When zero clear was performed, a charge
exceeding ±5%F.S of the default zero point was
present.
∗: After approx. 1 s. error display, the sensor
returns to measurement mode.
This time will depend on the deviation of the
product itself and ambient environment during
zero clear operation.
Return to a condition without
charge, and retry zero clear
operation.
Insertion o batteries
Slide the cover of the case open to insert the batteries.
Use 2 x AA (LR6) alkaline dry cell batteries.
When inserting the batteries, pay attention to the polarities and insert them in
the correct direction.
After insertion, ensure the cover is closed correctly.
Assembly
Error Display Error Type Troubleshooting Method
The sensor is broken. Stop using immediately and
contact SMC.
Error in internal data.
The charge has exceeded the lower limit of the
measurable voltage range, or the measuring
distance is not suitable.
Turn off the power supply
once, and turn it on again. If
the error persists contact
SMC.
Remove the charge until the
value is within the
measurable voltage range.
Also, check the measuring
distance.
The charge has exceeded the upper limit of the
measurable voltage range, or the measuring
distance is not suitable.
LCD
Charge amount
Peak/Bottom hold value
LIGHT button
Back light ON/OFF
Battery saving: "L"
Peak hold: "P"
Bottom hold: "b"
Power button
Switch on and off
the power supply
LR6 dry cell battery
(2 pcs.)
Ground terminal
Mounting o the high voltage measuring handle (option)
1. Insert the sensor into the handle in the
direction shown.
2. Fix the cable to the handle as shown.
3. The assembly is complete and the handle is
now ready for use.
Procedure
Setting / Adjustment
1. Insert the batteries.
2. Connect the ground wire.
Connect the ground wire to a point suitable for connection to an external
protective earthing system.
The position for attaching the ground wire to the Handheld meter is shown
below.
Stop using immediately and
contact SMC.
3. Press the button.
4. Move the sensor closer to the position 50 mm away from the measured target.
Gradually bring the sensor closer to the measured target from a distance, and
stop measurement immediately if the displayed value indicates overflow
(HHH) or underflow (LLL).
(The target has a high charged potential that is highly dangerous. The
measured value does not change even if the distance is shortened.)
5. Check the display.
Detecting port
Distance to a measured target
Detecting range
Reference of 50 mm
Protected earth
Ground wire
∗: The detecting range of the sensor is 180 mm (at a detection distance of 50 mm).
Handle
Names o individual parts
•Display
Function settings
Power ON
When the is pressed with the power
supply off, the power supply will be
turned on.
All indications are displayed for 1 s.
after the power supply is turned on.
1 s. after
All indications Instantaneous
value indicatio
Detecting port
Sensor label
Caution label for sensor
•Sensor
Power OFF
When the is pressed for 3 s. or more with the
power supply on, the power supply will be turned off.
Also, no button operation for a certain period will turn
off the power supply.
(For details, refer to Auto power-OFF function.)
Turn off the power supply after using the product to keep the life of batteries
as long as possible.
Auto power-OFF
If no button is pressed for 5 min. or more with the
power supply on, the power supply will turn off
automatically.
3 s. after
6 s.
All indications Instantaneous
value indication
Instantaneous
value indication
Flashing
Auto power-OFF extension
When the is pressed for 6 s. or more with the power supply off, the
continuous operating time with no button operation will extend to 15 min.
(When this auto power-off extension is activated, indications on the display
will keep flashing for 3 s.)
Peak / Bottom hold value
When the is pressed with the
power supply on, the indication
will be changed to instantaneous
value, peak hold value, bottom
hold value and instantaneous
value, in that order.
∗: Release the after "P" or "b" is displayed so as not to turn off the power supply.
•Peak hold value
The maximum charged potential and "P" are displayed.
The maximum charged potential is continuously detected and updated from
when the peak hold is started. If the value over the held maximum charged
potential is detected, the display will change.
•Bottom hold value
The minimum charged potential and "b" are displayed.
The minimum charged potential is continuously detected and updated from
when the bottom hold is started. If the value under the held minimum
charged potential is detected, the display will change.
Zero clear
A displayed value can be adjusted to zero at
measured charged potential in a range of ±5%F.S. of
default potential.
(There will be a slight displacement, depending on
the deviation of the sensor itself and ambient
environment of the sensor when the zero clear is performed.)
When the and are pressed for 6 s. or more with the power supply on,
the displayed value is reset to zero, and then the measuring mode is
recovered automatically.
Once the power supply is turned off, the offset value for zero clear is cleared.
Back Light
When the is pressed while the charged potential
is displayed, the back light will turn on.
Press the button again to turn off the back light.
Zero clear
error
Error ame
Sensor failure
System error
Cable broken
Measurement
error
The cable is broken and the product cannot
measure correctly. The charge amount is
detected, but not reflected in the displayed
value.
Battery low
When the battery voltage becomes low "L" will be
displayed. The display will vary depending on how
much battery is left.
Display resolution change
The display resolution changes depending on the measured charged potential
value.
L
(Flashing)
The batteries are low.
Prepare to replace with new batteries.
Display Content
L
(On continuously)
The batteries are very low.
Replace with new batteries immediately.
3 s. kV
5 min.
6 s.
Instantaneous
value indication
kV kV
POWER
LIGHT
POWER
LIGHT
kV
kV
POWER
LIGHT
POWER
LIGHT
Instantaneous
value indication
Peak hold value Bottom hold value
Handheld Electrostatic
Meter
Operation Manual
IZH10
Thank you for purchasing an SMC IZH10 Series Handheld Electrostatic Meter.
Please read this manual carefully before operating the product and make sure you
understand its capabilities and limitations.
Please keep this manual handy for future reference.
To obtain more detailed in ormation about operating this product, please
re er to the SMC website (URL http://www.smcworld.com) or contact SMC
directly.
Sa ety Instructions
These safety instructions are intended to prevent hazardous situations and/or
equipment damage.
These instructions indicate the level of potential hazard with the labels of
"Caution", "Warning" or "Danger". They are all important notes for safety and must
be followed in addition to International standards (ISO/IEC) and other safety
regulations.
CAUTIO indicates a hazard with a low level of risk
which, if not avoided, could result in minor or
moderate injury.
Operator
This operation manual is intended for those who have knowledge of machinery
using pneumatic equipment, and have sufficient knowledge of assembly,
operation and maintenance of such equipment. Only those persons are
allowed to perform assembly, operation and maintenance.
Read and understand this operation manual carefully before assembling,
operating or providing maintenance to the product.
Caution:
Warning:
Danger:
WAR I G indicates a hazard with a medium level of
risk which, if not avoided, could result in death or
serious injury.
DA GER indicates a hazard with a high level of risk
which, if not avoided, will result in death or serious
injury.
Sa ety Instructions
Do not disassemble, modi y (including change o printed circuit board) or repair.
An injury or failure can result.
Do not operate in an atmosphere containing lammable or explosive gases.
Fire or an explosion can result.
The product is not designed to be explosion proof.
Do not handle with wet hands.
This can cause electric shock.
Provide grounding to ensure sa ety and accurate measurement.
Improper grounding of the ground wire can cause electrical charge to build up at the sensor and ground terminal of
the product and discharge to the user.
Do not contact the sensor with the measured target.
It can cause failure.
Avoid strong impact on the product
Do not drop, hit or apply excessive shock to the product while handling.
It can cause failure.
Do not operate outside o the speci ication.
Fire, malfunction or damage can result.
Please use it after confirming the specification.
Warning
Caution
Summary o Product parts
Troubleshooting
Error Indication
This function is to display error location and content when a problem or an error occurs.
If the error cannot be reset after the above measures are taken, then please
contact SMC.
±1 to ±20 kV
Minimum display unit:
0.1 kV
0 to ±0.99 kV
Minimum display unit:
0.01 kV
<Display example>
Akihabara UDX 15F, 4-14-1, Sotokanda, Chiyoda -k u, Tokyo 101-0021, JAPAN
Phone: +81 3-5207-8249 Fax: +81 3-5298-5362
URL http://www.smcworld.com
ote: Specifications are subject to change without prior notice and any obligation on the part of the manufacturer.
© 2011 SMC Corporation All Rights Reserved
Speci ications
Outline with Dimensions (in mm)
Refer to the product catalogue or SMC website (URL http://www.smcworld.com) for
more information about the product specifications and outline dimensions.
Refer to the SMC website (URL http://www.smcworld.com) for more information
about troubleshooting.

Other manuals for IZH10 Series

6

Other SMC Networks Measuring Instrument manuals

SMC Networks ISA3-G series User manual

SMC Networks

SMC Networks ISA3-G series User manual

SMC Networks PME-500-TR User manual

SMC Networks

SMC Networks PME-500-TR User manual

SMC Networks PPA Series User manual

SMC Networks

SMC Networks PPA Series User manual

SMC Networks IZH10 Series User manual

SMC Networks

SMC Networks IZH10 Series User manual

SMC Networks PPA Series User manual

SMC Networks

SMC Networks PPA Series User manual

SMC Networks IN574-95 Instruction Manual

SMC Networks

SMC Networks IN574-95 Instruction Manual

SMC Networks ZSE10 Series User guide

SMC Networks

SMC Networks ZSE10 Series User guide

SMC Networks ML2B Series User manual

SMC Networks

SMC Networks ML2B Series User manual

SMC Networks PF2A3 Series User manual

SMC Networks

SMC Networks PF2A3 Series User manual

SMC Networks PFMV3 Series User manual

SMC Networks

SMC Networks PFMV3 Series User manual

SMC Networks IZH10 Series User manual

SMC Networks

SMC Networks IZH10 Series User manual

SMC Networks Prime-600 User manual

SMC Networks

SMC Networks Prime-600 User manual

SMC Networks IZH10 Series Manual

SMC Networks

SMC Networks IZH10 Series Manual

SMC Networks PME-500-TR User manual

SMC Networks

SMC Networks PME-500-TR User manual

SMC Networks IZH10 Series User manual

SMC Networks

SMC Networks IZH10 Series User manual

SMC Networks PFG3 0 Series User manual

SMC Networks

SMC Networks PFG3 0 Series User manual

SMC Networks PPA100 Series User manual

SMC Networks

SMC Networks PPA100 Series User manual

SMC Networks PFM3 Series User manual

SMC Networks

SMC Networks PFM3 Series User manual

SMC Networks PFG20 User manual

SMC Networks

SMC Networks PFG20 User manual

SMC Networks PFMV3 Series User manual

SMC Networks

SMC Networks PFMV3 Series User manual

SMC Networks IZE11 Series User manual

SMC Networks

SMC Networks IZE11 Series User manual

SMC Networks PPA100 Series User manual

SMC Networks

SMC Networks PPA100 Series User manual

SMC Networks PFMV301 User manual

SMC Networks

SMC Networks PFMV301 User manual

SMC Networks ZSE30AF series User manual

SMC Networks

SMC Networks ZSE30AF series User manual

Popular Measuring Instrument manuals by other brands

Titanium TRL-30RS Kit instruction manual

Titanium

Titanium TRL-30RS Kit instruction manual

AgSense Crop Link Lite Wiring instructions

AgSense

AgSense Crop Link Lite Wiring instructions

Rion NL-53 instruction manual

Rion

Rion NL-53 instruction manual

Sealey VS0563 instructions

Sealey

Sealey VS0563 instructions

Bamo BAMOWIZ OX 452 user manual

Bamo

Bamo BAMOWIZ OX 452 user manual

V-tech ULE DM251 installation guide

V-tech

V-tech ULE DM251 installation guide

GE PanaFlow HT quick start guide

GE

GE PanaFlow HT quick start guide

ACU-RITE 00325CADI instruction manual

ACU-RITE

ACU-RITE 00325CADI instruction manual

Fast SINGLE SENSOR KIT instructions

Fast

Fast SINGLE SENSOR KIT instructions

NKT Photonics SuperK FIANIUM Product guide

NKT Photonics

NKT Photonics SuperK FIANIUM Product guide

TechVision PS16P2N5 manual

TechVision

TechVision PS16P2N5 manual

FTS Radar Stage Sensor operating manual

FTS

FTS Radar Stage Sensor operating manual

Delta OHM HD 98569 manual

Delta OHM

Delta OHM HD 98569 manual

Felix Storch F-950 Operation manual

Felix Storch

Felix Storch F-950 Operation manual

AXIOMATIC UMAXTC4CO user manual

AXIOMATIC

AXIOMATIC UMAXTC4CO user manual

Endress+Hauser Micropilot S FMR540 quick start guide

Endress+Hauser

Endress+Hauser Micropilot S FMR540 quick start guide

Fluidwell F093-P manual

Fluidwell

Fluidwell F093-P manual

Kobold DON Series operating instructions

Kobold

Kobold DON Series operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.