
17
8. Trouble Shooting
If difficulties are encountered in the course of operation, and no major instrument functions can
initially be detected, please recheck the symptoms, referring to the tables provided below, before
contacting the service department representative.
1. Optical
Trouble Cause Remedy
Darkness at the periphery or
uneven view field brightness
Revolving nosepiece not in click stop
position (objective not centered in optical
path)
Revolve to click position (switch
the objective correctly into the
optical path)
Filament image not in center Centering
Dirt or dust on the lens (condenser,
objective, eyepiece, collector)
Cleaning
Dirt or dust in the view field Dirt or dust on the lens (condenser,
objective, eyepiece, collector)
Cleaning
Dirt or dust on the slide Cleaning
Condenser position too low Correct position
Poor image quality (low
resolution poor contrast)
No cover glass on the specimen Attach cover glass
Cover glass too thick or thin Use cover glass of specified
thickness (0.17mm)
Slide upside-down Turn over the slide
Immersion oil on dry objective (especially
40X)
Cleaning
Dirt or dust on the lens (condenser,
objective, eyepiece, collector)
Cleaning
No immersion oil used on immersion
objective
Use immersion oil
Air bubbles in immersion oil Remove air bubbles
Unspecified immersion oil used Use specified immersion oil
Condenser aperture and field Close properly
Dirt or dust on the entrance lens Cleaning
Condenser aperture closed too far Close properly
Condenser position too low Raise to the position where the
field diaphragm image is in focus
Condenser not in the center of the view
field or condenser inclined
Reinstall condenser and carefully
adjust with centering screw
Image shift on one side Revolving nosepiece not in click stop
position
Revolve to click-stop position
Floating specimen Fasten securely
Insufficient illumination
brightness
Specimen rise from stage surface Place it stable
Revolving nosepiece not in click stop Revolve to click-stop position
STEINDORFF®