Sonics VCX750 User manual

Sonics & Materials, Inc.
53 Church Hill Road
Newtown, CT 06470 USA
Phone: 203.270.4600
Rev. 10-21 www.sonics.com
ULTRASONIC PROCESSOR
Part No. VCX750
&
Part No. VCX500
OPERATION MANUAL

Page 2 of 32
TABLE OF CONTENTS
1. WARRANTY ............................................................................................................. 3
2. WARNINGS .............................................................................................................. 4
3. SPECIFICATIONS .................................................................................................... 5
4. PRINCIPLES OF OPERATION ................................................................................ 7
5. DESCRIPTIONS OF COMPONENTS / FUNCTIONS OF CONTROLS .................. 9
5.1. FRONT PANEL ................................................................................................ 9
5.2. REAR PANEL ................................................................................................... 9
5.3. FUNCTIONS OF KEYS CONTROLS & CONNECTORS ............................... 10
5.4. CONVERTER CLAMPING ............................................................................. 12
6. PREPARATION FOR USE ..................................................................................... 13
7. OPERATING INSTRUCTIONS............................................................................... 14
7.1. CAUTION ....................................................................................................... 14
7.2. SETUP ........................................................................................................... 15
7.3. OPERATION .................................................................................................. 16
8. TECHNIQUES FOR OPTIMIZING RESULTS ........................................................ 22
9. MAINTENANCE ..................................................................................................... 24
9.1. PROBE MAINTENANCE ................................................................................ 24
9.2. ATTACHING AND DETACHING PROBES AND MICROTIPS ....................... 25
9.3. SYSTEM CLEANING INSTRUCTIONS .......................................................... 26
10. TROUBLESHOOTING ........................................................................................... 27
10.1. OVERLOAD CONDITION .............................................................................. 27
11. RETURN OF EQUIPMENT .................................................................................... 28
11.1. SAFETY CERTIFICATION FORM .................................................................. 29
12. CONVERTER COOLING ....................................................................................... 30

Page 3 of 32
1. WARRANTY
Your Ultrasonic Processor is warranted and backed by the manufacturer for a period of 3 years
from the date of shipment against defects in material and workmanship under normal use as
described in this instruction manual. During the warranty period, the manufacturer will, at its
option, as the exclusive remedy, either repair or replace without charge for material and labor,
the part(s) which prove to be defective, provided the unit is returned to us properly packed with
all transportation charges prepaid.
Ultrasonic probes are guaranteed against defects for a period of one year from date of
shipment. A defective probe will be replaced once without charge, if failure occurs within the
warranty period. Wear resulting from cavitation erosion is a normal consequence of ultrasonic
processing, and is not covered by this warranty.
The manufacturer neither assumes nor authorizes any person to assume for it any other
obligations or liability in connection with the sale of its products. The manufacturer hereby
disclaims any warranty of either merchantability or fitness for a particular purpose. No person or
company is authorized to change, modify, or amend the terms of this warranty in any manner or
fashion whatsoever. Under no circumstances shall the manufacturer be liable to the purchaser
or any other person for any incidental or consequential damages or loss of goodwill, production,
or profit resulting from any malfunction or failure of its product.
This warranty does not apply to equipment that has been subject to unauthorized repair,
misuse, abuse, negligence or accident. Equipment which shows evidence of having been used
in violation of operating instructions, or which has had the serial number altered or removed, will
be ineligible for service under this warranty.
All probes are manufactured to exacting specifications and are tuned to vibrate at a specific
frequency. Using an out-of-tune probe will cause damage to the equipment and may result in
warranty nullification. The manufacturer assumes no responsibility for probes fabricated by
another party or for consequential damages resulting from their usage.
The aforementioned provisions do not extend the original warranty period of any product that
has either been repaired or replaced by the manufacturer.

Page 4 of 32
2. WARNINGS
Please read the manual in its entirety. Necessary instruction and guidance are provided to help
ensure the successful operation of this device. Observe the following:
High voltage is present in the power supply, converter and high frequency cable. There are no user-serviceable
parts inside any of these devices. Do NOT attempt to remove the power supply cover or converter case.
Do NOT touch any open cable connections on the unit while the power is turned ON.
Do NOT operate power supply with converter disconnected from high voltage cable. High voltage is present in
the cable and may pose a shock hazard.
Do NOT attempt to disconnect the converter high voltage cable while the unit is running.
The power supply must be properly grounded with a 3-prong plug. Test electrical outlet for proper grounding
prior to plugging in unit.
Install the ultrasonic power supply in an area free from excessive dust, dirt, explosive or corrosive fumes and
protected from extremes in temperature and humidity. (See page 5 for specifications.) Do not place the power
supply within a Fume Hood.
Hearing protection is highly recommended. It is recommended that a sound abating enclosure or ear protection
be used when operating the Ultrasonic Processor
NEVER immerse the converter in liquids of any kind, or let condensed moisture or liquid drip into the converter.
NEVER grasp an activated probe or touch the tip of a vibrating probe. It can cause severe burns and tissue
damage.
NEVER allow a horn/probe to vibrate in air.
NEVER hold or clamp the converter by the front driver or by the probe itself. This can cause permanent damage
to the system. Support the converter by only clamping around the converter housing (upper portion).
Do NOT allow the tip of a vibrating horn or probe to touch the counter top or any other hard surface. It could
damage the probe, overload the power supply, or damage the surface.
Avoid touching the bottom or sides of a sample vessel with an activated probe. It may crack or shatter the glass
or melt the plastic. Use glassware that is free from cracks or chips.
Turn OFF the power switch, unplug the power supply and disconnect the power cord from the back of the power
supply before attempting to replace the fuses.
Inspect high frequency cable for cracks in the protective outer jacket.
Do not operate unit with a damaged cable. Doing so may cause serious injury.
In case of AC power loss, wait 3 minutes minimum before reapplying power.
Do not turn off Main power switch while running a probe. Stop ultrasonic processing only by using the
START/STOP key
Symbols
Caution, Risk of electric shock, Hazardous voltage.
Caution, Risk of danger. Refer to User Manual.

Page 5 of 32
3. SPECIFICATIONS
Power Supply
Input Voltage 100 VAC – 132 VAC @ 50/60 Hz 198 VAC – 264 VAC @ 50/60 Hz
Rated Current 10 Amps max. 6.3 Amps max.
Fuse Rating 12 Amps Slow Blow 6.3 Amps Slow Blow
Weight 16 lbs. (6.8 Kg)
Dimensions 8"W x 15.25"L x 8.5"H
203 mm x 387 mm x 216 mm
Output Voltage 1000 VRMS (max.)
Output Frequency 20 kHz
Converter
Weight 2 lbs. (900 g)
Dimensions 7.25" L x 2.5" Dia.
(183 mm x 63.5 mm)
Materials Aluminum Alloy
Standard ½” Probe
Weight 0.75 lbs. (340 g)
Dimensions 5.375" L x .5" Dia.
(136 mm x 13 mm)
Materials Titanium Alloy

Page 6 of 32
Environmental
Pollution Degree
2
Installation Category
II
Operating Limits
Shipping/Storage
Temperature: 41 - 104ºF (5 - 40ºC)
Relative Humidity 10 - 95% (Non Condensing)
Altitude: 6,651 ft. (2000 m)
Temperature: 35 -120 ºF (2 - 49 ºC)
Relative Humidity 10 - 95% (Non Condensing)
Ambient Pressure Extremes: 40,000 ft. (12,192 m)
Restriction of
Hazardous
Substances (ROHS)
Relative humidity Maximum relative humidity 80% for temperatures up to 31ºC decreasing
linearly to 50% relative humidity to 40ºC
Other For indoor use only
The Power Cord supplied with the ultrasonic processor must be used. If the 220V plug is not
configured to match the wall receptacle, a properly grounded universal AC socket adapter must be
added.
Important: Universal adapters do not convert voltage or frequency. Manufacturer is not responsible for
damage caused by the use of an improper power cord or adapter. Transformers are not recommended.
WEEE Statement
This product contains electrical or electronic materials. The presence of these materials may, if
not disposed of properly, have potential adverse effects on the environment and human health.
Presence of this label on the product means it should not be disposed of as unsorted waste and
must be collected separately. As a consumer, you are responsible for ensuring that this product
is disposed of properly. To find out how to properly dispose of this product contact Customer
Service.

Page 7 of 32
4. PRINCIPLES OF OPERATION
The ultrasonic power supply transforms AC line power to a 20 kHz signal that drives a
piezoelectric converter/transducer. This electrical signal is converted by the transducer to a
mechanical vibration due to the characteristics of the internal piezoelectric crystals.
The vibration is amplified and transmitted down the length of the horn/probe where the tip
longitudinally expands and contracts. The distance the tip travels is dependent on the
amplitude selected by the user through the keypad. As you increase the amplitude setting the
ultrasonic processing intensity will increase within your sample.
In liquid, the rapid vibration of the tip causes cavitation - the formation and violent collapse of
microscopic bubbles. The collapse of thousands of cavitation bubbles releases tremendous
energy in the cavitation field. The erosion and shock effect of the collapse of the cavitation
bubble is the primary mechanism of fluid processing.
The probe tip diameter dictates the amount of sample that can be effectively processed. Probes
with smaller tip diameters deliver high intensity ultrasonic processing, but the energy is focused
within a small, concentrated area. Larger tip diameters can process larger volumes, but offer
lower intensity.
The choices of a power supply and horns/probes are matched to the volume, viscosity and other
parameters of the particular application. Horns/Probes are available for both direct and indirect
ultrasonic processing. The frequently asked questions (FAQ) section has more information on
this subject.
Please consult with a product specialist for assistance with selecting a probe for your
application.
Relationship of Amplitude and Wattage
Ultrasonic processing power is measured in watts. Amplitude is a measurement of the
excursion of the tip of the probe. (The probe is also known as a horn.)
Some ultrasonic processors have a wattage display. During operation, the wattage displayed is
the energy required to drive the radiating face of a probe at that specific amplitude setting,
against a specific load, at that particular moment. For example, the unit experiences a higher
load when processing viscous samples than when compared to aqueous samples.
The speed/cruise control on an automobile, can, to a certain extent, be compared to an
ultrasonic processor. The speed/cruise control is designed to ensure that the vehicle maintains
a constant rate of travel. As the terrain elevations change, so do the power requirements. The
cruise control senses these requirements, and automatically adjusts the amount of power
delivered by the engine in order to compensate for these ever changing conditions. The greater
the terrain rate of incline and greater the resistance to the movement of the vehicle, the greater
the amount of power that will be delivered by the engine to overcome that resistance and
maintain a constant speed.
The ultrasonic processor was designed to deliver constant amplitude to your liquid sample,
regardless of these changes in load (much like the vehicle’s cruise control described above).
As a liquid is processed, the load on the probe will vary due to changes in the liquid sample (i.e.

Page 8 of 32
viscosity, concentration, temperature, etc.). As the resistance to the movement of the probe
increases (increased load on the probe), additional power will be delivered by the power supply
to ensure that the excursion at the probe tip remains constant. The displayed wattage readings
will vary as the load changes; however, the amplitude will remain the same.
The resistance to the movement of the probe determines how much power will be delivered to
maintain amplitude. For example, a ½” probe at 100% amplitude will require approximately 5
watts to operate in air. The amplitude of this probe is approximately 120um. Insert the probe in
water and the wattage reading will increase to approximately 90 watts. The wattage required to
operate the probe will increase as the load increases but the amplitude remains the same.
The amplitude control allows the ultrasonic vibrations at the probe tip to be set to any desired
level. Although the degree of cavitation/ultrasonic energy required to process the sample can
readily be determined by visual observation, the amount of power required cannot be
predetermined. A sensing network continuously monitors the output requirements, and
automatically adjusts the power to maintain the amplitude at the preselected level. The greater
the resistance to the movement of the probe due to higher viscosity, deeper immersion of the
probe into the sample, larger probe diameter or higher pressure, the greater the amount of
power that will be delivered to the probe. Setting the amplitude control to its maximum will not
cause the maximum power rating of the unit to be delivered to the sample. The maximum power
that the Processor is capable of delivering will only be delivered when the resistance to the
movement of the probe is high enough to draw maximum wattage.
It is the intensity of cavitation that measures the effectiveness of the ultrasonic processing, not
the total power applied to the system. Intensity is directly related to the amplitude of the
radiating face of the tip or probe. It is amplitude that must be provided, maintained, and
monitored. The unit provides controlled amplitude under varying load conditions in order to give
reproducible results.

Page 9 of 32
Lab Stand
with Clamp
(Sold separately)
5. DESCRIPTIONS OF COMPONENTS / FUNCTIONS OF CONTROLS
5.1. VCX750/VCX500 FRONT PANEL
5.2. VCX750/VCX500 REAR PANEL
Power
Supply
Converter
Cable
½” Probe with
Replaceable
tip (#630-0220)
Converter

Page 10 of 32
5.3. FUNCTIONS OF KEYS CONTROLS AND CONNECTORS
FRONT PANEL
LCD display
Displays prompts and control parameters including:
• Amplitude selected
• Output power delivered to the probe in watts
• Selected duration of processing
• Actual processing time
• Elapsed time
• Set and read temperature
• Pulse on/off duration
• Accumulated amount of energy in Joules delivered to the probe
0 – 9 key Input digits.
ENERGY key Used to set the ENERGY SETPOINT
PAUSE key Suspends operation. Red indicator lights when the processing cycle is interrupted.
TEMP
key
Used with the numeric keys to set the high temperature limit – from 1°C to 99°C.
Red indicator lights when the temperature limit has been reached.
CLEAR key Clears the preceding entry.
ENTER
REVIEW key
Enters data into the program and selects various parameters for display on the
LCD display
TIMER
key
Used with the numeric keys to set the duration of ultrasonic application – from 1
second to 9 hours, 59 minutes, 59 seconds.
PULSER
key
Used with the numeric keys to set the pulse mode. The ON cycle and OFF cycle
can be set independently from 1 second to 59 seconds. Red indicator lights when
pulser is in the OFF portion of the cycle.
START/STOP
key Starts or stops the ultrasonics. In the STOP mode, the red indicator goes off.
I key Switches the main power on.
0 key Switches the main power off.
AMPL Controls the amplitude of vibration at the probe tip.
▲ ▼
key
Used with the AMPL key when the unit is on stand-by to set the amplitude of
vibration at the probe tip. Also used to increase or decrease the amplitude in small
increments while the unit is running. To accomplish this task, depress the
ENTER/REVIEW key twice to display AMPLITUDE CONTROL, then depress the
▲ or ▼ key as required.

Page 11 of 32
REAR PANEL
9 pin D-sub
connector
(IO Port)
Connects to external actuation device, and enable power and frequency
monitoring.
Footswitch
Connector Connects to the footswitch cable.
Temperature Probe
Connector Connects to the optional temperature probe.
Power Supply
Connector Connects to the electrical line cord and encases the fuse(s).
9-PIN D-SUB CONNECTOR
Pin No. Description
1 Not connected
2 Not connected
3 Not connected
4 Enables connection to a frequency counter.
5 Enables connection to an external power monitor (5 mv = 1 watt)
6 Ground
7 Energizes the ultrasonics when connected to ground.
8 and 9 Enables the intensity to be remotely adjusted using an external 10k potentiometer.
NOTE: To vary the
intensity remotely using a
variable DC power supply
(0-5V) instead of a 10 K
potentiometer, connect
positive to pin 8 and
negative to pin 6.

Page 12 of 32
5.4. CONVERTER CLAMPING
Improper clamping can damage the system and void the warranty. Using a sound enclosure
(part #830-00427) or stand (part #830-00459) will ensure a proper fit.
Proper Clamping Improper Clamping
Stand & Clamp 830-00459 Sound Enclosure 830-00427

Page 13 of 32
6. PREPARATION FOR USE
INSPECTION
Prior to installing the Ultrasonic Processor, perform a visual inspection to detect any evidence of
damage which might have occurred during shipment. Before disposing of any packaging material,
check it carefully for small items.
The Ultrasonic Processor was carefully packed and thoroughly inspected before leaving our
factory. The carrier, upon acceptance of the shipment, assumed responsibility for its safe delivery.
Claims for loss or damage sustained in transit must be submitted to the carrier.
If damage has occurred, contact your carrier within 48 hours of the delivery date. DO NOT
OPERATE DAMAGED EQUIPMENT. Retain all packing materials for future shipment.
ELECTRICAL REQUIREMENTS
The Ultrasonic Processor requires a fused, single phase 3-terminal grounding type electrical
outlet capable of supplying 50/60 Hz at 100 volts, 115 volts, 220 volts, or 240 volts, depending
on the voltage option selected. For power requirements, check the label on the back of the unit.
INSTALLING THE ULTRASONIC PROCESSOR
The Ultrasonic Processor should be installed in an area that is free from excessive dust, dirt,
explosive and corrosive fumes, and extremes of temperature and humidity. If processing
flammable liquids use an approved fume hood and do not place the power supply in the fume
hood.
When positioning the unit, be sure to leave adequate space behind the unit so that all
connections can be easily disconnected.
WARNING
For your personal safety, do not, under any circumstances, defeat the
grounding feature of the power cord by removing the grounding prong.

Page 14 of 32
7. OPERATING INSTRUCTIONS
7.1. CAUTION
Do not operate the power supply unless it is connected to the converter.
Never allow liquid to spill into the converter.
Do not allow a microtip to vibrate in air for more than 10 seconds.
Do not allow the vibrating microtip to contact anything but the sample.
Never place a washer between the probe and the converter.
Never apply grease to the mating surfaces or threads of the converter, probe or microtip.
Should it become necessary to remove a probe, use the wrenches supplied. Never attempt
to remove the probe by twisting the converter housing, as this may damage the electrical
connections within the housing.
Overheating will damage the converter. If continuous operation for more than 15 minutes is
required, see Section 12 for converter cooling instructions.
The converter should remain near room temperature at all times. The sample liquid
temperature should not exceed 60º C (140º F). If either of these temperature limits are
reached, shut the system off and allow it to cool.
Replaceable tip probes (1/2” – 1” diameter) are made for use with water based samples
only. If the liquid being processed is an organic solvent or any liquid with a lower surface
tension than water – always use a Solid tip probe. Solid probes can be used with any type of
liquid.
Replaceable tip Solid tip

Page 15 of 32
7.2. SETUP
1. Connect the power cord into the receptacle on the rear of the ultrasonic processor.
2. Make sure the unit is switched off. Plug the electrical line cord into the electrical outlet.
3. If the optional foot switch is used, insert the plug into the jack located on the rear panel.
4. For best results it is critical to use the appropriate size and type of accessory to
process your sample. If you are not sure that you have the proper probe for your sample
volume, please refer to the product brochure or call for assistance.
5. Horns/Probes must be properly tightened. Depending on the accessories purchased,
often the probe and the flat tip are attached to the converter at the factory. Check the
tightness of the probe and flat tip by using the wrench set. Please refer to images in
the Maintenance section of this manual. A loose probe or tip may cause damage to the
power supply circuitry or parts of the converter and probe. A loose probe may also show a
fluctuation in wattage readings. Always use the wrenches supplied with the unit.
6. If you will be using a microtip or extender, remove the flat tip on the end of the replaceable
tip probe, then attach the microtip or extender in its place.
7. Microtips must be used in pulse mode to prevent overheating - which could potentially
crack the tip. See page 16 for Microtip Limits and contact us with questions.
8. Horns and probe tips wear after normal usage. Using a severely worn probe tip can
damage internal power supply components.
9. If using a laboratory stand, mount the converter /probe assembly using a clamp. Be sure to
secure the clamp to the upper section of the converter housing only. Never secure the
clamp to any other portion of the converter/probe assembly. If you are using an acoustic
enclosure mount the converter properly in the converter collar.
10. Connect the converter cable to the power supply and then to the top of the converter. Push
the connectors in and turn the chrome rings clockwise ¼ turn to secure the connectors.
11. If the application requires long processing times, we recommended chilling the sample and
pulsing processing. If processing for over 15 minutes, the converter may get warm and
require cooling with dry, compressed air. See converter cooling instructions in the
Section 12.

Page 16 of 32
Time - : --: --
Pulse
--
--
Ampl
--
%
Time - : --: --
Pulse
--
--
Ampl
40 %
NOTE: To clear an erroneous entry press the CLEAR key.
Microtip Amplitude Setting Limits
When working with a Microtip, do not operate the equipment beyond the maximum amplitude
limits listed below. Ignoring this caution will void the tip warranty.
Size Maximum Amplitude
Tapered Microtip: 1/16” (2mm) 40%
1/8” (3mm) 40%
1/4” (6mm) 75%
7.3. OPERATION
START UP: Press the ON key. The screen will display the power rating of the Ultrasonic
Processor and the following control parameters.
AMPLITUDE: Desired amplitude must be set in order for the Ultrasonic Processor to be
operational. The other control parameters – Time and Pulse, do not have to be set for
continuous operation. AMPL displays the percentage of amplitude that was previously selected.
To set the amplitude at 40%, when the ultrasonics is off, press the AMPL key and the numeric
keys for a 40% reading on the screen, and then press the ENTER/REVIEW key. (Pressing the
AMPL key and the ▲ or ▼ key for a reading of 40% and then pressing the ENTER/REVIEW
key, will also achieve the same result.)
Note: The minimum amplitude setting is 20%.
The screen will display:
WARNING
1. Immerse the probe approximately halfway into the sample. If the probe is immersed to an
insufficient depth, air will be injected into the sample, causing the sample to foam. Also
ensure that the probe tip is not touching the wall of the sample vessel.
2. The Ultrasonic Processor is now ready for continuous operation. To energize the
ultrasonics, press the START key or the footswitch. To de-energize the ultrasonics, press
the STOP key or release the footswitch. If the Time or Pulse functions must be used, refer
to the appropriate paragraphs below.
Note: The START key and footswitch are mutually exclusive. If the process is initiated by
the START key, the footswitch becomes inoperative. If the process is initiated by the
footswitch, the STOP key becomes inoperative.

Page 17 of 32
Time Setting
Hrs: - Min: -- Sec: --
Time Setting
Hrs: 0 Min: 30 Sec: 00
Time 0:30:00
Pulse -- -- Ampl 40 %
Pulse on __ sec
Pulse off __sec
3. To increase or decrease the amplitude in small increments when the ultrasonics is on,
depress the AMPL to display Amplitude Setting on the screen, then depress the ▲ or ▼
key, as required. Since the amplitude required is application dependent and subject to the
volume and composition of the sample, it is recommended that the amplitude be selected
through experimentation, by increasing or decreasing the level of intensity as needed to
properly process the sample to achieve desired results.
4. Be sure to use the fittings provided, for the top of the Converter for air cooling if necessary.
Circulate clean dry compressed air through the Converter to cool the Converter during use.
See manual addendum for more information.
TIMER: In the pulsed mode, the processing time will be different from the elapsed time because
the processing time function monitors and controls only the ON portion of the duty cycle. For
example, for 30 minute processing time, the elapsed time will be 1 hour if the ON and OFF
pulse cycle is set for 10 seconds on and 10 seconds off.
1. To set the processing time, press the TIMER key.
The screen will display:
2. Using the numeric keys, set the processing time as required, for example:
3. Press the ENTER/REVIEW key. The screen will display:
PULSER: Ultrasonics generates heat. Pulsing ultrasonics on and off helps to prevent heat
build-up in temperature sensitive samples. In addition, pulsing may enhance processing by
allowing the material to settle back under the probe after each burst. The ON and OFF pulse
duration can be set independently from 01 second to 59 seconds. During the OFF portion of the
cycle, the red indicator on the PULSE key will illuminate. If the OFF portion of the cycle exceeds
three seconds, a cautionary message - Sonics in OFF Cycle - will warn the operator against
touching the ultrasonic probe.
1. To set the pulser, press PULSE key.
The screen will display:

Page 18 of 32
Pulse on 10 sec
Pulse off __ sec
Pulse on 10 sec
Pulse off 10 sec
Time 0:30:00
Pulse 10 10 Ampl 40 %
2. Using the numeric keys, set the ON portion of the cycle, then press the ENTER/REVIEW
key.
The screen will display:
3. Using the numeric keys set the OFF portion of the cycle, then press the ENTER/REVIEW
key.
The screen will display:
4. Press the ENTER/REVIEW key.
The screen will display:
On Cycle Off Cycle
REVIEW: The REVIEW function provides a “window” on the process by displaying various
operating parameters without process interruption. Pressing the ENTER/REVIEW key
repeatedly during processing will consecutively display the following information.
a) Selected amplitude:
e.g. Amplitude 40%
b) Selected processing time and elapsed processing time:
e.g. Set 0:30:00 Time 0:15:00
c) Selected pulsing cycle and actual pulsing cycle:
e.g. Pulse 10 10 / (actual time)
d) Amount of power in watts and accumulated amount of energy in JOULES delivered
to the probe (Note: The amount of energy displayed will be only for one cycle.
Initiating a new cycle will reset the display to zero.):
e.g. 20 watts / XXX Joules
e) Elapsed time since processing was initiated:
e.g. Elapsed time 0:15:00

Page 19 of 32
TEMPERATURE: The temperature function prevents overheating of the sample by
continuously monitoring the sample temperature and terminating the ultrasonics when the
temperature reaches a predetermined setpoint. The ultrasonics is automatically reinstated when
the temperature drops below the setpoint. If the temperature of the sample must be monitored
and/or controlled, insert the optional temperature probe forcefully info the small jack on the rear
panel and immerse the temperature probe in the sample.
1. Press the TEMP key.
The screen will display:
2. Using the numeric keys set the high temperature limit (setpoint).
The screen will display:
3. Press the ENTER/REVIEW key.
The screen will display:
ENERGY SETPOINT: The energy setpoint continuously monitors the amount of energy in
Joules (watts x seconds) that is being delivered to the probe, and terminates the ultrasonics
when the energy level reaches a predetermined setpoint.
1. To set the energy setpoint, press the ENERGY key.
The screen will display:
2. Using the numeric keys, set the energy setpoint.
The screen will display:
3. Press the ENTER/REVIEW key.
The screen will display:
Probe Temperature 27°C
Temperature Setpoint __°C
Probe Temperature 27°C
Temperature Setpoint 35°C
Time 0:30: 00 Temp 35°C
Pulse 10 10
Ampl 40 %
Energy Setting
_ _, _ _ _, _ _ _Joules
Energy Setting
600000 Joules
Time 0:30: 00 Temp 35°C
Pulse 10 10
Ampl 40 %

Page 20 of 32
SAVE: The save function retains in memory up to 9 (1-9) control parameters under a
storage identification (ID) number.
1. To store the parameters under an ID number, press
the SAVE key.
The screen will display:
2. Press the SAVE key again.
The screen will display:
3. Press the ENTER/REVIEW key.
The screen will display:
4. Using the numeric keys, enter the ID number.
The screen will display:
5. Press the ENTER/REVIEW key.
The screen will display:
*RECALL JOB
SAVE JOB
RECALL JOB
*SAVE JOB
S# T__: __: __ T__ __ °C
P _ _ _ _ E _ , _ _ _ , _ _ _ A _ _ _%
S6 T__: __: __ T__ __ °C
P _ _ _ _ E _ , _ _ _ , _ _ _ A _ _ _%
S6 Time 0:30: 00 Temp 35°C
Pulse 10 10
Ampl 40 %
This manual suits for next models
1
Table of contents
Other Sonics Computer Hardware manuals