7-FUNCTION MULTIMETER
MODEL NO: MM19.V3
Thank you for purchasing a Sealey product. Manufactured to a high standard, this product will, if used according to these instructions,
and properly maintained, give you years of trouble free performance.
IMPORTANT: PLEASE READ THESE INSTRUCTIONS CAREFULLY. NOTE THE SAFE OPERATIONAL REQUIREMENTS, WARNINGS & CAUTIONS. USE
THE PRODUCT CORRECTLY AND WITH CARE FOR THE PURPOSE FOR WHICH IT IS INTENDED. FAILURE TO DO SO MAY CAUSE DAMAGE AND/OR
PERSONAL INJURYAND WILL INVALIDATE THE WARRANTY. KEEP THESE INSTRUCTIONS SAFE FOR FUTURE USE.
1. SAFETY
When using this multimeter, please observe all normal safety rules concerning:
9Protection against the dangers of electrical current.
9Protection of the meter against misuse.
9Full compliance with safety standards can only be guaranteed if used with the test leads supplied. If necessary, they must be replaced
with genuine Sealey leads with the same electrical ratings. Failure to do so will invalidate the warranty.
8DO NOT use leads if damaged or if the wires are bared in any way.
1.1. GENERAL SAFETY INSTRUCTIONS
9Familiarise yourself with the application and limitations of the multimeter as well as the potential hazards.
IF IN ANY DOUBT CONSULT A QUALIFIED ELECTRICIAN. USE EXTREME CAUTION when working with high voltages.
9When the meter is connected to a circuit, do not touch unused meter terminals.
9When the magnitude of the value to be measured is unknown, set the range selector to the highest value available.
9Before commencing testing, follow instructions below and select the correct input sockets, function and range on the multimeter.
9Before rotating the rotary switch to change functions, disconnect the test leads from the circuit under test.
9Take care when working with voltages above 35V DC or 25V AC rms. These voltages are considered a shock hazard. Keep fingers
behind the probe barriers whilst measuring.
8DO NOT test voltages above 750V AC - 1000V DC - the circuitry of the multimeter may be destroyed.
WARNING! NEVER connect the multimeter to a voltage source / live circuit when the rotary switch is set to any other function apart from
Voltage testing.
WARNING! NEVER perform resistance, transistor, diode or continuity measurements on live circuits.
9ALWAYS discharge lter capacitors in power supplies and disconnect the power when making resistance or diode tests.
WARNING! Voltage checks on electrical outlets can be difcult and misleading because of the uncertainty of connection to the
recessed electrical contacts. Other means should be used to ensure that the terminals are not “live”.
8DO NOT use the multimeter in a potentially explosive atmosphere.
8NEVER operate the meter unless the back cover and the battery and fuse doors are in place and fastened securely.
9If any abnormal readings are observed, the multimeter must be checked out by an authorised technician.
9When not in use, store the multimeter carefully in a safe, dry, childproof location out of direct sunlight.
9Storage temperature range -20°C to 60°C.
9ALWAYS turn off the power and disconnect the test leads before opening the doors to replace the fuse or batteries.
9The user shall ensure that test probes are correctly selected in order to prevent danger. Probes shall be selected to ensure
that adequate barriers guard against inadvertent hand contact with live conductors under test and that probes have minimal
exposed probe tips. Where there is a risk of the probe tip short circuiting with other live conductors under test, it is recommended that
the exposed tip length shall not exceed 4mm.
The warnings, cautions and instructions referred to in this manual cannot cover all possible conditions and situations that
may occur. It must be understood that common sense and caution are factors which cannot be built into this product, but must
be applied by the operator.
1.2. SAFETY SYMBOLS
This symbol adjacent to another symbol, terminal or operating device indicates that the operator must refer
to an explanation in the Operating Instructions to avoid personal injury or damage to the meter.
This WARNING symbol indicates a potentially hazardous situation, which if not avoided, could result in death or serious
or serious injury.
This CAUTION symbol indicates a potentially hazardous situation, which if not avoided, may result damage to the product.
This symbol, adjacent to one or more terminals identies them as being associated with ranges that may, in normal use, be
subjected to particularly hazardous voltages. For maximum safety, the meter and its test leads should not be handled when
these terminals are energized.
This symbol indicates that a device is protected throughout by double insulation or reinforced insulation.
WARNING
CAUTION
Refer to
instruction
manual
MM19.V3 Issue:6 (H,F) 21/10/21
Original Language Version
© Jack Sealey Limited
WARNING! NEVER apply voltage or current to the meter that exceeds the specied
maximum stated.
WARNING! USE EXTREME CAUTION WHEN WORKING WITH HIGH VOLTAGES.
2. INTRODUCTION
High quality general-purpose multimeter with a large (46 x 25mm), clear and easy-to-read LCD display.
Supplied with test leads.
Measures:
- AC and DC Voltage
- DC Current
- Resistance
- Audible Continuity
- Diode/Transistor Verication Mode
3. SPECIFICATION
Model No:...............................................................MM19.V3
AC Voltage (Accuracy):............................. 200V, 750V (±2%)
DC Voltage (Accuracy):
............200mV, 2000mV, 20V, 200V (±1.5%), 1000V (±1.8%)
AC Current (Accuracy):..................................................... No
DC Current (Accuracy):
....200µA, 2000µA, 20mA (±2%), 200mA (±2.2%), 10A (±2%)
Resistance (Accuracy):
.............200Ω, 2000Ω, 20kΩ, 200kΩ (±1.8%), 2000kΩ (±2%)
Capacitance (Accuracy):................................................... No
Temperature (Accuracy):................................................... No
Frequency (Accuracy):...................................................... No
Duty Cycle:........................................................................ No
Continuity Audible: ...........................................................Yes
4. BATTERY INSTALLATION
WARNING! To avoid electric shock, disconnect the test leads from any source of voltage
before removing the battery cover.
4.2.1. Disconnect the test leads from the meter.
4.2.2. Open the battery cover by loosening the two cover screws using a cross head screwdriver.
4.2.3. Insert the battery into battery holder, observing the correct polarity.
4.2.4. Replace the battery cover. Secure with the screws.
WARNING! To avoid electric shock, do not operate the meter until the battery cover
is in place and fastened securely.
NOTE! If meter does not work properly, check the fuses and battery to make sure that
they are still good and that they are properly inserted.
5. FEATURES
Refer to diagram on right hand side
1. Function Switch
2. LCD Display
3. Common Jack
4. V.Ω.mA Jack
5. 10A Jack
NOTE: Fuse and battery compartment are at the rear of the unit.
Original Language Version
© Jack Sealey Limited
REMOVE SCREWS
TO ACCESS
BATTERYAND
FUSES
1
2
3
4
5
SYMBOLS AND ANNUNCIATORS
•))) Continuity
Diode Test
ΜMicro (amps)
MMilli (volts, amps)
KKilo (ohms)
ΩOhms
DCV Volts Direct Current
ACV Volts Alternating Current
DCA Amps Direct Current
HFE Transistor Measurement
Function Range Resolution Accuracy
DC Voltage
(V DC)
200mV 100µV
±(1.5% reading + 2 digits)
2000mV 1mV
20V 10mV
200V 100mV
1000V 1V ±(1.8% reading + 2 digits)
OVERLOAD PROTECTION: 220Vrms AC for 200mV range and 1000VDC or 750Vrms AC for other
ranges.
AC Voltage
(V AC)
200V 100mV ±(2% reading + 2 digits
(50/60Hz)
750V 1V
OVERLOAD PROTECTION: 1000 V DC or 750Vrms AC for all ranges.
RESPONSE: Average responding, calibrated in rms of a sine wave.
FREQUENCY RANGE: 45Hz - 450Hz
MM19.V3 Issue:6 (H,F) 21/10/21
6. a OPERATION
WARNING! Risk of electrocution. High-voltage circuits, both AC and DC, are very dangerous and should be measured with great care.
DO NOT measure voltages that might exceed 500 V above earth ground.
ALWAYS turn the function switch to the “OFF” position when the meter is not in use.
NOTE! On some low AC and DC voltage ranges, with the test leads not connected to a device, the display may show a random
reading. This is normal and is caused by the high-input sensitivity. The reading will stabilise and give a proper measurement when
connected to a circuit.
6.1. DC VOLTAGE MEASUREMENTS
CAUTION: DO NOT measure DC voltages if a motor on the circuit is being switched ON or OFF. Large voltage surges may occur that
can damage the meter.
6.1.1. Set the function switch to the highest VDC position.
6.1.2. Insert the black test lead banana plug into the negative (COM) jack. Insert the red test lead banana plug into the positive (V) jack.
6.1.3. Touch the black test probe tip to the negative side of the circuit. Touch the red test probe tip to the positive side of the circuit.
6.1.4. Read the voltage in the display. Reset the function switch to successively lower VDC positions to obtain a higher resolution reading.
The display will indicate the proper decimal point and value.
6.2. AC VOLTAGE MEASUREMENTS
WARNING! Risk of Electrocution. The probe tips may not be long enough to contact the live parts inside some 240V outlets for
appliances because the contacts are recessed deep in the outlets.
As a result, the reading may show 0 volts when the outlet actually has voltage on it. Make sure the probe tips are touching the metal
contacts inside the outlet before assuming that no voltage is present.
CC CCAUTION: DO NOT measure AC voltages if a motor on the circuit is being switched ON or OFF. Large voltage surges may occur that
can damage the meter.
6.2.1. Set the function switch to the highest VAC position.
6.2.2. Insert the black test lead banana plug into the negative (COM) jack. Insert red test lead banana plug into the positive (V) jack.
6.2.3. Touch the black test probe tip to the negative side of the circuit. Touch the red test probe tip to the positive side of the circuit.
6.2.4. Read the voltage in the display. Reset the function switch to successively lower VAC positions to obtain a higher resolution reading.
6.3. DC CURRENT MEASUREMENTS
CAUTION: DO NOT make current measurements on the 10A scale for longer than 30 seconds. Exceeding 30 seconds may cause
damage to the meter and/or the test leads.
6.3.1. Insert the black test lead banana plug into the negative (COM) jack.
6.3.2. For current measurements up to 200mA DC, set the function switch to the highest DC mA position and insert the red test lead banana
plug into the (mA) jack.
6.3.3. For current measurements up to 10A DC, set the function switch to the 10A range and insert the red test lead banana plug into the
(10A) jack.
6.3.4. Remove power from the circuit under test, then open up the circuit at the point where you wish to measure current.
6.3.5. Connect test leads in series with the circuit under measurement.
6.3.6. Apply power to the circuit.
6.3.7. Read the current in the display. For mA DC measurements, reset the function switch to successively lower mA DC positions to obtain a
higher resolution reading. The display will indicate the proper decimal point and value.
6.4. RESISTANCE MEASUREMENTS
WARNING! To avoid electric shock, disconnect power to the unit under test and discharge all capacitors before taking any resistance
measurements. Remove the batteries and unplug the line cords.
6.4.1. Set the function switch to the highest Ω position.
4.4.2. Insert the black test lead banana plug into the negative (COM) jack. Insert the red test lead banana plug into the positive Ω jack.
Original Language Version
© Jack Sealey Limited
DC Current
(A DC)
200µA 100nA
±(2.0% reading + 2 digits)2000 µA 1 µA
20mA 10µA
200mA 100µA ±(2.2% reading + 2 digits)
10A 10mA ±(2.0% reading + 2 digits)
OVERLOAD PROTECTION: 200mA /250V fuse (10A range unfused)
MEASURING VOLTAGE DROP: 200mV.
Resistance
200Ω 0.1Ω ±(1.8% reading + 5 digits)
2000Ω 1Ω
±(1.8% reading + 2 digits)20kΩ 10kΩ
200kΩ 100Ω
2000kΩ 1kΩ ±(2% reading + 2 digits)
MAXIMUM OPEN CIRCUIT VOLTAGE: 2.8V
OVERLOAD PROTECTION: 15 seconds maximum 220Vrms on all ranges.
Diode Test Display Voltage Drop Approximation
OVERLOAD PROTECTION: 15 seconds maximum 220Vrms Sounds alarm.
•)))
Audible Continuity Built-in buzzer sounds if resistance is less than 100Ω
OVERLOAD PROTECTION: 15 seconds maximum 220Vrms Sounds alarm.
hFE
Transistor hFE
Data Test
0 ~ 1000 Basic Current is Approx. 10µA., Vce is
Approx. 3V
Overload Protection: 10 seconds max. 220Vrms.
MM19.V3 Issue:6 (H,F) 21/10/21
Sealey Group, Kempson Way, Suffolk Business Park, Bury St Edmunds, Suffolk. IP32 7AR
01284 757500 01284 703534 sales@sealey.co.uk www.sealey.co.uk
ENVIRONMENT PROTECTION
Recycle unwanted materials instead of disposing of them as waste. All tools, accessories and packaging should be sorted, taken to
a recycling centre and disposed of in a manner which is compatible with the environment. When the product becomes completely
unserviceable and requires disposal, drain any fluids (if applicable) into approved containers and dispose of the product and fluids
according to local regulations.
Note: It is our policy to continually improve products and as such we reserve the right to alter data, specifications and component parts without prior
notice. Please note that other versions of this product are available. If you require documentation for alternative versions, please email or call
our technical team on technical@sealey.co.uk or 01284 757505.
Important: No Liability is accepted for incorrect use of this product.
Warranty: Guarantee is 12 months from purchase date, proof of which is required for any claim.
WEEE REGULATIONS
Dispose of this product at the end of its working life in compliance with the EU Directive on Waste Electrical and Electronic Equipment
(WEEE). When the product is no longer required, it must be disposed of in an environmentally protective way. Contact your local solid
waste authority for recycling information.
6.4.2. Touch the test probe tips across the circuit or component under test. It is best to disconnect one side of the component under test so
the rest of the circuit will not interfere with the resistance reading.
6.4.3. Read the resistance in the display and then set the function switch to the lowest Ω position that is greater than the actual or any
anticipated resistance. The display will indicate the proper decimal point and value.
.
Original Language Version
© Jack Sealey Limited MM19.V3 Issue:6 (H,F) 21/10/21

Other Sealey Measuring Instrument manuals

Sealey GA70CR User manual

Sealey

Sealey GA70CR User manual

Sealey TM105 User manual

Sealey

Sealey TM105 User manual

Sealey AK9636D.V2 User manual

Sealey

Sealey AK9636D.V2 User manual

Sealey VS602 User manual

Sealey

Sealey VS602 User manual

Sealey GA450 User manual

Sealey

Sealey GA450 User manual

Sealey AK9623EV User manual

Sealey

Sealey AK9623EV User manual

Sealey TA090 User manual

Sealey

Sealey TA090 User manual

Sealey TA101 User manual

Sealey

Sealey TA101 User manual

Sealey AK9621EV User manual

Sealey

Sealey AK9621EV User manual

Sealey DBG5011 User manual

Sealey

Sealey DBG5011 User manual

Sealey VAG VS862 User manual

Sealey

Sealey VAG VS862 User manual

Sealey VS0563 User manual

Sealey

Sealey VS0563 User manual

Sealey VSE5863 User manual

Sealey

Sealey VSE5863 User manual

Sealey AK4013 User manual

Sealey

Sealey AK4013 User manual

Sealey DBG5010 User manual

Sealey

Sealey DBG5010 User manual

Sealey Auto Service Series User manual

Sealey

Sealey Auto Service Series User manual

Sealey VS8196 User manual

Sealey

Sealey VS8196 User manual

Sealey SA9302 User manual

Sealey

Sealey SA9302 User manual

Sealey AK9637D.V2 User manual

Sealey

Sealey AK9637D.V2 User manual

Sealey AK9636D.V2 User manual

Sealey

Sealey AK9636D.V2 User manual

Sealey PP5 User manual

Sealey

Sealey PP5 User manual

Sealey MM405 User manual

Sealey

Sealey MM405 User manual

Sealey PP7 User manual

Sealey

Sealey PP7 User manual

Sealey TP67DT.V2 User manual

Sealey

Sealey TP67DT.V2 User manual

Popular Measuring Instrument manuals by other brands

SATO KEIRYOKI PC-7700II instruction manual

SATO KEIRYOKI

SATO KEIRYOKI PC-7700II instruction manual

Digitax F1 plus user manual

Digitax

Digitax F1 plus user manual

Black Box FOVFL quick start guide

Black Box

Black Box FOVFL quick start guide

ClimeMET CM2030 user guide

ClimeMET

ClimeMET CM2030 user guide

VMI VIBER X2 Pro manual

VMI

VMI VIBER X2 Pro manual

Thermo Orion AQUAfast II Operation

Thermo

Thermo Orion AQUAfast II Operation

Promax PROPOWER - 5 user manual

Promax

Promax PROPOWER - 5 user manual

Emtec BioProTT operating instructions

Emtec

Emtec BioProTT operating instructions

Accudata GVM-3 manual

Accudata

Accudata GVM-3 manual

AEMC CA811 user manual

AEMC

AEMC CA811 user manual

Metrol TFM1186 user manual

Metrol

Metrol TFM1186 user manual

Protimeter MMS3 instruction manual

Protimeter

Protimeter MMS3 instruction manual

ATAGO PAL-03CS instruction manual

ATAGO

ATAGO PAL-03CS instruction manual

Quark-Elec IS20 Setup guide

Quark-Elec

Quark-Elec IS20 Setup guide

Siko SO100 User information

Siko

Siko SO100 User information

Sontex Supercal 5 installation guide

Sontex

Sontex Supercal 5 installation guide

SENSIT Technologies ULTRA-TRAC APL Quick start instructions

SENSIT Technologies

SENSIT Technologies ULTRA-TRAC APL Quick start instructions

sauter TN-US instruction manual

sauter

sauter TN-US instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.