SHEKEL MERAV-6000 User manual

HL318LC Weighing Indicator
User Manual
SH.01

1
Warning
Ask professional personnel to debug, detect and repair
controller.
Warning
Please keep good grounding of controller.
Warning
In electrical connection of controller, please cut off the power
supply in advance. Wait for 30 seconds between power-on of the
controller for 2 times.
Pay Attention to Static Electricity

2
Contents
!"
TECHNICAL INDICATORS ..................................................................................................... 3
#"
MAIN FUNCTION ............................................................................................................... 3
$"
BOUNDARY DIMENSION .................................................................................................... 4
%"
INTRODUCTION TO PANEL ................................................................................................. 5
&"
SETTING OF PARAMETERS .................................................................................................. 7
ENTRY SETTING .........................................................................................................
F1 SETTING OF SCALE ............................................................................................................................. 7
F2 SETTING OF APPLICATION FUNCTION ..................................................................................................... 9
F3 SETTING OF ENERGY SAVING PARAMETERS ........................................................................................... 10
F4 SETTING OF SERIAL PORT .................................................................................................................. 11
F5 MAINTENANCE AND SERVICE ............................................................................................................. 11
'"
FUNCTION DESCRIPTION .................................................................................................. 12
APPENDIX 1. CONTINUOUS OUTPUT FORMAT SPECIFICATION ............................................. 1616
APPENDIX 2. PRINTING FORMAT SPECIFICATION ..................................................................... 16

3
1
、
Technical Indicators
● 6-digit 1.6-inch LCD display. Long service life and shock resistance
● 7 function keys. Operation is simple and convenient
● Protection level: IP5X
● Excitation voltage: +5VDC
● Load capacity of sensor: at most 4 350Ω simulation sensors
● Input signal range of null point: 0-5mV
● Input signal range of full scale: 1-10 mV
● Inner resolution: 1 million
● Weight upgrading rate: 40 times per second
● Power supply mode
Battery: Lithium battery 7.4V4Ah
Adapter: voltage 100-240VAC Current 0.1A Frequency 50-60Hz.
● 2 RS232 ports
● Operating temperature: -10℃-40℃, relative humidity is below 85 %
● Storage temperature: -20℃-60℃, relative humidity is below 85 %
● Conforming to standard :GB/T 7724-1999
2
、
Main Function
● Basic weighing function: resetting, removing the peel and clearing the peel
● Weight detection function, counting function, animal scale function
● Weight keeping function, weight accumulation function, percentage
display
● Set redundant backup function of parameters
● Automatic screen protection and automatic shutdown energy-saving
function
● Rich printing formats and communication protocol

4
3
、
Boundary Dimension
Instrument size: detailed in the following figure (mm); instrument weight:
1.5kg
4
、
Connector
4.1 RS232-1 and RS232-2
2Pin: TXD 3Pin: RXD 5Pin: GND

5
4.2 Load cell Connector
1Pin: +E 2Pin: +S 3Pin: SHLD 4Pin: -S
5Pin: +E
5
、
Introduction to Panel
● Introduction to indicator lamps
Identification
Analysis
Sorting and weight check status
indicator
Extension indicator
Counting scale indicator
Animal scale indicator
Accumulating scale indicator
Accumulating scale adjustment
display indicator
Weight hold indicator
Net weight indicator
Scale in dynamic status
Scale at zero indicator
Key pressed indicator
Battery indicator
Weight unit

6
Counting scale unit
Weight % indicator
`
● Introduction to operation keys
Operation without special specification refers to short press on keys.
Key
symbol
Normal weighing state
Set stage
Weight maintenance key->『Hold』
Short press→ F2.1 = 1, keep/cancel.
F2.1 = 2, switch between percentage and
weight.
F2.1 = 5, switch between quantity and
weight.
Long key→ enter setting menu.
Return to the
last menu.
Accumulation key->『Total』
Short key → F2.1 = 4, include display weight
in accumulation value.。
F2.1 = 0 ,enter into/exit the
accumulation model.
Long key → F2.1 = 3, select scale to sample
target weight.
F2.1 = 4, accumulate weight of
scale display.
F2.1 = 5, count sampling of scale.
No definition.
Printing key->『Print』
Short key → start up of print.
Flicker bit is on the
left.

7
Function key->『Func』
Long key → Enter into setting menu.
Flicker bit is on the
right.
Tare key->『Tare』
Short key →gross weight turns to net weight.
Indication lamp of net weight “Net” is on.
Conduct skin removal operation for multiple
times.
Digit flicker
position reduces.
ZERO Key->『Zero』
Gross weight state resets weight. When the
scale is in net weight, dynamic state, saving
state and out of resetting range, zero clearing
operation is invalid.
In setting, digit of
flicker position
increases. In
adjustment of display,
accumulation is
cleared.
Short key → Start up。
Long key → shut down.
Confirm operation,
to save setting
data.
6
、
Parameter Setting
Setting entry:
Press the 『Func』 button on the operating panel in the state of
normal weighing.
If F1.14 = 0, you can set all the parameters within F1~F5.
If F1.14 = 1, you can only set all parameters within F2~F5.
If F1.14 = 1 and you need to set the parameters within F1 menu, you
can press the calibration switch button until the F1 menu is entered.
F1 Parameter Setting
F1.1 Measuring Range
Selectable parameters: 3~200000 (default value: 6)
F1.2 Decimal Places
Selectable parameters: 0 ---- no decimal point
0.0 ---- 1 decimal place

8
0.00 ---- 2 decimal places
0.000 ---- 3 decimal places (default value)
0.0000 ---- 4 decimal places
F1.3 Number of Divisions
Selectable parameters: 1 (default value), 2, 5, 10, 20, 50
F1.4 Calibration Unit
Selectable parameters: 0 ---- kg (default value)
1 ---- lb
F1.5 Gravitational Acceleration
Selectable parameters: 9.70000~9.99999. Default value = 9.79455.
F1.6 Null-point Calibration
【 】 Keeping empty the scale
Remove the weights on the weighing platform
to guarantee the scale is in the empty state.
Press the『ON/OFF』。key and the indicator will
display【】,The displayed digits will
reduce slowly until the indicator displays
【】,In the end it will display【 】
for one second, which indicates the end of
null-point calibration.
F1.7 Load-point Calibration
【 】 Loading weights
Load weights on the weighing platform to ensure
that 10% of full-scale value ≤ weight of weights
≤ full-scale value, and then press the『ON/OFF』
key to start the next step.
【 】 Entering the same weight value as that of the
loaded weights.
Entering the same weight value as that of the
loaded weights, please press the『ON/OFF』key
after the scale becomes stable, and the meter
will display [ ]. After that, the displayed
digits will reduce slowly until the meter displays

9
[ ]. In the end it will display [ ] for
one second, which indicates the end of
null-point calibration.
F1.8 Automatic Null Tracking
Selectable parameters: OFF, 1 d, 2 d, 3 d, 5 d (default value)
F1.9 Automatic Reset Range at Startup
Selectable parameters: OFF, 2 %, 10 %, 20 % (default value)
F1.10 Button Reset Range
Selectable parameters: OFF, 2 %, 10 % (default value), 20 %
F1.11 Digital Filter
Selectable parameters: 0 ---- Mild Filtering
1 ---- Moderate Filtering (default value)
2 ---- Severe Filtering
F1.12 Steady Range
Selectable parameters: 1 d, 2 d, 3 d (default value)
F1.13 Overload Display Range
Selectable Parameters: 9d, 5% (default value), 10%, 20% ±120%
F1.14 F1 Menu Protection
Selectable Parameters: 0 ---- Enter F1 menu by keyboard
operation
1 ---- Enter F1 menu by pressing the calibration button
F1.15 Strong filter mode
Selectable Parameters:0 ---- Not enabled
1 ---- start using
F1.16 Restoring Factory Default
Set the parameters within F1~F4 as the defaults, which can’t
impact the parameters of standard scale.
F2 Application Function Setting
F2.1 Function Selection
Selectable Parameters: 0 ---- ×10 functions

10
1 ---- Weight keeping function
2 ---- Percentage display function
3 ---- Weight checking and sorting function
(default value)
4 ---- Accumulative scale function
5 ---- Counting scale function
6 ---- Animal scale function
7 ---- Peak holding function
F2.2 Empty-scale threshold value
Selectable Parameters: 0~ full range (default value: 0.001)
F2.3 Target Weight for Weight Checking and Sorting
Selectable Parameters: 0~ full range (default value: 2.000)
F2.4 Positive Error for Weight Checking and Sorting
Selectable Parameters: 0~ full range (default value: 0.100)
F2.5 Negative Error for Weight Checking and Sorting
Selectable Parameters: 0~ full range (default value: 0.100)
F2.6 Access to Target Weight for Weight Checking and Sorting, and
Counting Sample Weight
Selectable Parameters: 0 ---- Access to Platform Weighing
(default value)
1 ---- Manual Input Access
F3 Energy-saving Parameter Setting
F3.1 Automatic backlight energy-saving time setting
Parameters selectable: 0 ~ 999。
0: Backlight NC;
1: Backlight NO;
2~999: 2~999s automatic backlight (default 10s)
F3.2 Automatic OFF energy-saving time setting
Parameters settable: 0 ~ 60min. (default 5 minutes)
“0” stands for deactivation of this function.
F3.3(DATE)
F3.4(TIME)

11
F4 Serial-port Setting
F4.1 UART0 Communication Interface Parameter
F4.1.1 Communication Mode
Selectable Parameters:
0 ---- command out
1 ---- continuous output
2 ---- Key output
F4.1.2 Setting of Data and Check Pit
Selectable Parameters::8_N_1 ---- 8-pit no parity check (default value)
7_E_1 ---- 7-pit odd parity check
7_O_1 ---- 7-pit even parity check
8_E_1 ---- 8-pit odd parity check
8_O_1 ---- 8-pit even parity check
F4.1.3 Baud Rate
Selectable Parameters:600, 1200, 2400,4800、9600(default value)
F4.2 Parameter Setting of UART1 Printing Interface
F4.2.1 Whether to connect the printer
Selectable Parameters: 0 ---- not connected to the miniprinter (default)
1 ---- connected to the miniprinter
F4.2.2 Printing Carriage Return Character
Selectable Parameters: 0~9 carriage return characters
(default value: 3)
F4.2.3 Printing Setting of Accumulative Scale
Selectable Parameters: 0 ---- total accumulative data for printing (default value)
1 ---- printing detail + total accumulative
data
F4.2.4 Print language setting
Selectable Parameters: 0 ---- in Chinese (default)
1 ---- in English
(Eng,CH1)
F5 Maintenance and Service

12
F5.1 Key test
Instrument display【 】, press『Print』、 『Zero』、『 Tare』、
『Gross』、『 lb/kg』、『 Total』in order, and the instrument displays
【】、【 】、【 】、【 】、【 】、
【 】,press 『Hold』 to quit key test.
F5.2 Display screen test
All strokes of meter display will have self-inspection, to observe
whether there is lacks of strokes.
Press 『Hold』 or 『ON/OFF』 to quit test of display screen.
F5.3 Display current internal code
The display will show internal code of current instrument after
smoothing. Press 『Hold』 or 『ON/OFF』 to quit the interface.。
7
、
Function Description
● ×10 function F2.1 = 0
Operation
In normal weighing mode, press 『Total』 on the operation panel once,
the display accuracy will promote for 10 times and icon ×10 gets on at
the same time; press the key again it returns to normal weighing mode
and the icon of ×10 gets off.
Note: 1. In net weight status, it’s forbidden to change over to ×10 function status;
2. If there have been 4 decimal places, it’s forbidden to change over to ×10 function
status;
3. Under ×10 function status, both serial output and printing output are forbidden.
4. Under ×10 function status, it’s forbidden to tare.
● Weight hold function F2.1 = 1
How to operate In normal weighing mode, press 『Hold』 on the operation
panel once, the weight presently displayed can be locked and “Hold”
indicator gets. Weight Hold operation is valid only when the displayed
weight is ≥F2.2 set value; otherwise, it displays invalid operation with
【 】 for one second and the scale goes back to normal weighing
mode.

13
If the weight is locked, press 『Hold』 again to unlock and it goes back to
normal weighing mode and “Hold” indicator goes off.
In weight locked status, it refuses to tare, clear tare and zero.
● Percentage display F2.1 = 2
Display introduction
Displaying 【 】 indicates 20.5 %.
Pr = present actual weight / measuring range ×100%
Pressing 『Hold』 key to change over between percentage and weight
● Weight check, sorting scale function F2.1 = 3
Function description:
Parameters as F2.2 = A, F2.3 = B, F2.4 = C, F2.5 = D etc shall be set.
Presently displayed weight is X.
If X ≤ A, no weight check or sorting will be performed.
If X < (B – D), it stands for short weight; the short-weight icon goes
on.
If (B – D) ≤ X ≤ (B – C), it stands for PASS; PASS icon goes on.
If X > (B – C). It stands for overweight; overweight icon goes on.
Target value acquisition
Hold 『Total』 until 【TARGET】 is displayed, and then press『ON/OFF』;
the present target value goes on and flashes.
If F2.6 = 0, press 『ON/OFF, the present weight on the scale will be taken as
new target value and it exits setting interface.
If F2.6 = 1, it displays 【 】, requiring modifying the target value
manually. After the modification, press『ON/OFF』 to save the data and exit
the setting interface.
Press 『Hold』 to exit target value acquisition.
● Accumulating scale function F2.1 = 4
How to operate
In normal weighing state, the scale is at zero; apply the scale with
load and press 『Total』 on the operation panel; if it displays
【 】, it means that the present weight has been counted in

14
the accumulated value; then it goes back to normal weighing status.
If it displays 【 】 for one second and then goes back to normal
weighing mode, it means invalid operation! Causes: 1. The scale must
be zeroed between two successive accumulation operations,
otherwise, accumulation will be refused. 2. Accumulation is valid
only when the displayed weight is ≥F2.2 setting. 3. The scale is in
dynamic status.
Adjusted display, clearing, and printing of accumulated value
In normal weighing mode, hold 『Total』 on the operation panel for
over two seconds, it displays 【 】 for one second; then the
monitor shows the present accumulated value 【 】 or times
of accumulation 【 】; press『Gross』,『Ib/kg』to change over;。
now the adjustment display icon goes on. Now if it’s necessary to
clear the accumulated value, press 『Zero』 to have the display
becomes into “0”. If F4.2.1=1, pressing 『Print』 can print the
accumulated data. To exit the interface, press 『Hold』.
Note: whether to be over the specific data and accumulated data can
be set in F4.6.
● Counting scale function F2.1 = 5
Instrument display
【 】,displays the present counts
Sampling method
1. Check whether the scale is at Zero, otherwise, press 『Zero』.
2. Put the counted material onto the scale platform.
3. Hold 『Total』 until 【 】 is displayed; then press『Print』.
If F2.6 = 0, it displays【 】. Enter the value just counted and then
press 『Print』 to confirm, the sampled data will be saved and the
system exits sampling interface. If F2.6 = 1, it displays【 】; enter
the sample weight and press 『Print』; the system saves the setting
data and exits sampling setting interface.
4. In this function, press『Hold』 to switch between counts and weight.

15
● Animal scale functions F2.1 = 6
Operation method
In normal weighing state, place the animal on the weighing platform and its
weight must be ≥threshold value set in F2.2. Press 『Total』, instrument will
collect data sampling. After sampling, the average value of sampling data
will be locked, showing X.XXX kg; animal state indicator flashes. Press『Print』
to print; press 『Hold』 or 『Total』 to quit the interface.
● 8
、
Prompt Message of Instrument
The instrument has extremely high stability and reliability, thus is not easy to
have error in general situation. Once an error occurs, please make clear the error
first and observe whether the instrument still has error after power-on. Do not
hurry to repair the scale body or instrument. Repair the instrument according to
error code of the instrument as possible
No.
Symbol
Analysis
Treatment Method
1
【 】
【 】
Unable to reset after
startup
1、Determine it is no-load state
in startup;
2、Make zero calibration again.
2
【 】
The weighed object is over
full range for 9d
Reduce weight on weighing
platform
3
【 】
The weighed object is
below 0 for 5d
Press 『Zero』 to reset
4
【 】
【 】
Out of zero clearing range
Check whether the weighing
platform has weight.
Remove weight.
5
【 】
Invalid operation
6
【 】
EEPROM checksum and
error
Press 『Print』 reprint factory
value. Start up again. If the
information occurs again,
return to factory for repair.
Please calibrate the scale again
if the situation does not occur;
Attention: this place is provided
with all parameters of

16
instruments of the factory.
7
【 】
The calibration input
weight is too small
Input≥10 % weight of full range
8
【 】
The weight in calibration is
too light
Load≥10% weight of full range
9
【 】
The scale is dynamic in
scale
Inspect the scale body
10
【 】
Setting error of date and
time
Set according to specification of
date and time
11
【 】
Error of AD initialization
If the error occurs after
restarting, return it to the
factory for repair
12
【 】
In loading scale, it
indicates to loading
weight;
Load weight according to
requirements;
13
【 】
It has enters menu setting
Press 『Print 』 to continue
setting.
14
【 】
End of zero point and
loading point calibration
15
【 】
Including current display
weight in accumulated
value
16
【 】
Accumulated weight
overflows
Clear accumulated weight in
time
17
【 】
Loading default value
18
【 】
Printing
Appendix 1
、
UART0 Output Format Specification
1
、
UART0 output MT format
No.
mark
Bit
analyze
Re-mark
1
(
1
0x28
2
+ OR -
1
0x2b or 0x2d
3
)
1
0x29
4
XXXXXXX
7
7-bit ASCIIcode
Contains decimal

17
points
5
\r
1
0x0d
example:
1、0x28 0x2b 0x29 0x30 0x30 0x31 0x32 0x38 0x2e 0x35 0x0d show 128.5
2、0x28 0x2d 0x29 0x30 0x30 0x31 0x32 0x38 0x2e 0x35 0x0d show -128.5
3、0x28 0x2b 0x29 0x30 0x30 0x31 0x32 0x38 0x31 0x35 0x0d show 12815
4、0x28 0x2d 0x29 0x30 0x30 0x31 0x32 0x38 0x31 0x35 0x0d show -12815
2
、
Output description
F4.1.1 = 0,command out。“W” 57 H from host.
Output format refer to UART0 output MT format。
F4.1.1 = 1,continuous output。
Output format refer to UART0 output MT format。
F4.1.1 = 2,Key output。
Output format refer to UART0 output MT format。
Zero Command: “Z”(0x5a)。
Tara Command: “T”(0x54)。
Zero Command: “C”(0x43)。
Appendix 2. Printing Format Specification
F2.1 = 0、1、4、6,print current resetting, press『Print』
F2.1 = 1 weight maintenance function:
Weight is not in maintenance state: Weight is in maintenance state:
or
REPORT
------------------------
Gross 0.200kg
Tare 0.000kg
Net 0.200kg
REPORT
----------------------
Gross 25.000kg
Status Hold
REPORT
------------------------
Gross 0.200kg
Tare 0.000kg
Net 0.200kg
REPORT
----------------------
Net 25.000kg
Status Hold

18
F2.1 = 3 selection, check weight, press 『Print』:
Underweight Qualified Overweight
F2.1 = 4 accumulation scale, print detailed statement or format of total weight:
Print details and total weight Only print total weight
F2.1 = 5 counting scale, press『Print』:
REPORT
-----------------------
Gross 1.980kg
State Less
REPORT
-----------------------
Gross 2.000kg
State OK
REPORT
-----------------------
Gross 2.020kg
State Over
REPORT
----------------------
Gross 0.547kg
Amount 55
REPORT
-------------------------
1 0.200kg
2 0.175kg
3 0.347kg
4 0.375kg
-------------------------
Count 4
Total 1.097kg
REPORT
-------------------------
Count 4
Total 1.097kg
This manual suits for next models
1