SANKO EDY-5000 User manual

EDDY CURRENT TYPE
DIGITAL COATING THICKNESS METER
EDY‑5000
INSTRUCTION MANUAL
CAUTIONS:
●Before using the Meter, read this INSTRUCTION MANUAL thoroughly
and use the Meter correctly.
●Keep this INSTRUCTION MANUAL carefully and refer to this, when
necessary.
●In the event of any doubt arising, the original INSTRUCTION MANUAL
in Japanese is to be final authority.
SANKO ELECTRONIC LABORATORY CO., LTD.
(株式会社 サンコウ電子研究所)

‑1‑
CONTENTS
ITEM PAGE
1.PRINCIPLE 2
2.APPLICATIONS 2
3.SPECIFICATIONS 3
4.PARTS IDENTIFICATION 4
5.PREPARATION 5
5‑1. Preparation of the zero plate 5
5‑2. Connecting (Disconnecting) the probe 5
6.OPERATING INSTRUCTIONS 6
6‑1. Power ON 6
6‑2. Power OFF 6
6‑3. Zero adjustment 7
6‑4. Standard calibration(CAL) 8
6‑5. Resetting the working curve 9
6‑6. Measurement 10
7.ADDITIONAL FUNCTIONS 10
7‑1. Convertingthekeylockmode 10
8.REPLACING THE BATTERY 11
9.MAINTENANCE AND INSPECTION 11
10. CARES TO RAISE AN ACCURACY OF THE MEASUREMENT 12

‑2‑
Before using the Meter, read this INSTRUCTION MANUAL thoroughly and use
the Meter correctly.
Keep this INSTRUCTION MANUAL carefully and refer to this, when necessary.
1.PRINCIPLE
The thickness of non-conductive films treated on
non-magnetic metallic substrate can be measured non-
destructively, simply, speedily, and accurately utilizing
the electric correlation between the eddy current
induced on metal surface by a high frequency electric
field and the thickness of the surface film.
When a metal is moved toward or away from
a coil which carries electric current supplied
by a high frequency electric source, the current increases or decreases in accordance with a distance
from the coil.
This principle is utilized to measure the thickness (D) of non-conductive film on non-magnetic metallic
substrate.
2.APPLICATIONS
The Meter is used for non-destructive measurement of the thickness of a non-conductive film e.g. ALMITE,
paintings, linings, etc. applied to non-magnetic metallic substrate e.g. aluminum, aluminum alloy, copper, etc.
and austenitic stainless(non-magnetic stainless) steel substrate.
The Meter can be applicable for various domestic and foreign standards and rules.
●Anodic oxide films Measurement of the thickness of ALMITE(anodic oxide films)
applied to aluminum sashes, kitchen utenciles, electric appliances, etc.
●Paintings Measurement of the thickness of coatings applied to housing materials,
machines, tanks, etc. made of aluminum or stainless steel.
●Linings Measurement of the thickness of linings applied to machines, parts,
chemical plants, etc.
●Resin films Measurement of the thickness of films, papers, etc. based on aluminum plate.

‑3‑
3.SPECIFICATIONS
・Name and type Eddy current type digital coating thickness meter EDY-5000
・Measuring method Eddy current type
・Measuring range 0 to 5.00 mm
・Accuracy ±0.01mm on uniform surface or ±2 %of reading
・Indication Digital display on large size Liquid Crystal Display (LCD)
With hold function
・Resolution 0.01mm (1.00 mm to 5.00 mm)
1μm(0to999μm)
・Range of CAL 10 μmto5.00mm
・Probe One point contact constant pressure type with V-groove
φ18 × 55mm
・Additional function 1) Conversion of key lock mode
2) Auto power off (about 5 minutes)
・Power source Dry batteries R03(1.5V) × 4 pcs
With auto power off function
・Operatingtemp. 0to40℃(except dew condensing condition)
・Dimensions 80(W) × 35(H) × 150(D) mm
・Weight Approx. 330g (including batteries)
・Accessories Standard thickness plate, Zero plate for checking the Meter, Dry battery,
Carrying bag
※Specifications and appearance are subject to modify for improvement without prior notice.

‑4‑
4.PARTS IDENTIFICATION ⑤⑥
⑦
①
②
③
④
⑧
①POWER key Key for power ON/OFF
②ZERO key Key for zero adjustment
③LOCK key Key for locking all keys except ON/OFF key
④
≪
・
≫
key Key for standard calibration with the standard thickness plate
⑤Printer connector Connector for connecting the printer
⑥Probe connector Connector for connecting the probe
⑦LCD display Indicating part of the measured values
⑧Hand strap
Be sure to pass the hand strap around the wrist to prevent the Meter from dropping.

‑5‑
5. PREPARATION
5‑1. Preparation of the zero plate
Prepare the substrate which is the same kind, thickness and shape with an object
to be measured.
The attached zero plate for checking the Meter, Aluminum 1050, is used for the
operation check of the Meter.
Prepare the substrate of an actual object to be measured.
・Same kind The same kind of material as the substrate of an object to be measured
・Same thickness Nearly the same thickness as the substrate of an object to be measured
・Same shape The same shape in pipe diameter, curvature, geometrical shape, etc. as the substrate
of an object to be measured
・Size Substrate with an enough area on which the probe can be easily operated
・Surface condition Untreated (unoxidized etc.) substrate having smooth surface
Remove rust, dirt or dust, if any.
5‑2. Connecting(Disconnecting) the probe
Carry out the connection (disconnection) of the probe by the following procedures.
・Be sure to keep the Power OFF.
・The probe cannot be substituted with other probe.
Use the probe which is the same number with the Meter number.
◎ Connecting method
Insert the probe into the probe
connector keeping the wide cut
right ward and tighten it by turning
A clockwise.
◎ Disconnecting method
Release by turning A counter‑
clockwise and calmly pull the probe
out of the connector.
Don't pull the cord as it
causes breaking of cord.
※When the probe is not connected to the Meter or any system is out of order even if the probe
is connected, the indication changes as follows after power ON .
If the probe is disconnected in power ON condition, 「PE」is indicated on the display and power
is automatically cut off. (「PE」is abbreviation of Probe Error.)
↓
→
The buzzer emits a beeping sound. Power is automatically cut off.
−−− μm
Non‑Fe
PE
Non‑Fe Non‑Fe

‑6‑
6. OPERATING INSTRUCTIONS
6‑1. Power ON
Press the 「ON/OFF」key ①after connecting the probe.
[Indication on the LCD ⑦changes as follows.]
↓
Inside of the Meter is being initialized.
↓Keep the probe turn towards the air during this period.
↓←The buzzer emits a beeping sound.
The Meter is ready for measurement.
・When power ON, the last measured value of
the previous measurement is indicated.
・When the first power ON or no measured data
is in the Meter, [-]is indicated.
6‑2. Power OFF
Press the「ON/OFF」key ①.
[Indication on the LCD ⑦changes as follows]
↓←The buzzer emits a beeping sound.
↓←The buzzer emits a beeping sound.
Power is automatically cut off.
−−− μm
Non‑Fe
−− μm
Non‑Fe
3.20 mm
Non‑Fe
−μm
Non‑Fe
μm
Non‑Fe

‑7‑
6‑3. Zero adjustment
Press the probe against the substrate for adjustment (zero plate).
[Indication on the LCD ⑦changes as follows.]
↓←The buzzer emits a beeping
sound.
Measured value [XX] is Zero plate
indicated on the LCD ⑦.
↓
Press the「ZERO」key ②.
(Both the probe contact to zero plate or not is possible.)
↓←The buzzer emits a beeping sound.
[CAL]is indicated on the upper left of the
LCD ⑦, [0]is indicated.
↓←The buzzer emits a beeping sound.
Press the probe against the zero plate several times, when the [0] or closed values are indicated
on the LCD ⑦, the adjustment is acceptable.
When the indications are away from [0], repeat above procedures several times.
※When 「LLL」is indicated while zero adjusting, its adjusting point is widely deviated
from a right position.
Repeat the zero adjustment 2 to 4 times after confirming no coating is applied to the
substrate and confirm zero is stably indicated.
XX μm
Non‑Fe
CAL 0
Non‑Fe
0μm
Non‑Fe

‑8‑
6‑4. Standard calibration (CAL)
Place the standard thickness plate on the zero plate,
and press the probe against it.
Select a little thicker standard thickness plate than
a film thickness to be measured.
[Indications on the LCD ⑦change as follows.]
↓←The buzzer emits a
beeping sound. Standard
thickness plate
Zero plate
(Example: Standard thickness
plate of 2.00 mm th.)
↓
Match the indicated value on the LCD ⑦to the thickness of the standard thickness plate
by pressing the「
≪
」・「
≫
」key ④.
(This operation is possible for both the probe is contacting to the standard thickness plate or not.)
↓←When the「
≪
」・「
≪
」key ④is pressed, [CAL]
is indicated on the upper left of the LCD ⑦with a
beeping sound of the buzzer and indication
[YYY]of the left figure changes.
The Indicated values traverse quickly by keeping on
the「
≪
」・「
≪
」key ④pressing.
↓←When the indication matches to the thickness of the
standard thickness plate, stop the operation of
the 「
≪
」・「
≪
」key ④.
↓
The Meter is ready for measurement.
Press the probe several times against the standard thickness plate placed on the zero plate.
When the indication on the LCD ⑦is the same thickness or near with the standard thickness plate,
the calibration is acceptable.
When the indication deviates from the thickness of the standard thickness plate, repeat above procedures
several times.
【NOTICE 】
●Wait 10 to 15 minutes ※after power ON for the adjustment or calibration to raise the accuracy.
※Power is cut off in about 5 minutes by the working of the auto power OFF function.
In this case, turn ON the POWER key ①again and fully warm up the Meter.
●Carry out the「zero adjustment」and the「standard calibration」to confirm the accuracy,
even if on the way of the measurement.
●When the new values are set for both zero adjustment and standard calibration , the last characteristic
of substrate (working curve) is erased and new set value (working curve) is memorized.
●Pay attention to avoid mis-operation by using the lock key as shown in item
「7-1. Converting the key lock mode」after completing the adjustment or calibration.
2. 15 mm
Non‑Fe
CAL YYY mm
Non‑Fe
2.00 mm
Non‑Fe

‑9‑
6‑5. Resetting the working curve
When the batteries were replaced, indication was locked or measurement, zero adjustment,
standard calibration (CAL) became impossible for operation, reset the Meter by following method.
The Meter is in power ON condition.
↓←Press the「
≪
」key ④continuously 5 times while
keepingonthe「ZERO」key ②pressing.
[CAL] is indicated on the upper left of the LCD ⑦and
[0] is indicated blinking. The buzzer emits a beeping
sound for every press of the 「
≪
」key ④.
↓
←The buzzer emits beeping sounds.
[-E-] is indicated for 3 seconds.
↓←The Meter is restored to the possible condition of
measurement.
Carry out the zero adjustment and standard
calibration again, if necessary.
【NOTICE 】
●Applicable range of the standard calibration (CAL) is 10 μmto5.00mm.
When it is out of the applicable range, [LLL] or [HHH] is indicated.
When the indication is in locked condition, operate the Meter according to
「6-5. Resetting the working curve」.
3. 20 mm
Non‑Fe
CAL 0
Non‑Fe
CAL ―E―
Non‑Fe
―μm
Non‑Fe

‑10‑
6‑6. Measurement
Be sure to pass the hand strap ⑧ around the wrist to prevent the Meter
from dropping.
When the above mentioned zero adjustment and standard calibration are completed, the Meter
is ready for the measurement.
Hold the mid-part of the probe as shown in left figure and press it against
the surface of an object to be measured vertically, quickly and calmly.
The Meter emits a beeping sound and the measured value is indicated on
the LCD ⑦.
If no sound is emitted, press again after removing the probe about 5 to 10 cm
from the surface.
※Make use of the key lock mode function to prevent from the mis-operation
while measuring.
◆When the measurement is not carried out for about 5 minutes or more after power ON,
the power will be cut off by the working of the auto power OFF function.
The last measuring condition is restored by working of the resuming function at the succeeding
power ON.
Correct usage Incorrect usage
Hold the mid‑part of the probe and Don't tilt the probe Don't slide the probe Don't press the probe
press it down vertically and calmly against an object to side ways while pressing. against an object to be
against the surface of an object to be measured. Both the probe and an measured too slowly.
be measured. The measured value is The accurate measured object to be measured Measurement error will
indicated with a beeping sound. value cannot be will be damaged. cause to be enlarged.
If no sound is emitted, measure obtained.
again after removing the probe
about 5 to 10cm from the surface.
7. ADDITIONAL FUNCTIONS
7‑1. Converting the key lock mode
Press the「LOCK」key ③in power ON condition.
↓
Then the buzzer emits three beeping sounds.
All keys except the「power」key ①can not be operated to prevent from mis-operation.
↓
To release the locking function, power OFF once and power ON again.

‑11‑
8.REPLACING THE BATTERY
When the battery closes to the limit of use
due to consumption,「LOBAT」is indicated on
the lower left of the LCD ⑦.
As it causes malfunction under this condition,
replace the batteries earlier.
Replace the batteries after surely power OFF.
When the batteries are replaced under power ON, (Indication of battery's voltage drop)
it causes out of order of the Meter.
Replace all 4 batteries with fresh specified dry
batteries.
(Battery: Dry battery R03( 1.5V) × 4 pcs)
9.MAINTENANCE AND INSPECTION
●Use the Meter within the range of 0 to 40 ℃.
Avoid to expose the Meter to dew condensation, wet with water, dust, intense heat, vibration, etc. in use.
●Handle the probe cautiously not to damage on its tip.
●Keep the Meter in a dust-free place where high temperature and high humidity can be avoided.
●In case the Meter is left on no use for 1 month or more, take the batteries out of the Meter.
●To ensure accurate measurement, regular inspection of the Meter at least once a year is recommended.
XXX μm
LOBAT Non‑Fe

‑12‑
10. CARES TO RAISE AN ACCURACY OF THE MEASUREMENT
①Zero plate
For use in the zero adjustment and standard calibration(CAL), prepare the zero plate which is
same kind, thickness and shape as the substrate of an object to be measured.
It will result in an inaccurate measurement to use the zero plate which is different
specifications from an object to be measured.
※Attached zero plate (For eddy current: Aluminum 1050) for test is for operation
test of the Meter.
Prepare the substrate of an object to be actually measured.
②Standard thickness plate
Carry out the standard calibration (CAL) with a little thicker standard thickness plate than a film
thickness of an object to be measured.
※It will cause an erroneous measurement to use a too thicker or too thinner
standard thickness plate than a film thickness of an object to be measured.
When the standard thickness plate has been damaged or bent, renew with a new plate.
In case the standard thickness plates other than the attached plates are necessary, please contact us.
(15μm th. or more)
③Characters of a film
The measurement of the film having a metal substance causes a measurement error.
With respect to an elastic film, place a standard thickness plate with 30 to 50 μmonitandsubtract
the thickness of the standard thickness plate from the total thickness, then the measurement error
due to film dent can be prevented.
④Effect of edge and corner
The edges, corners and their surroundings of an object to be measured are the places where the
condition of a magnetic flux is not uniform.
Generally, measure inside taking 15 to 20 mm or more away from the edges.
The same care is needed for a projection, bend, sharp deformed portion and surrounding.
⑤Effect of surface roughness
The surface roughness of both substrate and film of an object to be measured affects the measurement
results.
Take several spots to measure and calculate the mean value.
⑥Effect of rolling
In some cases, rolling strains exist in a substrate and it causes an incorrect measurement
depending on the spots to be measured.
Then take several spots to measure and calculate the mean value.
⑦Effect of the temperature
The range of the operating temperature is within 0 to 40 ℃.
Especially, when the temperature of the probe much differs from the Meter, it causes an erroneous
measurement.
Ver.4 April 2007

Show rooms:
You are welcomed to the show rooms located at the
following places.
Tokyo show room near the Otemachi station of the subway・Osaka show room at Tenjinbashi-kitazume・Nagoya show room near the Kurokawa station of the subway・Fukuoka show room near the Gofukucho station of the subway・
Products sold: ,,Sales of Coating thickness meter Pinhole detector
,, ,Condensator Viscosity cup Moisture meter
Needle detector Iron piece detector,
Manufacturer:
Sanko Electronic Laboratory Co., Ltd.
.,264, , ,Tokyo branch Shibata Bldg - - Uchikanda Chiyoda-ku
Tokyo - Japan101 0047,
81 3 3254 5031 81 3 3254 5038Tel--- Fax---
Osaka branch Konishi Bldg - Sugawara-cho., 2 3,
Kita-ku Osaka - Japan, 530 0046,
81 6 6362 7805 81 6 6365 7381Tel--- Fax---
., 31127, ,Nagoya branch Meihoku Bldg - - Kinjo
Kita-ku Nagoya - Japan, 462 0847,
81 52 915 2650 81 52 915 7238Tel - - - Fax - - -
Fukuoka branch - Naraya-cho Hakata-ku11 11 ,
Fukuoka - Japan812 0023,
81 92 282 6801 81 92 282 6803Tel - - - Fax - - -
1677 , ,Head office Hisasue Takatsu-ku
Kawasaki - Japan213 0026,
81 44 751 7121 81 44 755 3212Tel - - - Fax - - -
Table of contents
Other SANKO Measuring Instrument manuals

SANKO
SANKO SWT-7000 III User manual

SANKO
SANKO SAMAC-F User manual

SANKO
SANKO MR-200 II User manual

SANKO
SANKO SP-3300D User manual

SANKO
SANKO PM-101 User manual

SANKO
SANKO SWT-7200 II User manual

SANKO
SANKO SWT-7000 III User manual

SANKO
SANKO SWT-7200IV User manual

SANKO
SANKO MR-300 User manual

SANKO
SANKO AQ-30 User manual