Siemens RD150 User manual

Operating Instructions • 07/2018
Indication and adjustment
RD150
4 … 20 mA/HART
External display and adjustment unit

2MAN-100472RD150 - Operating Instructions
57030-EN-180717

3
MAN-100472 RD150 - Operating Instructions
57030-EN-18071757030-EN-180717
Contents
1 About this document............................................................................................................... 5
1.1 Function ........................................................................................................................... 5
1.2 Target group ..................................................................................................................... 5
1.3 Symbols used................................................................................................................... 5
2 For your safety ......................................................................................................................... 6
2.1 Authorised personnel ....................................................................................................... 6
2.2 Appropriate use................................................................................................................ 6
2.3 Warning about incorrect use............................................................................................. 6
2.4 General safety instructions............................................................................................... 6
2.5 EU conformity................................................................................................................... 7
2.6 NAMUR recommendations .............................................................................................. 7
2.7 Installation and operation in the USA and Canada ........................................................... 7
3 Product description................................................................................................................. 8
3.1 Conguration.................................................................................................................... 8
3.2 Principle of operation........................................................................................................ 9
3.3 Packaging, transport and storage................................................................................... 10
3.4 Accessories and replacement parts ............................................................................... 11
4 Mounting................................................................................................................................. 12
4.1 General instructions ....................................................................................................... 12
4.2 Mounting instructions ..................................................................................................... 12
5 Connecting to power supply.................................................................................................15
5.1 Preparing the connection ............................................................................................... 15
5.2 Connection technology and steps .................................................................................. 16
5.3 Wiring plan ..................................................................................................................... 18
5.4 Connection HART standard ........................................................................................... 18
5.5 Connection HART multidrop........................................................................................... 19
5.6 Connection signal conditioning instrument/four-wire sensor .......................................... 19
5.7 Connection example ...................................................................................................... 20
5.8 Switch-on phase............................................................................................................. 20
6 Set up with the display and adjustment module................................................................21
6.1 Short description ............................................................................................................ 21
6.2 Insert display and adjustment module............................................................................ 21
6.3 Adjustment system......................................................................................................... 22
6.4 Measured value indication - Selection of national language........................................... 23
6.5 Start menu...................................................................................................................... 24
6.6 Parameter adjustment - RD150...................................................................................... 24
6.7 Parameter adjustment - Sensors via Generic HART....................................................... 29
7 Setup via PACTware...............................................................................................................31
7.1 Connect the PC.............................................................................................................. 31
7.2 Parameter adjustment with PACTware............................................................................ 32
7.3 Saving the parameterisation data................................................................................... 33
8 Diagnostics and servicing .................................................................................................... 34
8.1 Maintenance .................................................................................................................. 34
8.2 Asset Management function........................................................................................... 34
8.3 Rectify faults................................................................................................................... 35

4MAN-100472RD150 - Operating Instructions
57030-EN-180717
8.4 Software update ............................................................................................................. 36
8.5 How to proceed if a repair is necessary.......................................................................... 36
9 Dismount.................................................................................................................................37
9.1 Dismounting steps.......................................................................................................... 37
9.2 Disposal ......................................................................................................................... 37
10 Supplement ............................................................................................................................38
10.1 Technical data ................................................................................................................ 38
10.2 HART communication .................................................................................................... 41
10.3 Dimensions .................................................................................................................... 42
10.4 Trademark ...................................................................................................................... 45
Safety instructions for Ex areas
TakenoteoftheExspecicsafetyinstructionsforExapplications.
These instructions are attached as documents to each instrument
with Ex approval and are part of the operating instructions manual.
Editing status: 2018-07-12

5
MAN-100472 RD150 - Operating Instructions
57030-EN-180717
1 About this document
1.1 Function
This operating instructions manual provides all the information
you need for mounting, connection and setup as well as important
instructionsformaintenance,faultrectication,theexchangeofparts
and the safety of the user. Please read this information before putting
the instrument into operation and keep this manual accessible in the
immediate vicinity of the device.
1.2 Target group
This operating instructions manual is directed to trained personnel.
Thecontentsofthismanualmustbemadeavailabletothequalied
personnel and implemented.
1.3 Symbols used
Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can result.
Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result.
Danger: If this warning is ignored, serious injury to persons and/or
destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
• List
The dot set in front indicates a list with no implied sequence.
→ Action
This arrow indicates a single action.
1 Sequence of actions
Numbers set in front indicate successive steps in a procedure.
Battery disposal
This symbol indicates special information about the disposal of bat-
teries and accumulators.

6MAN-100472RD150 - Operating Instructions
57030-EN-180717
2 For your safety
2.1 Authorised personnel
All operations described in this documentation must be carried out
only by trained specialist personnel authorised by the plant operator.
During work on and with the device, the required personal protective
equipment must always be worn.
2.2 Appropriate use
The RD150 is suitable for measured value indication and adjustment
of 4 … 20 mA/HART sensors.
Youcannddetailedinformationabouttheareaofapplicationin
chapter "Product description".
Operational reliability is ensured only if the instrument is properly
usedaccordingtothespecicationsintheoperatinginstructions
manual as well as possible supplementary instructions.
2.3 Warning about incorrect use
Inappropriate or incorrect use of this product can give rise to applica-
tion-specichazards,e.g.vesseloverllthroughincorrectmounting
or adjustment. Damage to property and persons or environmental
contamination can result. Also, the protective characteristics of the
instrument can be impaired.
2.4 General safety instructions
This is a state-of-the-art instrument complying with all prevailing
regulations and directives. The instrument must only be operated in a
technicallyawlessandreliablecondition.Theoperatorisresponsi-
ble for the trouble-free operation of the instrument. When measuring
aggressive or corrosive media that can cause a dangerous situation
if the instrument malfunctions, the operator has to implement suitable
measures to make sure the instrument is functioning properly.
During the entire duration of use, the user is obliged to determine the
compliance of the necessary occupational safety measures with the
current valid rules and regulations and also take note of new regula-
tions.
The safety instructions in this operating instructions manual, the na-
tional installation standards as well as the valid safety regulations and
accident prevention rules must be observed by the user.
For safety and warranty reasons, any invasive work on the device
beyond that described in the operating instructions manual may be
carried out only by personnel authorised by the manufacturer. Arbi-
traryconversionsormodicationsareexplicitlyforbidden.Forsafety
reasons,onlytheaccessoryspeciedbythemanufacturermustbe
used.
To avoid any danger, the safety approval markings and safety tips on
the device must also be observed and their meaning read in this oper-
ating instructions manual.

7
MAN-100472 RD150 - Operating Instructions
57030-EN-180717
2.5 EU conformity
ThedevicefullsthelegalrequirementsoftheapplicableEUdirec-
tives.ByaxingtheCEmarking,weconrmtheconformityofthe
instrument with these directives.
2.6 NAMUR recommendations
NAMUR is the automation technology user association in the process
industry in Germany. The published NAMUR recommendations are
acceptedasthestandardineldinstrumentation.
ThedevicefullstherequirementsofthefollowingNAMURrecom-
mendations:
• NE 21 – Electromagnetic compatibility of equipment
• NE53–Compatibilityofelddevicesanddisplay/adjustment
components
For further information see www.namur.de.
2.7 Installation and operation in the USA and
Canada
This information is only valid for USA and Canada. Hence the follow-
ing text is only available in the English language.
Installations in the US shall comply with the relevant requirements of
the National Electrical Code (ANSI/NFPA 70).
Installations in Canada shall comply with the relevant requirements of
the Canadian Electrical Code.

8MAN-100472RD150 - Operating Instructions
57030-EN-180717
3 Product description
3.1 Conguration
Thetypelabelcontainsthemostimportantdataforidenticationand
use of the instrument:
2
3
1
5
6
4
7
11
10
9
8
Fig. 1: Layout of the type label (example)
1 Instrument type
2 Product code
3 Field for approvals
4 Electronics/Voltage supply
5 Protection rating
6 Order number
7 Identicationcode
8 Serial number of the instrument
9 Reminder to observe the instrument documentation
10 ID number, instrument documentation
11 Device protection class
TheRD150isavailableindierenthousingmaterials,seechapter
"Technical data".
The instrument is optionally available with or without display and
adjustment module.
This operating instructions manual applies to the following instrument
versions:
• Software from 1.10.00
• Hardware from 1.00.00
The scope of delivery encompasses:
• RD150
• Display and adjustment module (optional)
• Mounting accessories (optional)
• Documentation
– This operating instructions manual
– Ex-specic"Safety instructions" (with Ex versions)
– Ifnecessary,furthercerticates
Information:
In this operating instructions manual, the optional instrument features
are also described.The actual range of functions is determined by the
orderspecication.
Type label
Instrument versions
Scope of this operating
instructions manual
Scope of delivery

9
MAN-100472 RD150 - Operating Instructions
57030-EN-180717
3.2 Principle of operation
RD150 is suitable for measured value indication and adjustment of
sensors with HART protocol. The instrument is looped directly into
the 4 … 20 mA HART signal line at any location. Separate external
energy is not necessary. RD150 also operates exclusively as an
indicating instrument in a 4 … 20 mA current loop.
The sensor adjustment is carried out in the RD150 integrated in the
display and adjustment module.
4
5
3
1
2
Fig. 2: Connection of the RD150 to the sensor, adjustment via the display and
adjustment module
1 Voltage supply/Signal output sensor
2 RD150
3 Display and adjustment module
4 4 … 20 mA/HART signal cable
5 Sensor
The sensor adjustment is carried out via a PC with PACTware/DTM.
4
5
6
3
2
1
Fig. 3: Connection of the RD150 to the sensor and the PC, adjustment via PC
with PACTware
1 Voltage supply/Signal output sensor
2 RD150
3 Interface adapter
4 4 … 20 mA/HART signal cable
5 Sensor
6 PC with PACTware/DTM
Application area
Sensor adjustment
Sensor adjustment via PC
with PACTware

10 MAN-100472RD150 - Operating Instructions
57030-EN-180717
4 … 20 mA mode: when connected to a 4 … 20 mA signal cable,
RD150 operates exclusively as a display instrument.
Adjustment range: Indication scaling RD150
HART mode: when operated with a 4 … 20 mA/HART sensor, the
RD150 operates as display and HART adjusment instrument.
The parameter adjustment of the sensor is carried out via HART com-
munication. During the parameter adjustment, the RD150 operates as
a Primary or Secondary Master to the sensor.
Adjustment range: Sensor adjustment, indication scaling RD150
HART multidrop: the RD150 can be also used as an indicating
device for a bus participant in a HART multidrop system.
Adjustment range: Sensor adjustment for a bus participant, indication
scaling RD150
3.3 Packaging, transport and storage
Your instrument was protected by packaging during transport. Its
capacity to handle normal loads during transport is assured by a test
based on ISO 4180.
The packaging of standard instruments consists of environment-
friendly, recyclable cardboard. For special versions, PE foam or PE
foil is also used. Dispose of the packaging material via specialised
recycling companies.
Transport must be carried out in due consideration of the notes on the
transport packaging. Nonobservance of these instructions can cause
damage to the device.
The delivery must be checked for completeness and possible transit
damage immediately at receipt. Ascertained transit damage or con-
cealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and
stored according to the orientation and storage markings on the
outside.
Unless otherwise indicated, the packages must be stored only under
the following conditions:
• Not in the open
• Dry and dust free
• Not exposed to corrosive media
• Protected against solar radiation
• Avoiding mechanical shock and vibration
• Storage and transport temperature see chapter "Supplement -
Technical data - Ambient conditions"
• Relative humidity 20 … 85 %
With instrument weights of more than 18 kg (39.68 lbs) suitable and
approved equipment must be used for lifting and carrying.
Modes
Packaging
Transport
Transport inspection
Storage
Storage and transport
temperature
Lifting and carrying

11
MAN-100472 RD150 - Operating Instructions
57030-EN-180717
3.4 Accessories and replacement parts
The display and adjustment module LG Local Display Interface is
used for measured value indication, adjustment and diagnosis. It can
be inserted into the sensor or the external display and adjustment unit
and removed at any time.
Youcanndadditionalinformationintheoperatinginstructions
manual "LG Local Display Interface" (Document-ID 43838).
Display and adjustment
module

12 MAN-100472RD150 - Operating Instructions
57030-EN-180717
4 Mounting
4.1 General instructions
RD150 functions in any installation position.
Protect your instrument against moisture ingress through the following
measures:
• Use a suitable connection cable (see chapter "Connecting to
power supply")
• Tighten the cable gland or plug connector
• Whenmountinghorizontally,turnthehousingsothatthecable
gland or plug connector point downward
• Lead the connection cable downward in front of the cable entry or
plug connector.
This applies mainly to outdoor installations, in areas where high
humidity is expected (e.g. through cleaning processes) and on cooled
or heated vessels.
To maintain the housing protection, make sure that the housing lid is
closed during operation and locked, if necessary.
Makesurethatthedegreeofcontaminationspeciedinchapter
"Technical data" meets the existing ambient conditions.
4.2 Mounting instructions
The RD150 in all available housing materials is suitable for wall
mounting.
7 mm
(0.28")
82 mm
(3.23")
10 mm
(0.39")
Fig. 4: Drilling dimensions for RD150 for wall mounting
The RD150 with plastic housing is suitable for direct carrier rail
mounting according to EN 50022.
Installation position
Protection against mois-
ture
Wall mounting
Carrier rail mounting

13
MAN-100472 RD150 - Operating Instructions
57030-EN-180717
12
Fig. 5: RD150 with plastic housing for carrier rail mounting
1 Base
2 Carrier rail
The versions with aluminium or stainless steel housing for carrier rail
mounting according to EN 50022 are supplied with unassembled
mounting accessories.The kit consists of an adapter plate and four
mounting screws M6 x 12.
The adapter plate is screwed to the base of RD150 by the user.
3
21
Fig. 6: RD150 with aluminium and stainless steel housing for carrier rail mount-
ing
1 Base
2 Adapter plate with screws M6 x 12
3 Carrier rail
The RD150 for tube mounting is supplied with unassembled mounting
accessories.The kit consists of two pairs of mounting brackets and
four mounting screws M6 x 100.
The mounting brackets are screwed to the base of RD150 by the user.
Tube mounting

14 MAN-100472RD150 - Operating Instructions
57030-EN-180717
21 3
Fig. 7: RD150 for tube mounting
1 4 screws M6 x 100
2 Mounting brackets
3 Tube (diameter 1" to 2")
The RD150 is also available with a plastic housing for panel mount-
ing.The housing is fastened to the rear of the panel by means of the
supplied screw clamps.
4
5
123
Fig. 8: RD150 for panel mounting
1 Inspection window
2 Front panel
3 Screw clamp
4 Housing
5 Plug connector
Front panel mounting

15
MAN-100472 RD150 - Operating Instructions
57030-EN-180717
5 Connecting to power supply
5.1 Preparing the connection
Always keep in mind the following safety instructions:
• Carry out electrical connection by trained personnel authorised by
the plant operator
• If overvoltage surges are expected, overvoltage arresters should
be installed
Warning:
Connect only in the complete absence of line voltage.
Power supply and current signal are carried on the same two-wire
cable.Thevoltagesupplyrangecandierdependingonthesensor.
Thedataforpowersupplyarespeciedinchapter"Technical data".
Make sure that the supply circuits are separated from the mains
circuits and of an energy-limited voltage supply, e.g. of "Class 2" (acc.
to UL 1310, NEC 725 or CAN/CSA C22.2 No. 223), according to inter-
nationallyharmonizedstandardIEC61010-1.
Keepinmindthefollowingadditionalfactorsthatinuencetheoperat-
ing voltage:
• Output voltage of the power supply unit can be lower under nomi-
nal load (with a sensor current of 20.5 mA resp. 22 mA in case of
fault message)
• Voltage loss on the RD150 (see supply circuit in chapter "Techni-
cal data")
Youcanndinformationontheloadresistanceinchapter"Technical
data", (voltage supply of the respective sensor)
The instrument is connected with standard two-wire cable without
screen. If electromagnetic interference is expected which is above the
test values of EN 61326-1 for industrial areas, screened cable should
be used.
We generally recommend the use of screened cable for HART mode.
Usecablewithroundcross-section.Toensurethesealeectofthe
cablegland(IPprotectionrating),ndoutwhichcableouterdiameter
thecableglandissuitablefor.Useacableglandttingthecable
diameter.
Youcanndanoverviewofthecableglandsinchapter"Technical
data".
Metric threads
In the case of instrument housings with metric thread, the cable
glands are screwed in at the factory.They are sealed with plastic
plugs as transport protection.
You have to remove these plugs before electrical connection.
Safety instructions
Voltage supply
Connection cable
Cable glands

16 MAN-100472RD150 - Operating Instructions
57030-EN-180717
NPT thread
In the case of instrument housings with self-sealing NPT threads, it is
not possible to have the cable entries screwed in at the factory. The
free openings for the cable glands are therefore covered with red dust
protection caps as transport protection.
Prior to setup you have to replace these protective caps with ap-
proved cable glands or close the openings with suitable blind plugs.
On plastic housings, the NPT cable gland or the Conduit steel tube
must be screwed into the threaded insert without grease.
Max. torque for all housings, see chapter "Technical data".
If screened cable is necessary, we recommend connecting the cable
screen on both ends to ground potential. In the RD150, the screen
should be connected directly to the internal ground terminal.
In Ex systems it must be ensured that the grounding complies with the
installation regulations.
In electroplating plants as well as plants for cathodic corrosion protec-
tionitmustbetakenintoaccountthatsignicantpotentialdierences
exist.This can lead to unacceptably high currents in the cable screen
if it is grounded at both ends.
5.2 Connection technology and steps
The voltage supply and signal output are connected via the spring-
loaded terminals in the housing.
Connection to the display and adjustment module or to the interface
adapter is carried out via contact pins in the housing.
Information:
The terminal block is pluggable and can be removed from the
electronics.To do this, lift the terminal block with a small screwdriver
and pull it out. When reinserting the terminal block, you should hear it
snap in.
Proceed as follows:
1. Unscrew the housing lid
2. If a display and adjustment module is installed, remove it by turn-
ing it slightly to the left
3. Loosen compression nut of the cable gland and remove blind
plug
4. Remove approx. 10 cm (4 in) of the cable mantle, strip approx.
1 cm (0.4 in) of insulation from the ends of the individual wires
5. Insert the cable into the sensor through the cable entry
Cable screening and
grounding
Connection technology
Connection procedure

17
MAN-100472 RD150 - Operating Instructions
57030-EN-180717
Fig. 9: Connection steps 5 and 6
6. Insert the wire ends into the terminals according to the wiring plan
Information:
Solidcoresaswellasexiblecoreswithwireendsleevesareinsert-
eddirectlyintotheterminalopenings.Incaseofexiblecoreswithout
end sleeves, press the terminal from above with a small screwdriver,
the terminal opening is then free.When the screwdriver is released,
the terminal closes again.
Youcanndfurtherinformationonthemax.wirecross-sectionunder
"Technical data - Electromechanical data".
7. Check the hold of the wires in the terminals by lightly pulling on
them
8. Connect the screen to the internal ground terminal, connect the
external ground terminal to potential equalisation
9. Tighten the compression nut of the cable entry gland. The seal
ring must completely encircle the cable
10. Reinsert the display and adjustment module, if one was installed
11. Screw the housing lid back on

18 MAN-100472RD150 - Operating Instructions
57030-EN-180717
5.3 Wiring plan
Sensor
12
+
( ) (-)
power supply
34
+
( ) (-)
R
off on
HART
14
2
3
Fig. 10: Wiring plan RD150 4 … 20 mA/HART
1 To the sensor
2 Switchforcommunicationresistor(on=activated,o=deactivated)
3 Terminal for connection of the cable screen
4 For power supply
13
2
4
3
12
Fig. 11: Wiring plan RD150 for 4 … 20 mA sensors - panel mounting
1 To the sensor
2 Ground terminal in the switching cabinet for connection of the cable screen
3 For power supply
5.4 Connection HART standard
ThefollowingillustrationshowsinasimpliedwaytheuseofRD150
in conjunction with a HART sensor.
~
=
1
23
Fig. 12: Installation example RD150 in conjunction with an individual sensor
1 Sensor
2 RD150
3 Voltage supply/Processing
Wiring plan
Wiring plan - Panel
mounting

19
MAN-100472 RD150 - Operating Instructions
57030-EN-180717
5.5 Connection HART multidrop
ThefollowingillustrationsshowinasimpliedwaytheuseofRD150
in conjunction with several HART sensors.
1
~
=
23
Fig. 13: Installation example with one RD150 for several sensors in a Multidrop
system
1 Sensor
2 RD150
3 Voltage supply/Processing
5.6 Connection signal conditioning instrument/
four-wire sensor
Thefollowingillustrationshowsthesimpliedtheconnectionof
RD150 to a four-wire sensor with active 4 … 20 mA output.
For this, terminals 1 and 4 on RD150 must be bridged.
Four-wire sensor with ac-
tive 4 … 20 mA output

20 MAN-100472RD150 - Operating Instructions
57030-EN-180717
3
1
2
Sensor
12
+
( ) (-)
power supply
34
+
( ) (-)
R
+
-
off on
HART
Fig. 14: Bridge between terminals 1 and 4 on the RD150
1 Bridge
2 RD150
3 Active sensor
5.7 Connection example
1
3
24
5
12
+
( ) (-)
678
4...20mA Sensor
12
+
( ) (-)
power supply
34
+
( ) (-)
R
off on
HART
Fig. 15: Connection example 4 … 20 mA/HART
1 Voltage supply
2 RD150
3 Connection cable
4 Sensor
5.8 Switch-on phase
After connecting the instrument to power supply or after a voltage
recurrence, the instrument carries out a self-check for approx. 10 s:
• Internal check of the electronics
• Indication of the instrument type, hardware and software version,
measurement loop name on the display or PC
• Indication of a status message on the display or PC
The duration of the warm-up phase depends on the connected sen-
sor.
Thentheactualmeasuredvalueisdisplayed.Youcanndfurther
information on the display in chapter "Measured value indication -
Selection national language".
Table of contents
Other Siemens Industrial Monitor manuals

Siemens
Siemens SIMATIC Industrial Flat Panel IFP2200 User manual

Siemens
Siemens SIMATIC Industrial Flat Panel IFP1500 User manual

Siemens
Siemens INOX PRO SIMATIC IFP1900 User manual

Siemens
Siemens SIMATIC Industrial Flat Panel IFP2200 User manual

Siemens
Siemens Simatic MP 377 User manual

Siemens
Siemens Operator Display Panel II User manual

Siemens
Siemens SIMATIC Industrial Flat Panel IFP1500 User manual

Siemens
Siemens SIMATIC IPC477E User manual

Siemens
Siemens INOX PRO SIMATIC IFP1900 Administrator guide

Siemens
Siemens MTP700 Unified Comfort User manual