When FPM reading is displayed but unit is not flashing,
flash tube may need to be replaced.
1. Unplug line cord from power line. Turn power switch
OFF. Wait a few minutes until stroboscope is cool
before replacing flash tube.
2. Remove protective window by removing the 4 screws.
3. Use a rag and pull the tube out of its socket by rocking
it slightly up and down. Do not use bare hands to
remove tube, it may break and cause injury.
4. Insert new tube using the technique mentioned above.
Make sure that the tube is placed properly in the socket
otherwise it will touch the reflector. Tube should be set
symetrically within the neck of the reflector.
5. Replace protective window.
Flash Tube Replacement
MODELMODEL
MODELMODEL
MODEL DTDT
DTDT
DT- 721(AC)- 721(AC)
- 721(AC)- 721(AC)
- 721(AC)
FLASHING RANGEFLASHING RANGE
FLASHING RANGEFLASHING RANGE
FLASHING RANGE 40 — 12,500 FPM (Flashes Per Minute)
ACCURACYACCURACY
ACCURACYACCURACY
ACCURACY ±0.02% of reading
RESOLRESOL
RESOLRESOL
RESOLUTIONUTION
UTIONUTION
UTION 0.1 , 40.0 – 4,999.9 FPM
0.2 , 5,000.0 – 7,999.8 FPM
0.5 , 8,000.0 – 9,999.5 FPM
1.0 , 10,000.0 – 12,500.0FPM
DISPLDISPL
DISPLDISPL
DISPLAA
AA
AYY
YY
Y 5-Digit LED, 0.3" (8 mm) height
RR
RR
RAA
AA
ATETE
TETE
TE Divide by 2, Multiply by 2
UPDAUPDA
UPDAUPDA
UPDATE TIMETE TIME
TE TIMETE TIME
TE TIME Ext. mode: varies with flashing rate
OPEROPER
OPEROPER
OPERAA
AA
ATING TIMETING TIME
TING TIMETING TIME
TING TIME Continuous
FLASH TUBE POWER/LIFEFLASH TUBE POWER/LIFE
FLASH TUBE POWER/LIFEFLASH TUBE POWER/LIFE
FLASH TUBE POWER/LIFE Xenon, 10 W, 100 million flashes
FLFL
FLFL
FLASH DURASH DUR
ASH DURASH DUR
ASH DURAA
AA
ATIONTION
TIONTION
TION 10 - 15 µs
SIGNAL OUTPUTSIGNAL OUTPUT
SIGNAL OUTPUTSIGNAL OUTPUT
SIGNAL OUTPUT NPN Open Collector (24 VDC max.,
50 mA) 150 µs typical
PHASE SHIFPHASE SHIF
PHASE SHIFPHASE SHIF
PHASE SHIFTT
TT
T Internal mode only by using the “+”
and “–” switches
EXTEXT
EXTEXT
EXT. TRIGGER. TRIGGER
. TRIGGER. TRIGGER
. TRIGGER a) 12-24 VDC thru a 3-wire NPN output
sensor (requires external power supply)
INPUT SIGNALINPUT SIGNAL
INPUT SIGNALINPUT SIGNAL
INPUT SIGNAL b) 12 VDC thru a 2-wire proximity
sensor (leakage current 1 mA max.,
Load current 8 mA min.)
c) Switch or Relay contact
POWER REQUIREMENTSPOWER REQUIREMENTS
POWER REQUIREMENTSPOWER REQUIREMENTS
POWER REQUIREMENTS 115 VAC ±10% 60/50 Hz
OPEROPER
OPEROPER
OPERAA
AA
ATING TEMPERTING TEMPER
TING TEMPERTING TEMPER
TING TEMPERAA
AA
ATURETURE
TURETURE
TURE 32° – 104° F (0 – 40° C)
DIMENSIONSDIMENSIONS
DIMENSIONSDIMENSIONS
DIMENSIONS 9" L x 4.3" W x 8.5" H
(Height includes handle)
WEIGHTWEIGHT
WEIGHTWEIGHT
WEIGHT 1.5 lbs. (0.7 Kg)
ACCESSORIES AVAILABLEACCESSORIES AVAILABLE
ACCESSORIES AVAILABLEACCESSORIES AVAILABLE
ACCESSORIES AVAILABLE Carrying Case
Specifications