Sentry Hybrid ST-643 User manual

INFRARED THERMOMETERINFRARED THERMOMETER
Intelligent Infrared Thermometer with Emissivity
Smart and Color Identification Signal
INFRARED THERMOMETERINFRARED THERMOMETER
RoHS
COMPLIANT
Hbrid
y
Hbrid
y
3
ST-643


Instruction Manual
Version-01 07/NOV.
HybridHybrid
Infrared Thermometer
RoHS
COMP LIANT


1
Electronic trigger lock function.

2

EN61326: Electrical equipment for measurement,
control and laboratory use.
IEC61000-4-2: Electrostatic discharge immunity
test.
IEC61000-4-3: Radiated, radio-frequency,
Electromagnetic field immunity test.
IEC61000-4-8: Power frequency magnetic field
immunity test.

Restrict to use of six substances within
electrical and electronic equipment (EEE)
,
thereby contributing to the protection of
human health and the environment.
RoHS
The device may not be disposed of with
the trash. It promotes the re-use recycling
and other forms of recovery of used materials
and components, and to improve the
environmental performance of all operators
(manufacturers ,traders ,treatment facilities)
involved in the life cycle of products.
Dispose of the product appropriately in
accordance with the regulations in force in
your country.
4

3. Specification
5
Audio Alarm
Dual Display
Lock Function
Max/Min/Avg
Distance/Spot Ratio
Temperature Range
Accuracy
Thermopile
12:1
-32~760 °C(-25~1400°F)
±3 C(±5 F)
within -32~-20 C(-25~-4 F)
° °
° °
9V(006P, IEC6F22, NEDA1604)
195g Approx.
(@ ambient temperature
of 25 C/77 F) ° °
±2°C(±3°F)
within-20~100°C(-4~212 F)°
±2%
within 100~760 C(212~1400 F)° °
9V Battery, Instruction manual,
Carrying case.
Repeatability
Resolution
Response Time
Emissivity
℃/℉ Switchable
LCD Backlight
Auto Power Off
Laser Sight Switchable
Dimensions
Weight
E-Smart
Operation Temp.
Thermocouple K Type
CIS
±1 °C (±2 °F)
0.1°C (0.1 °F)
500 ms.
Automatically after approx. 6 sec.
180x130x40mm (7.09′′x5.12′′x1.57′′)
0~50°C(32~122°F),10~95%RH
YES
YES
YES
YES
Adjustable 0.1~1.0
YES
Thermocouple Range
-200 ~1380 C°
5~14μm
Thermocouple Accuracy
±1.5%+1degree
Battery Type
Accessory
YES
YES
YES
YES
YES
10 point Memory
YES

4. Operations of Instrument
4-1 Quick Start
4-2 Unit Diagram
6

LCD & Control panel
7
SCAN / HOLD / AUTO
°C / °F Indication
Primary Display
Max / Min / Avg
Secondary Display
Lock Key
Up Button
Function Key
Low Battery
Laser / Buzzer On / Off
Data Log
Thermocouple / Emissivity
High / Low Alarm
Memory Key
Down Button
1
2
3
4
5
6
8
9
10
11
12
7
13
14
15
8
9
10
11
12
13
14
15
1
2
3
4
5
6
7

EMS
8

Operation Remarks
Thermocouple: Activate the functions just by
connecting to the connector
and switch to TC function.
E-smart: 1. In Emissivity mode, push the button
to enter the E-smart function.
2. Contact the target surface with
thermocouple.
3. Point the target with infrared
thermometer and push the measure
trigger till the buzzer beep.
The cross-reference Emissivity
value will be memorized in the unit.
4. Press the “F” button to leave
the E-smart function.
CIS: The color of backlight will change when the
target temperature exceeds the setting point
of alarm temperature.
Memory: Activate the record function by push
the “M” button. To delete all the
records, press up or down button to
DATA 0 and press “M” button.
Lock: Push the button to continuously measure
and display the temperature without pull the
measuring trigger.
Function: The above functions can be activated
always in any step of operations
mode in flow-chart.
Scan/Hold: In SCAN mode, the LCD displays both
the current temperature in Celsius
or Fahrenheit. The unit will HOLD the
last reading for 6 seconds after the
trigger is released.
When the battery is low, the battery
icon shows and the unit will
continue to function.
Data: While DATA# flashes on the left button ,
Dis
the value on the Main Temperature play
can be recorded in “#” log. Simply press
“M” button. 9

10

Field of View (FOV) ratio
=Distance to Spot Ratio
The field of view is the angle of vision at which
the instrument operates, and is determined by
the optics of the unit.The FOV is the ratio of the
distance from the target to the target diameter.
The smaller the target,the closer you should be
to it. When the target diameter is small, it is
important to bring the thermometer closer to the
target to insure that only the target is measured,
excluding the surroundings.
11
5.Techniques of Infrared Thermometer
5-1

12
Emissivity
Emissivity is the ability of an object to emit or
absorb energy. Perfect emitters have an emissivity
of 1, emitting 100% of incident energy. An object
with an emissivity of 0.8 will absorb 80% and
reflect 20% of the incident energy. Emissivity is
defined as the ratio of the energy radiated by an
object at a given temperature to the energy emitted
by a perfect radiator at the same temperature. All
values of emissivity fall between 0.0 and 1.0.
6. Maintenance
Cleaning the lens: Blow off loose particles using
clean compressed air. Gently brush remaining
debris way with a camel's hair brush. Carefully
wipe the surface with a moist cotton swab. The
wab may be moistened with water.
NOTE:
DO NOT use solvents to clean the lens.
Cleaning the housing:
Use soap and water on a damp sponge or soft
cloth.
a
s
5-2

Emissivity Table (for reference)
Material Temp C/ F ° ° Emissivity
Gold(pure highly polished) 227/440 0.02
Aluminum foil 27/81 0.04
Aluminum disc 27/81 0.18
Aluminum household(flat) 23/73 0.01
577/1070 0.06
Aluminum(rough plate) 26/78 0.06
Aluminum(oxidized @599 C)°199/390 0.11
599/1110 0.19
Aluminum surfaced roofing 38/100 0.22
Tin(bright tinned iron sheet) 25/77 0.04
Nickel wire 187/368 0.1
Lead(pure 99.95-unoxidized) 127/260 0.06
Copper 199/390 0.18
599/1110 0.19
Steel 199/390 0.52
599/1110 0.57
Zinc galvanized sheet iron(bright)28/82 0.23
Brass(highly polished): 247/476 0.03
Brass(hard rolled-polished w/lines): 21/70 0.04
Iron galvanized(bright) - 0.13
Iron plate(completely) 20/68 0.69
Rolled sheet steel 21/71 0.66
Oxidized iron 100/212 0.74
Wrought iron 21/70 0.94
Molten iron 1299-1399/3270-2550 0.29
Copper(polished) 21-117/70-242 0.02
Copper(scraped shiny not mirrored) 22/72 0.07
Copper(Plate heavily oxidized) 25/77 0.78
Enamel(white fused on iron) 19/66 0.9
Formica 27/81 0.94
Frozen soil - 0.93
Brick(red-rough) 21/70 0.93
Brick(silica-unglazed rough) 1000/1832 0.8
Carbon(T-carbon 0.9% ash) 127/260 0.81
Concrete - 0.94
Glass(smooth) 22/72 0.94
Granite(polished) 21/70 0.85
Ice 0/32 0.97
Marble(light gray polished) 22/72 0.93
Asbestos board 23/74 0.96
Asbestos paper 38/100 0.93
371/700 0.95
Asphalt(paving) 4/39 0.97
Aluminum (polisned prate 98.3%) 227/400 0.04

PA05-000064X-01