for use in specific situations.
One-port calibration
OSL (OSM) Open-Short-Load (Open-Short-Match)
When only reflection measurements are needed, the fastest and easiest calibration tech-
nique is OSL. It requires 3 connections of calibration standards on the reference plane.
Two-port calibrations
OSLT (TOSM) Open-Short-Load-Thru (Thru-Open-Short-Match)
OSLT is the most commonly used two-port calibration method. The Thru connections are
performed by using well-known, precision calibration adapters in male-male or female-
female configuration or as a direct connection of the male and the female test ports.

User Manual
SMP Calibration Kit 19CK010-150
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.com
Tel. : +49 8684 18-0
Email : [email protected]
Page
8 / 16
Dieses Dokument ist urheberrechtlich geschützt
● This document is protected by copyright ● Rosenberger Hochfrequenztechnik GmbH & Co. KG
TRM (7-term) Thru-Reflect-Match
The male and female Short standards of this kit are designed with the same offset length.
So the TRM calibration technique can be used. It requires high reflects with a symmetrical
reflection during VNA calibration on both test ports. TRM needs less connections during
calibration than OSLT.
Recommendation:
The TRM calibration is the recommended calibration method to use as two-port calibra-
tion. It can be used in a SMP male-male, female-female and male-female test port config-
uration.
The standard definition files for the VNAs include a special ‘Reflect’ device, please use it
for a TRM calibration. If no file for the specific VNA is available the standard definitions for
all calibration standards have to be entered manually. The ‘Reflect’ device is derived from
the short calibration standard but the L0, L1, L2 and L3 are all set to zero. This keeps the
symmetry of the ‘Reflect’ that is necessary for the TRM calibration.
Calibration Procedure
The procedures to perform a VNA calibration are very different between the VNA models. Mod-
ern VNA often guide you thru the settings and actions. For details please consult the manual of
your specific VNA.
4.2 Calibration Check
It is advisable to check the calibration after error correction has been performed, particularly
when test port cables and adapters have been used.
During the calibration
Calibration standards can easily be mixed up
Connections may not be tightened or aligned correctly
There might be defects
VNA, e.g. mechanically overloaded test port
Instable test port cables
Mechanically overloaded load
Bent or broken contact fingers
There is no Verification Kit available for SMP connectors. So additional loads and adpaters can
be used to verify the quality of reflection and transmission measurements or short cable assem-
blies of high quality may be used to check for plausible measurement results. This should be
done from time to time to identify defects.

User Manual
SMP Calibration Kit 19CK010-150
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.com
Tel. : +49 8684 18-0
Email : [email protected]
Page
9 / 16
Dieses Dokument ist urheberrechtlich geschützt
● This document is protected by copyright ● Rosenberger Hochfrequenztechnik GmbH & Co. KG
5. Calibration Kit Service
5.1 Calibration interval
Rosenberger suggests a calibration interval of 12 months. Depending on the frequency of use,
the wear and the actual requirements a different calibration interval might be suitable. This lies
in the responsibility of the user.
5.2 Spare parts
Each part of this kit can be ordered separately as spare part with reference to the part number
shown in Table 1-1 / Table 1-2 and the quantity desired. Telephone or send your order to your
local Rosenberger sales office.
5.3 Where to send a Kit for Re-Calibration / Repair
In case of re-calibration or repair contact your local Rosenberger sales office.
5.4 Extent of Rosenberger Re-Calibration Services
Every calibration kit sent for re-calibration or repair the following services are performed
Cleaning of connector interfaces
Optical check of connector mating planes, contact pins and contact fingers
Check of connector gauge for impermissible protrusion and significant recession
Electrical measurement of all calibration standards
Issuing of calibration certificates including measurement values and uncertainties
Individual optimizing of capacitive coefficients C0 to C3 for all Open calibration standards
and short inductance coefficients L0 to L3 for all Short calibration standards for best fitted
model based standard definitions.
In case of any defect or abnormality a cost estimation will be issued before continuing repair
work.
If a calibration standard has to be replaced or modified, the incoming status is additionally re-
ported in the calibration certificate. This service is excluded when this measurement may dam-
age the Rosenberger measurement setup.

User Manual
SMP Calibration Kit 19CK010-150
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.com
Tel. : +49 8684 18-0
Email : [email protected]
Page
10 / 16
Dieses Dokument ist urheberrechtlich geschützt
● This document is protected by copyright ● Rosenberger Hochfrequenztechnik GmbH & Co. KG
6. Use, Maintenance and Care
6.1 Connector Care
It is of particular importance to note that mechanical damage can be inflicted on a connector
while making a connection or a disconnection.
6.1.1 Connector Inspection
Optically check the connector mating planes, contact pins and contact fingers. Specially the
outer conductor of the female SMP connector are very sensitive for bent or broken contact fin-
gers, resulting in very unstable or bad RF performance.
6.1.2 Connector Cleaning
To ensure a long and reliable connector life, careful and regular inspection of connectors is
necessary and cleaning of connectors is essential to maintain good performance.
Connectors should be inspected initially for dents, raised edges, and scratches on the mating
surfaces. Connectors that have dents on the mating surfaces will usually also have raised edg-
es around them and will make less than perfect contact; further to this, raised edges on mating
interfaces will make dents in other connectors to which they are mated. Connectors should be
replaced unless the damage is very slight.
Awareness of the advantage of ensuring good connector repeatability and its effect on the
overall uncertainty of a measurement procedure should encourage careful inspection, interface
gauging and handling of coaxial connectors.
Prior to use, a visual examination should be made of a connector or adaptor, particularly for
concentricity of the centre contacts and for dirt on the mating planes. It is essential that the axial
position of the centre contact of all items offered for calibration should be checked because the
adjoining surfaces of mated centre contacts must not touch. If the centre contacts do touch,
there could be damage to the connector or possibly to other parts of the device to which the
connector is fitted.
Small particles, usually of metal, are often found on the inside connector mating planes and on
the threads. They should be removed to prevent damage to the connector surfaces. The items
required for cleaning connectors and the procedure to be followed is described below.

User Manual
SMP Calibration Kit 19CK010-150
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.com
Tel. : +49 8684 18-0
Email : [email protected]
Page
11 / 16
Dieses Dokument ist urheberrechtlich geschützt
● This document is protected by copyright ● Rosenberger Hochfrequenztechnik GmbH & Co. KG
6.1.3 Cleaning procedure
Items required:
Low pressure compressed air (solvent free).
Cotton swabs.
Lint free cleaning cloth.
Isopropanol.
Illuminated magnifier or microscope (factor 5 to 10)
NOTE: Isopropanol that contains additives should not be used for cleaning connectors
as it may cause damage to plastic dielectric support beads in coaxial and micro-
wave connectors. It is important to take all necessary safety precautions when
using chemicals or solvents.
Remove loose particles on the mating surfaces and threads etc. using low-pressure com-
pressed air.
Clean surfaces using Isopropanol on cotton swabs or lint free cloth. Use only sufficient sol-
vent to clean the surface. When using swabs or lint free cloth, use the least possible pres-
sure to avoid damaging connector surfaces. Do not spray solvents directly on to the con-
nector surfaces or use contaminated solvents.
Use the low-pressure compressed air once again to remove any remaining small particles
and to dry the surfaces of the connector to complete the cleaning process before using the
connector.
6.1.4 Cleaning connectors on static sensitive devices
Special care is required when cleaning connectors on test equipment containing electrostatic
sensitive devices. When cleaning such connectors always wear at least a grounded wrist strap
and observe correct procedures.
6.2 Interface Dimensions and Gauging
It is of the utmost importance that connectors do not damage the test equipment interfaces to
which they are connected. Poor performance of many coaxial devices and cable assemblies
can often be traced to poor construction and non-compliance with the mechanical specifica-
tions. The mechanical gauging of connectors is essential to ensure correct fit and to achieve the
best performance. This means that all coaxial connectors fitted on all equipment, cables and
terminations etc. should be gauged on a regular basis in order to detect any out of tolerance
conditions that may impair the electrical performance.
6.2.1 Gauging connectors
A connector should be gauged before it is used for the very first time or if someone else has
used the device to which it is connected.

User Manual
SMP Calibration Kit 19CK010-150
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.com
Tel. : +49 8684 18-0
Email : [email protected]
Page
12 / 16
Dieses Dokument ist urheberrechtlich geschützt
● This document is protected by copyright ● Rosenberger Hochfrequenztechnik GmbH & Co. KG
If the connector is to be used on another item of equipment, the connector on the equipment to
be tested should also be gauged.
Connectors should never be forced together when making a connection since forcing often indi-
cates incorrectness and incompatibility. There are some dimensions that are critical for the me-
chanical integrity, non-destructive mating and electrical performance of the connector.
The mechanical gauging of coaxial connectors will detect and prevent the following problems:
Inner conductor protrusion
This may result in buckling of the socket contacts or damage to the internal structure of a
device due to the axial forces generated.
NOTE: At no time should the pin depth of the connector be protruding.
Inner conductor recession
This will result in poor reflection and possibly unreliable contact.
6.2.2 Gauging
There are no special Rosenberger gauges available for SMP connectors.
It is recommended to use a Caliper (e.g. Mitutoyo ID-F150H)
.
Checking devices for protrusion or excessive recession only a single measurement is sufficient.
More precise measurements or checking devices very close to critical limits should be done in
different orientations averaging up to 4 single measurements.

User Manual
SMP Calibration Kit 19CK010-150
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.com
Tel. : +49 8684 18-0
Email : [email protected]
Page
13 / 16
Dieses Dokument ist urheberrechtlich geschützt
● This document is protected by copyright ● Rosenberger Hochfrequenztechnik GmbH & Co. KG
6.3 Handling and Storage
Install the protective end caps and store the calibration devices in the foam-lined storage
case when not in use.
Never store connectors loose in a box, or in a desk or bench drawer. This is the most com-
mon cause of connector damage during storage.
Keep connectors clean.
Do not touch mating plane surfaces or the contact fingers. Natural skin oils and microscopic
particles of dirt are easily transferred to a connector interface and are very difficult to remove.
The wearing of protective gloves is recommended.

User Manual
SMP Calibration Kit 19CK010-150
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.com
Tel. : +49 8684 18-0
Email : [email protected]
Page
14 / 16
Dieses Dokument ist urheberrechtlich geschützt
● This document is protected by copyright ● Rosenberger Hochfrequenztechnik GmbH & Co. KG
7. Connections
Good connections require a skilled operator. The most common cause of measurement error is
bad connections. The following procedures illustrate how to make good connections.
The wearing of protective gloves is strongly recommended for every single connection. Even the
smallest amount of skin lipids/particles that gets into the connector interface can have a strong
influence on the measurements and even make them unusable.
Adapting Test Ports
The calibration kit comes with two RPC-2.92 to SMP adapters to adapt the test port of the VNA
or the 2.92 mm test port cable to SMP interfaces.
How to connect
Pick up the calibration standard. Make sure that the test port cable and the calibration standard are
aligned with each other. Carefully guide it to the test port cable until you are fully centred and you
feel the male and female SMP connectors touching each other. Now the force can be gradually in-
creased until the SMP connection snaps into place.
The test port cable and the calibration standard
are aligned with each other. The test port cable and the calibration standard
are not aligned with each other. A strong me-
chanical stress will be caused.
Draft Date Approved Date Rev. Engineering change number Name Date
Herbert Babinger 24/04/12 Markus Hantschel 21/12/20 a00 20-1058 Lars Ramtke 21/12/20

User Manual
SMP Calibration Kit 19CK010-150
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.com
Tel. : +49 8684 18-0
Email : [email protected]
Page
15 / 16
Dieses Dokument ist urheberrechtlich geschützt
● This document is protected by copyright ● Rosenberger Hochfrequenztechnik GmbH & Co. KG
Make sure that the mated connection is kept free of mechanical stress. Even slight me-
chanical stresses can then lead to different contact situations at one or more contact la-
mellae, which in turn can result in different measurement results.
This mated connection is kept free of me-
chanical stress
.
This mated connection is bent too much
and
results in mechanical stress
.
If a mated connection is bent too much, contact lamellae are permanently deformed or in
the worst case they break out. The contact force of the lamellae concerned decreases
considerably or there is no contact at all.
This mated connection is kept free of me-
chanical stress
.
This mated connection is bent too much
and
results in mechanical stress
.
Therefore, with every connection, care must be taken to ensure that all components are
aligned with each other and remain so.

User Manual
SMP Calibration Kit 19CK010-150
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.com
Tel. : +49 8684 18-0
Email : [email protected]
Page
16 / 16
Dieses Dokument ist urheberrechtlich geschützt
● This document is protected by copyright ● Rosenberger Hochfrequenztechnik GmbH & Co. KG
Example for contact lamellae
without
permanently defor-
mation
.
Examples for contact lamellae with permanently defor-
mation.
All four lamella
e have the
correct distance to each other
and to the inner conductor
.
The upper lamella is
permanently bent inwards.
One Lamella is completely
broken out.
How to disconnect
The SMP connector is very small but the full detent version needs a high force to disconnect. Care
should be taken not to get hurt during this process and to keep the test port cable and the calibra-
tion standard under control.
Take the test port cable with one hand and the calibration standard in the other hand. Touch the
forefingers and the thumbs. Gradually increase the disconnecting force between these fingers.
Draft Date Approved Date Rev. Engineering change number Name Date
Herbert Babinger 24/04/12 Markus Hantschel 03/05/21 a00 20-1058 Lars Ramtke 03/05/21
While the information has been carefully compiled to the best of our knowledge, nothing is intended as representation or warr
anty on
our part and no statement herein shall be construed as recommendation to infringe existing patents. In the effort to improve our prod-
ucts, we reserve the right to make changes judged to be necessary.
Table of contents
Other Rosenberger Test Equipment manuals