SellEton Scales SL-904 User manual

www.selletonscales.com
SL-904 PANEL MOUNT
INDICATOR USER’S MANUAL

TABLE OF CONTENTS
Safety Precautions 1
Features 2
Specications 3
Display and Key Descriptions 4
Operating Instructions 5
Rear Panel 6
General Functions 7
Function Settings 8
Parameter Settings 9-10
Calibration Settings 11-14
Check Weighing Conguration 15
Cable Connection Manual 16-18
Troubleshooting 19
Contact Us 19

1
SAFETY PRECAUTIONS
For safe operation of the weighing indicator, please follow these
instructions:
●Calibration inspection and maintenance of the indicator are prohibited by non-pro-
fessional staff
●The indicator is a piece of static sensitive equipment; Please cut off power during
electrical connections
●Touching the internal components by hand is prohibited
●DO NOT exceed the rated load limit of the unit
●DO NOT step on the unit
●DO NOT jump on the scale
●DO NOT use this product if any of the components are cracked
●DO NOT use for purposes other then weight taking
●To avoid damaging the battery do not keep charger plugged in once battery is fully
charged
●Make sure the weight is not over the Max capacity as it could damage the load cell
inside
● Material that has a static electric charge could inuence the weighing. Discharge
the static electricity of the samples, if possible. Another
solution to the problem is to wipe both sides of the pan and the top of the case with
an anti-static agent
●Plug into a wall outlet to avoid interference with other wirings
●Calibration may be required before weighing when the scale is initially installed or
moved from a location

2
FEATURES
●LED 6 digit display
●Multiple weighing units: kg/lb/t
●Gross/Tare/Hold/Zero
●Check weighing feature
Technical Parameters
Hardware construction features
●Power supply: 24vDC
●Load cell excitation voltage: 5vDC±5%
● Load cell number: up to eight 350Ω
●Load cell sensitivity: 0~3.0mv/v
●Load cell connections: six wire
●Keyboard: six key
●Display: six digits red 7 segment LED display
●Relay output: 4 output, AC250v 5A
●Input: isolation voltage 2500V
●Analog output: 4~20mA/0~5v
●Serial port: RS232/RS485, baud rate 600~19200bit/s
●Operation temperature: -10 °C ~ +40 °C
● Operation humidity: ≤90%RH
●Storage temperature: -40 °C ~ +70 °C (32-104°F)
●Housing dimension: 92 x 45mm
●Front panel dimension: 102 x 55mm
●Trepanning dimension: 93 x 46mm
Software features
●Max sampling speed: 120SPS
● AD digital lter
●Digital calibration
●Batching or dosing functions

3
SPECIFICATIONS
FIGURE 1: INDICATOR MEASUREMENTS
2.16”
4”
5.46”
1.78”
5.6”
6.1”

4
DISPLAY AND KEY DESCRIPTION
|/ Powers the Indicator On or Off if held for 2 seconds
ZERO Zero’s the scale
TARE 1. Resets the scale to zero when there is something on the scale
(ex. Tare out the weight of a pallet to weigh only the product on it)
2. Clears the tare to see the gross weight (pallet + product)
M1 Displays Gross/Net weight
MODE Setpoint parameter set
ENTER Enter key
Zero Indicates that you have zero’d the scale
Motion The weight on the scale is unstable
Gross Shows you are in Gross weight mode (includes tare); default mode
Net Shows you are in Net weight mode (weight without tared weight)
kg The weight is shown in kilograms
t The weight is shown in tons
lb The weight is shown in pounds
Power Flashes red = low battery, Solid red = charging, Green = fully charged
Over Flashes when weight is higher than set alarm parameter
Accept Flashes when weight is within the set alarm parameters
Under Flashes when weight is lower than set alarm parameter
Save and Exit
Arrow keys
Return/Enter

5
OPERATING INSTRUCTIONS
Power On
●Turn on the power by pressing the power button for 2 seconds. Once on, the scale
will ash the voltage and then begin to auto-check and count down from 0-9
sequentially before entering the weighing mode
Note: Anything on the scale before powering on will automatically be tared out.
Zeroing
●The zero function is used only when the scale is empty and is not at gross zero due
to material build up
●Pressing the ZERO key will reset your scale to 0
●Depending on what your manual zero range parameter is set to, you can zero out
any number within your set selection, after that you will receive an error and will
need to tare out the weight
Tare Function
●The Tare function is used when you only wish to see the current change in weight,
not the entire amount of weight that is on the scale
●When the indicator is in gross mode (gross light is shown) pressing the TARE key
will Tare the current weight on the scale and enter the net mode (net light shown)
●For example if you are using a container add the container to the scale, press tare
and the display will show the gross light and reset back to 0
●Add your product to the scale to weigh without the weight of the container
●To exit Tare mode press the TARE key again to enter gross mode and you will see
the total weight of the container and the product
Note: If you remove the container the scale will show the minus weight of the container

6
REAR PANEL
Connection denition:
Power: POW+ POW- for 24VDC, AC1 AC2 for 220VAC
Load cell: EXC+ SEN+ EXC- SEN- SIG+ SIG
Serial communications: TXD RXD GND for RS232, A B for RS485
Analog output: Vout Iout GND
Relay output: J1 J1 J2 J2 J3 J3 J4 J4
Input: YX1 YX2 YX3 YX4 YXCOM 0vTo exit Tare mode press the TARE key again to enter
gross mode and you will see the total weight of the container and the product
Note: If you remove the container the scale will show the minus weight of the container

7
GENERAL FUNCTIONS
Function setup and operation procedure:
Function Operation Display Remark
Enter calibration mode Turn the calibration switch to “ON” 01 CSP 4
Enter function setting Weight mode, press and 01 FnC 3.3
Check weighing setpoint
parameter setting
Weight mode, press 1.FinAL 5
Enter to test mode Turn power on, press and 1. dsp 7.1
Reset all parameters back
to default
Turn power on, turn the calibration
switch to “ON”, press and
i.ALL 7.2
Reset general function
parameters back to default
Turn power on, press and 1 FnC 7.2
Parameter Settings, Key meaning
Key Meaning
Save and Exit
Move the ashing digit to the left
Move the ashing digit to the right
Increase the ashing digit
Decrease the ashing digit
Conrm

8
FUNCTION SETTINGS
Press and at the same time to enter the function settings
The screen will display “01 FnC” for function setting
Press to display “02 232” for serial port interface
Press to display “03 Sq ” for weight comparison procedures
Press to display “04 AnL” for analog current output
Press to display “05 In ” for external input interface
Press to enter the next menu level
Press to save and exit out of this menu and restart
01 FnC
01 FnC
01 FnC
01 FnC
01 FnC
FnC 01
In 01
AnL 01
Sq 01
232 01
Weighing Mode
and

9
INDICATOR PARAMETER SETTINGS
To enter parameter settings, follow the procedure below:
1. Press ENTER and MODE at the same time for 2 seconds to enter the function settings
2. Navigate through the settings as shown in the table below by using the arrow keys and return
keys as labeled under each indicator button
3. Press the ENTER key to enter/edit the parameter setting
Press the key to save and exit settings at any time
01 FnC Function Parameter Code
Function Item Parameter Description Default
Digital Filter 1 FnC 01 00
01
...
10
Greater
Less
03
Digital Filter 2 FnC 02 00
01
02
03
Greater
Less
02
Rate for display
rewrite FnC 03 01
05
10
20
120
1 times/sec
5 times/sec
10 times/sec
20 times/sec
120 times/sec
10
02 232 Serial Port Interface
Function Item Parameter Description Default
Baud Rate 232 01 00
01
02
03
04
05
600bit/s
1200bit/s
2400bit/s
4800bit/s
9600bit/s
19200bit/s
04
Transmit Mode 232 02 00
01
02
03
04
close serial port
continuous
print mode
stable send
command mode
01
Data Format 232 03 00
01
02
format 1
format 2
print format
00
Transmit Time 232 04 00
01
02
03
open
1 sec
2 sec
5 sec
00
Address for Multi-
computer 232 05 00
01~99
Only one
Address (bcd)
00

10
03 Sq Weight Comparison Procedures
Function Item Parameter Description Default
Batching Mode sq 01 00
01
02
03
04
normal batch
less-in weight
comparison
01
Control Mode sq 02 01
02
03
manual start
auto
for custom
01
Comparison Format sq 03 01
02
net
gross
01
Start Delay sq 04 00
01
...
10
no delay
1 sec delay
…
10 sec delay
01
Stable Time sq 05 00
01
...
10
no wait
1 sec wait
…
10 sec wait
03
04 Analog Current Ouput
Function Item Parameter Description Default
Signal output AnL 01 00
01
02
4~20mA current
0~20mA current
0~5v voltage
00
Data format AnL 02 00
01
display weight
gross weight
00
Maximum weight AnL 03 010000 when max analog output weight 010000
Zero adjust AnL 04 4mA/0mA/0v Press or to adjust output 4
Linearity adjust AnL 05 20mA/5v Press or to adjust output 20
Output range limit AnL 06 00
01
no limit
limit
00
05 External Input Interface
Function Item Parameter Description Default
INPUT1 IN 01 00
01
02
03
04
05
06
no function
zero
tare
gross
hold
total
print
01
INPUT2 IN 02 02
INPUT3 IN 03 05
INPUT4 IN 04 06

11
CALIBRATION SETTINGS
To enter calibration settings, follow the procedure below:
1. In weighing mode, make sure calibration switch is set to “ON”
2. The display will show “01 CSP” meaning you entered the calibration parameter code
●Press to enter the next step level
●Follow the steps in the parameters to set up your calibration
3. Press to display “02 CAL” General calibration
●Follow the steps to calibrate your scale
4. Press to display “03 CUo” Sensitivity calibration
● Follow the steps to set lters and manage the sensitivity of your scale
5. When you are done with calibrating, make sure to turn the switch to “OFF”
6. The screen will display “End” , save the data and restart
02 CAL
Calibration
01 CSP
Calibration
Set-up
03 CUo
Sensitivity
Calibration
Weighing Mode
Save Data & Restart
Calibration switch set to “ON”
Calibration switch set to “OFF”

12
“01 CSP” Calibration Parameter Code
01 CSP Calibration Parameter Code
Function Item Parameter Description Default
Unit CSP 01 01
02
03
kg
lb
t
01
Decimal point CSP 02 00
01
02
03
04
none
1 decimal point
2 decimal point
3 decimal point
4 decimal point
00
Division CSP 03 01
02
05
division size 02
Max Capacity CSP 04 010000 Max capacity 1000
Zero-Setting range CSP 05 00
01
02
0
±1%
±2%
02
Initial zero-setting
range CSP 06 00
01
02
05
10
0
±1%
±2%
±5%
±10%
10
Automatic
zero-setting range CSP 07 00
05
10
20
0
0.5d
1d
2d
05
Automatic
zero-setting time CSP 08 00
01
02
03
0
1 sec
2 sec
3 sec
01
Stable time CSP 09 00
01
02
fast
medium
slow
01
Stable range CSP 10 01
02
05
10
1d
2d
5d
10d
02
Automatic zero CSP 11 00
02
05
10
20
no
-2d
-5d
-10d
-20d
02
Preserved Menu CSP 12

13
“02 CAL” Calibration
In weighing mode set calibration switch to “ON”
Calibration press
To enter calibration press
Zero calibration press
Skip zero calibration press
Display zero code
3 seconds
Weight calibration press
Skip weight calibration press
Use to set the weight
Once the weight is stable press
Sensitivity
3 seconds, calibration done
Set calibration switch to “OFF”
You are nished
Note: Zero calibration, can be no more than 5mv.
01 CSP
02 CAL
ZERO
288888
SPAN
01000kg
120000
02 CAL
END

14
“03 CUo” Sensitivity Calibration
In weighing mode set calibration switch to “ON”
Sensitivity Calibration press
To enter calibration press
Zero calibration press
Skip zero calibration press
Auto gather zero code
Or press to set zero code
to enter zero calibration
Sensitivity calibration press
Skip sensitivity calibration press
Use to set the sensitivity
enter press
Sensitivity calibration done
Set calibration switch to “OFF”
You are nished
Sensitivity = 100000 x load cell full scale output x weigh-
ing meter capacity /(N x Load cell capacity)
01 CSP
03 CUo
ZERO
288888
SPAN
120000
02 CUo
END

15
CHECK WEIGHING CONFIGURATION
Weighing mode press to go enter parameters
Display Function
1. FinAL Final value
2. SP1 SP1 value
3. SP2 SP2 value
4. SP3 SP3 value
F. FALL Free fall value
6. oVer Over value
7. UndEr Under value
8. Z.bAnd Zero band
Warning: set FinAL > SP1 > SP2 > SP3 > F. Fall
set Sq01=1: net weigh
Signal Output Condition Relay Output
SP1 Net ≥ Final-SP1 J1
SP2 Net ≥ Final-SP2 J2
SP3 Net ≥ Final-SP3 J3
Free Fall Net ≥ Final-F.FALL J4
Set Sq01=3:
Signal Output Condition Relay Output
HI Net ≥ SP1 J1
OK SP1 ≥Net ≥ SP2 J2
LO Net < SP2 J3

16
CABLE CONNECTION MANUAL
Power
POW+ 24VDC+
POW- 24VDC-
AC1 220VAC
AC2 220VAC
Load cell
EXC+
SEN+
EXC-
SEN-
SIG+
SIG-
NOTE: use 4-wire load cell need EXC+ and SEN+ short connect, EXC- and SEN- short
connect.
Serial Port
TXD RS232 transmission RXD RS232 receive GND RS232 GND A RS485 A B RS485 B
Analog Output
Vout 0~5v voltage, load more than 1kΩ
Iout 4~20mA current, load span 100~500Ω
GND GND

17
Relay Output
J1 J1 rst group relay output
J2 J2 second group relay output
J3 J3 third group relay output
J4 J4 fourth group relay output
Relay Output Connecction diagram
Common connection diagram
Note: Use common connection, put one of J1, J2, J3, J4 short connect.

18
Input
YX1 rst group input
YX2 second group input
YX3 third group input
YX4 fourth group input
YXCOM input common
0v out connect power GND
Input Diagram
Out connect power input diagram
Note: Using out elctric power needs more than 3V battery charge, Out electric power
less than 24V.
Table of contents
Other SellEton Scales Measuring Instrument manuals

SellEton Scales
SellEton Scales OP-907 Series User manual

SellEton Scales
SellEton Scales SL-7561 User manual

SellEton Scales
SellEton Scales OP-901 Series User manual

SellEton Scales
SellEton Scales OP-903 User manual

SellEton Scales
SellEton Scales SL-7515 User manual

SellEton Scales
SellEton Scales SL-918 User manual

SellEton Scales
SellEton Scales OP-902 User manual

SellEton Scales
SellEton Scales SL-7510 Series User manual

SellEton Scales
SellEton Scales OP-900 Series User manual

SellEton Scales
SellEton Scales SL-905 User manual