Shimadzu MUX-100D User manual

Manual No. : M503-E017
Revision : C
Mobile X-ray System
MUX-100D
OPERATION MANUAL
Medical Systems Division
Read this manual carefully for correct product use.
Keep it with the equipment.

Mobile DaRt Operation Manual 1
Thank you for purchasing a SHIMADZU medical system. Prior to using this unit,
be sure to read these “Safety Instructions” carefully and operate the equipment
within its design specifications.
The meanings of the following precaution and prohibition terms used in the
operation manual are defined as below:
Safety Instruction Meanings
DANGER
Indicates a direct danger that may cause death
or serious injury if it is not avoided.
WARNING
Indicates an indirect or potential danger that
may cause death or serious injury if it is not
avoided.
CAUTION
Indicates a danger that may cause slight or
medium degree of injury, or may cause
equipment damage or fire if it is not avoided.
NOTE
Indicates information useful for correct system
operation.
Indications for Use
All procedures that use the Mobile DaRt must be prescribed by a physician and
treatment determined by qualified personnel (i.e. Radiological Technician).
The MUX100D is not labeled or intended for home use. It is intended to be used
in a hospital environment by qualified personnel.
The MUX100D is not shipped as a sterile device, as indicated by labeling.
Original version is approved in English.

Mobile DaRt Operation Manual
2
Safety Instructions
Both users and professional service engineers should be sure to follow the safety-related
instructions given here. To fully understand the safety instructions related to each product, be
sure to carefully read each related Operation Manual and/or Installation Manual.
(Items 2. and 3. are intended only for professional service engineers.)
Operating Precautions when Using Electrical Medical Equipment
(for Both Safety and the Prevention of Danger)
1. Only qualified radiographic staff should operate the equipment.
2. During installation:
(1) Do not install the equipment near water supplies or equipment using water.
(2) Do not install the equipment where extreme atmospheric pressures,
temperatures or humidity, drafts, direct sunlight, dust, chlorine or sulfur gas
are present.
(3) During transportation and operation of the equipment, avoid tilting, vibration
and strong impacts.
(4) Do not install the equipment where chemicals or gases are stored.
(5) Be sure to use only the specified electrical power source with the
appropriate frequency, voltage and current (or power consumption).
(6) Check the condition of the battery’s power source (power and polarity)
before operating the equipment.
(7) Properly ground the equipment.
3. Prior to operating the equipment:
(1) Check the conditions of switch contacts, polarity, dial settings, and meters,
and confirm that the equipment performs correctly.
(2) Confirm that the ground is properly connected.
(3) Check that all cords, cables, and other wiring are correctly connected.
(4) Be careful when simultaneously using more than one unit, since it can be
dangerous and lead to an incorrect diagnosis.
(5) Check all external electrical circuits, which are directly connected to a
patient.
(6) Check the condition of the battery’s power source.

Mobile DaRt Operation Manual
3
4. While operating the equipment:
(1) Do not over-exceed time or the amount of equipment use needed for
diagnosis or therapy.
(2) Observe the equipment and patient continuouslyfor early detection of
problems.
(3) When a problem is detected with the equipment or patient, take proper
action to stop the equipment without harming the patient.
(4) Do not let the equipment touch the patient.
5. After operating the equipment:
(1) Turn off all switches and return all dials to their original positions in the
prescribed order. Then, turn off the main power switch.
(2) Do not pull the power cable forcibly from the outlet.
(3) When storing the equipment, pay attention to the following factors:
(i) Do not install the equipment near water supplies or equipment using
water.
(ii) Do not install the equipment where extreme atmospheric pressures,
temperatures or humidity, drafts, direct sunlight, dust, chlorine or sulfur
gas are present.
(iii) During transportation and operation of the equipment, avoid tilting,
vibration and strong impacts..
(iv) Do not install the equipment where chemicals or gases are stored.
(4) Clean all attachments, cables and contacts, and store them in one place.
(5) Keep the equipment clean to avoid problems during the next use.
6. When the unit does not operate correctly, do not try to repair it. Place a
sign on the unit to indicate that it is out of order and contact a certified
repair technician.
7. Do not modify any part of the equipment.
8. Preventive maintenance
(1) The equipment and its parts should be periodically checked.
(2) If the equipment has not been operated for an extended period of time, test it
prior to actual operation to make sure it works correctly and safety.
9. For other related items, be sure to follow all instructions given in the
equipment’s operating manual.

Mobile DaRt Operation Manual
4
WARNING
The responsibility for correct management of use and maintenance
of medical equipment lies with the user.
The equipment should not be used by other than a qualified person
or under the direct supervision of a qualified person.
Unauthorized repair and inspection of the equipment may be
dangerous. All service should be carried out by an authorized
Shimadzu Service center.
NEVER MODIFY THE EQUIPMENT!
The Drugs, Cosmetics and Medical InstrumentsAct of Japan makes
it compulsory that a manufacturer should apply and get approval for
any modifications to the equipment. Modifications without approval
are prohibited.
PERFORM PERIODIC INSPECTION!
Preventive maintenance is required to maintain operational safety
and long-term performance of this system.
This manual gives detailed description of occasional and periodic
maintenance and inspections that a user should perform.
As to the maintenance and inspections that specially trained
specialists must perform, utilize the maintenance agreement offered
by your nearest authorized Shimadzu service center.
CAUTION
Federal law restricts this device to sale by or on the order of a
physician.
(U.S.A. Federal Regulations require this information.)

Mobile DaRt Operation Manual
5
GUARANTEE
The guarantee period of this system is one year from the date of purchase.
Shimadzu Corporation is not responsible for loss/damage of data and the following
failures or damage of the system.
1. Failure/damage of the system caused by improper installation, movement,
maintenance, or repair performed by anyone other than Shimadzu and
authorized Shimadzu service personnel.
2. Failure/damage to a product of Shimadzu caused by another company’s product
not delivered by Shimadzu.
3. Failure/damage of the system resulting from maintenance, repair, etc., using
spare parts other than genuine parts specified by Shimadzu.
4. Failure/damage of the system resulting from not following notes or operating
procedures described in this operation manual.
5. Failure/damage of the system caused by ambient or environmental conditions
beyond the recommended service conditions applicable to this system described
in this operation manual, such as power source and installation requirements.
6. Failure/damage of the system caused by environmental pollution or any natural
disaster such as a fire, earthquake, flooding, or lightning.
7. Failure/damage of the system resulting from installing the system in vehicles,
ships, or aircrafts without Shimadzu’s consent.
8. Failure/damage of the system resulting from installing or connecting software
programs or peripheral devices not specified by Shimadzu.
9. Failure/damage of the system resulting from changing or deleting configuration
files (autoexec.bat, config.sys, etc.) of the basic software or DR system files.
10. Failure/damage of the system caused by use of the system in countries not
specified by Shimadzu.
11. Failure/damage of the system when the system has been purchased from
dealers other than Shimadzu and authorized Shimadzu distributors.
12. Failure/damage of the system caused by hitting or dropping the unit or exposing
it to a shock.
Servicing after the expiration of the guarantee period will be offered at a charge.
Contact your nearest Shimadzu service center.

Mobile DaRt Operation Manual
6
WARNING
NEVER REPAIR OR CHECK INSIDE THE EQUIPMENT !
Responsibility for maintenance of medical equipment is the responsibility of the
user. This equipment should only be operated by, or under the direct supervision
of, a suitably qualified person.
For any repair or maintenance of this equipment contact your nearest authorized
Shimadzu service center.
WARNING
NEVER MODIFY THE EQUIPMEHT !
Modification of the equipment by a user is prohibited as detailed in paragraph 7 of
the “Operating Precautions (for both the Safety and Prevention of Danger) in the
Use of Medical Equipment” (the Notification No.495 by the Chief of the
Pharmaceutical Affairs Bureau, the Ministry of Health and Welfare dated June 1,
1972).
Please Note: The Pharmaceutical Affairs Law of Japan prevents unauthorized
modification of this equipment without authorisation.
WARNING
PERFORM PERIODICAL INSPECTIONS !
Preventive maintenance is essential to ensure the long term operational safety
and performance of this system.
This instruction manual gives detailed description of the periodical maintenance
and inspections that should be performed on this equipment.
For the maintenance and inspections that require qualified specialists to perform
the “Maintenance Agreement” offered by authorized Shimadzu service center is
recommended.
WARNING
PRECAUTIONS FOR X-RAY IRRADIATION
When using X-ray equipment incorrectly, the operator and patient may become
exposed to unnecessary irradiation. Special care to ensure that only the patient
stays in the examination room when X-rays are exposed. If it is necessary for
another person, other than patient, to be in the examination room, ensure that
appropriate protective measures are provided.
WARNING
PRODUCT SAFETY
Do not operate this equipment if there is any uncertainty as to the proper
functioning of the system. Refer all servicing to qualified service personnel.

Mobile DaRt Operation Manual
7
WARNING
CONSIDERATION FOR MECHANICAL SAFETY
Be careful when operating this equipment. Be aware operators and patient’s body
- hand, foot, etc. - may not be got jammed between moving parts of the system or
between the unit and other equipment when moving a part of the unit.
WARNING
NEVER USE THIS SYSTEM AT A PLACE WHERE FLAMMABLE OR
EXPLOSIVE GASES MAY BE PRESENT.
This system is not explosion-proof. Never use this equipment at a place where
flammable or explosive gases may be present.
WARNING
ABOUT ELECTRICAL EARTHING
Connect the power cord plug to a socket with an electrical earthing terminal.
CAUTION
DO NOT SPLASH WATER ON THIS EQUIPMENT !
Do not splash water on this equipment, because there is a risk of electric shock.
To clean the equipment, wipe the surfaces with a cloth dampened with
antibacterial solution.
CAUTION
THE SYSTEM MAY FALL OVER, IF IT IS OPERATED ON A SURFACE
INCLINED MORE THAN 5 DEGREES.
When operating the tube arm of the equipment make sure the equipment is
located on level ground or floor where the slope is less than 5 degrees.
CAUTION
The operator should set the focal spot to skin distance as large as practible to
ensure the absorbed dose to the patient is as low as possible.
CAUTION
PAY ATTENTION TO OBSTACLES DURING DRIVING THE UNIT!
If the unit hits obstacles during driving and the force applied to the bumper switch
by the hitting is less than 2 kg, the front bumper switch will not detect the obstacle.

Mobile DaRt Operation Manual
8
CAUTION
PLEASE BE SURE TO AGING THE X-RAY TUBE !
When using an X-ray tube, a running-in operation called aging is required. Please
aging the X-ray tube with reference to "6.3 Aging the X-ray tube."
CAUTION
Protection from ionizing radiation : Use protective devices and protective clothing
as necessary.
CAUTION
The software (Operating system, OS) installed in this device runs underWindows
XP. Do not change any of the OS settings. If the OS settings are changed, device
operation after those changesis not guaranteed.
CAUTION
Accidents due to operation errors or unforeseen causes can result in deleting or
damage to stored hard disk data (i.e. images and data). Thus, be sure to backup
(archive) important data to an external file system, or print out the file data.
Additionally, since external file systems also have the danger of being damaged,
be sure to perform multiple backups to prevent data deletion, damage or loss.
Also, Shimadzu Corporation assumes no liability for the loss or damage to any
customer data.
CAUTION
Do not turn the main power breaker OFF during the normal operation of this device.
Turning the breaker OFF can result in device damage, an operation error or data
loss or damage.
CAUTION
Be sure to follow the procedures described in this manual to start or stop this
product. Also, do not turn the main power breaker OFF during normal operation of
this device, and especially during the operation of this device’s internal magnetic
disk. Turning the breaker OFF can result in device damage, an operation error or
data loss or damage.
CAUTION
Connect peripheral units (external recording media, etc.) which conform to
IEC60601-1 to USB connector on the top cover.
Use of a non-conforming unit can cause a device accident, operation error, smoking,
overheating, electric shock, and/or damage or loss of data.

Mobile DaRt Operation Manual
9
CAUTION
Shimadzu Corporation assumes no liability in the event of data deletion or
damage, device accident or damage, or loss occurs due to the installation of
software or the connection of a peripheral unit that is not designated by Shimadzu
Corporation.
CAUTION
Shimadzu Corporation assumes no liability in the event of data deletion or
damage, device accident or damage, or loss occurs due to the modification or
deletion of this software’s setting files (Autoexec.bat, config.sys, etc.) and DR
system software files.
CAUTION
Shimadzu Corporation assumes no liability in the event of device data deletion,
damage or loss, due to an operation error or sudden accident.
Software License Agreement
The software used in this product is copyrighted by Shimadzu Corporation,
who possesses all rights, including sublicenses for those rights (for
copyrights, etc.) held by third parties.
The Customer and Shimadzu have the following agreement concerning the
usage of this software.
1. Application
i. This software (hereafter “the software”) includes all software programs,
additional data and the Installation Guide used by this device.
2. Rights
i. The Customer is allowed to use this software with a single (1) device.
ii. If the customer wishes to transfer all of their rights to this software, the
creation of a copy of this software is prohibited and this software in its
entirety (the program, and all its media and user documentation,
including this Installation Guide) must be transferred. The assignor is
then assumed to be legally bound by the terms and conditions of this
agreement.
iii. The Customer shall not, in all or in part, loan, lease, sub-license,
reproduce, modify, edit, reverse-engineer, reverse-compile, or
reverse-assemble this software. These actions also are not allowed by
third parties.
3. Copyright
i. Shimadzu Corporation retains all this software’s ownership, copyright, and
software-related intellectual property rights.

Mobile DaRt Operation Manual
10
NOTE
Use the following cassettes for the bucky device.Other cassettes should not
be used.
CM or INCH size cassettes based on IEC standard 60406.
Attenuation equivalent not greater than 1.6mm Al.
Less than 15mm thickness.
All parts of the cassette such as hinges, locks and marking device shall be
within the minimum radii for carners and edges.
Flatness shall be less than 1mm warp.
Language of Approved Version
NOTE
Original version of this manual is approved in English.
Software Revision
This manual applies to the software revision 1.0
Disposal of the Products and Parts
This system contains materials that can contaminate the environment not displosed
of properly. For the disposal of this product or any of its attachments, contact your
nearest authorized Shimadzu service center.

Mobile DaRt Operation Manual
11
Contents
Chapter 1 Introduction
Introduction 1-2
Chapter 2 Outline
2.1 Purpose 2-2
2.2 Features 2-2
2.3 Use Conditions 2-4
2.3.1 Environmental Conditions 2-4
2.3.2 EMC (Electro-Magnetic Compatibility) 2-5
2.4 Conformity to IEC Standards 2-6
2.5 Classification of Equipment 2-7
2.6 Safety Symbols for System 2-8
Chapter 3 Descriptions of Each Part
3.1 Outside View 3-2
3.2 X-ray Control Panel 3-3
3.3 Power Supply Panel 3-6
3.4 Touch Panel Monitor 3-8
3.5 Hand Switch 3-9
3.6 Driving Handle Section 3-10
3.7 X-ray Tube Operating Section and Collimator 3-11
3.8 Operation Handle 3-13
3.9 Tube Arm Joint Section 3-14
3.10 Main Circuit Breaker and Additional Protective Electrical Earthing Wire 3-15
3.11 CXDI-50G Imaging Unit 3-17
3.12 Imaging Unit / Grid Storage Bin 3-19
3.13 Apron Hanger 3-20
3.14 Switch for activating the emergency brake release function 3-21
3.15 Operational State Indicator Lamp 3-22
3.16 LAN Cable for external connection 3-23
Chapter 4 How to Operate
4.1 Operation Table 4-2
4.1.1 Operation Table 4-2
4.2 Starting and Stopping the System 4-4
4.2.1 Starting the System 4-4
4.2.2 Stopping the System 4-5
4.2.3 Starting Up the DR System 4-7
4.2.4 Stopping the DR System 4-8
4.3 Moving the System 4-9
4.3.1 Imaging Unit Storage During Movement 4-10
4.3.2 Normal Travel Mode 4-11

Mobile DaRt Operation Manual
12
4.3.3 Slow Travel Mode 4-11
4.3.4 Small Movement Switch 4-12
4.3.5 Bumper Switch 4-13
4.4 Emergency Stop Switch 4-15
4.5 Tube Arm Operations 4-16
4.5.1 Tube Arm Lock 4-16
4.5.2 Positioning the X-ray Tube Assembly 4-18
4.5.3 Setting the Irradiation Field 4-19
4.5.4 Positioning of X-ray Field and Image Receptor 4-20
4.6 How to Change Exposure Condition 4-21
4.6.1 How to Set Exposure Condition 4-21
4.6.2 Manual Setting Method 4-21
4.6.3 Setting Method with Anatomical Programs 4-22
4.6.4 Registration of Anatomical Programs 4-23
4.6.5 Delete of Anatomical Programs 4-25
4.7 Making an X-ray Exposure 4-27
4.7.1 Archiving an Exposure 4-28
4.7.2 Thumb Nail Images 4-30
4.7.3 To transfer images from the PC hard drive to either a DICOM Printer,
DICOM workstation or to a DICOM PACS system 4-30
4.7.4 Saving an Image to an External Media 4-31
4.7.5 Filming Procedures 4-31
4.7.6 Acquiring an Image and Filming with Insufficient Battery Charge 4-33
4.8 Charging the Battery 4-34
4.9 User Setting 4-37
4.9.1 Setting and Registering User Settings 4-38
4.9.2 Power-assist setting 4-40
4.9.3 Acceleration setting 4-40
4.9.4 Steer control setting 4-41
4.9.5 Backward power-assist setting 4-41
4.9.6 Auto-power-off time setting 4-42
4.9.7 Buzzer settings 4-42
4.9.8 Setting collimator lamp control 4-43
4.9.9 Setting APR at the time of starting the apparatus 4-44
4.9.10 Setting apparatus ID 4-44
4.10 Emergency Brake Release Function 4-45
4.10.1 How to activate the emergency brake release function 4-45
4.10.2 Releasing the emergency brake 4-45
Chapter 5 Options
5.1 Remote Controller 5-2
5.2 Protective Screen 5-5
5.3 Optional Grid 5-6
5.4 Dose Area Meter 5-7

Mobile DaRt Operation Manual
13
Chapter 6 Maintenance and Inspection
6.1 Inspection 6-2
6.2 Cleaning 6-12
6.3 Aging X-ray Tubes 6-13
6.4 Expendable Parts List 6-15
Chapter 7 Self-Diagnostic Functions
7.1 Warning Indication Lamp 7-2
7.2 Fault Indication Lamp 7-4
7.3 CXDI 6.0 Operation Manual 7-5
Chapter 8 Others
8.1 Name Plate and Labels 8-2
8.2 Regulatory Information (For Europe) 8-14
Chapter 9 Specifications
9.1 Specifications 9-2
9.2 Dimensional Drawing of the System 9-6
9.3 X-ray Reference Axis and Focal Spot Position 9-7
9.4 Exposure Condition 9-8
9.5 Imaging Unit 9-10
9.6 Grid (Option) 9-11
9.7 Anatomical Program Number 9-12

Mobile DaRt Operation Manual
14
No Text

Mobile DaRt Operation Manual
1-1
Chapter 1
Introduction
Chapter Contents
Introduction

Chapter 1 Introduction
Mobile DaRt Operation Manual
1-2
Introduction
This manual provides information pertaining to the system’s description,
operation, and specifications.
Review the Canon (CXDI-50G) Manual to understand the operation of the
acquisition system.
Read this manual carefully to ensure correct use. Be sure that the warnings
(NOTES and CAUTIONS) detailed in this instruction manual are fully understood
when operating this product.
After reading this manual, retain it in a convenient location so that it may be
referred to at any time.
This manual does not cover the X-ray tube assembly. Read the following manual
for the X-ray tube assembly and Heating & Cooling Curve.
X-ray tube model name 0.7U163CS-36
X-ray tube document
number M535-E219
Heating & Cooling Curve
document number M535-E274

Mobile Art Plus Operation Manual
2-1
Chapter 2
Outline
Chapter Contents
2.1 Purpose
2.2 Features
2.3 Use Conditions
2.4 Conformity to IEC Standards
2.5 Classification of Equipment
2.6 Safety Symbols for System

Chapter 2 Outline
Mobile DaRt Operation Manual
2-2
2.1 Purpose
The mobile X-ray system has been designed to deliver exceptional mobile X-ray
performance and image quality. It will also ensure optimum operator and patient
safety.
CAUTION
Do not use the unit for any other purpose. Do not connect the unit to any other
equipment by electrical or mechanical means. Unauthorized modification to any
part of this equipment may compromise the patient and operator safety and will
void all the manufacturer's warranty conditions.
2.2 Features
Smooth and quiet movement
The system may be moved quietly and smoothly with its power assisted
drive mechanism.
Compact design
The unit, with its exceptional forward view, can be moved into limited
spaces confidently and with complete control.
Easy positioning
The unit may be moved into position easily and speedily with the unique
positioning central switches (Inch –mover) mounted on the x-ray tube head.
Immediate response
Three seconds after an exposure is taken a thumb nail image appears on
the display, allowing on-site image confirmation.
High throughput
Improves workflow by both eliminating cassette preparation, image
development, and reading in of CR, and using the hospital’s internal
network to send image data to an imager or PACS.
High-frequency inverter
Use of Shimadzu's high-frequency inverter technology, with a maximum
frequency of 60 kHz, generates the high voltage with the lowest ripple
factor available, ensuring an efficient X-ray generation.
Anatomical programs
Anatomical programs are installed as standard, facilitating the setting of
optimum exposure conditions.

2.2 Features
Mobile DaRt Operation Manual
2-3
Cordless system
The built-in battery power-supply enables X-ray exposures to be generated
without requiring the equipment to be connected to the main electrical
supply.
Plain indicator
The large indicator provides a plain indication of proceeding from the
preparation to the completion of an exposure.
.
Other manuals for MUX-100D
2
Table of contents
Other Shimadzu Medical Equipment manuals
Popular Medical Equipment manuals by other brands

ECOPOSTURAL
ECOPOSTURAL C5990 instructions

MX HEALTH
MX HEALTH MX86200GR user manual

Care of Sweden
Care of Sweden Optimal Evacuation sheet Instructions for use

DHG
DHG RoMedic StretcherBar 70200006 Instructions for use

Bedfont
Bedfont GastroCH4ECK Gastrolyzer user manual

Storz
Storz AIDA Interface II instruction manual