Maintenance
If the chek-mate is not going to be used for an extended period, remove the battery from the battery compartment. Deep discharge of alkaline
batteries can result in the eventual leakage of battery electrolyte even from quality batteries and can corrode the battery compartment
terminals and potentially the flowmeter printed circuit board.
Replace the Battery
• SKC recommends using quality alkaline batteries to ensure long battery life and prevent damage due to leaking battery electrolyte. A quality battery
such as the Duracell Industrial model supplied with the chek-mate should provide as many as 30 hours of operation.
• If using a rechargeable battery, it must be 8.4-volt NiMH (see Performance Profile for specifications).
• Failure to follow these guidelines will void the warranty.
When the baery voltage drops below 8 volts as it nears the end of its life, LOW BATTERY will be displayed in the upper left corner of
the LCD. Replace the baery as follows:
1. Unscrew the two captive screws securing the baery compartment cover using a small Phillips head screwdriver and remove the cover.
2. Lift the old baery straight up and out of the baery compartment.
3. Insert the replacement baery into the compartment, ensuring that the terminals are oriented correctly as shown on the label at the
boom of the compartment.
4. Insert the baery compartment cover and secure it with the two captive screws. Do not overtighten the screws.
When the battery voltage drops below 7 volts, the LOW BATTERY message displayed on the LCD will flash on and off and the
chek-mate will switch off automatically. If the chek-mate is turned on again with the battery still in this condition, the LCD will display
bAtt LO ten times and then turn off automatically.
Calibrate the chek-mate Flowmeter
Although SKC recommends a minimum calibration interval of one year, users are responsibile for determining the most suitable
interval to meet their quality assurance system requirements; they should also consider the frequency of use and operating
environment.
Calibrate the chek-mate Flowmeter under vacuum.
Because the chek-mate is designed specifically for use with air sample pumps, which operate under vacuum, it is factory-
calibrated under vacuum, connecting the vacuum flow source alternately to the chek-mate Flowmeter and reference flowmeter
to ensure its accuracy for this application. If a calibration is to be performed other than as just specified and/or using a positive
pressure flow source, contact SKC technical support for advice on calibration methodology.
SKC CAL Service offers calibration and documentation for the chek-mate and other flowmeters. The SKC Calibration Laboratory
will calibrate the chek-mate at flow rates of 750 ml/min and 1, 2, 4, and 5 L/min. NIST-traceable and ISO/IEC 17025:2005 services
are available. Order SKC CAL Service at www.skcinc.com.
Replacement Parts
Battery, 9-volt alkaline P37500