SKC Leland Legacy User manual

Operating Instructions
Leland Legacy
SKC Inc.
863 Valley View Road
Eighty Four, PA 15330
Form #40075 Rev 1001

Leland Legacy Quick Guide
Terms »
Star button Ò
• Scrolls through run time data and Setup options
Up and down arrow buttons ST
• Toggle between display choices and increase or decrease sampling parameters in Setup
Button sequence
TÒ= press buttons individually
[ST] = press simultaneously
ÒSTÒ= security code, always press in sequence
Security code ÒSTÒ
• Prevents unauthorized changes to the pump’s sampling program
Programming Sequences »
• To activate pump (e.g., to change pump from Sleep to Hold):
Press any button.
• To change pump from Hold to Run or Run to Hold:
Press [ST].
• To reset accumulated data:
Press [ST], then ÒSTÒ. Press Òuntil CLr displays then press [ST]; press Òuntil End
displays then press [ST].
• To set pump flow rate:
Press [ST], then ÒSTÒ. Flow rate and SET flash. Press Sor Tto change flow rate. Press Ò
until End appears then press [ST] to save setting and place pump in Hold.
• To calibrate flow rate with standard calibrator:
Press [ST], then ÒSTÒ. Flow rate and SET flash. Press Sor Tto change flow rate. Press Ò
once. ADJ displays. Press Sor Tuntil desired flow rate is indicated on calibrator. When finished,
press Òuntil End displays then press [ST] to save new setting and place pump in Hold. For CalChek
Calibration, see operating instructions.
• To change temperature scale from F to C or C to F:
Press [ST], then ÒSTÒ. Press Òuntil temperature displays. Press Sor Tto switch units;
press Òuntil End displays then press [ST] to save new setting.
• To change atmospheric pressure scale (mm, mb, In):
Press [ST], then ÒSTÒ. Press Òuntil pressure displays then press Sor Tto switch units;
press Òuntil End displays then press [ST] to save new setting.
• To change time scale (12 Hr/24 Hr/Dela):
Press [ST], then ÒSTÒ. Press Òuntil 12 Hr, 24 Hr, or Dela displays then press Sor Tto
switch units; press Òuntil End displays then press [ST] to save new setting. To set delayed start
(Dela), see operating instructions.
• To change clock:
Press [ST], then ÒSTÒ. Press Òuntil clock displays then press Sor Tto change flashing
hour; press Òto move to minutes and Sor Tto change setting. Press Òuntil End displays then
press [ST] to save new setting.
• To change the sampling time function:
Press [ST], then ÒSTÒ. Press Òuntil ST L/min displays then press Sto change flashing digit;
press Òuntil End displays then press [ST] to save new setting. To delete, follow above steps and
press Tuntil 0 appears. Exit Setup.
Note: When in Setup, choosing Esc instead of End will exit Setup without saving new settings.
SKC Inc., 863 Valley View Road, Eighty Four, PA 15330 • www.skcinc.com

Table of Contents
Description............................................................................................... 1
Performance Profile................................................................................. 2
Battery Installation/Charging ................................................................... 4
Installing Battery Pack.................................................................................................. 4
Battery Status............................................................................................................... 5
Low Battery Fault ......................................................................................................... 5
Charging the Battery .................................................................................................... 5
Reading the Charging Status LED on the Single Charger ........................................... 6
Pump Setup............................................................................................. 7
Keypad Basics.............................................................................................................. 7
Turning the Pump On/Off ............................................................................................. 7
Entering and Navigating Setup..................................................................................... 7
Setup Options............................................................................................................... 7
Resetting Run Time Data ............................................................................................. 9
Setting a Sampling Time .............................................................................................. 9
Setting a DataTrac Program......................................................................................... 9
Setting a Delayed Start .............................................................................................. 10
Setting Pump Flow Rate............................................................................................. 11
Calibration ............................................................................................. 12
Verifying Flow Rate Using a Primary Standard Calibrator ......................................... 12
Verifying Flow Rate Using CalChek Automatic Calibration Feature........................... 13
Sampling................................................................................................ 18
Scrolling Through Data............................................................................................... 18
Resetting Run Time Data ........................................................................................... 18
Deleting a DataTrac Program or Delayed Start.......................................................... 19
Deleting a Sampling Time .......................................................................................... 19
Flow Fault................................................................................................................... 19
Battery Replacement............................................................................. 20
Removing and Replacing the Battery Pack................................................................ 20
Programming the Pump Using a PC ..................................................... 21
Optional Accessories............................................................................. 22
Service Policy........................................................................................ 23
Warranty................................................................................................ 24
Indicates a reminder or note
Indicates a warning or caution
Notice: This operating instruction may not address all safety concerns (if any) associated with this prod-
uct and its use. The user is responsible for determining and following the appropriate safety and health
practices and regulatory limitations (if any) before using the product. The information contained in this
document should not be construed as legal advice, opinion, or as a final authority on legal or regulatory
procedures.

1
The Leland Legacy®dual diaphragm sample pump is designed specifically
to provide constant airflows from 5 to 15 L/min with minimum power
requirements and low noise. Diaphragm and valve design minimizes power
requirements and reduces noise. Incoming and outgoing airflow is pulsation
dampened. The lightweight Leland Legacy is housed in a thermoplastic
material for strength and features an overmolding of sorubber that protects
against damage and reduces noise. Powered by a rechargeable Li-Ion baery
pack, the Leland Legacy provides 24-hour run times at 10 L/min and 12 inches
water back pressure. The pump’s patented internal flow sensor measures flow
directly and acts as a secondary standard, constantly maintaining the set flow
rate. Built-in sensors automatically correct flow for variations in temperature
and atmospheric pressure. Advanced programming features are available when
used with a PC and DataTrac®for Leland Legacy Soware. The performance of
the Leland Legacy pump with the Sioutas Impactor has been verified through
EPA-ETV.
Leland Legacy Sample Pump
Description
Inlet Port
with protective filter
Soft Rubber
Overmolding
protects against
damage
Liquid
Crystal
Display
(LCD)
Keypad
with large
operating
buttons
Flashing LED
Run Indicator
Battery
Status
Icon
Li-Ion
Battery
Pack
for 24-hour
run times
(included)
Not shown:
Beltclip (back)
Battery Charging
Jack (top)
Computer Interface
Port (top)

2
Flow
Flow Range: 5 to 15 L/min
Flow Control System: Closed loop with patented* internal flow sensor
Compensation Range: 15 L/min at 5 inches water back pressure
10 L/min at 12 inches water back pressure
5 L/min at 20 inches water back pressure
Typical Back Pressure of Sampling Media (inches water)
Compare the information in this table to pump compensation range to determine appropriate applications.
Accuracy: Flow Rate: ± 5% of set-point after calibration to desired flow
Timing: 1 min/month at 25 C
Atmospheric Pressure: ± 0.3 in Hg
Flow Fault: If the pump is unable to compensate for > 15 seconds due
to excessive back pressure, a flow fault icon displays and
flashes, the pump enters Hold mode, and the pump retains
historical data. Auto-restart is attempted every 20 seconds
up to 10 times. Adjustable with DataTrac for Leland Legacy
Software. See pages 19 and 21.
Operation
Display: LCD displays pump serial number, pump software revision
level, flow rate, volume, temperature, atmospheric
pressure, time of day, run time, and pump status, i.e., Hold
and run as well as Setup information.
Volume Display: Continually updated, based on corrected flow rate
multiplied by sampling time
Time Display: Time of day in hours and minutes (12 or 24-hour clock)
with AM and PM indicators
Timer Display Range: 1 to 99999 minutes (69 days). If the run time exceeds
69 days, the timer display rolls over.
Operating Temp. Range: 32 to 113 F (0 to 45 C)
Protect sample pump from weather when
in use outdoors.
Altitude: Do not use pump beyond 7500 ft.
Operating Humidity: 0 to 95% non-condensing
Typical Run Time†: • Sioutas Impactor (approx. 13 in water bp): 24 hrs at
9 L/min
• PEM with 37-mm, 2.0-μm PTFE filter 225-1709‡:
24 hrs at 10 L/min
• Low-volume PUF Tube: 24 hrs at 5 L/min
• DPS Sampler (PM2.5 or PM10): 24 hrs at 10 L/min
• For extended run times, the pump may be operated
while attached to the approved charger.
Noise Level: 62.5 dBA - pump without case
52 dBA - pump housed in noise-reducing case (optional
accessory Cat. No. 224-89, see page 22)
Measured 3-ft (1-m) distance from pump operating at
10 L/min and 12 inches water back pressure
Performance Profile
Flow Rate (L/min) 5.0 8.0 10.0 12.0 15.0
Filter/Pore Size (μm)
37-mm MCE/0.8 11 18 22 28 36
37-mm PVC/5.0 4 7 9 11 15

3
Performance Profile
User-adjustable Values: Sample run time, calibration, clock display, flow rate,
time of day, delayed start, and temperature and
atmospheric pressure display
Recorded Values: Start date and time, stop date and time, total sample
time, flow rate, sample volume, temperature,
atmospheric pressure, and pump mode transitions
Adjustable Logging Records pump history from 3 seconds (15.4 min of data)
Interval: up to 8 hours (over 102 days of data) depending on setting.
Option available when using DataTrac Software.
Power
Power Supply: • Battery: Removable, rechargeable lithium-ion (Li-Ion)
(see Li-Ion Battery Shipment below),
7.4 V, 12-Ah capacity,
88.8 Wh
• Charger/AC adapter: input voltage 100 to 240 V AC
Battery Recharge Time: 15 hrs
(with SKC-approved chargers;
varies with baery capacity and
level of discharge)
Charging Temp. Range: 32 to 113 F (0 to 45 C)
Storage Temp. Range: -4 to 95 F (-20 to 35 C)
Physical
Size: 8 x 3.9 x 2.6 in (20 x 10 x 7 cm)
Weight: 36 oz (1 kg)
Case: Thermoplastic with soft rubber overmolding
RFI/EMI Shielding CE marked
Approvals: Leland Legacy with Sioutas Impactor performance has
been verified by EPA-ETV.
* U.S. Patent No. 5,892,160
† Results when tested with a new pump and new fully charged battery. Pump performance may vary.
‡ Back pressure on PTFE (Teflon) filters can vary within the same lot. PTFE filter 225-1709 contains a PMP
support ring with a maximum operating temperature of 464 F (240 C).
Cautions:
• Use only SKC-approved parts to ensure reliable performance and to
maintain any warranty.
• Failure to follow warnings and cautions voids any warranty.

4
Ensure that the longest of the three screws is placed in the top
screw hole. Do not overtighten screws.
Cautions:
• Do not charge or operate the pump with or without the charger in
hazardous locations.
• Using a repaired or rebuilt battery pack VOIDS ANY WARRANTY.
• Use only an SKC-approved charger and battery pack designed for the
Leland Legacy Sample Pump to ensure reliable performance.
• Failure to follow warnings and cautions voids any warranty.
Installing the Battery Pack
SKC ships Leland Legacy baery packs separate from the pump.
Once installed, completely charge the baery pack before operating
the pump. It may be necessary to charge the baery a few times
before maximum capacity is achieved.
1. Align connector of baery
pack with connector in pump
body.
2. Gently press baery pack into pump body until it is flush with the
pump case. Insert and tighten three screws.
Battery
connector
Belt clip
Battery
pack
Screws
(top screw
is longer
than
bottom
screws)
Battery Installation/Charging

5
Battery Status
Three bars indicate a full charge (normally appears aer charging),
approximately 75 to 100%.
Two bars indicate that the baery is charged enough to operate the
pump, approximately 25 to 75%.
One bar indicates baery charge is low (charge baery),
approximately 1 to 25%.
Low Battery Fault
No bars and a flashing outline indicate a Low Baery Fault mode
(pump will go into Hold).
Charging the Battery
Shown with single charger (Cat. No. 223-241). A five-station charger is
available; see Optional Accessories on page 22. Follow charger instructions.
For a complete charge, ensure the pump is not running during charging.
• Use of a non-approved charger may damage the battery and pump and
VOIDS ANY WARRANTY.
• Do not charge or operate pump with or without the charger in hazardous
locations.
• Ensure proper orientation of charging cable before plugging it into the
charging jack. Improper orientation/contact will short circuit the battery
and VOID ANY WARRANTY.
• Short circuiting the battery pack will render it immediately inoperative.
• Failure to follow warnings and cautions voids any warranty.
1. Insert the plug on the charging unit into the
baery charging jack on top of the pump
(underneath the protective cover).
2. Insert the plug on the power supply into the
jack on the charging unit.
3. Slide the appropriate wall plug into the
power supply and plug the power supply
into a wall outlet. The baery will recharge
in approximately 15 hours.
Aer charging the baery pack, it is good practice to run the pump for
approximately 5 minutes before calibrating. This ensures the baery is
in more steady-state conditions and improves the agreement in pre and
post-sampling calibrations.
For more information on SKC pump baery packs,
go to www.skcinc.com instructions/1756.pdf.
The baery pack may be kept on the SKC-approved charger for an
indefinite time.
Leland Legacy charging train with
single charger
Battery charging jack
Power
supply
Charging
unit
Jack
Battery Installation/Charging

6
Battery Installation/Charging
Reading the Charging Status LED on the Single Charger
The Li-Ion Charging Unit (Cat. No. P22300) indicates baery charge status
via an LED on the unit that blinks in specific paerns. Observe the LED
steadily for > 5 seconds to read charge status.
LED Action Charge Status
ON
Ò
steady
Charge in progress
ON
Ò
2 sec
OFF
{
.25 sec
ON
Ò
2 sec
(Repeats) Approximately
80% charged
OFF
{
2 sec
ON
Ò
.25 sec
OFF
{
2 sec
(Repeats) Charge completed
The baery pack may be kept on SKC-approved Li-Ion baery
chargers for an indefinite time.
Caution:
• Use only the SKC-approved charger for this pump. Use of an
unapproved charger may damage the battery and pump and
VOID ANY WARRANTY.
• Tampering with the battery pack VOIDS ANY WARRANTY.
• Do not charge in hazardous locations.
• Do not operate the pump with or without the charger in hazardous
locations.
• Do not open, disassemble, short circuit, crush, incinerate, or expose
the battery to fire or temperatures in excess of 212 F (100 C).
• Failure to follow warnings and cautions voids any warranty.
For more information on SKC pump baery packs,
go to www.skcinc.com/instructions/1756.pdf.
Power supply
jack
Charge status
LED

7
Keypad Basics
ÒScrolls through run time data
and Setup options
SIncreases values such as flow
rate
TDecreases values such as flow
rate
[ST] When pressed simultaneously,
displayed item is selected or
entered.
ÒSTÒSecurity code that must be
pressed in sequence to enter
Setup
Turning the Pump On/Off
• Press any buon to turn on the power.
• Press [ST] to run the pump or to place a running pump in Hold.
• Manual Off: from Hold, press and hold Ò.
• Auto Offturns offthe pump aer 5 minutes in Hold.
Entering and Navigating Setup
Entering: Press [ST], then press the security code
ÒSTÒin sequence. Setup should appear
briefly on the LCD.
Navigating: Press Òto scroll through parameters. Once
the LCD shows End, parameters will repeat
until the user exits Setup.
Exiting: Press Òuntil End appears on the LCD. Press
[ST]. The pump is now in Hold.
Setup Options
Aer entering Setup, go to:
1. Flow Set: Press Sor Tto increase or decrease pump
flow rate. Pump will start running. Press Òto move to
next parameter.
2. ADJ: Used during calibration with primary standard
calibrator (not for use with CalChek feature). Press S
or Tto increase or decrease flow adjustment until
desired flow is indicated on calibrator. Press Òuntil End
appears. Press [ST] to save new flow and adjustment
seings and exit Setup.
TDown
arrow
button
Ò
Button
SUp
arrow
button
Pump Setup

8
If changing other parameters, do not press [ST] but continue
pressing Òaer End appears and the remainder of the menu
items will appear. Once all changes are entered, press Òuntil End
appears, then press [ST] to save new seings and exit Setup.
Pressing [ST] when Esc appears will exit Setup without saving
new seings.
3. CALCh: Use for CalChek calibration feature only.
Pressing [ST] initiates single-point calibration.
Pressing Sseven times initiates a full calibration.
See CalChek Calibration instructions on pages 13 to 17.
4. 12 Hr/24 Hr Clock and Delayed Start (factory
default is 12 Hr clock): Press Sor Tto move
between standard (12 hour), military (24 hour), and
Dela (delayed start). Press Òto select. If Dela (delayed
start) is selected, follow instructions on page 10.
5. Time of day: Press Sor Tto increase or decrease
flashing hour. Press Òto move from hours to
minutes. Press Sor Tto increase or decrease
flashing minutes. Press Òto move to next parameter.
6. ST (Sampling Time): Allows the user to program
a specific run time. Press Sor Tto increase or
decrease the time in minutes (up to 99999 minutes).
Press Òto move to next parameter. See pages 9 and 19
for Seing and Deleting a Sampling Time.
7. Temperature (factory default is Celsius): Press S
or Tto toggle between Fahrenheit (F) and Celsius
(C). Press Òto move to next parameter.
8. Atmospheric Pressure (factory default is mm):
Press Sor Tto toggle between inches of mercury
(In), millibars (mb), and millimeters of mercury
(mm). Press Òto move to next parameter.
9. CLr: Press [ST] to reset accumulated run time and
volume data to zero
(see Reseing Run Time Data on page 9).
Pump Setup

9
10. ESC: Press [ST] to exit Setup without saving new
seings.
11. End: Press [ST] to save new seings and exit Setup.
PrOFF: Appears only when a program is loaded into pump memory.
See DataTrac for Leland Legacy Soware Operating Instructions
(Form #40085, included on soware CD) for seing a program.
See page 19 for Deleting a DataTrac Program or Delayed Start.
Resetting Run Time Data
To reset accumulated volume and run time data to zero:
1. Press [ST], then press the security code ÒSTÒin sequence. Setup
will display briefly.
2. Press Òuntil Clr appears, then press [ST].
3. Press Òuntil End appears, then press [ST] to exit Setup. The pump is
now in Hold.
CLr does not clear previously set sampling time (ST). See Deleting a
Sampling Time on page 19.
Setting a Sampling Time (ST)
Program the Leland Legacy from the integral keypad or a PC using DataTrac
soware to sample from 1 to 99999 minutes.
1. Press [ST], then press the security code ÒSTÒin
sequence. Setup will display briefly.
2. Repeatedly press Òuntil ST L/min and a flashing time
and Set appear on the display.
3. Set the sampling time by pressing Sor Tto increase
or decrease it to the desired time in minutes.
4. Press Òrepeatedly until End appears.
5. Press [ST] to save the new sampling time and exit
Setup.
6. Press [ST] to begin sampling. The time display will
count down in minutes and the pump will go to Hold.
The total sampling time will display.
7. To delete a set sampling time, see Deleting a Sampling
Time on page 19.
Setting a DataTrac Program
See DataTrac for Leland Legacy Software Operating Instructions (included on
software CD).
Pump Setup

10
Setting a Delayed Start
A delayed start can be programmed using the pump
keypad or from a PC using DataTrac Soware. The
following instructions are for keypad only. See DataTrac
for Leland Legacy Operating Instructions (included on soware
CD) for programming from a PC.
When seing the pump for sampling from 1 to 99999
minutes to begin within the next 12-hour period, follow this
procedure:
1. Press [ST], then press the security code ÒSTÒin
sequence. Setup will display briefly.
2. Press Òuntil the display reaches the 12 Hr/24 Hr clock.
If delayed start is already programmed, the display will show Dela
(delayed start) in place of 12 Hr. If no delay is programmed, press Sor
Tuntil the display shows a flashing Dela (delayed start).
3. Press Òuntil the time of day (flashing hours) displays. Select the hour
(time of day) that the pump is to begin sampling (within the next 12
hours) by pressing Sor Tuntil the desired hour displays. Press Òand
the minutes will flash. Press Sor Tuntil the desired minutes display.
The time of day entered will be the next occurrence of this time
within the next 12-hour period aer the delayed start is entered.
There is no a.m. or p.m. designation.
4. Press Òuntil the ST displays. Press Sor Tto set the desired run time
in minutes. A delayed start cannot be run unless a sampling time (ST)
is programmed.
5. Press Òuntil End appears.
6. Press [ST] to save seings and exit Setup.
7. Prog and a flashing Hold will appear in the upper
lecorner of the display. The pump is now set for
delayed start.
Once a program is set in the pump, the pump cannot be run
manually. To return to manual pump operation, let the program
run its course or delete the program (see Deleting a DataTrac Program
or Delayed Start on page 19).
Display for
delayed start
Pump Setup

11
Resetting Run Time Data
1. Press [ST], then press the security code ÒSTÒin
sequence. Setup will display briefly.
2. Press Òuntil Clr appears, then press [ST].
3. Press Òuntil End appears, then press [ST] to exit
Setup. The accumulated data is cleared and the pump
is now in Hold.
CLr does not clear previously set sampling time (ST). See Deleting a
Sampling Time (ST) on page 19.
Setting Pump Flow Rate
1. Press [ST], then press the security code ÒSTÒin sequence.
2. The flow rate and Set will flash on the LCD. Press Sto increase flow
rate. Press Tto decrease flow rate. The pump will run while flow is set.
3. Once the desired flow rate is displayed, press Òuntil End appears on
the display. The pump will stop running.
4. Press [ST] to save the new flow rate and exit Setup.
Pump Setup
Flow Rate and Volume Display
hFlow Rate displayed on the pump LCD is the flow to which the pump has been
calibrated. To maintain flow as displayed, the pump automatically adjusts flow
during sampling for changes in temperature and atmospheric pressure that may
differ from the temperature and atmospheric pressure present at the time of
calibration. The flow rate display does not change from the calibrated flow rate.
The pump will fault if it is unable to maintain the calibrated flow rate.
hVolume displayed on the pump LCD is “corrected” in that it is the result of a
continual calculation of corrected flow rate multiplied by sample time.

12
Verifying Flow Rate Using a Primary Standard Calibrator
Before use, allow pump to equilibrate after moving it from one
temperature extreme to another.
1. Ensure the pump has run for 5 minutes before performing calibration.
2. Connect the pump inlet to a calibrator with representative media in line
(see photo below).
3. Press [ST], then press the security code ÒSTÒin sequence. The flow
rate and Set will flash.
4. Set the flow on the pump display by pressing Sor T
to increase or decrease flow to the desired rate.
5. Press Ò. Adj will appear.
6. If the calibrator reads a higher flow rate than the
pump is set for,
press Tuntil they
are in agreement
(within 10 ml).
If the calibrator
reads a lower flow
rate, press Suntil
they agree (within
10 ml). When
pressing Sor T,
the pump display
will indicate the
adjustment (or correction) made in L/min.
7. Press Òuntil End appears.
8. Press [ST] to save new flow rate and Adj and exit Setup. Reset run
time data (see page 11).
If the pump has been programmed with DataTrac Soware and
switched to manual operation, a program may remain in pump
memory. Prog will display in the upper lecorner of the pump
display. See page 19 for Deleting a DataTrac Program or Delayed Start.
Calibration train with sample
medium in line
Calibration

13
Verifying Flow Rate Using the CalChek Automatic
Calibration Feature
The CalChek automatic calibration
feature is available when calibrating
a Leland Legacy with a Defender
calibrator (Cat. No. 717-510H). A
CalChek Communication Cable
(Cat. No. 210-502) is required
for communication between the
pump and the calibrator. Optional
DataTrac for Leland Legacy
Soware can be used to expand
the documentation capabilities of
this feature. The CalChek feature
provides single-point calibration to set and verify flow at a single point before
and aer sampling and multiple-point (full) calibration to calibrate flow to a
primary standard at multiple flow rates within the pump flow range. Both
calibration options bring flow to within ± 5%.
For optimum accuracy, do not perform single-point or multiple-point
calibration until the pump has remained at ambient temperature for
several hours.
Single-point Calibration Using CalChek
The CalChek feature provides correction at a single flow seing and usually
takes less than one minute to complete. Use it to set the desired flow rate
before sampling and to verify flow aer sampling.
Before use, allow pump to equilibrate after moving it from one
temperature extreme to another.
Calibrate with representative sampling media in line.
1. Ensure the pump has run for 5 minutes before starting calibration. Leave
the pump on.
2. Use two pieces of 1/4-inch tubing, one to connect the Defender suction
port to the inlet of the representative sample medium and the other to
connect the medium outlet to the Leland Legacy pump inlet.
3. Select the Defender data port:
a. Press and hold the Defender power buon to turn on the flowmeter.
b. Press the right arrow to highlight Setup; press Enter.
c. Press the right arrow to highlight Preferences; press Enter.
d. Press the down arrow to navigate to Data Port.
e. Press the leor right arrow to toggle to SKC.
f. Press the down arrow to highlight Confirm; press Enter.
Calibration
Single-point calibration train with CalChek

14
4. Enter Defender calibration mode:
a. Press the right arrow and then the down arrow to highlight Measure;
press Enter.
b. Press the right arrow to highlight Cont.; press Enter.
5. Aach the female end of the CalChek Communication Cable to the serial
port (RS-232) on the back of the Defender calibrator.
6. Insert the male end of the CalChek Communication Cable into the data
port on the pump.
7. Press the security code ÒSTÒin sequence on the pump keypad to enter
Setup.
8. Set the pump to the desired flow rate.
9. Press Òon the pump keypad until CALCh appears on
the pump display.
Note: If “no” is flashing on the pump LCD, the pump
has not equilibrated. Wait until the “no” display
disappears before proceeding with calibration.
10. Press [ST] to initiate single-point calibration.
11. The Defender calibrator will begin to automatically
calibrate the pump. Initial flow measurements are
taken without flow from the pump and the pump flow
rate is adjusted automatically. The pump will display
1Cal. During calibration, the pump will briefly display
the flow rates that it is reading from the calibrator.
12. When calibration is completed, the Leland Legacy will
continue to run. If the calibration was successful, the
pump LCD will revert to displaying pump run time as
0.0. If there was failure during the calibration process,
an error code of E4[x] will appear (see CalChek Error
Chart on page 17).
Note: To remove a CalChek error code from the LCD, press Ò.
13. Place the pump in Hold. Disconnect the pump from the representative
sampling medium and the calibrator.
14. Allow the pump to go to sleep.
15. Place a fresh sampling medium in line and sample when ready.
Successful single-point calibration will provide an entry in the
pump history that can be viewed using DataTrac for Leland Legacy
Soware.
Allow pump to go to Sleep mode to write calibration data to pump
history.
Calibration

15
Multiple-point (Full) Calibration Using CalChek
This type of calibration using a Defender calibrator provides flow correction
across the complete operating range of the Leland Legacy (5 to 15 L/min)
in approximately 4 minutes. The operation calibrates each flow rate to a
primary standard. It can also provide a record of calibration for maintenance
and quality purposes if DataTrac for Leland Legacy Soware is used. SKC
recommends that a full calibration be performed during pump maintenance
and aer non-factory repairs.
Do NOT place sampling media in line for full calibration. Ensure the
battery pack is completely charged before starting a full calibration.
Before use, allow pump to equilibrate after moving it from one
temperature extreme to another.
1. Use 1/4-inch tubing to connect the Defender suction port to the Leland
Legacy pump inlet.
2. Follow Steps 1 and 3 through 9 of Single-point CalChek Calibration on
pages 13 and 14.
3. Verify that the baery icon on the pump display shows at least two bars.
If it does not, charge the baery before proceeding.
4. Press Son the pump keypad 7 times to place pump in full calibration
mode.
5. The Defender calibrator will begin to automatically
calibrate the pump. Initial flow measurements are
taken without flow from the pump and the pump
flow rate is adjusted automatically. The pump will
display FCAL, CS1, and a brief flow rate. The pump
will continue to display CS2, then a flow rate, CS3, then
a flow rate, etc. until calibration is completed at all flow
rates between 5 and 15 L/min.
Note: CalChek full calibration can be aborted by pressing [ST]. The pump
will go into Hold.
Calibration
Multiple-point calibration train
with CalChek; do not place
sample medium in train.

16
6. CCAL will display during Calibration Check mode and will count down
to 1. The pump will stop running.
7. When calibration is completed, the Leland Legacy will
go to Hold. If the calibration was successful, the pump
LCD will revert to displaying pump run time as 0.0.
If there was failure during the calibration process, an
error code of E4[x] will appear (see CalChek Error Chart
on page 17).
Note: To remove a CalChek error code from the LCD, press Ò.
8. Allow the pump to go to Sleep mode to write calibration data to pump
memory.
CalChek Full Calibration Data
Requires DataTrac for Leland Legacy Software
Full calibration completely clears pump history, run time parameters, and the
DataTrac Scheduler. Full calibration data can be viewed and printed by going to
the DataTrac Pump Manager window in DataTrac for Leland Legacy Soware
and clicking on the View menu. Choose Calibration Info. This will display
calibration results, pump serial number, and date of the last full calibration. A
buon allows this data to be printed. The printed report contains pump version,
date printed, and a validation code to perform data verification.
CalChek Full Calibration Data Verification
Requires DataTrac for Leland Legacy Software
To ensure that printed calibration data has not been tampered with, pull
down the Tools menu in the Calibration Info window and choose Confirm
Validation Code. Enter the data from the printed report, including the
validation code. DataTrac Soware will indicate whether the information is
completely valid or if a parameter has been changed.
When entering data to confirm the validation number, enter the
date in the following format: mmm, dd, yyyy (e.g., Aug 18 2009).
Calibration

17
Single-point Calibration Errors
Error Problem Troubleshooting
E41 Correction required too large. A Perform a full calibration. If this fails,
gross mismatch between the flow call SKC Technical Support at 724-941-
setting on the pump and the reading 9701 (800-752-8472 [U.S. only]).
generated by the Defender calibrator
has occurred.
E48 Could not get a successful single-point Try the calibration again. If problem
calibration within five flow readings. persists, perform a full calibration.
Multiple-point (Full) Calibration Errors
Error Problem Troubleshooting
E44 First flow reading greater than Check pressure sensor tubing to ensure
5 L/min. The pump is flowing faster that it is not pinched or blocked, or call
than it should, even though the SKC Technical Support at 724-941-9701
calibration routine delivered only a (800-752-8472 [U.S. only]).
very small voltage to the pump.
E45 Pump unable to achieve flow rate of Check pump inlet filter for debris and
15 L/min possibly due to a blocked inlet flow tube for blockage, or call SKC
filter or flow tube or an air leak inside Technical Support at 724-941-9701
the pump. (800-752-8472 [U.S. only]).
E46 Analysis error in the data (rare). Try full calibration again. If problem
or E49 persists, call SKC Technical Support at
724-941-9701 (800-752-8472 [U.S.
only]).
E47 Less than two bars appear in the battery Recharge the battery.
icon on the pump display indicating that
the battery is too low. There must be at
least two bars to begin a full calibration.
—— At conclusion of full calibration, pump Pump not at ambient conditions for at
does not verify to within 5%. least 2 hours. Retry calibration after
pump has been at ambient conditions
for 2 hours.
Pump not running for 5 minutes
prior to calibration. Run pump for
5 minutes and retry calibration.
Errors That Can Occur During Both Calibration Modes
Error Problem Troubleshooting
E42 Unstable average. There is too Try the calibration again. If problem
much variation in the flow readings. persists, contact SKC Technical Support
at 724-941-9701 (800-752-8472 [U.S.
only]).
E43 Serial time out. The calibrator is Check adapter connection. If loose or
not communicating with the pump. disconnected, connect properly.
E4A Calibration has been initiated before Press Ò. Allow pump to run until “no”
pump has equilibrated. disappears from display.
CalChek Error Chart
Other manuals for Leland Legacy
3
Table of contents
Other SKC Measuring Instrument manuals

SKC
SKC chek-mate 375 series User manual

SKC
SKC chek-mate 375 series User manual

SKC
SKC 701-001NBS User manual

SKC
SKC Split 2 User manual

SKC
SKC SoundCHEK User manual

SKC
SKC NoiseCHEK User manual

SKC
SKC 224-6580 User manual

SKC
SKC SoundCHEK 702 Series User manual

SKC
SKC NoiseCHEK User manual

SKC
SKC chek-mate User manual