SciCan HYDRIM M2 User manual

HYDRIM®M2/M2 G4
HANDPIECE CLEANING SYSTEM (LCS)
• Operator Manual
• Manuel de l’opérateur
• Bedienungsanleitung
• Manual de usuario
• Manuale operatore
• 操作手册
• Brugermanual
• Bedieningshandleiding
• Instrukcja obsługi
• Manual de Utilização
• Manual de utilizare
• Användarmanual
• Керівництво з експлуатації
95-114408 EU ML R4 HYDRIM M2/M2 G4 Handpiece Cleaning System Operator Manual.
Copyright 2021 SciCan Ltd. All rights reserved


ENGLISH
HYDRIM®M2/M2 G4 Handpiece Adaptor (LCS) Operator Manual
Coupling
MULTIflex - p/n 01-113930
or
INTRAmatic - p/n 01-113931
with adaptor
(not included)
Luer Lock Fittings
Handpiece
(not included)
Connecting handpieces to the LCS adaptor
Insert the handpiece coupling into the handpiece. Push the handpiece into one
of the ports.
1 2
2
3
Multiflex
Intramatic
Push tab
to release
Program settings
To ensure maximum wash performance results of the
handpieces and the other instruments, the Technician
needs to increase the wash time to the maximum setting
of 15 minutes for the cycle.
Note: The overall cycle time will increase accordingly by
about 6 minutes.
1
Please note: As recommended by leading manufacturer/suppliers of handpieces, the hollow parts of instruments should be dried immediately after cleaning and disinfection using
medical compressed air before additional processes or sterilization. This is to be followed by recommended regular maintenance procedures and application of maintenance
agents/oils. Please follow all applicable health & safety regulations for your region.
Using the touchscreen menu
The HYDRIM will
automatically detect
the presence of the
LCS adaptor. Select
the desired program
and follow the prompts.
Upon completion, open
the door and remove
the handpieces from the
adaptor by pushing on the
tab at the side of the port.
Example: Regular cycle Please note: As recommended
by leading manufacturer/suppliers
of handpieces, the hollow parts
of instruments should be dried
immediately after cleaning and
disinfection using medical compressed
air before additional processes or
sterilization. This is to be followed by
recommended regular maintenance
procedures and application of
maintenance agents/oils. Please
follow all applicable health & safety
regulations for your region.
12 3

ENGLISH
Maintenance and troubleshooting
CLEANING THE LCS WATER FILTER | Please Note: The LCS Water Filter should be periodically checked and cleaned.
Return the filter to the
housing and close the cap
tightly.
Remove the filter. Rinse the filter and, if required, clean with a
soft brush or using compressed air. Check
and clean the O-ring.
Remove any debris inside the housing.
Unscrew the cap at
the front of the
LCS adaptor cylinder.
1 2 43
LCS system Messages What it Means What to Do – User
REPLACING THE BACTERIOLOGICAL AIR FILTER
Replace the bacteriological filter
every 500 cycles or if dirty.
To change the filter:
1. Power the unit OFF,
disconnect tube A and remove
the filter from the bracket.
2. Disconnect tube B from the
back of the filter.
3. Before installing the replacement filter
(01-102119S) check that the arrow mark
on the filter matches the direction of
the arrow on the bracket. (Arrow should
be pointing to the right).
4. Insert filter into bracket and connect
tube B. Re-connect tube A.
5. After replacing the filter, go to the “Reset
LCS Counter” screen in the Setup menu
and reset the counter to 0.
Replace Water Filter
Replace Air Filter
CF 12 Water
Filter Clogged
CF 18 Water Filter
Pressure Sensor Stuck
CF 26 Air Valve error
CF 30 LCS adaptor
disconnected
Water filter in the LCS adaptor
is dirty or clogged.
The bacteriological air filter in the
kickplate is dirty.
Problem with Lumens Cleaning System.
Problem with Lumens Cleaning System.
Problem with air suppy pressure in air line.
Incoming air pressure must be between 1 bar
(15 psi) - 10 bar (145 psi).
Problem with Lumens Cleaning System.
• Follow instructions above:
“Cleaning the water filter.”
• Follow instructions above:
“Replacing the air filter.”
• Check LCS adaptor is correctly positioned.
• You selected LCS at program start but
LCS is not installed.
• Replace Water Filter, see
“Cleaning the water filter.”
• If cleaning doesn’t resolve the problem,
replace the water filter (01-113742S)
• If message persists, call for service.
• Call for service.
• Call for service.
• Check that LCS adaptor is correctly positioned.
Tube A Tube B (behind)

FRANÇAISE
Adaptateur de pièce à main HYDRIM®M2/M2 G4 Manuel de l’opérateur
Couplement
MULTIflex - p/n 01-113930
ou
INTRAmatic - p/n 01-113931
avec l’adaptateur
(non comprise)
Raccords Luer-Lock
Pièce à main
(non comprise)
Connexion des pièces à main à l’adaptateur LCS
Visser le LCS allant
avec le couplage
de pièce à main
approprié.
Introduire le couplage de pièce à main
dans celle-ci.
Pousser la pièce à main
dans l’une des fixations.
1 2 3
Multiflex
Intramatic
Appuyer
sur l’attache
afin de
libérer.
Paramétrage du programme
Pour optimiser les résultats des performances de lavage des
pièces à main et des autres instruments, le technicien doit aug-
menter le délai de lavage à la valeur de paramétrage maximum
de 15 minutes pour le cycle.
Remarque : le délai du cycle général augmentera en
conséquence d’environ 6 minutes.
2
1
Remarque : Conformément aux recommandations des principaux fabricants/fournisseurs de pièces à main, les parties creuses des instruments doivent être séchées immédiatement après le
nettoyage et la désinfection en utilisant de l’air comprimé médical avant d’effectuer des procédés supplémentaires ou une stérilisation. Cela doit être suivi par les procédures d’entretien régulier
recommandées et par l’application d’agents/huiles d’entretien. Veuillez vous conformer à tous les règlements en matière de santé et sécurité en vigueur dans votre région.
Utilisation du menu d’écran tactile
L’HYDRIM va détecter
automatiquement la
présence de l’adaptateur
LCS. Sélectionner le
programme désiré et
suivre les instructions.
Une fois terminé, ouvrir la
porte et retirer les pièces
à main de l’adaptateur en
poussant sur l’attache sur
le côté de la fixation.
Exemple : cycle régulier
1 2 3
Remarque : Conformément aux
recommandations des principaux
fabricants/fournisseurs de pièces
à main, les parties creuses des
instruments doivent être séchées
immédiatement après le nettoyage
et la désinfection en utilisant de l’air
comprimé médical avant d’effectuer
des procédés supplémentaires ou
une stérilisation. Cela doit être suivi
par les procédures d’entretien régulier
recommandées et par l’application
d’agents/huiles d’entretien. Veuillez
vous conformer à tous les règlements
en matière de santé et sécurité en
vigueur dans votre région.
Other manuals for HYDRIM M2
3
This manual suits for next models
1
Table of contents
Languages:
Popular Cleaning Equipment manuals by other brands

Suevia
Suevia 130.5011 EASYCLEANER Mounting instructions

i-MO
i-MO Öko 2000 user guide

unGer
unGer Hydro Power Ultra UNP01 operating instructions

Black & Decker
Black & Decker BHPC130 Original instructions

Uni-ram
Uni-ram UG5000E operating manual

Axi
Axi MTC HC-300 Installation, operating and maintenance manual