manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. SATO KEIRYOKI
  6. •
  7. Measuring Instrument
  8. •
  9. SATO KEIRYOKI 640S-1 User manual

SATO KEIRYOKI 640S-1 User manual

This manual suits for next models

1

Other SATO KEIRYOKI Measuring Instrument manuals

SATO KEIRYOKI S270WP Series User manual

SATO KEIRYOKI

SATO KEIRYOKI S270WP Series User manual

SATO KEIRYOKI PC-5110 User manual

SATO KEIRYOKI

SATO KEIRYOKI PC-5110 User manual

SATO KEIRYOKI PC-5400TRH User manual

SATO KEIRYOKI

SATO KEIRYOKI PC-5400TRH User manual

SATO KEIRYOKI 1410-00 User manual

SATO KEIRYOKI

SATO KEIRYOKI 1410-00 User manual

SATO KEIRYOKI SK-620PH User manual

SATO KEIRYOKI

SATO KEIRYOKI SK-620PH User manual

SATO KEIRYOKI SK-130ITH User manual

SATO KEIRYOKI

SATO KEIRYOKI SK-130ITH User manual

SATO KEIRYOKI SK-50TCH User manual

SATO KEIRYOKI

SATO KEIRYOKI SK-50TCH User manual

SATO KEIRYOKI Sophia User manual

SATO KEIRYOKI

SATO KEIRYOKI Sophia User manual

SATO KEIRYOKI SK-5RAD-SP User manual

SATO KEIRYOKI

SATO KEIRYOKI SK-5RAD-SP User manual

SATO KEIRYOKI PC-7700 II User manual

SATO KEIRYOKI

SATO KEIRYOKI PC-7700 II User manual

SATO KEIRYOKI 7720-00 User manual

SATO KEIRYOKI

SATO KEIRYOKI 7720-00 User manual

SATO KEIRYOKI SK-M460-TRH User manual

SATO KEIRYOKI

SATO KEIRYOKI SK-M460-TRH User manual

SATO KEIRYOKI PC-7700II User manual

SATO KEIRYOKI

SATO KEIRYOKI PC-7700II User manual

SATO KEIRYOKI 7542-00 User manual

SATO KEIRYOKI

SATO KEIRYOKI 7542-00 User manual

SATO KEIRYOKI SK-660PH Series User manual

SATO KEIRYOKI

SATO KEIRYOKI SK-660PH Series User manual

SATO KEIRYOKI SK-93F-II User manual

SATO KEIRYOKI

SATO KEIRYOKI SK-93F-II User manual

SATO KEIRYOKI WVAT6D0-3-10 User manual

SATO KEIRYOKI

SATO KEIRYOKI WVAT6D0-3-10 User manual

SATO KEIRYOKI SK-181GT User manual

SATO KEIRYOKI

SATO KEIRYOKI SK-181GT User manual

SATO KEIRYOKI PC-5000TRH II User manual

SATO KEIRYOKI

SATO KEIRYOKI PC-5000TRH II User manual

SATO KEIRYOKI PALMA 2 User manual

SATO KEIRYOKI

SATO KEIRYOKI PALMA 2 User manual

SATO KEIRYOKI SK-28V User manual

SATO KEIRYOKI

SATO KEIRYOKI SK-28V User manual

SATO KEIRYOKI SK-100R User manual

SATO KEIRYOKI

SATO KEIRYOKI SK-100R User manual

SATO KEIRYOKI 1518-00 User manual

SATO KEIRYOKI

SATO KEIRYOKI 1518-00 User manual

SATO KEIRYOKI 1722-00 User manual

SATO KEIRYOKI

SATO KEIRYOKI 1722-00 User manual

Popular Measuring Instrument manuals by other brands

Powerfix Profi 278296 Operation and safety notes

Powerfix Profi

Powerfix Profi 278296 Operation and safety notes

Test Equipment Depot GVT-427B user manual

Test Equipment Depot

Test Equipment Depot GVT-427B user manual

Fieldpiece ACH Operator's manual

Fieldpiece

Fieldpiece ACH Operator's manual

FLYSURFER VIRON3 user manual

FLYSURFER

FLYSURFER VIRON3 user manual

GMW TG uni 1 operating manual

GMW

GMW TG uni 1 operating manual

Downeaster Wind & Weather Medallion Series instruction manual

Downeaster

Downeaster Wind & Weather Medallion Series instruction manual

Hanna Instruments HI96725C instruction manual

Hanna Instruments

Hanna Instruments HI96725C instruction manual

Nokeval KMR260 quick guide

Nokeval

Nokeval KMR260 quick guide

HOKUYO AUTOMATIC UBG-05LN instruction manual

HOKUYO AUTOMATIC

HOKUYO AUTOMATIC UBG-05LN instruction manual

Fluke 96000 Series Operator's manual

Fluke

Fluke 96000 Series Operator's manual

Test Products International SP565 user manual

Test Products International

Test Products International SP565 user manual

General Sleep Zmachine Insight+ DT-200 Service manual

General Sleep

General Sleep Zmachine Insight+ DT-200 Service manual

Sensa Core Lacto Spark user manual

Sensa Core

Sensa Core Lacto Spark user manual

VOLTCRAFT UTS-1980 operating instructions

VOLTCRAFT

VOLTCRAFT UTS-1980 operating instructions

EBCHQ 94915 Operation manual

EBCHQ

EBCHQ 94915 Operation manual

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Balluff

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Pulsar IMP Lite Series instruction manual

Pulsar

Pulsar IMP Lite Series instruction manual

ETCR 6800 user manual

ETCR

ETCR 6800 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Instruction Manual
Standard Probe Model 640S-1
Introduction
Thank you for purchasing the standard probe Model 640S-1.
-This probe is designed to measure hydrogen ion concentration (pH) and temperature (liquid temperature)
connecting with the pH meter Model SK-640PH or SK-650PH. Do not use this product for any other
purposes.
-Read this instruction manual thoroughly before using and keep it in a safe place for your future references.
Names of Section
①Electrode membrane ⑤Internal reference electrode ⑨pH sensor connection plug
②Glass internal electrode ⑥Electrode holding tube ⑩Tem. sensor connection plug
③Liquid junction ⑦Refill port for the inner solution ⑪Electrode cap
④Temperature sensor ⑧Sensor cord ⑫Electrode protection cap
Precautions before use
When the unit is first used or after it has been stored for several days, the pH measurement response may
be slower. Allow the unit to adjust to the ambient environment before use as follows.
(1) Remove the electrode cap from the unit.
Note: The electrode cap is filled with pure water to prevent the glass electrode membrane from drying.
The pure water may spill when removing the cap.
(2) Immerse the glass electrode in the pH standard solution (pH7) or water for two hours or more before use.
Using the 640S-1
(1) Remove the electrode cap from the unit.
- Hold the unit upright to avoid spilling the inner solution.
- To protect the electrode, it is recommended to attach the protective cover to the tip of the unit.
(2) Flip open the cap for the refill port.
(3) Connect the unit to the indicator, and then turn on the indicator.
(4) Immerse the glass electrode in the measurement sample to a depth of about 8 cm.
- Calibrate the unit before measurement as necessary.
11
12
1
2
3
4
5
6
7
8
9
10
(5) Upon completion of measurement, turn off the indicator.
(6) Close the cap for the refill port.
(7) Before storing, thoroughly wash the glass electrode with distilled, pure or tap water and cover the
electrode with the cap filled with water.
* For details on handling the indicator, refer to the instruction manual provided with the indicator.
Maintenance
1. Glass electrode: When the glass electrode becomes dirty, clean it as follows.
- For organic stains
Lightly wipe the electrode with gauze or cotton wool damped with a diluted neutral detergent, and
then rinse the electrode well with pure water.
- For inorganic stains
Immerse the glass electrode in an approximately 0.1N hydrochloric acid solution or a neutral detergent,
and then rinse the electrode well with pure water. Do not leave the electrode in the solution for a long time.
Note: Make sure that the neutral detergent does not come in contact with the junction. Otherwise, the
performance of the inner solution could be adversely affected if the neutral detergent penetrates inside
the sensor.
2. Reference electrode
(1) When the inner solution level falls below the position of the internal reference electrode, refill the
electrode with the inner solution. When refilling, pour an appropriate amount of the provided inner
solution through the refill port on the unit.
(2) When the inner solution becomes dirty, first remove the dirty inner solution from the holding tube,
thoroughly clean the inside of the tube with new inner solution, and then fill the electrode with new inner
solution. The inner solution should be replaced at regular intervals.
Important: The unit cannot be repaired due to its characteristics and structure. If the glass electrode is
damaged or its performance is deteriorated, replace with a new sensor probe.
Caution
For proper use of the unit, be sure to observe the following:
- The sensor is made of glass and is a precision instrument; be careful not to drop or knock it.
- If the unit is used in an environment where electrical noise is generated, the display may become unstable
or the measurement error may increase.
- Do not alter, pull forcibly, bend or bind the cord, as doing so may cause a break or malfunction in the probe.
- Never use the unit outside the measurement range, otherwise it may malfunction.
- When measuring, do not let the sensor probe come in contact with the human body.
- The glass electrode could be damaged or its service life could be shortened if it is used for measurement
samples such as organic solvents, oils, adhesives and surface-active agents.
- The pH reading may become unstable when pure water is measured.
- If the pH standard solution contacts the skin (hand), immediately wash the contaminated skin with running
water. If the solution gets into the eyes, immediately rinse with water and consult a doctor.
- Keep the pH standard solution out of reach of children. If accidentally ingested, consult a doctor immediately.
- If the unit is left in a car in the hot summer sun, it will become extremely hot and may malfunction. Do not
leave the unit in such a place.
Specifications
Product
Standard probe
Cat. No.
6424-00
Model
640S-1
pH
Temperature
Measuring Range
2.00 to 13.00pH
0.0 to 50.0°C
Accuracy
±0.03pH
±0.5°C
Sensing Element
Glass electrode
Thermistor
Liquid junction
Sleeve type
Inner solution
3.3mol/L (KCL) (refillable)
Operation ambient
Temperature: 0.0 to 50.0°C
Materials
Electrode holding tube: glass (lead-free), Cord: PVC
Dimensions
approx. 12mm dia. x 152mm (cord length: 1m)
Weight
About 78g
Accessories
Cap for electrode 1 pc.
Electrode protection cap 1 pc.
Refill of internal solution (3.3mol/L KCL 35mL) 1 ea.