SANKO SWT-7000 III User manual

SANKO Coating Thickness Meter
SWT-7000Ⅲ/7100Ⅲ
Instruction Manual
CAUTIONS:
Before using the Meter, read this INSTRUCTION MANUAL
thoroughly and use the Meter correctly.
Keep this INSTRUCTION MANUAL carefully and refer to this
when necessary.
In the event of any doubt arising, the original
INSTRUCTION MANUAL (Japanese) is to be of final authority.
SANKO ELECTRONIC LABORATORY CO., LTD.
Tokyo・Osaka・Sendai・Nagoya・Fukuoka・Kawasaki
December 2014

Contents Pages
Attention for safety ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 1
Notes ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 3
Get started ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 4
・ Contents in a package ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 4
・ Names of part ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 5
・ How to fit batteries ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 6
・ How connect, disconnect probes ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 7
・ How to hold probes ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 8
・ How to press a probe to an object ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 8
How to operate ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 9
(1) How to switch power source ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 9
(2)How to switch off ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 9
(3)Zeroing ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 10
(4)Calibration standard (CAL) ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 12
(5)Zeroing ―in special cases ―・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 14
(6)2-point calibration in case it is difficult to perform zeroing・・・・・・・・・・・・・・・・・・・・・ 17
(7)How to delete calibration ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 22
Measuring ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 23
Function switching ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 24
(1)Switching to Non-interupt measurement mode ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 24
(2)Switching to resolution ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 26
(3)Setting to Auto-Power-OFF ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 27
(4)Switching to a mode for an optional substarate with FN-325 probe・・・・・・・・・・・・・・ 29
(5)Switching Backlight (ON, OFF) ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 30
Transferring data (SWT-7100Ⅲ)・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 31
Note to improve measuring accuracy・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 32
Trouble shooting ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 33
Specifications ◆ main unit ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 35
◆ probe (optional)・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 36
References(measurement principles) ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 39

Attention for safety(to use in correct ways)
To prevent you and your properties from damaging please take some time to read thoroughly this “Attention
for Safety” and correct uses keep these instructions attentive to read when necessary.
Warning
● Do not dump nor wet the gauge unit in water, otherwise it may cause damages.
Keep off water Please contact our distributor or sales office should submerged water into the unit.
● Keep metals or foreign substances from the unit, otherwise it may cause damages.
Prohibition Please contact our distributor or sales office should put any materials or foreign substances in the unit.
● Do not insert a screwdriver into the connector, otherwise that it may cause damages.
Prohibition
● Do not throw, smash, drop the unit, otherwise it may cause damages.
Prohibition
● Never dismantle or modify the gauge unit by yourself, otherwise it may cause errors or damages.
Dismantling prohibited
● Do not use AC adaptors other than an exclusive adaptor for this unit, (SWT-7100Ⅲ only)
Prohibition otherwise it may cause damages, electric shocks, fires.
● Please use the exclusive adaptor with a designated Voltage only, otherwise it may cause
Prohibition damages, electric shocks, fires.
● Keep the terminals of the adaptor free from metal pieces or dust, otherwise it may cause short circuit,
Prohibition electric shocks, fires.
● Do not handle the AC adaptor with wet hands, otherwise it may cause electric shocks.
Prohibition
● Do not damage, brake, modify, forcefully bend or twist the cord of the AC adaptor. Or, do not load it
Prohibition with heavy staff or pinch it forcefully, otherwise it may cause breaking wires, short circuit, fires.
-1-

Attention For Safety(to use safely and correctly)
Warning
! ● Never fail to remove batteries from the unit when not in use for a long time.
Must Leakages occurred from deterioration of batteries may cause erroneous reactions or damages.
!
● Be sure to read this book on the item of 「How to fit batteries」 to replace batteries.
Must
!
● Store batteries in a place where children and pets are incapable of handling them.
Must Please call a doctor like in a case that a battery is swallowed.
!
● Do not put batteries into fire or water. Store them in a cool, dry and dark place avoiding flames, high
Must temperature and moisture.
!
● Do not get batteries shocked and dismantled, and soldered for processing
Must
!
● Do not short or recharge batteries and handle with metallic tools like pliers.
Must
! ● Replace with new (unused) batteries according to the procedure of this Operating Instructions.
Must
! ● Be sure of paying attention on battery polarity marks, (+、-) to place the batteries.
Must
! ● In case a battery has leakage please clean up the place with clothes to replace batteries.
Must And do not touch the leaked liquid and wash skins or clothes in case they are contaminated.
! ● Comply to regulations and laws in your neighbors when disposing of them.
Must
! ● Insert a plug of the AC adaptor to the full end, otherwise it may cause electric shocks and fires with
Must burns. And do not use faulty or loose receptacles.
! ● Switch to OFF and unplug the AC adaptor from the receptacle to avoid electric shocks and
Must damages when inspecting or cleaning the gauge.
-2-

Attention For Safety(to use safely and correctly)
Attention
● Do not use Benzene or Thinner for cleaning and spray pesticides on the meter, otherwise it may
Prohibition cause cracks or malfunctions.
● Do not store the meter in places getting high in temperatures such as in a car in strong sunlight or near
Prohibition heaters, otherwise it will be hazardous to the meter and may cause malfunctions.
● Do not step, trample down nor put anything on the meter.
Prohibition
● Keep the meter away off rubber-made articles or vinyl articles. A lengthy contact between
Prohibition meter and them may cause stickiness and it may be difficult to get rid of them.
Notes:
● Please read this manual thoroughly for correct operations before getting started.
● This meter is a precision gauge. Please handle with care.
● Do not tug, bend, fold or curl up forcefully the cables of probes.
● Do not knock or scratch objects with the tip of a probe.
● Keep the tip of a probe clean. A slight amount of dust may cause errors in measurements.
● Clean the meter and store it in free from dust and moisture after operation.
● To keep precision with a gauge please contact our distributor or our sales office once a year for inspection
● Keep the meter away off electric noises, shocks or magnetic fields when in a use.
-3-

Get started
◆ Contents in a package Please make sure if the following items sre included.
・ Main unit
SWT-7000Ⅲ, or SWT-7100Ⅲ
・ Dry batteries LR6 (2 pieces)
・ Carrying case
・ Instruction manual (this manual)
・ AC adaptor (SWT-7100Ⅲ only)
・ USB cable (2.0m) (SWT-7100Ⅲ only)
・ USB driver (CD) (SWT-7100Ⅲ only)
● In case there are optional probes included.
・ Probes
For ferrous(Fe), non-ferrous(NFe), for both Ferrous/non-ferrous (FN-325)
・ Zero boards for testing (for Fe:ferrous substrates/for NFe:non-ferrous substrates
/for FN:either ferrous or non-ferrous substrates, 2 pieces.)
・ Thickness standards (films :2 foils、bake:1 sheet)
-4-

◆ Names of part
Probe socket
Display
Power ON/OFF key
Hand strap
Keys
Battery compartment (the back)
Socket for AC adaptor connector
(SWT-7100Ⅲ only)
Socket for USB cable connector
(SWT-7100Ⅲ only)
● Probe socket
Connect an optional-exclusive SWT probe to the probe socket.
(1) To measure a film thickness of coated, plated, lining layer on substrates made of ferrous material
please use a probe of (Fe) series for the connection.
(2) To measure a film thickness of coated, plated, lining layer on substrates made of non-ferrous materials
such as Aluminum, Copper, etc. please use a probe of (NFe) series for the connection
(3) To measure a film thickness on either ferrous or n on-ferrous substarate please use a probe of (NF-325)
f or connection.
● Reading display
It indicates measurement results, operation guides, or malfunction status. With a backlight.
● Power ON/OFF key
It switches On or Off.
● Keys
(1) 「ZERO」key、「▲ 」key、「▼」key
They are adjusting keys to be pre-used before measuring to obtain correct results.
(2) 「CAL/DELETE」key
CAL: Initiate or finish adjustment by using standard folils.
DELETE: Deletes incorrect or unnecessary measuring results for adjustment.
(works only when 「ZERO」、「CAlLIBRATION」 is processed.)
● Battery compartment
It contains 2 pieces of dry battery (LR6).
● Hand strap
Hang the meter through a strap over your wrist never to drop it.
● Socket for AC adaptor (SWT-7100Ⅲ only)
This is a socket connected to the exclusive AC adaptor (accessory).
● Socket for USB cable (SWT-7100Ⅲ only)
It is a socket connected to a USB cable (accessory)
-5-

◆ How to fit batteries
① Open the battery lid on the back of the unit.
Press down and slide the lid in direction of arrow to open.
② Insert batteries.
Ensure correct battery polarity ○
+、 ○
- for placement.
③Close the lid.
Caution
● Use designated and new (check battery-life)) batteries or ones supplied in this package.
● An incorrect use of batteries may cause leakages, bursts. Do not intermingle new ones
with old ones.
● Take out batteries to store when not in use for a long absence, and that may avoid
Leakages.
● Keep batteries off children and pets.
● Comply to the laws and rules in your Local Authorities when disposing of batteries.
◎ About reading display
When placing batteries in the unit, the messages and warning below on the screen may be indicated.
And these are not breakdowns, wait until the reading disappears with a beeping sound.
Batteries have run out when the display on the unit indicates the mark listed below.
Replace with new batteries.
In case of using FN-325 probe:
After Power source switched to OFF at Automode,
AUTO is indicated on the display and when Fe or NFe substrate
is set at mode corresponding to substrate, “ ” for a
ferrous substrate, “ ” for a non-ferrous substarate is
indicated on the display.
A use of exclusive Fe-probes shows “Fe” and of exclusive
NFe-probes does “NFe ” on the display.
-6-
(AUTO)
-
μm
START UP
PHASE
Hold Probe
in the air.
Don‘t touch it
on any
metallic object.
ERROR!
Connect a probe
before
switched on.
BAT
NFe
Fe

◆ How to connect or disconnect a probe
◆ Connect an optional, exclusive SWT probe to the main unit.
Select one of the probes suited for your application.
Insert a probe connector into the probe socket of the main unit.
Make sure of aligning the keyway leading to a smooth joint without doing by force.
Insert and push it untill it is locked.
Clasp
Probe connector
Probe socket
Probe socket in the main unit
◇ Remove the exclusive probe from the unit.
Pull off the probe carefully by bending inward clasp springs at the both ends of the probe connector to
release the clasps.
Do not pull off by force or it may cause damages.
Caution
Make sure that Power switches to off when connecting or disconnecting the exclusive
SWT probes.
Or else, it may cause damages to connect or disconnect while Power is on.
-7-

Measurement object Measurement object
◆ How to hold probes
Cable
Pinch and hold this part using fingers (thumb and forefinger)
Black resinous ring
Tip to press against a measurement object
◆ How to press a probe to an object
・ Keep the probe 5 cm or over away off metallic objects when not in use of measuring.
・ Press the tip of the probe perpendicularly against a measurement object.
Tilting may cause large errors.
○ ×
Probe
Press the probe quickly and smoothly to objects.
A slow-acting press may cause large errors.
Quickly and calmly press perpendiculary against the object by grabing the probe as illustrated.
It beeps and the reading screen shows the measuring result.
When it does not beep, lift it 5~7 cm above the object and try again to take measurements.
Caution
◆ Do not smash or hit the probe against objects, or it may cause damages to probes and to
objects
◆ Do not scrape, scrub objects with the probe except in a special measurement.
Or it may break the tip of the probe and cause damages to the tip and surface of objects
-8-

How to operate
(1) How to switch Power source
Press ON/OFF key.
This message lasts for about 3 seconds.
Caution
Hold the probe in air without operation when
the reading is on display. Or it may indicate
「ERROR」 and automatically switch off
Power.
The buzzer emits a beeping sound. The Backlight mark is indicated when ON.
In case of connecting FN-325:
After Power source switched to OFF at Automode,
AUTO is indicated on the display and when Fe or NFe
substrate is set at mode corresponding to substrate,
“ ” for a ferrous substrate, “ ” for a
non-ferrous substarate is indicated on the display.
A use of exclusive Fe-probes shows “Fe” and of exclusive
NFe-probes does “NFe ” on the display.
(2) How to switch off
Press ON/OFF key.
The buzzer beeps.
The message lasts for about 2 seconds.
Caution
Never remove a probe from a unit when
Power is ON. Or the electric shocks may
damage the probe and the unit
The buzzer beeps and this unit is switched off.
-9-
POWER
OFF
Terminating
phase
START UP
PHASE
Hold Probe
in the air
Don‘t touch it
on any
metallic object.
(AUTO)
-
μm
NFe
Fe

(3) Zeroing
It is capable of getting started on measurements immediately after the message of 「START UP PHASE」
has disappeared as described on page 9.
However, it makes errors depending on material formation and shapes to be measured. To minimize measurement
errors and obtain as accurate results as possible please be sure of carrying out 2 points of adjustments of
「Zeroing」 and 「Calibration standard」 before measuring process.
※ Please prepare for a Substrate plate the identical material, quality and size to a measuring object.
(This substrate plate should be designated as a 「Zero Plate」)
◆ Press ZERO key.
The buzzer emits a beep sound. CAL/DELETE key changes to one data “deletion” function.
Press the probe to the Zero Plate.
The buzzer beeps.
Zeroing is in process.
The reading shows that the unit is set to
measure on a substrate material of Ferrous by
connecting a 「Fe-325」 probe, or a 「Fe」 series
probe.
Measured a meanl value
When the measuring value results in largely off
the “0”, press CAL/DELETE + ▼ key to
delete the latest data out of measured values.
It indicates “—“to delete all data.
Remove the probe from 「Zero plate」 ● Repeat a measuring process 1~20 times by pressing
the probe to 「Zero plate」. (A mean value is
displayed whenever a probe is pressed)
◆ When a measuring process reaches the 20th time, the
buzzer beeps, beeps 2 times and afterward do not accept
any entry.
Press ZERO key when finishing adjustment.
-10-
【ZERO Cal.】
1.Measure subst
-rate Several
times.
2.press ZERO key.
ZERO cal (Fe)
1
μm

The buzzer beeps. It stopes its DELETION function of CAL/DELETE key
The reading shows that 「Zeroing」 has completed and
it becomes possible to take measuring and adjusting
operations of this unit.
● It is correct that numerical values measured by pressing the probe to the 「Zero Plate」 indicates 「0」 or in the
neighborhood of 「0」.
When the measured value results in largely off 「0」μm, please try again zeroing from the beginning.
● [LLLL] indicated on display during a time of zeroing means that the calibration point heavily deviates from
the standard please make sure that the material is not in process of being built with others and repeat the
zeroing in 2~4 times until a stable 「0」 is obtained.
※ In case of using a dual type probe 「FN-325」, perform 「calibration by ferrous substrate」 and 「calibration by
nonferrous」 at the same time as possible as you can.
Note : The latest measured value replaces the previous ones and the new value of 「Zeroing」 is stored.
-11-
(Fe)
0
μm

(4) Calibration standard (CAL)
● Pepare 「Zero Plate」 used for 「ZEROing」
● Prepare 「Thickness standard」、foiles that are the same thick as the measuring film or thicker than that.
● Place the 「Thickness standard」 on the 「Zero Plate」.
● Press the CAL/DELETE
The buzzer beeps. CAL/DELETE key changes to one data “deletion” function.
Press again CAL/DELETE key to interrupt Calibration
standard (CAL).
It beeps whenever pressing.
Calibration standard is in process.
The reading shows that the unit is set tomeasure
on a substrate material of Ferrous by connecting a
「Fe-325」 probe, or a 「Fe」 series probe.
Measured a meanl value
When the measuring value results in largely off
the “0”, press CAL/DELETE + ▼ key to
delete the latest data out of measured values.
It indicates “—“to delete all data.
● Repeat a measuring process 1~20 times by pressing
the probe to 「Zero plate」. (A mean value is
Remove the probe from 「Zero plate」 displayed whenever a probe is pressed)
◆ When a measuring process reaches the 20th time, the
buzzer beeps, beeps 2 times and afterward the unit does not
accept any entry.
To adjust the displayed reading until it is correct relative
to thickness standared (in this case,100μm) use.
▲ , ▼ key. Pressing ▲ , or, ▲ key interrupt the delete function of
CAL/DELETE key.
Pressing ▲ key increases the displayed value.
Pressing ▼ key decreases the displayed value.
(Press and holding key goes fast forward)
Adjust the reading value to the Standard thickness.
-12-
Cal. w. fo (Fe)
102
μm
Cal. w. foil (Fe)
100
μm
【 CAL. W. foil】
1.Measure thick
-ness of foil
several times.
2.With ▼ or ▲ , Ad
-just to thick
-ness of foil
3.press CAL key.

After adjusting the reading value to the Standard thickness, press CAL/DELETE key.
The buzzer beeps, 【Calibration Standard】 on the upper left disappears and
it returns to a measuring mode.
This is a measusring mode.
● It is correct that numerical values measured by pressing the probe a few toimes to the 「Standard theickness」 on
the 「Zero Plate」 indicates the thickness iin the neighborhood of the 「Standard thickness」.
● When the measured value results in largely off the 「Standard thicknesws」 please try again zteh calibration from
the beginning.
※ In case of using a dual type probe 「FN-325」, perform 「calibration by ferrous substrate」 and 「calibration by
nonferrous」 at the same time as possible as you can.
Note:
The lastest measured value replaces the previous ones and the new value of 「Calibration standard」 is stored.
-13-
(Fe)
100
μm

Substrate material
(1
st layer)
(2
nd layer)
(3
rd layer)
(5)Zeroing in special cases (Multi-layers)
◎ In case of being painted as shown with multi-layers on the substrate there may be needs to measure thicknesses of
each layer. For example, measuring only the thickness of the 4th layer please zero as an assumed ZERO at the surface
of the 3rd layer stacked on the substrate.
Measuring the thickness of the 4
th layer
by zeroing at the 3
rd layer.
Coating films
◆ Releasing of special-case zeroing
When zeroing again on the substrate after having finished the above measurements and if the combined thickness of
3 coating layers from 1st to 3rd exceeds 50 μm, please zero the meter on the following procedures.
If the thickness of 3 combined layers is bellow 50μm, take the same procedure as usual zeroing to release.
● Prepare the identical material quality, plate size to a measuring object.
(This is a designated as a Zero Plate)
Press
ZERO key.
The buzzer emits a beeping sound. CAL/DELETE key changes to one data “deletion” function.
Press the probe to the Zero Plate.
The buzzer beeps, beeps, beeps.
-14-
(4
th layer)
【ZERO Cal.】
1.Measure subst
-rate Several
times.
2.press ZERO key.

Press ZERO key . -- ①
The buzzer beeps.
Press ZERO key. -- ②
The buzzer beeps.
Press the probe to Zero plate.
The buzzer beeps.
The reading shows that it is process of zeroing
adjustments
Measured a mean value
When the measuring value results in largely off
the “0”, press CAL/DELETE + ▼ key to
delete the latest data out of measured values.
It indicates “—“to delete all data.
Remove the probe from 「Zero plate」 ● Repeat a measuring process 1~20 times by pressing
the probe to 「Zero plate」. (A mean value is
displayed whenever a probe is pressed)
To finish press Zero key ◆ When a measuring process reaches the 20th time, the
buzzer beeps, beeps 2 times and afterward the unit does not
accept any entry.
-15-
OFFSET.
To continue,
press ZERO twice.
【ZERO Cal.】
1.Measure subst
-rate Several
times.
2.press ZERO key.
ZERO cal (Fe)
1
μm

The buzzer beeps. It stopes the DELETE function of CAL/DELETE key.
「Zeroing」 has completed (the zeroing letters
disappeared) and it becomes possible totake measuring
and adjusting operations of this unit.
● It is correct that numerical values measured by pressing the probe to the 「Zero Plate」 indicates 「0」
or in the neighborhood of 「0」.
When the measured value results in largely off 「0」μm, please try again zeroing from the beginning.
● [LLLL] indicated on display during a time of zeroing means that the calibration point heavily deviates from
the standard please make sure that the material is not in process of being built with others and repeat the
zeroing in 2~4 times until a stable 「0」 is obtained.
Note:
The latest measured value replaces the previous ones and the new value of 「Zeroing」 is stored.
-16-
(Fe)
0
μm

(6) 2-foil calibration when “Zeroing “ is difficult to perform.
In case zeroing is difficult to perform such as measuring the thickness of the film on the rough surface of Blast-steel plates,
a calibration method using 2 different thicknesses of standard plates pinching a thickness of the object is defined as
「JIS K5600」Standard. This calibration method complies to the regulations.
Caution
It is not possible to use both this calibration method and other calibration ones together, or mixing them together.
Should were the methods taken, measuring results could be the wrong values.
Prepare the same blast-steel-plate in material as the objective base or, a rough face on non-ferrous base like
aluminum and 2 different thicknesses of Thickness standards.
Please choose the suitable difference of thickness standards from the list below.
Predicting film thickness Difference of thickness standard
~ 49.9μm 10μm or over
50.0 ~ 99.9μm 25μm or over
100.0 ~ 499.9 μm 50μm or over
500 ~ 999 μm 199μm or over
1.00 ~ 3.00 mm 0.5mm or over
3.01mm ~ 2.0mm or over
Press and hold the ZERO key for 3 seconds.
CAL/DELETE key changes to one data “DELETE” function.
The buzzer beeps.
Stack the thicker 「Standard thickness」 on the substrate and
press the probe on it.
The buzzer beeps whenever pressing the probe.
It reads that 2-foil calibration is under process.
It reads that a probe of 「FN-325」 or (Fe) series is connected to
measure on a ferrous substrate.
Mean value
In case a measured value abnormally deviates, a press of
CAL/DELETE key + ▼ deletes the latest entry and displays
the second latest.
All deletion indicates “ - “ on the reading.
- 17 -
【2-foil’s cal】
Meseure thinner
foil on metal
several times.
Afterwards,
press ZERO key.
【2 - f. c. 】 (Fe)
27
μm

●Try measuring plural times in range of 1―20 times by pressiing
the probe on a thinner 「Thickness Standard」.
( a mean value is indicated whenever the probe is pressed )
Press ZERO key afterwards. ◆When repeating measurements 20 times, the buzzer beeps, beeps
2 times and afterward the unit does not accept any entry.
The buzzer beeps, beeps.
A press o f ▲ or ▼ key makes the buzzer
beep and a last mesuaring result appear on the reading.
Mean value
Equate this reading value with thickness Standard (in this example, 25μm)
by pressing ▲ or ▼ key.
A press of ▲ key increase a reading value.
A press of ▼ key decreases a reading value.
(Press and holding key goes fast forward)
Equate this value with the Thickness Standard.
After equating the reading value with the thickness standard , press ZERO key.
The buzzer beeps.
- 18–
【2-foil’s cal】
Use ▲ or ▼ ,
To set foil
thickness.
Afterwards,
press ZERO key.
【2 - f. c. 】 (Fe)
25
μm
【2 - f. c. 】 (Fe)
27
μm
Other manuals for SWT-7000 III
1
This manual suits for next models
1
Table of contents
Other SANKO Measuring Instrument manuals

SANKO
SANKO SWT-7000 III User manual

SANKO
SANKO SWT-7200 II User manual

SANKO
SANKO MR-200 II User manual

SANKO
SANKO SAMAC-F User manual

SANKO
SANKO MR-300 User manual

SANKO
SANKO SWT-7200IV User manual

SANKO
SANKO EDY-5000 User manual

SANKO
SANKO PM-101 User manual

SANKO
SANKO SP-3300D User manual

SANKO
SANKO AQ-30 User manual