Alarms
Controller Manual
4 Alarms
The following table explains messages that the electronic controller displays and related
alarms. Alarms signal unexpected operational changes in the cabinet and stop when action is
taken to resolve the problem.
Note that cabinets with twin refrigeration cartridges have two electronic controllers, and an
alarm might be coming from the unseen controller.
Table 2: Controller alarms
Code Alarm Action
P1
Probe alarm.
Press any button on the electronic controller to reset the alarm.
If the alarm persists, arrange a service call.
P2
P3
P4
HA Maximum temperature
alarm
Press any button on the controller to mute the audible alarm.
The alarm will stop once the temperature has reached an acceptable level.
Check the installation meets specification. Ensure there are no blockages in front or
behind the cabinet, and that the cabinet is installed within the operating temperature
range.
If the alarm persists, arrange a service call.
LA Minimum temperature alarm
HA2 Condenser high temperature
alarm
Press any button on the controller to mute the audible alarm.
The alarm will stop once the condenser temperature has reached an acceptable level.
Clean the condenser coil. Check the installation meets specification. Ensure there are no
blockages in front or behind the cabinet, and that the cabinet is installed within the
operating temperature range.
If the alarm persists, arrange a service call.
LA2 Condenser low temperature
alarm
Press any button on the controller to mute the audible alarm.
The alarm will stop once the condenser temperature has reached an acceptable level.
Clean the condenser coil. Check the installation meets the guidelines. Ensure there are
no blockages in front or behind the cabinet, and that the cabinet is installed within the
operating temperature range.
If the alarm persists, arrange a service call.
dA Door open Close the door.
EA External alarm
Press any button on the electronic controller to reset the alarm.
If the alarm persists, arrange a service call.
CA External alarm or pressure
switch alarm
Press any button on the electronic controller to reset the alarm.
If the alarm persists, arrange a service call.
rtC Real time clock alarm Only applicable when the real time clock is enabled.
Press any button on the electronic controller to reset the alarm.
If the alarm persists, arrange a service call.
rtF Real time clock board alarm