Sigma 3-16PK User manual

CONTROLLED COPY
Centre for Biological Engineering
Standard Operating Procedure SOP122
Title: Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK
Location: CBE Laboratory Unit (H21/22)
Version 003 Effective Date: 26th June 2018 Review 26th June 2020
Written by: Victoria Workman
Reviewed by: P.Hourd
Approved by: R/I.Temple
Page 1 of 22
1PURPOSE
To describe procedures for the safe operation and maintenance of the bench standing Sigma 3-16PK
refrigerated centrifuge.
2 SCOPE
This SOP applies to the use and maintenance of Sigma 3-16PK centrifuge by authorised CBE
personnel within CBE laboratories. The centrifuge is used for separating substances or mixtures with a
density of up to max 1.2g/cm3. The SOP covers the use of the centrifuge using the following
designated rotors; swing-out rotor for sealable buckets (Cat.-No. 11180) and swing-out rotor for plate
centrifugation (Cat.-No. 11222). This SOP does not cover the use of any other type of rotor with this
centrifuge. Certified and hermetically sealed accessories (e.g. rotors, buckets or safety cups) MUST be
used for applications involving the centrifugation of infectious, toxic or pathogenic material
SPECIAL NOTES: HEALTH & SAFETY
2.1 Hazards associated with use of centrifuge
•Mechanical failure of rotating parts (often violent).
•Contact with rotating parts.
•Sample leaks causing aerosols, stress corrosion, contamination.
•Sample imbalance causing machine movement/walking.
•Fire/explosion.
•Contact with contaminated components/vapours.
2.2 Safe Working Practices (also refer to Safety Note in Operator Manual)
Users MUST read all Precautionary Statements in Section 2 of the Operators Manual and should
refer to the Guidelines in Annex 1 for further recommendations on the safe use of centrifuges.
(i) Ensure that the centrifuge is placed on a horizontal level surface and there is a clearance of at
least 30 cm around the centrifuge when in operation. Make sure that hazardous materials are
not stored or placed within this area and there is sufficient ventilation.
(ii) The centrifuge should not be placed in a safety cabinet unless an operator protection factor (KI
Discus) test has been carried out with it running in situ and it shown not to compromise operator
protection.
(iii) Always observe the start-up checks detailed in this SOP before using the centrifuge

CONTROLLED COPY
Centre for Biological Engineering
Standard Operating Procedure SOP122
Title: Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK
Location: CBE Laboratory Unit (H21/22)
Version 003 Effective Date: 26th June 2018 Review 26th June 2020
Written by: Victoria Workman
Reviewed by: P.Hourd
Approved by: R/I.Temple
Page 2 of 22
(iv) Do not centrifuge toxic, pathogenic or radioactive material without taking proper precautions for
aerosol containment.
(v) Do not centrifuge substances which could damage or affect the mechanical strength of the
material of the centrifuge, the rotor or the buckets. See Operators Manual for details of chemical
resistance to plastics.
(vi) Do not centrifuge materials capable of developing flammable or explosive vapours.
(vii) Never attempt to slow rotor by hand.
(viii) Only use the rotors and accessories specified by the manufacturer. If unsure of which rotor or
tube to use, ask Responsible Person or call the manufacturer.
(ix) Never exceed the maximum speed for the rotor. NOTE: The centrifuge is equipped with an
automatic rotor identification system. The system performs an automatic rotor check after the
start. If a rotor other than the preselected one is installed, the system will display the installed
rotor and the pre-selection speed will be corrected if necessary.
(x) Use only the correct designated tubes for the rotor. Do not modify them to make them fit
unless using attachments authorized by the manufacturer. NEVER USE GLASS CONTAINERS
IN THE CENTRIFUGE.
(xi) Avoid overfilling tubes or bottles. .
(xii) Place centrifuge vessels symmetrically so that the rotor is loaded evenly. Load vessels outside
the centrifuge. Fill vessels carefully to same weight. Imbalances will result in increased wear of
bearings.
(xiii) Stay at the centrifuge until it is running smoothly. Shut the machine down immediately if there is
any unusual noise or vibration. Report non-resolvable issues to the Laboratory Manager.
(xiv) Do not open the door until the rotor stops spinning.
(xv) Always check for spills. If you find one, clean the centrifuge and rotor thoroughly (Refer to
SOP038).
(xvi) Always clean the equipment after it is used with salts or corrosives.
(xvii) Check that the replacement of time-expired components are made according to the
manufacturer’s Instructions - Do not use rotors or buckets which have exceeded their maximum
operating life.
(xviii) When centrifuging infectious materials, use capped or sealed containers and aerosol
containment sealable buckets

CONTROLLED COPY
Centre for Biological Engineering
Standard Operating Procedure SOP122
Title: Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK
Location: CBE Laboratory Unit (H21/22)
Version 003 Effective Date: 26th June 2018 Review 26th June 2020
Written by: Victoria Workman
Reviewed by: P.Hourd
Approved by: R/I.Temple
Page 3 of 22
(xix) Check that the replacement of time-expired components are made according to the
manufacturer’s Instructions - Do not use rotors or buckets which have exceeded their maximum
operating life.
3.RESPONSIBILITES
3.1. CBE Laboratory Personnel:
(i) Shall ensure that they are familiar with the centrifuge, its controls, requirements and emergency
procedures by reference to this SOP, and the Manufacturer’s Operating Instructions.
(ii) Shall be an experienced centrifuge user or will have received appropriate training. Before a new
user is allowed to operate a centrifuge on their own, competence to operate the centrifuge must
have been assessed by the responsible person or another experienced user.
(iii) Shall ensure that the centrifuge is suitable for the work they intend to carry out.
(iv) Shall record alarms, adverse events, non-conformances or malfunction on the Preventative
Maintenance of Centrifuges Record (Section 6) and notify the Laboratory Manager/Responsible
Person.
(v) Shall complete the cleaning procedures as required and record in the relevant sections on the
Preventative Maintenance of centrifuges Record (refer to section 6)
3.2. The Responsible Person/Laboratory Manager:
(i) Shall ensure laboratory personnel are given suitable information, instruction, training or
supervision in the safe use and maintenance of the equipment, providing the following:
•information on the hazards and risks to health;
•instruction in safe procedures;
•training, where necessary; and effective supervision to ensure, so far as is reasonably
practicable, that centrifuges are operated without risks to the health of employees and other
persons, i.e. including students and visiting research workers
(ii) Shall ensure that the centrifuge and rotors are properly installed and set properly according to the
Operators Manual.
(iii) Shall coordinate weekly maintenance duties to be performed by laboratory personnel (according
to SOP004).
(iv) Shall regularly review maintenance logs and investigate and reported alarms, adverse events or
non-conformances associated with the centrifuge.

CONTROLLED COPY
Centre for Biological Engineering
Standard Operating Procedure SOP122
Title: Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK
Location: CBE Laboratory Unit (H21/22)
Version 003 Effective Date: 26th June 2018 Review 26th June 2020
Written by: Victoria Workman
Reviewed by: P.Hourd
Approved by: R/I.Temple
Page 4 of 22
(v) Shall schedule servicing and preventative maintenance with authorized service representatives.
External maintenance and servicing of the equipment can only be performed after it has been
suitably disinfected (see SOP003) and a 'Decontamination Certificate' has been issued (a
proforma is available on the CBE website).
4. EQUIPMENT AND MATERIALS
1. Sigma 3-16PK refrigerated centrifuge: Ser. No. 137665.
2. Rotors: Swing-out rotor (Cat.-No. 11180), Plate rotor (Cat.-No. 11222).
3. Centrifuge tubes: Corning plug seal cap, sterile, polypropylene, non-pyrogenic, RCF:9400xg,
max 50mL (Fisher, CFT-643-021J).
4. 70% IMS for cleaning.
5. 2% Detergent solution (Neutracon).
5.PROCEDURE
5.1 Pre-run checks
NOTE: A record for the use of the centrifuge and the outcome of routine inspections MUST be
maintained. A safety inspection outlined below must be carried out before each use.
NOTE: Lifetime of the Rotor: After 10 years the rotors and accessories must be inspected by the
manufacturers. The rotor must be scrapped if it has done 50,000 cycles for reasons of safety.
CAUTION: Centrifuging can produce aerosols and therefore bio-hazardous material MUST NOT be
centrifuged in open containers. Sealed containers provide initial protection but should be reinforced by
the use of sealed buckets or rotors, which reduce the hazard if the container collapses.
Before each centrifuge run:
(i) Inspect all the all safety relevant parts of the centrifuge i.e. rotor and accessories (e.g. cover
seal, adapters and rotor tie down screw) for cleanliness and visible signs of damage, or wear,
abrasion or discolouration or wear e.g. cracks, nicks, chemical degradation, pitting or corrosion.
(ii) Inspect buckets and centrifuge tubes for visible signs of damage, malformation or wear (e.g.
cracks, nicks, chemical degradation, pitting or corrosion).
CAUTION: If any sign of visible damage or wear is detected, DO NOT USE the centrifuge.
Consult the Responsible Person and record the outcome of the inspection on the Preventative
Maintenance of centrifuges Record (Section 6), otherwise proceed as follows:
NOTE: Slight scratches and cracks can cause severe inner damage to the rotor materials,
which are difficult to detect with the eye. If in doubt, DO NOT USE the rotor.

CONTROLLED COPY
Centre for Biological Engineering
Standard Operating Procedure SOP122
Title: Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK
Location: CBE Laboratory Unit (H21/22)
Version 003 Effective Date: 26th June 2018 Review 26th June 2020
Written by: Victoria Workman
Reviewed by: P.Hourd
Approved by: R/I.Temple
Page 5 of 22
(iii) Make certain the rotor tie down nut is secure before starting the centrifuge. Check that the rotor
is seated on the drive hub correctly.
(iv) Inspect the supporting lugs on the rotor; lubricate if required (use Sigma grease Part no. 70
284).
(v) Fill sample containers only within the range recommended by the manufacturer. Never fill
sample containers when they are mounted in the centrifuge unless contained in a head
specifically designed for this method of filling.
(vi) Distribute the load symmetrically around the rotation assembly so that it is evenly balanced and
complies with the manufacturer's instructions in this respect. Always consider all components
when balancing.
(vii) Select a centrifuging speed up to the maximum recommended by the manufacturer according to
the head and accessories to be used or to the recommended reduced speed if the sample has
a high density.
(viii) Follow the manufacturer's instructions for starting and (where the control is not automatic) for
increasing the speed of rotation.
(ix) Make certain the rotor tie down nut is secure before starting the centrifuge. Check that the rotor
is seated on the drive hub correctly.
(x) Make sure the bowl is dry and that the drive spindle is clean. In the event of condensation water
formation, dry the centrifugal chamber by wiping out with an absorbent cloth
5.2 Inserting and removing the rotor
NOTE: It is not permitted to use any other rotor other than those designated by the
manufacturer.
(i) Open the centrifuge by pressing the illuminated lid key
(i) Unscrew rotor tie-down screw from the motor shaft (anticlockwise).
(ii) Lower the rotor straight down onto the motor shaft.
(iii) Tighten the rotor tie-down screw clockwise with the supplied rotor wrench. In doing so, hold the
rotor at its outer rim.
(iv) Once a day or after approx. 20 cycles the rotor must be removed and then re-inserted. This
ensures a proper connection between rotor and shaft.
5.2.1 Placing the tubes in the swing-out rotor

CONTROLLED COPY
Centre for Biological Engineering
Standard Operating Procedure SOP122
Title: Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK
Location: CBE Laboratory Unit (H21/22)
Version 003 Effective Date: 26th June 2018 Review 26th June 2020
Written by: Victoria Workman
Reviewed by: P.Hourd
Approved by: R/I.Temple
Page 6 of 22
(i) Swing out rotor (part number: 11180) takes a maximum of 1100g in each of the four buckets.
Maximum speed permitted is 4700rpm
(ii) Distribute the load symmetrically in the inserts within the buckets so that the rotor is evenly
balanced. It is essential that opposing buckets carry an identical load, and in particular are filled
with a liquid of density not grossly dissimilar to that being centrifuged. Always consider all
components (e.g. inserts and trunnion rings) when balancing. There should be no more than
100g difference between buckets next to each other
(iii) Avoid contaminating the outside of a sealed bucket. Never fill sample containers when they are
mounted in the centrifuge.
(iv) Place the buckets correctly on the trunions, at a slight angle, pointing into the middle. ALWAYS
insert all four buckets to ensure correct running of the centrifuge
(v) Close the lid of the centrifuge by firmly pressing it down. Ensure that this is accompanied by a
clicking sound. The lid will be locked once the centrifuge is put into operation.
5.2.2 Placing plates in the plate holder rotor
(i) Each plate holder takes a maximum of three plates, up to a total of 600g. The maximum speed
permitted with the rotor (part number: 11222) is 3000rpm
(ii) Place plates in holders so that they are diametrically (diagonally) balanced, not mirror images of
each other, i.e.:

CONTROLLED COPY
Centre for Biological Engineering
Standard Operating Procedure SOP122
Title: Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK
Location: CBE Laboratory Unit (H21/22)
Version 003 Effective Date: 26th June 2018 Review 26th June 2020
Written by: Victoria Workman
Reviewed by: P.Hourd
Approved by: R/I.Temple
Page 7 of 22
(iii) Insert the holders into the rotor, ensuring that the holders fit correctly into the trunions. Always
use both holders
5.3 Display/ program options
1 Start key
2 Time display
3 Temperature scale
4 Speed display
5 RCF display
6 Lid key
7 Stop key
8 Display for rotor, deceleration curve and
programs
9 Rotary knobs

CONTROLLED COPY
Centre for Biological Engineering
Standard Operating Procedure SOP122
Title: Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK
Location: CBE Laboratory Unit (H21/22)
Version 003 Effective Date: 26th June 2018 Review 26th June 2020
Written by: Victoria Workman
Reviewed by: P.Hourd
Approved by: R/I.Temple
Page 8 of 22
The centrifuge display has the following display fields:
1 Time field
2 "Lock" symbol for lockdown
3 Temperature field
4 Field for rotor selection, imbalance indication, run mode, and program selection
5 Speed field
6 RCF field
5.3.1 Setting the rotor
(i) Turn the left rotary knob until “set” appears in front of the option “Rotor” on the display
(ii) Select the option by pressing or turning the right rotary knob. "Set" will now flash.
(iii) Turn the right rotary knob until the desired rotor number is displayed.
(iv) Press the right rotary knob to confirm the input. If this is not done, the value will be automatically
reset to the last setting.
5.3.2 Setting the time
(i) Turn the left rotary knob until “set” appears in the lower left of the display
(ii) Select the option by pressing or turning the right rotary knob. “Set” will now flash.
(iii) Turn the right rotary knob until the desired duration is displayed.
(iv) Press the right rotary knob in order to confirm the input. If this is not done, the value will be
automatically reset to the last setting.
(v) After 11:59 min, the display changes from “min:sec“ to “hrs:min“. The time can then be changed
in steps of 10 minutes.

CONTROLLED COPY
Centre for Biological Engineering
Standard Operating Procedure SOP122
Title: Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK
Location: CBE Laboratory Unit (H21/22)
Version 003 Effective Date: 26th June 2018 Review 26th June 2020
Written by: Victoria Workman
Reviewed by: P.Hourd
Approved by: R/I.Temple
Page 9 of 22
(vi) To change the time in steps of 1 second (instead of in steps of 10 seconds in the min:sec mode)
or in steps of 1 minute (instead of in steps of 10 minutes in the hrs:min mode): Keep the stop key
pressed while setting the desired runtime with the right rotary knob.
(vii) To run in Short Run or Continuous Run mode see the user manual
(viii)It is possible to change the runtime whilst the centrifuge is running
5.3.3 Setting the speed
(i) To preselect a speed turn the left rotary knob until “set” appears in front of the parameter “Speed”
on the display.
(ii) Select the option by pressing or turning the right rotary knob. “Set” will now flash.
(iii) Turn the right rotary knob until the desired value is displayed.
(iv) Press the right rotary knob to confirm the selected value. If this is not done, the value will be
automatically reset to the last setting.
(v) When the maximum permissible speed of the rotor is reached, the “max” display lights up.
(vi) It is possible to change the speed during the centrifugation run.
(vii) To change the speed in steps of 1 rpm (instead of in steps of 100 rpm): Keep the stop key
pressed while setting the desired speed with the right rotary knob.
5.3.4 Setting the RCF value
(i) To preselect the RCF value turn the left rotary knob until “set” appears in front of the parameter
“rcf” on the display.
(ii) Select the option by pressing or turning the right rotary knob. “Set” will now flash.
(iii) Turn the right rotary knob until the desired value is displayed.
(iv) Press the right rotary knob to confirm the selected value. If this is not done, the value will be
automatically reset to the last setting.
(v) When the maximum permissible RCF value of the rotor is reached, the “max” display lights up.
(vi) It is possible to change the RCF value during the centrifugation run.
(vii) To change the speed in steps of 1 x g (instead of in steps of 10 x g): Keep the stop key pressed
while setting the desired RCF value with the right rotary knob.

CONTROLLED COPY
Centre for Biological Engineering
Standard Operating Procedure SOP122
Title: Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK
Location: CBE Laboratory Unit (H21/22)
Version 003 Effective Date: 26th June 2018 Review 26th June 2020
Written by: Victoria Workman
Reviewed by: P.Hourd
Approved by: R/I.Temple
Page 10 of 22
5.3.5 Setting the temperature
The temperature of the centrifuge is controlled by a refrigeration unit. Temperatures between -10 °C
and + 40 °C can be preselected.
To preselect the temperature in steps of 1 °C:
(i) Turn the left rotary knob until “set” appears in the display field in front of the parameter "Temp".
(ii) Select the option by pressing or turning the right rotary knob. "Set" will now flash.
(iii) Turn the right rotary knob until the desired value is displayed.
(iv) Press the right rotary knob to confirm the selected value. If this is not done, the value will be
automatically reset to the last setting.
The temperature scale shows the actual temperature in the rotor chamber.
• If the selected temperature (= set temperature) is below the actual temperature, the set
temperature will be represented by a flashing bar on the temperature scale.
• If the set temperature is above the actual temperature, it will be represented by one single bar on
the temperature scale.
• If the set temperature and the actual temperature are identical, the line of bars of the scale will be
shown completely.

CONTROLLED COPY
Centre for Biological Engineering
Standard Operating Procedure SOP122
Title: Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK
Location: CBE Laboratory Unit (H21/22)
Version 003 Effective Date: 26th June 2018 Review 26th June 2020
Written by: Victoria Workman
Reviewed by: P.Hourd
Approved by: R/I.Temple
Page 11 of 22
5.3.5.1 Using the Rapid cooling facility
Pre-cooling at a standstill may distort the measurement results and subsequently caused increased
wear of the mechanical components. The rapid cooling protocol pre-cools the centrifuge rapidly under
defined conditions.
(i) Turn the left rotary knob until “run” appears in the display
(ii) Select the option by pressing the right rotary knob. “Run” will now flash
(iii) Turn the right rotary knob until “PrC” is displayed
(iv) Press the right rotary knob or the start key to confirm the input. The display shows 1/3 of the
maximum rotor speed and the corresponding RCF value. The runtime field indicates “cont” and
the symbol “∞” for continuous run
(v) The program will only be loaded if the actual temperature is above the set temperature
(vi) The program runs until the set value is reached. Then, a sound signal is issued and the program
that was set previously will be re-loaded
5.4 Starting and Stopping the Centrifuge
Once the rotor is in place, the main switch turned on and the lid closed the centrifuge can be started.
(i) The centrifuge is ready for operation when the start key is illuminated.
(ii) Press the start key in order to start a centrifugation run.
(iii) Once the pre-selected time has come to an end the centrifuge slows down, comes to a stop,
opens the lid and emits an audible alarm
NOTE: Refer to the operators manual for procedures for emergency (manual) lid release.
5.5 Preventative Maintenance
5.5.1. Rotor Care and Use
(i) If the rotor is not kept clean and chemicals remain on the rotor, corrosion will result. Moisture left
for extended periods of time can also initiate corrosion. It is important that the rotor is left clean
and dry after use. (Wash with mild detergent and warm water using a nylon bottle brush, if
necessary). Dry the rotor thoroughly and store upside down with the cover and tubes removed.

CONTROLLED COPY
Centre for Biological Engineering
Standard Operating Procedure SOP122
Title: Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK
Location: CBE Laboratory Unit (H21/22)
Version 003 Effective Date: 26th June 2018 Review 26th June 2020
Written by: Victoria Workman
Reviewed by: P.Hourd
Approved by: R/I.Temple
Page 12 of 22
(ii) To avoid corrosion, do not expose aluminium rotor components to strong acids or bases, alkaline
lab detergents, or salts (chlorides) of heavy metals (e.g., caesium, lead, silver or mercury).
(iii) Check that the centrifuge chamber, drive spindle, and tapered mounting surface of the rotor are
clean and free of scratches or burrs.
CAUTION: Damaged rotors MUST NOT be used.
(iv) Wipe drive surfaces prior to installing the rotor.
(v) Make sure rotor, tubes, and spindle are dry and that the rotor is properly seated and secured to
the drive hub. Do not operate the centrifuge without the appropriate rotor cover securely fitted
with seals in place.
(vi) If the temperature of the chamber is below room temperature, pre-cool the rotor to the lower
temperature before securing the rotor (this will minimize the chance of it seizing to the tapered
spindle).
(vii) Never exceed the stated maximum speed for any rotor.
(viii)Balance the rotor to within the limits specified (take care that materials of similar densities are in
opposite positions of the rotor).
(ix) A service of the centrifuge is performed every 2 years. The report is filed in the equipment file.
5.5.2. Tube Care
(i) Before use, tubes should be checked for cracks. The inside of cups should be inspected for
rough walls caused by corrosion and adhering matter should be removed. Metal or plastic tubes
(other than nitrocellulose) should be used whenever possible.
(ii) Make sure each tube compartment is clean and corrosion free.
(iii) Tubes must be properly balanced in the rotor.
(iv) Check compatibility of the tube material to the solvent medium (some solvents may cause the
tubes to swell or crack in the rotor).
(v) Never fill centrifuge tubes above the maximum recommended by the manufacturer.
(vi) Use only correctly fitting tubes.

CONTROLLED COPY
Centre for Biological Engineering
Standard Operating Procedure SOP122
Title: Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK
Location: CBE Laboratory Unit (H21/22)
Version 003 Effective Date: 26th June 2018 Review 26th June 2020
Written by: Victoria Workman
Reviewed by: P.Hourd
Approved by: R/I.Temple
Page 13 of 22
5.5.3. Routine Checks of Rotors and Accessories
EVERY MONTH:
(i) This is the responsibility of the housekeeper as part of the housekeeping duties.
(ii) Check all of the safety-relevant parts of the centrifuge i.e. rotor and accessories, for any visible
signs of damage, wear/abrasion, malformation or discolouration (e.g. cracks, corrosion). If
visible damage of the surface, a crack, or corrosion, is detected the part (rotor, etc.) must be
replaced immediately.
NOTE: Chemical reactions as well as stress-corrosion (combination of oscillating pressure and
chemical reaction) can affect or destroy the metals. Hardly detectable cracks on the surface
expand and weaken the material without visible signs.
(iii) Check the concentricity of the motor shaft:
•Visual inspection: Slowly rotate the rotor by hand without the rotor fastening nut. If the motor
shaft does not turn around on a perpendicular axis, the motor and motor shaft must be
replaced.
•Refit the rotor correctly after visual inspection (Refer to Operators Manual - Installation of
Rotors and Accessories).
(iv) Check the unit for atypical running noises.
(v) Check that all screw connections are tight.
(vi) Record the outcome of the operator inspection in the Preventative Maintenance of centrifuges
Record (Section 6). Consult the Responsible Person/Laboratory Manager if any signs of wear
and tear, damage or corrosion are detected
5.5.4. Centrifuge Cleaning
CAUTION: Unsuitable cleaning agents or disinfecting procedures not recommended by the
manufacturer may damage the centrifuge and its accessories.
NOTE: Before any cleaning or decontamination process other than that recommended by the
manufacturer is applied, the user should contact the Laboratory Manager and/or the manufacturer to
ensure that the planned procedure does not damage the equipment.
(i) Any liquid spilt in the centrifuge should be removed immediately (refer to SOP038).
(ii) At the end of each day's use, clean and dry the inside of the centrifuge chamber, all parts of the
rotation assembly and any head accessories used. In order to reduce corrosion, inserts and

CONTROLLED COPY
Centre for Biological Engineering
Standard Operating Procedure SOP122
Title: Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK
Location: CBE Laboratory Unit (H21/22)
Version 003 Effective Date: 26th June 2018 Review 26th June 2020
Written by: Victoria Workman
Reviewed by: P.Hourd
Approved by: R/I.Temple
Page 14 of 22
adaptors should be removed after use to allow adequate cleaning and drying. Ensure that the
manufacturer's recommendations on cleaning materials are followed.
EVERY MONTH:
(i) This is the responsibility of the responsible person ( or any laboratory user if the centrifuge is in
need of cleaning)
(ii) Before cleaning, pull the mains plug
(iii) Wipe the centrifuge casing, venting slots, the rotor chamber, the rotor and the accessories with
mild (pH 6 to 8) detergent (e.g. Fisherbrand or Neutracon).
(iv) During cleaning, make sure that cleaning liquids and especially organic solvents DO NOT come
into contact with the drive shaft and the ball bearing (WARNING: organic solvents may
decompose the lubricant of the motor bearing and the drive shaft may block.)
(v) After cleaning, dry the holders with tissue. Add a spot of grease the size of a small grain of rice to
the load-bearing bolts of the rotor and the buckets
(vi) After completion of the routine clean, record the event the Preventative Maintenance of
centrifuges Record. (Section 6). Consult the Responsible Person/Laboratory Manager if any
signs of wear and tear, damage or corrosion are detected during the cleaning process.
NOTE: This procedure increases life time and reduces corrosion. Corrosion or resultant damages
which are caused by insufficient care do not constitute a warranty claim.
5.5.5. Sterilization and Disinfection of Rotor Chamber and Accessories
Where a spill or leak is detected within the centrifuge, the centrifuge MUST be disinfected immediately.
The procedure will depend upon the risk group of the agent involved (refer to the Risk assessment). If
there is reason to believe that a breakage may have occurred whilst the centrifuge was running, adopt
the procedure in SOP038.
NOTE: The centrifuges and the accessories consist of different materials. Before using detergents or
decontamination agents other than those recommended by the manufacturer is applied, the user
should contact the Laboratory Manager and/or the manufacturer to ensure that the planned procedure
does not damage the equipment.
If autoclaving:
(i) Check the cleaning records for the centrifuge rotor BEFORE autoclaving. The rotor MUST NOT
be autoclaved more than 10 times.

CONTROLLED COPY
Centre for Biological Engineering
Standard Operating Procedure SOP122
Title: Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK
Location: CBE Laboratory Unit (H21/22)
Version 003 Effective Date: 26th June 2018 Review 26th June 2020
Written by: Victoria Workman
Reviewed by: P.Hourd
Approved by: R/I.Temple
Page 15 of 22
(ii) Check Operator’s Manual to make sure that individual materials are resistant to this steam
treatment and for recommended sterilisation cycles.
(iii) Notify the Laboratory Manager and record the action in the Preventative Maintenance of
centrifuges Record, refer to section 6.
After the Autoclave cycle:
(i) Check all of the safety-relevant parts of the centrifuge i.e. rotor and accessories, for any visible
signs of damage, wear/abrasion, malformation or discolouration (e.g. cracks, corrosion). .
CAUTION: The life of the accessories essentially depends on the frequency of autoclaving and
use. Whenever the parts show changes in colour or structure or in the event of leaks etc., the
affected accessories MUST be replaced.
5.5.6. Glass Breakage
In case of glass breakage all glass particles must be carefully removed. Rubber inserts must be
carefully cleaned or replaced. If a problem has occurred, the following must be considered:
•Glass particles in the rubber cushion will cause glass breakage again.
•Glass particles in the centrifuge chamber will cause metal abrasion due to the strong air circulation.
This dust will not only pollute the centrifuge chamber, the rotor and the material to be centrifuged
but also damage the surfaces of the accessories, the rotors and the centrifuge chamber.
To completely remove the glass particles and the metal dust from the rotor chamber:
(i) Grease the upper part of the centrifuge chamber.
(ii) Then rotate the rotor for 5 minutes at a moderate speed. The glass and metal particles
will now collect at the greased part and can easily be removed with a cloth together with
the grease. If necessary repeat this procedure.
5.6 Centrifuge Malfunction/Faults
(i) If any part of the equipment fails or malfunctions, including faults or defects, indicated by
vibration, noise or by failure to operate, the user should contact the Laboratory
Manager/Responsible Person. With permission of the Laboratory Manager or Responsible
Person the user should consult the Operator Instruction Manuals to access fault finding, error
displays and troubleshooting procedures.
CAUTION: DO NOT ATTEMPT TO REPAIR THE CENTRIFUGE YOURSELF

CONTROLLED COPY
Centre for Biological Engineering
Standard Operating Procedure SOP122
Title: Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK
Location: CBE Laboratory Unit (H21/22)
Version 003 Effective Date: 26th June 2018 Review 26th June 2020
Written by: Victoria Workman
Reviewed by: P.Hourd
Approved by: R/I.Temple
Page 16 of 22
(ii) All problems and corrective actions should be recorded in the Preventative Maintenance of
centrifuges Record (Section 6).
(iii) If the equipment fails to work or malfunctions and cannot be rectified according to
troubleshooting procedures detailed in the Operator and Users Manuals the Laboratory
Manager must be informed and the centrifuge must be tagged and locked-out or “Do Not Use”
notice posted on the equipment. Contact the manufacturer for advice and coordinate with the
Lab Manager for external maintenance and servicing.
NOTE: Centrifuges in need of repair should be tagged and locked-out (or “Not in Use” label
applied) while awaiting service.
(iv) External maintenance and servicing of the equipment can only be performed after it has been
suitably disinfected (refer to SOP003 for further details) and a 'Decontamination Certificate' has
been issued (a proforma is available on the CBE website). NOTE: A ‘Declaration of
decontamination’, available in the Operators Manual may also be required.
5.7 Centrifuge or Rotor Decommissioning Checklist
(i) If the centrifuge or rotor needs to be decommissioned, for example due to relocation or change
of use, it is necessary to ensure no hazardous materials are left behind and that the unit has
been decontaminated and made safe for future use or removal. A checklist that can be used to
record that the centrifuge or rotor has been suitably decommissioned is provided in Section 6.
(ii) On completion, the form should be forwarded to DSO or other responsible person to request its
disposal. If multiple items need to be disposed of together from the same room, it may be
possible to use one form – as long as each item is listed and each has a “Safe for Disposal”
sticker / note on it to confirm that appropriate cleaning / disinfecting has been carried out. A
completed decommissioning checklist precludes the need for maintenance staff and contractors
to be issued with a Laboratory Permit to Work.
6. DOCUMENTATION
The following records are outputs of this SOP:
6.1. QS-FORM-003 Preventative Maintenance of centrifuges Record
6.2 QS-FORM-006 Centrifuge decommissioning checklist
6.3. QS-FORM-009 Generic equipment decontamination certificate
These records shall be filed in the Equipment File and stored in the CBE Office or otherwise archived for
future review or retrieval.

CONTROLLED COPY
Centre for Biological Engineering
Standard Operating Procedure SOP122
Title: Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK
Location: CBE Laboratory Unit (H21/22)
Version 003 Effective Date: 26th June 2018 Review 26th June 2020
Written by: Victoria Workman
Reviewed by: P.Hourd
Approved by: R/I.Temple
Page 17 of 22
Annex 1: Safe Use of Centrifuges (Extracts from BS4402)
Moving parts in centrifuges can reach very high kinetic energies. Breakage of a rotor or accessories inside
the centrifuge releases large amounts of energy, and debris can be expelled at high speeds from the
centrifuge; bodily movement of centrifuges for a distance of several metres can also occur in this situation.
Modern centrifuges must be designed and manufactured to British Standard BS 4402:1982, Safety
Requirements for Laboratory Centrifuges. This Standard is intended to ensure that the casing of the
centrifuge will contain the debris of components which break inside it and to ensure that centrifuges are
fitted with necessary interlocks and other safety devices.
It is important that all users of centrifuges are properly trained in the safe use and maintenance of
the equipment.
Appendix B of BS 4402 gives advice on the use of centrifuges and part of it reads:
'RECOMMENDATIONS ON THE USE OF CENTRIFUGES’
B.1 Recommendations applying to all centrifuges
B.1.1. Introduction: When running, a centrifuge has considerable kinetic energy and even a small 'bench
centrifuge' not complying with the requirement of this British Standard can expel debris, or even
move bodily, for a distance of several metres in the event of mechanical failure. While considerable
effort goes into the design and construction of centrifuges to try to minimise the hazards, sensible
use is also important for the safety of operators and others in the laboratory.
B.1.2 Installation: It is essential that all centrifuges be fixed to the working surface, following the
specifications and recommendations of the manufacturer for suitable fixings and fixing details. Care
should be taken to ensure that the working surface and its substrate are strong enough to withstand
the maximum forces of rotation and translation expected to act on the fixings in the event of a
disruption of the rotation assembly (see instruction manual). A centrifuge may move abruptly in the
event of such a disruption, in spite of the fixings, but the distance will be limited if the centrifuge is
correctly fixed.
B.1.3 Training of Operators: Training in the operation of the centrifuge should be carried out by an
experienced user or (by arrangement) by the manufacturer's representative, and should be given to
all operators of the centrifuge before they are permitted to use the centrifuge.
B.1.4 Precautions to be taken when using centrifuges: NOTE: attention is also drawn to
recommendations given in BS 5345 and to the 'User Guide for the Safe Operation of Centrifuges'
published by the Institution of Chemical Engineers.
B.1.4.1. Inspect each sample container and head accessory and all seals (if any) before filling.
Reject any which are damaged or not authorised by the manufacturer, since such
components could cause unsatisfactory sealing, breakage or jamming in the centrifuge head.

CONTROLLED COPY
Centre for Biological Engineering
Standard Operating Procedure SOP122
Title: Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK
Location: CBE Laboratory Unit (H21/22)
Version 003 Effective Date: 26th June 2018 Review 26th June 2020
Written by: Victoria Workman
Reviewed by: P.Hourd
Approved by: R/I.Temple
Page 18 of 22
B.1.4.2 Fill sample containers and head accessories only within the range recommended by the
manufacturer and avoid contamination outside the seal of a sealed head or sealed bucket.
Never fill sample containers when they are mounted in the centrifuge unless contained in a
head specifically designed for this method of filling.
B.1.4.3 Distribute the load symmetrically around the rotation assembly so that it is evenly balanced
and complies with the manufacturer's instructions in this respect. If more than one sample
container is placed in a bucket, use an appropriate insert to ensure their correct positioning.
It is essential that balancing buckets carry an identical load, and in particular are filled with a
liquid of density not grossly dissimilar to that being centrifuged. Always consider all
components (e.g. inserts and trunnion rings) when balancing.
B.1.4.4 Select a centrifuging speed up to the maximum recommended by the manufacturer
according to the head and accessories to be used or to the recommended reduced speed if
the sample has a high density (i.e. above 1200 kg/m3).
B.1.4.5 Follow the manufacturer's instructions for starting and (where the control is not automatic)
for increasing the speed of rotation.
B.1.4.6 Allow the rotation assembly to come to rest before attempting to open the centrifuge lid.
NEVER attempt to slow down a head by hand.
B.1.5 Maintenance
B.1.5.1 any liquid spilt in the centrifuge should be removed immediately.
B.1.5.2 At the end of each day's use of the centrifuge, clean and dry the inside of the centrifuge
chamber, all parts of the rotation assembly and any head accessories used. In order to
reduce corrosion, inserts and adaptors should be removed after use to allow adequate
cleaning and drying. Ensure that the manufacturer's recommendations on cleaning materials
are followed.
B.1.5.3 periodically inspects the clean and dry rotation assembly for any signs of corrosion or
mechanical defect. If such signs are seen, do not use it until advice has been obtained from
the manufacturer or his representative.
B.1.5.4 It is advisable to have the centrifuge heads, buckets and accessories checked by the
manufacturer's representative, at the recommended intervals.
B.1.6 Procedures after disruption of the rotation assembly.
B.1.6.1 Whenever a mechanical failure (other than the simple breakage of sample containers) of any
part of the rotation assembly is suspected to have occurred, the procedure detailed in
B.1.6.2 to B.l.6.5 is the simplest procedure that should be followed. (For hazardous
materials, a more extensive procedure applies.)
B.1.6.2 Immediately isolate the centrifuge from the electrical supply. Do not attempt to open the
centrifuge chamber until at least 30 minutes after all motion has ceased.

CONTROLLED COPY
Centre for Biological Engineering
Standard Operating Procedure SOP122
Title: Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK
Location: CBE Laboratory Unit (H21/22)
Version 003 Effective Date: 26th June 2018 Review 26th June 2020
Written by: Victoria Workman
Reviewed by: P.Hourd
Approved by: R/I.Temple
Page 19 of 22
B.1.6.3 Avoid moving defective and damaged components, unless it is necessary for
decontamination of hazardous materials, and in any event retain all components and debris
for inspection by the manufacturer's representative.
B.1.6.4 Note all the details of the incident and any measures subsequently taken and inform the
relevant authority of the laboratory.
B.1.6.5 Do not use the centrifuge or the rotation assembly again until the necessary repairs and
replacements have been completed and the manufacturer's representative has given an
assurance that its safety is not impaired.’ The Appendix of BS 4402 goes on to deal with
extra precautions required when centrifuging hazardous materials (see below).
Rotors
It is essential that rotors are properly handled, used and maintained.
An ultracentrifuge rotor, for example, experiences 600,000 times the force of gravity. Each gram of it will
effectively weigh over 600kg. Under these conditions a small flaw in part of it may lead to failure and the
virtual explosion of the rotor within the centrifuge.
The centrifugal forces cause rotors to stretch. At low and normal speeds the change of size is elastic and
the rotor returns to its original size when stationary. At high or over speeds the elastic limit of the material
may be exceeded and plastic deformation occurs. If this process recurs micro cracks may be formed and
grow at high stress points within the rotor and subsequently cause catastrophic failure.
Corrosion of the rotor by moisture or chemicals greatly reduces the level of stress at which micro cracks
form and grow. This increases the probability of early failure (by corrosion fatigue or stress-corrosion
cracking). For these reasons it is essential to conform to the manufacturer's conditions on maximum rotor
speeds and maximum running times, and to follow closely their cleaning and maintenance procedures.
Useful advice on the care of rotors is given in the cited references. The publication Centrifuges by L.W.
Price, University of Cambridge, is particularly useful and contains the following list of some causes of rotor
failure.
1. Misplaced trunnions on MSE Minors
2. Use of wrong accessories e.g. with horizontal rotors. In a laboratory where several makes of centrifuge
are in use, it is possible to mix up accessories which are not interchangeable. Keep them separate.
3. Failure to reduce speed above sample specific gravity of 1.2.
4. Powdered glass in rotor.
5. Failure to use speed restriction curves to avoid recyrstallisation with CsC1 gradients.
6. Mis-hooked buckets on swinging bucket rotors.

CONTROLLED COPY
Centre for Biological Engineering
Standard Operating Procedure SOP122
Title: Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK
Location: CBE Laboratory Unit (H21/22)
Version 003 Effective Date: 26th June 2018 Review 26th June 2020
Written by: Victoria Workman
Reviewed by: P.Hourd
Approved by: R/I.Temple
Page 20 of 22
7. Over speeding because de rated over speed disc not fitted.
8. Running badly corroded and weakened rotors or buckets; stress corrosion failure.
9. Loose lid on angle rotor.
10. Incorrect installation of rotor on drive shaft.
Bench Centrifuges
Although the kinetic energy attained by the components of bench centrifuges is lower than that involved in
larger models their improper use can lead to dangerous failures.
Breakages of glass tubes or improper balancing can cause considerable vibration and 'wandering' of an
unfixed centrifuge over the bench top. This may, in extreme cases, allow the mains cable to become
wrapped around the centrifuge body with consequent strain and damage to the cable. The centrifuge may
move completely off the bench-top to the floor. This may cause the chamber lid to open and the spinning
rotor or buckets to be ejected at high speed.
A visual check should always be made to see that all tubes are filled to the same level and are not overfilled,
that glass tubes are not cracked, and that a cushion is installed in each bucket.
Centrifuging Hazardous Materials
Hazardous materials must be centrifuged in the proper type of sealed container. The basic principle of safe
containment is an air-tight, leak proof seal.
Most equipment is unsafe for the centrifuging of flammable materials. A sealed chamber is required and the
necessary design features are laid down in BS 4402 which allows only approved centrifuges to be marked
with 'Sc' (sealed chamber).
Advice on sealed containers for centrifuging hazardous chemicals is given in BS 4402 and in the book by
Price. These also mention biologically hazardous materials and these are dealt with in greater detail in
Laboratory-Acquired Infections, C.H. Collins, Butterworths, Second Edition, 1988.
If pathogenic materials are used it may be necessary to sterilise parts of a centrifuge and its accessories.
Price gives the following relevant advice. Before sterilising always remove and dissemble tube cap
assemblies. Tubes and bottles for autoclaving should be separated from cap parts to avoid damage and
deformation. Cellulose tubes cannot be autoclaved and they may explode if autoclaving is attempted.
Polyallomer tubes distort.
Polycarbonate containers with join seams should not be autoclaved and this method of sterilisation is
possible only for seamless polycarbonate containers. Even then severe distortion will probably render the
container useless. Cold sterilisation is recommended.
Cold methods of sterilisation that can be used on container materials include ultra violet radiation, zephiran
chloride, ethylene oxide gas and 70% ethanol (except cellulose)
Table of contents
Other Sigma Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

MOGlabs
MOGlabs MSA manual

Hanna Instruments
Hanna Instruments Groline HI981413 instruction manual

Bego
Bego Gelovit 200 Translation of the original instructions

NEW BRUNSWICK SCIENTIFIC
NEW BRUNSWICK SCIENTIFIC innova 4900 Guide to operations

novocure
novocure Optune Patient Information and Operation Manual

Applied Biosystems
Applied Biosystems A30207 user guide

dosatron
dosatron DM11F Troubleshooting

Quantifi Photonics
Quantifi Photonics LASER 1100 Series user manual

Thermo Scientific
Thermo Scientific FTIR user guide

IKA
IKA HS 260 Basic operating instructions

CEM
CEM Discover SP-D 10/35 Cleaning and Maintenance Tips

Yamato
Yamato NeoCool Circulator CF301 instruction manual