Verifying the temperature indication
To ensure that the results of the adjustment operations
described below are reliable, the following minimum
immersion depths should be maintained:
• 60 mm for bimetallic thermometers-types TS, TH, TM, TK
• 70, 120, 220 mm (ls+20) – according to bulb type –
for gas pressure thermometers (types TF, TG).
These lengths are indicated by a welding seam around
the bulb.
The reference temperature
If the temperature of the medium (liquid, gas) in which the
stem or the bulb is immersed is not known, and neither a
correctly calibrated reference thermometer is placed nearby, or
a calibration bath (filled with oil or water and stirred)
is available, the instrument can be verified by either of the
following two simple methods:
a. Immerse the thermometer stem or the bulb in melting ice
(ratio ice: water approx. 2:1) and stir it around. After about
5 minutes – the time to allow the measuring system to
stabilize – the thermometer should indicate 0°C.
b. Immerse the thermometer stem or the bulb in boiling
water. After about 5 minutes – the time to allow the
measuring system to stabilize – the thermometer should
indicate:
100 °C at 0 m (sea level)
99 °C at 327 m
98 °C at 654 m
The temperature of boiling water at other altitudes can be
found by assuming a drop of 1°C per 327 m height increase.
In this approximate but often useful method, temporary
atmospheric pressure variations can be neglected.
Important ! Verification and adjustment at ambient
temperature are not recommended.
Verifying, setting and readjusting Rüeger Thermometers
All these thermometers, designed to meet the most demanding requirements, are provided with one or several adjust-
ment devices. These devices allow :
• setting to a given working temperature, for accurate indication and the compensation of errors due to the
measuring point
• readjustment, for example after long use under severe vibrations or frequent shocks
• rotation of the dial (type TMH).
• resetting after rotating the dial (types TMV... and TMI...), for example to orient the stem in a different direction
• alignment of a thermometer mounted in a row
• setting switchpoints of electrical switching thermometers from the outside; types TM (H, V, I) K, TG (H, V, I) K,
TF (H, V) K
• external resetting of thermometers with pierced case; types TH (H, I)
Instruction manual for adjustment and recalibration Thermo-Standard TS
of thermometers Hit TH
Modul TM
Kontakt TK
Flexible TF
Gas TG
IM-TH-E-7.10
Thermo
Instruction manual for adjustment and recalibration of thermometers
A. ADJUSTMENT BY DIAL ROTATION, 360°
for thermometers in: thermo-MODUL Group, type TMH
a. Twist the bezel counter clockwise to release, and remove the glass
b . With the fingertip, and without pressing inwards, turn the dial
gently until the graduation corresponding to the true or re f e r ence tem-
p e r a t u r e is opposite the pointer
c. Fit the sealing gasket against the dial, replace the glass, and twist-
lock the bezel, hold the assembly
External adjustment (pierced glass), ± 360°
available for TMH group
a. Unscrew the plug on the fixed device. Important : this plug is a seal,
so the O-ring with it must not be lost !
b. By means of the setting key supplied, turn the adjustment screw to
the left or right, until the dial graduation corresponding to set-value
c. Screw the plug back in, not forgetting the O-ring
External adjustment, ± 10°, available for TH (H, I)
a. Remove plug from the back of the housing
b. Insert a small screw driver into the dial holes
c. Adjust dial until indicated temperature corresponds to set-value
d. Refit the plug firmly in the case
B. ADJUSTMENT BY POINTER REPOSITIONING,
360° for thermometers in: thermo-STANDARD Group,
types THH, THV, THI pierced glass
a. Remove the plug from the glass
b. Insert the setting tool (225-002)
c. Turn the grey ring (2) to bring the dogs onto the faces of the
pointer (4)
d. Turn the black knob (1) to engage the blade in the slots (3)
of the pointer (4)
e. Holding the black knob (1) stationary, turn the grey ring (2) until the
pointer indicates the temperature required
f. Take the tool out, replace the plug in the glass
C. ADJUSTMENT BY POINTER REPOSITIONING,
360° for thermometers in:
thermo–MODUL, types TM (V, I) and TG (H, V, I)
thermo–FLEXIBLE, types TF (H, V)
t h e r m o – K O N TA K T, types TM (H, V, I), TG (H, V, I)
and TF (H, V) K
a. Twist the bezel counter clockwise to release, and remove the glass
b. Engage black knob (1) in slots (3) of the pointer
c. Hold pointer (4) stationary with two fingers on sides
d. Turn knob (1) to set pointer
e. Refit the glass and bezel, not forgetting the dial/glass seal
• The tool N° 225-002 may also be used for this adjustment. The grey ring (2) can be taken off if desired, to simplify operations
• On thermometers of group TK, take care not to lift the pointer, as this could cause the drive pin to disengage from the plate.
Setting tool no code 225-002
1) Black knob with 2-tooth blade, for engaging in the slots (3)
of the pointer (4)
2) Grey ring with 2 dogs, for engaging in the faces of the pointer (4)
Setting tool no code 225-003
1) Black knob with 2-tooth blade, for engaging n the slots (3)
of the pointer (4)
G r o u p s Ty p e s R o t a t i o n I n d i c a t o r Setting tool With external device S e t t i n g E x t e r nal adjustment
S e t t i n g 3 6 0 ° 3 6 0 ° N° 225-002 N° 225-00 fixed to glass key supplied ment ±10° TH (I, H)
T h e r m o - S TA N D A R D TH, (H,V, I ) • • p i e r ced glass + plug • (I, H)
T h e r m o - K O N TA K T TM, (H,V,I) K • • • • for setting targ e t •
TG, (H,V,I) K • • • values only •
T F, (H,V) K • • •
T h e r m o - M O D U L T M H • • • allows ro t a t i o n
T M V • • • of dial •
T M I • • • •
TG (H, V, I) • • • •
T h e r m o - F L E X I B L E TF (H,V) • • •
A, E, B, F, T, U
Instruction manual for adjustment and recalibration of thermometers
D. ADJUSTMENT BY DIAL ROTATION WITH EXTERNAL
DEVICE FIXED TO GLASS
(device supplied on request, at extra charge)
Rotation of dial types TMH
a. Unscrew the plug on the fixed device. Important: this plug is a
seal, so the O-ring with it must not be lost!
b. By means of the setting key supplied, turn the adjustment screw
to the left or right, until the dial graduation corresponding to the
true or required temperature is opposite the pointer
c. While pressing the adjustement screw in, turn it to bring the metal
driver into dial with the switchpoint temperature required
d. Screw the plug back in, not forgetting to fit the O-ring
E. ADJUSTING THE SWITCHPOINTS ON ELECTRICAL
SWITCHING THERMOMETERS,
Group Thermo-MODUL, types TM (H,V,I) K and TG (H, V, I,) K
Group Thermo-FLEXIBLE, types TF (H, V) K
a. Unscrew the plug on the fixed device. Important: this plug is a seal, so the O-ring with it must not be lost !
b. By means of the setting key supplied, turn the adjustment screw to the left or right, to bring the metal driver near to the red
switchpoint indicator
c. While pressing the adjustement screw in, turn it to bring the metal driver into contact with the switchpoint temperature required
d. Turn the driver to a position where it does not hinder reading the temperature (e.g. on one of the dial tabs)
e. Replace the plug, not forgetting the O-ring
N.B. If the thermometer is not provided with the adjustment device on the glass, the switchpoint may also be set as described in method A.
F. BOTTOM-CONNECTED BIMETALLIC THERMOMETERS, types TMV
On these thermometers, types T M V (data sheet TM 1), the position
of the dial relative to the stem may be rotated to any angle.
Rotate the dial as described in method A. To readjust, proceed as
in method B.
ADJUSTMENT METHODS
RÜEGER SA
Ch. de Mongevon 9
Case postale 98
1023 CRISSIER 1
SWITZERLAND
Tel + 41 (0)21 637 32 32
Fax + 41 (0)21 637 32 00
E-mail info@rueger.ch
RÜEGER GmbH
Plieninger Strasse 58
70567 STUTTGART
GERMANY
Tel + 49 (0)711 16-163-0
Fax + 49 (0)711 16-16333
E-mail rgmbh@rueger.com
RÜEGER Sdn Bhd
No 22-5, Jalan Wangsa Delima 10
D’ Wangsa, Wangsa Maju
53300 Kuala Lumpur
MALAYSIA
Tel + 603 - 4142 3808
Fax + 603 - 4142 3909
E-mail sales@rueger.com.my
Manufacturers of Sensors and Gauges for Temperature & Pressure
www.rueger.com www.instrugate.com
Tilting and rotating Groups TH 1 Types THI
TM 1, TMK TMI
TG 1, TGK TGI
Tilting
Tilting the head (0-90°)
1Slacken screws (a) on both sides.
2Press the two parts of the bracket (A) together and tilt
to required angle.
3Retighten screws (a) on both sides.
Rotating
Rotating the head (360°)
1Slacken locknut (b) at top and bottom.
2Hold bracket (A) with one hand, and turn head
with the other.
3Retighten locknut (b) at top and bottom.
Storage of thermometers
Rüeger thermometers should remain in their original packaging until
installation. They should not be exposed to any mechanical shocks or
vibrations and be protected against dust and humidity
Storage temperature should not exceed -20°C or + 60°C, unless speci-
fied otherwise.
BEIJING RÜEGER PRECISION
INSTRUMENT CO. , LTD
No. A135 Chengshousi Road Nansanhuan,
Chaoyang District
100164 Beijing P.R. CHINA
Tel + 86 10 8767 7502 / 3379
Fax + 86 10 8761 3727
E-mail sales@bjrueger.com

This manual suits for next models

5

Popular Thermometer manuals by other brands

Omega Engineering HH806AW instruction sheet

Omega Engineering

Omega Engineering HH806AW instruction sheet

Exergen TemporalScanner TAT-5000S-RS232-CORO Operator's manual

Exergen

Exergen TemporalScanner TAT-5000S-RS232-CORO Operator's manual

Velleman DVM8810 user manual

Velleman

Velleman DVM8810 user manual

TFA THD2FE instruction manual

TFA

TFA THD2FE instruction manual

Maverick DT-09 instruction manual

Maverick

Maverick DT-09 instruction manual

Fluke 66 Specifications

Fluke

Fluke 66 Specifications

HoldPeak 981D instruction manual

HoldPeak

HoldPeak 981D instruction manual

Extech Instruments RH200W user manual

Extech Instruments

Extech Instruments RH200W user manual

SKF TKTL 20 INSTRUCTIONS FOR US

SKF

SKF TKTL 20 INSTRUCTIONS FOR US

Kinsa QuickCare Thermometer Instructions for use

Kinsa

Kinsa QuickCare Thermometer Instructions for use

Medisana TM 750 manual

Medisana

Medisana TM 750 manual

Microlife MT 600 user manual

Microlife

Microlife MT 600 user manual

Silverline 675133 quick start guide

Silverline

Silverline 675133 quick start guide

Ashcroft S5500 Installation and maintenance instruction manual

Ashcroft

Ashcroft S5500 Installation and maintenance instruction manual

Wahl 2500MVX quick start guide

Wahl

Wahl 2500MVX quick start guide

Alecto BC-07 user manual

Alecto

Alecto BC-07 user manual

Perel CWT20 manual

Perel

Perel CWT20 manual

Tescoma PRESTO 420910 Instructions for use

Tescoma

Tescoma PRESTO 420910 Instructions for use

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.