SJÖBLOMS 010-8251 User manual

Ver 200320
USER MANUAL
ULTRASOUND TABLE
For models
010-8251, 011-8251
NOTE!
The operator must read the user manual
carefully BEFORE the product is used,
to avoid unnecessary risks to the patient and operator.

Ultrasound table Table of contents
TABLE OF CONTENTS
Foreword
Warranty
Important information
1. Function Description
1.1 General description
2. Operation
2.1 Connection
2.2 Height adjustment
2.3 Adjusting the backrest
2.4 Adjusting the Trendelenburg position
2.5 Replacing the paper roll
2.6 Paper dispenser
2.7 The function of the castors
2.8 Cut-out pad section
3. Maintenance & Cleaning
3.1 Routine check-ups
3.2 Cleaning
3.3 Lubrication scheme
3.4 Spare parts, hydraulic
3.5 Spare parts, electric
3.6 Accessories
3.7 Solutions for lubrication, cleaning and disinfecting
4. Technical specification
4.1 Dimensions
4.2 Motors and control units
4.3 Battery backup
4.4 Labels
5. EU-Declaration of Conformity
6. Contact details

Ultrasound table Foreword, important information
FOREWORD
Welcome as the owner of a product manufactured by Sjöbloms Sjukvårdsutrustning
AB. We hope that you will be pleased and satisfied with our quality product.
If a problem should arise with this or any of our products, you are always welcome
to contact us, and our service department will assist you with the expertise needed
to solve the problem.
The user manual contains all the information needed to use, maintain and service
the product. The document is continuously updated as the products are modified.
Therefore we reserve against possible changes in the user manual.
WARRANTY
Sjöbloms provides a two-year warranty for material, design and manufacturing
defects, provided the products are not subjected to tampering or excessive wear.
For the guarantee to apply, the product must have been used in accordance with
the user manual, and the maintenance performed as specified.
IMPORTANT INFORMATION
This product is intended for professional medical use. The user manual should be
read carefully before using the product, to avoid unnecessary risks for the patient
and operator.
The user manual contains warnings where there is a risk of personal injury, which
should be noted. Improper handling can be fatal.
A warning triangle is shown where particularly important information can be found.
Warning triangle

Ultrasound table Function description
1. FUNCTION DESCRIPTION
This product is meant for professional use in health care, for examination of the
patient.
It is designed to be a functional and user-friendly tool for the operator and patient.
Note that the product must only be used by competent health professionals who
read the user manual, for correct handling.
General description
2-sectioned ultrasound table, electrically or hydraulically adjustable, chromium-
plated or coated all-welded steel tube chassis. Backrest and Trendelenburg
adjusted by gas spring. The cushions are stuffed with foamed plastic and
upholstered with imitation leather.
The cut-out section is adjusted manually.
125mm swivel castors, two lockable, one directional locking. Various accessories
such as central locking castors, neck pillow, IV stand etc. are also available.

Ultrasound table Operation
Connection, height adjustment
2. OPERATION
2.1 Connection
The plug should be connected to a grounded 230V outlet. The product is now ready
for use.
2.2 Height adjustment
Electric, hand control
(standard)
Hydraulic
To raise the product
Pump the longer part of
the pedal to raise the
chassis to the desired
height.
To lower the product
Press the shorter part of
the pedal to lower the
chassis to a desired
height.
Electric, foot controll
(optional)
NOTE!
Ensure that no unauthorized persons are
in the couch's working range, as a pinch
hazard exists when raising and lowering
the product.
Follow the symbols and
arrows to regulate the
couch.
Follow the symbols and
arrows to regulate the
couch.

Ultrasound table Operation
Adjusting the backrest
2.3 Adjusting the backrest
Electric, hand control
Gas spring
Raising the backrest
Press at least one of the
handles positioned on the
side of the backrest.
The gas spring helps to
raise the backrest
(variably adjustable).
Lowering the backrest
Press the handle again
while pushing down the
backrest.
Electric, foot control
For exam table with two
motors: The right pair
regulates the backrest.
The left pair regulates the
height of the table.
For exam table with three
motors: The right pair
regulates the backrest
and/or the Trendelenburg
position. Press the middle
switch to change between
the two functions.
The left pair regulates the
height of the table.
NOTE!
Ensure that no unauthorized persons are in the couch's working range, as a
pinch hazard exists when raising and lowering the product.
Follow the symbols and
arrows to regulate the
couch.

Ultrasound table Operation
Adjusting the Trendelenburg
2.4 Adjusting the Trendelenburg position
Electric, hand control
(standard)
Follow the symbols and
arrows to regulate the
couch.
NOTE!
Ensure that no unauthorized persons are in the couch's working range, as a
pinch hazard exists when raising and lowering the product.
Gas spring
To raise the
Trendelenburg position,
stand at the foot-end
side of the table. Press
one of, or both, the
handles.
The gas spring helps
adjust the tables’
Trendelenburg function.
To lower the
Trendelenburg position,
press the handle again
while pressing down the
table top.
Electric, foot control
(optional)
The right pair regulates
the backrest and/or the
Trendelenburg position.
Press the middle switch to
change between the two
functions.
The left pair regulates the
height of the table.

Ultrasound table Operation
Paper-roll holder, dispenser
2.5 Replacing the paper roll (placed at the foot-end side)
2.5A Mounting and de-mounting a paper roll holder
2.6 Paper dispenser
A. Unscrew the black ball.
B. Push the metal rod to the left.
C. Pull the rod gently towards you
until it comes loose on the right
side.
D. Pull the rod gently to the right.
The rod is now loose and you can
put on a new paper roll.
E. Put the rod back on the left side.
F. Put the rod back on the right side,
in the score.
G. Push the rod to the right.
H. Screw the ball tight again.
The pair of attachments is mounted onto
two square tubes underneath the pad on
the foot-end side of the trolley.

2-sectioned exam table Operation
Castors
2.6 The function of the castors
The product is equipped with either central locking castors or separately locking
castors.
Both versions include a directional-locking function, which helps transporting the
chassis straight in long corridors.
Central locking castors
Separately locking castors
Free castors
Directional locking mode
Braked mode
Unbraked mode
Central locking castor
Free mode
When the directional lock is activated,
it is important to have the castor
locked as shown to the right (the
arrow shows a forward transportation
direction).
Wrongly positioned, the exam table
transportation will feel wobbly.
Locked mode
Braked mode
Free castors

Ultrasound table Operation
Cut-out section
2.7 Cut-out pad section
Pull out the pad section.
NOTE!
Don’t leave the section
hanging, risk of pinching.

Fold the pad all the way
around as shown.
Make sure the pin is inserted
in the hole as shown.
Done!

Ultrasound table Maintenance & cleaning
Routine check-ups
3. MAINTENANCE & CLEANING
3.1 Routine check-ups
Routine check-ups should be made to ensure proper functionality.
Important!
Failure to perform the following routine can cause the patient and/or operator to
be exposed to unnecessary risks and the couch may have a reduced life expectancy.
Please check all functions and parts on a regular basis.
Check the following:
Any unusual sounds.
Gas spring functions.
Castors and castor locking mechanism.
On electric tables:
Damages to electric cables.
The motor functions.
That joints and fittings are lubricated according to the lubrication chart.
3.2 Cleaning
Cleaning of imitation leather
During normal cleaning, warm water and a mild soap solution should be used, or
our own “Cleaning solution for artificial leather”, art no. R-1. Stronger cleaning
agents should be avoided whenever possible.
If it nonetheless becomes necessary to use stronger agents, it is important to wipe
with a damp cloth.
Disinfection
If the product needs to be disinfected, use a diluted alcohol solution or our own
“DAX Ytdesinfektion”, art no. C068. Too strong an alcohol solution can have a
drying effect, which means that the imitation leather's properties deteriorate and
cracks can form.

Ultrasound table Maintenance & Cleaning
Lubrication
3.3 Lubrication
Lubrication according to the scheme should be performed every 6 months or as needed.
1. First lock the castors.
2. Enlist the help of another person and tip the couch on the side, in order to
lubricate the moving joints.
The spray can in the pictures above indicates where to lubricate.
HYDRAULIC
ELECTRIC

Ultrasound table Maintenance & Cleaning
Spare parts, hydraulic
3.4 Spare parts, hydraulic
No. Art no Article
1-2 G0001 Gas spring
3 010-8540 Paper roll holder including ball
4 12532142120LPZ Castor, standard
4a 12532142122LPZ Castor, braked
5 12532142123LPZ Castor, directional locked
6 125321630T2LPZ Castor, central-locking
6a 125321630T3LPZ Castor, central-locking, directional-locking
7 HY001 Hydraulic pump
1
1
2

Ultrasound table Maintenance & Cleaning
Spare parts, electric
3.5 Spare parts, electric
4
5
6
7
9
8
No. Art no Article
1-2 G0001 Gas spring
3 010-8540 Paper roll holder including ball
4 12532142120LPZ Castor, standard
4a 12532142122LPZ Castor, braked
5 12532142123LPZ Castor, directional locked
6 125321630T2LPZ Castor, central locked
6a 125321630T3LPZ Castor, central locked, directional locked
7 HB814500001+01 Hand control, one motor, Linak
8 HB824500004+11 Handkontroll två motorer, Linak
9 HB834500003+11 Handkontroll tre motorer, Linak
1
1
2

Ultrasound table Maintenance & Cleaning
Spare parts, electric
12
13
14
15-16
No. Art no Article
12 FS1001 Fotkontroll en motor, Linak
13 FS2001 Fotkontroll två motorer, Linak
14 FS3001 Fotkontroll tre motorer, Linak
15 CB9140AE3+000F0 Styrenhet Linak, nätfrånskiljare
16 314100+1115004E Lyftmotor Linak
17 312100+3115004C Ryggstödsmotor Linak
18 312100+1110004C Chocklägesmotor Linak
17
18

Ultrasound table Maintenance & Cleaning
Acessories
3.6 Accessories
Central locking castors Back support
Not in picture
Battery backup (for electric ultrasound tables only)
Electrically adjustable backrest (for electric ultrasound tables only)
Electrically adjustable Trendelenburg (for electric ultrasound tables only)

Ultrasound table Maintenance & Cleaning
Solutions for lubrication and cleaning
3.7 Solutions for lubrication and cleaning
R-1, Artificial leather cleaning solution
5557001, Lubrication for medical furniture
C068, Disinfection solution
Lubrication Cleaning solution Disinfection solution

Ultrasound table Technical specifications
Dimensions
4. TECHNICAL SPECIFICATION
4.1 Dimensions
Length 200 cm
Length backrest 75 cm
Width 70 cm
Weight 67 kg
Height 53-91 cm
Max patient weight 200 kg
SWL 220 kg
Angle backrest 71o
Angle Trendelenburg -14o

Ultrasound table Technical specifications
Motors, control units
4.2 Motors and control units
The motors comply with EMC directives in accordance with standard EN 60601-1-2.
At full load, the motors can be used a maximum of 6 minutes/hour. The actuator
manufacturer guarantees at least 10 000 cycles performance.
Control units
Max. usage time 10% or 6 min/h at continuous operation. Complies with directives
in accordance with standard EN 60601-1.