Stiegelmeyer Sicuro pesa User manual

Instruction Manual
Hospital bed for intensive care
Sicuro pesa
Date: 2016-12-02 V4_211952

Instruction Manual for Sicurao pesa Hospital Bed
Page 2
Hospital bed Sicuro pesa: Operational controls
Illustration with Multiflex safety sides and Elano head and footboards
(Illustration shows optional equipment)
1
Headboard, can be folded down to extend the X-ray area, removable
1a
Footboard, removable
2
Grab handles / side mattress guides / sliding rail for backrest and lower leg rest
3
Patient handset (optionally: backlit)
4
Multiflex safety side (on both sides)
5
Grab handles / side mattress guides for lower leg rest
6
Location sleeve for infusion stand (head/foot end, 2 at each end)
7
Linen holder (with control box for extended electric locking and adjusting
functions, removable)
8
Release buttons for extending the bed (pull down on both sides)
9
Foot pedal, on both sides, for operating castor brakes or for steering
10
Sliding rails, on both sides (optionally: at head end in addition)
11
3x Pull levers, on both sides: 1. Quick height adjustment of backrest / 2. Emergency
release for the mattress base (CPR) / 3. Quick height adjustment of lower leg rest
12
Location sleeve at the head end for patient lifting pole/infusion stands with
adapter (concealed, 2x)
Note:
In this instruction manual, numbers which appear in bold type and square brackets [ ]
refer to features described for the Sicuro pesa hospital bed.
11
10
1
10
4
8
6
2
7
9
3
4
5
1a
12

Instruction Manual for Sicuro pesa Hospital Bed
Page 3
Contents
1FOREWORD.................................................................................................................. 5
2GENERAL INFORMATION ............................................................................................ 6
2.1 Definition of the Groups of Persons Involved...................................................... 7
2.2 Safety Information................................................................................................... 8
2.2.1 Explanation of the Safety Symbols Used .................................................. 8
2.2.2 Safety Information for the Operator ......................................................... 9
2.2.3 Safety Information for the User .................................................................. 9
2.3 Product Description .............................................................................................. 12
2.3.1 Intended Use: ............................................................................................. 12
2.3.2 Designated Use .......................................................................................... 12
2.3.3 Contraindications ...................................................................................... 13
2.3.4 Side Effects.................................................................................................. 13
2.3.5 Special Features......................................................................................... 14
2.3.6 Materials Used ............................................................................................ 14
2.3.7 Structural Design ........................................................................................ 15
3PUTTING INTO SERVICE ............................................................................................. 16
3.1 Checklist: Inspection by the User........................................................................ 17
3.2 Location Requirements ........................................................................................ 18
4OPERATION................................................................................................................ 19
4.1 Moving and Braking The Bed .............................................................................. 19
4.2 Electrical Drive System.......................................................................................... 20
4.2.1 Special Safety Information on the Electrical Drive System.................. 20
4.2.2 Specifications for the use of Additional Electromedical Equipment. 22
4.2.3 Handset ....................................................................................................... 23
4.2.4 Control Box.................................................................................................. 25
4.2.4.1 Battery Charge Indicator ...................................................................................28
4.2.4.2 Positioning / Removing the Control Box...........................................................28
4.2.4.3 Enabling/Locking/Actuating Adjustment Functions......................................28
4.2.4.4 Enabling the Foot Pedal (Optional Feature)...................................................29
4.2.4.5 Trendelenburg and Reverse-Trendelenburg Positions ...............................29
4.2.4.6 Lateral Tilting (Optional Feature)......................................................................30
4.2.4.7 Shock Position (Trendelenburg Position) .........................................................31
4.2.4.8 CPR .........................................................................................................................31
4.2.4.9 Restricting the Maximum Mattress Base Height..............................................32
4.2.4.10 Sitting Position .....................................................................................................33
4.2.4.11 Normal Position ....................................................................................................33
4.2.4.12 Overwriting the Factory Default Settings .........................................................33
4.2.5 Battery Operation...................................................................................... 34
4.3 Manual Adjustments ............................................................................................. 35
4.3.1 Safety Information ..................................................................................... 35
4.3.2 Quick Adjustment of the Backrest .......................................................... 36
4.3.3 Quick Adjustment of Lower Leg Rest ..................................................... 37
4.3.4 Linen Holder................................................................................................ 38
4.3.5 Extending the Mattress Base .................................................................... 38
4.4 Attachments .......................................................................................................... 40
4.4.1 Location Sleeves for Patient Lifting Pole / Infusion Stands................... 40
4.4.2 Grab Handle (Triangular Handle)............................................................ 41
1/8
5
4
7
6
3
11
10

Instruction Manual for Sicurao pesa Hospital Bed
Page 4
4.4.3 Resuscitation Support................................................................................ 42
4.4.4 Head and Footboards .............................................................................. 43
4.4.5 X-Ray Cassette Holder (Optional Feature)............................................ 45
4.4.6 Mains Cable Holder .................................................................................. 45
4.4.7 Sliding Rails/Supply Rails............................................................................ 46
4.5 Safety Sides ............................................................................................................ 47
4.5.1 Safety Information on Safety Sides ......................................................... 47
4.5.2 Multiflex Safety Sides ................................................................................. 49
5CLEANING AND DISINFECTION.................................................................................51
5.1 Safety Information on Cleaning and Disinfection ........................................... 51
5.2 Manual Cleaning .................................................................................................. 52
5.3 Machine Cleaning................................................................................................ 53
6MAINTENANCE...........................................................................................................55
6.1 Servicing ................................................................................................................. 55
6.1.1 Service Interval........................................................................................... 55
6.1.2 Servicing Plan ............................................................................................. 56
6.2 Regular Safety Checks on Hospital Beds .......................................................... 58
6.3 Replacement Parts ............................................................................................... 61
6.4 Service Address ..................................................................................................... 61
6.5 Electrical Adjustment System .............................................................................. 62
6.5.1 Safety Information ..................................................................................... 62
6.5.2 Control Unit................................................................................................. 63
6.5.2.1 Terminal Assignments ..........................................................................................63
6.5.2.2 Resetting Control Unit CB16 to Original Settings (RESET)...............................64
6.5.2.3 Initialising Drives (Alignment)..............................................................................65
6.5.2.4 Restore Factory Settings......................................................................................67
6.5.3 General ....................................................................................................... 68
6.5.3.1 Earth Wire Connection (Earth Terminal) ..........................................................68
6.5.3.2 Replace Handset .................................................................................................68
6.5.3.3 Replace Mains Cable ......................................................................................... 68
6.5.3.4 Battery Replacement..........................................................................................69
7TROUBLESHOOTING ...................................................................................................70
7.1 Troubleshooting Guide......................................................................................... 70
7.2 Error Codes Displayed.......................................................................................... 72
8ACCESSORIES.............................................................................................................73
9TECHNICAL DATA.......................................................................................................74
9.1 Dimensions and Weights...................................................................................... 74
9.2 Adjustment Options .............................................................................................. 74
9.3 Electrical Data....................................................................................................... 74
9.4 Ambient Conditions.............................................................................................. 79
9.5 Technical Information on Electromagnetic Compatibility (EMC)................. 79
9.6 Classification.......................................................................................................... 83
10 DISPOSAL INSTRUCTIONS...........................................................................................84
11 EC DECLARATION OF CONFORMITY ........................................................................86

Instruction Manual for Sicuro pesa Hospital Bed
Page 5
1Foreword
Dear Customer,
Stiegelmeyer has built this bed to give you the best possible help with the
challenges posed by nursing and caregiving. We passionately pursue the goal of
developing high-quality and durable products. Our products should make
patients and residents feel as safe and comfortable as possible during their stay in
bed and also maximise the safety and lighten the workload of nursing and care
staff. For this reason, the electrical safety and all functions are tested prior to
delivery. Each bed leaves our factory in perfect condition.
Correct operation and care are necessary to keep the bed in excellent condition
during long-term use. Please therefore read and observe these instructions
carefully. They will help you to put the bed into service for the first time and to use
it on a daily basis. This instruction manual contains all the information you will need
to make controlling and handling this bed as easy and safe as possible, both for
you as the operator and for your users. It is a practical reference book and should
be kept close to hand at all times.
Even after purchasing a bed, Stiegelmeyer is still on hand to help at any time. Our
Assist business division can provide you with customised solutions in all matters
relating to inspection and maintenance, repair and process optimisation. You
can contact our service centre in Germany by phone at +49 (0) 5221 185 - 777.
Customers outside Germany can contact our sales distributors in their particular
country if they have any questions. Contact details can be found on our website
www.stiegelmeyer.com.
We wish you and your users every success and satisfaction in caring for your
patients and residents.
Stiegelmeyer GmbH & Co. KG
Disclaimer
This product is not licensed for use on the North American market. This applies
particularly to the United States of America. The distribution and use of the
hospital bed in these markets, including via third parties, is forbidden by the
manufacturer.

Instruction Manual for Sicurao pesa Hospital Bed
Page 6
2General Information
This instruction manual describes several different models of the Sicuro pesa
hospital bed.
It is possible that certain functions or features are described which are not
incorporated in your particular bed.
In the following text, the Sicuro pesa hospital bed will be referred to simply as the
bed.
Overview of standard features (●) and optional equipment/models (○).
Model
Feature
Safety sides
Multiflex (MSG), fixed to mattress base frame
Electrical equipment
4 motors: electric adjustment of backrest, thigh
rest and mattress base height; tilting to
Trendelenburg and reverse-Trendelenburg
positions;
5 motors: as for 4 motors, but with lateral tilting in
addition
o
Patient handset, 10 functions
Patient handset, 10 functions; with two-stage
backlight
o
CPR activation, electrically but with additional
operating lever on both long sides of the bed
Control box:for extended control of all electric
adjustments by nursing staff
Rechargeable battery for emergency operation,
independent of mains supply
Acoustic alarm that sounds if the brakes are
released and the mains cable is still connected.
o
Under bed light for orientation at night
o
Foot pedal for adjusting the mattress base height
o
Mattress base:
Hygiene mattress base of high-quality plastic
Wire mesh mattress base
o
Holder for X-ray cassettes * under backrest
o
Castors
Double castor Ø 150mm
Double castor Ø 150mm, conductive
o
Miscellaneous
Sliding rails on both long sides under the
mattress base frame
Sliding rails at head end
o
Sliding rails on backrest
o
Wall deflection rollers, including vertically
effective
o
Mechanical cleaning
o
Colours
Argentum
o
Crystal white
* Not available in a model suitable for mechanical cleaning

Instruction Manual for Sicuro pesa Hospital Bed
Page 7
2.1 DEFINITION OF THE GROUPS OF PERSONS INVOLVED
In this instruction manual, the following groups of persons are defined as:
Operator
An operator (e.g. clinic, hospital, hospital administration) is every natural and
legal person with property rights to the Sicura pesa hospital bed. The operator is
responsible for the safe operation of this medical device.
Users
Users (e.g. medical specialists, nursing staff, doctors, carers, care workers) are
persons who, based on their training, experience or briefing, are qualified to
operate the bed on their own authority or to carry out work on it, or have been
instructed how to handle the bed. Furthermore, they can recognise and avoid
potential dangers and assess the clinical condition of the patient.
Patient
A patient is defined as an adult who is in need of care, ill, infirm or disabled, and
occupies this bed.
It is strongly recommended that the operator or user instructs each new patient in
the bed functions that are important for him/her.
Instructions for the Operator:
-This hospital bed fulfils all the requirements of the 93/42/EEC Medical Device
Directive. It is classified as a Class 1 active medical device in accordance with
the Medical Devices Act (Medizinproduktegesetz, German abbreviation: MPG)
§ 13.
-Please observe your obligations as the operator in accordance with the
Medical Devices Operator Ordinance (Medizinprodukte-Betreiberverordnung,
German abbreviation: MPBetreibV), in order to ensure the permanently safe
operation of this medical product with no risk of danger to patients, users or
third parties.
-Any item of technical equipment, electrical or otherwise, can prove hazardous
if used improperly.
Instructions for the User:
-Before using a hospital bed, the user must check that the bed is fully functional
and in perfect working order, and must observe the instructions in the manual
in accordance with the Medical Devices Operator Ordinance (MPBetreibV)
§ 2.
-The same applies for accessories that are attached to the bed.
Before using the hospital bed for the first time:
Remove all transport securing devices and packaging film.
Clean and disinfect the hospital bed prior to first-time use.

Instruction Manual for Sicurao pesa Hospital Bed
Page 8
2.2 SAFETY INFORMATION
At the time of leaving the factory, this hospital bed represents state-of-the-art
technology.
Only use the hospital bed if you are absolutely certain that it is in perfect working
order!
The most important objective of the safety information is to prevent personal
injuries. Property damage is also to be avoided.
2.2.1 Explanation of the Safety Symbols Used
In this instruction manual, the following safety symbols are used:
Risk of injury to persons
Dangerous electrical voltage. There is danger to life.
General danger. There is danger to life and health.
Risk of damage to property
Possible damage to the drive unit, materials or the environment.
Other advice
Useful tip. Makes it easier to handle the hospital bed or improves
understanding.
The safety symbols used are not a substitute for the written safety information.
Therefore read the safety information and follow the instructions precisely!
All persons who work on or with this bed must be familiar with the contents of this
instruction manual and follow all the relevant safety advice.

Instruction Manual for Sicuro pesa Hospital Bed
Page 9
2.2.2 Safety Information for the Operator
In order to ensure the permanently safe operation of this medical product, with
no risk to patients, users or third parties, you must observe your obligations in
accordance with the Medical Devices Operator Ordinance (MPBetreibV)!
Using this instruction manual, which must be provided with the bed, ensure that
every user is instructed in the safe operation of the Sicuro pesa hospital bed
before using it for the first time.
Draw every user’s attention to the possible hazards that can arise if the bed is
improperly used. This applies in particular to the use of electrical drives and
safety sides!
Only allow this bed to be used by persons who have been instructed in its safe
operation!
Make sure that substitute staff are also sufficiently well instructed in the safe
operation of the bed!
Check to ensure that the safety instructions are adhered to!
Ensure that users know where this instruction manual is kept, in accordance
with the Medical Devices Operator Ordinance (MPBetreibV) § 9.
If the hospital bed is in long-term use, test the functions and check for any
visible damage (see Chapter 6.2) after a reasonable period of time.
If any other equipment is attached to the bed, (e.g. compressors for
positioning systems, etc.), ensure that this is securely fastened and is
functioning properly.
Pay particular attention to:
-Chapter 4.2.2 Specifications for the use of Additional Electromedical
Equipment
-Safe routing of all loose connector cables, tubing, etc.
-Ensure that no multiple socket outlets are located under the bed
If you have any queries or concerns, consult the manufacturer of the equipment
or Stiegelmeyer.
2.2.3 Safety Information for the User
Ensure that the operator instructs you in the safe operation of this bed.
Read this instruction manual in order to prevent any injury or damage resulting
from incorrect operation.
Check each time before using the hospital bed to ensure that it is in perfect
working order (see Chapter 3.1)!

Instruction Manual for Sicurao pesa Hospital Bed
Page 10
Ensure that no obstacles such as bedside cabinets, supply rails or chairs could
impede adjustments to the bed.
If other equipment (e.g. compressors for positioning systems, etc.) is attached,
ensure that all items of equipment are fixed securely and function properly.
Pay particular attention to:
-Safe routing of all loose connector cables, tubing, etc.
-If you have any queries or concerns, consult the manufacturer of the
equipment or Stiegelmeyer.
If any damage or malfunction is suspected, take the bed out of service:
-Unplug the bed from the mains supply immediately.
-Indicate clearly that the bed is “Out of Order”
-Report this immediately to the operator responsible.
Route the mains cable, and also all other cables, in such a way
that they cannot be pulled, driven over or damaged by moving
parts when the bed is operated. Hazards due to electric shock
and malfunctions can be avoided in this way.
Before moving the bed, always make sure that you have
unplugged it from the mains supply. Hang the mains cable in the
mains cable holder to ensure that it will not fall out or trail on the
floor.
At regular intervals, carry out a visual inspection of the mains
cable to check for mechanical damage (scuffing, exposed wires,
kinks, pressure points, etc.). Such a check should be performed:
-Whenever the cable has been subjected to any mechanical
load, e.g. has been driven over by the bed itself or by an
equipment trolley;
-Whenever the cable has been bent, stretched or violently pulled,
e.g. due to the bed rolling away while it is still plugged into the wall
socket.
-Whenever the bed has been moved or relocated and before
plugging it into the mains supply;
-Regularly, at least once a week, if the bed is in continuous
operation.
Check the strain relief of the mains cable regularly to ensure that
the screws are tight and secure.
Make sure that the mattress base has travelled to its lowest position
before leaving the patient unattended. In this way, you greatly
reduce the risk of the patient injuring himself/herself as a result of
falling when getting in or out of bed.
Always ensure that the castors brakes are applied before leaving
the patient unattended in the bed.
When not in use, always stow the handset in such a way that it
cannot fall onto the floor (hang it on the hook). Make sure that the

Instruction Manual for Sicuro pesa Hospital Bed
Page 11
cable cannot be damaged by moving parts of the bed.
Do not place multiple socket outlets under the bed. This could
cause electrical hazards due to damaged mains cables or
penetrating fluids.
Before carrying out any adjustments, make sure that there are no
limbs or objects in the way, in order to avoid hazards due to
trapping and/or damage to property. This applies particularly
when mattress base sections are adjusted to a lower height.
Before using patient lifts, or cleaning equipment, pay attention to
the limited clearance under the bed, to prevent damaging the
enclosures surrounding the electric adjustment drives.
To safeguard the patient, and children in particular, against
unintentional electrical adjustments, place the handset out of their
reach (e.g. at the foot end of the bed).
In these cases, adjustments must only be performed by, or in the
presence of, a trained person!
Lock the handset using the control box or locking box if:
-The patient is unable to operate the bed safely or to free
himself/herself from potentially dangerous situations;
-The patient could be at risk from inadvertent adjustment of the
electric drives;
-The safety sides are raised. There is a risk of limbs being crushed or
trapped when adjusting the backrest or thigh rest;
-Children are left unsupervised in a room with the bed;
Observe the safety information found in chapters:
4.2.1 This instruction manual describes all the available options. It is
possible that certain functions or features are described which are
not incorporated in your bed.
Special Safety Information on the Electrical Drive System
4.5.1 Safety Information on Safety Sides
5 Cleaning and Disinfection
6.5.1 Safety Information
8 Special Safety Information for the Use of Accessory Safety sides
-as well as the safety information found in the various chapters of
this instruction manual!

Instruction Manual for Sicurao pesa Hospital Bed
Page 12
2.3 PRODUCT DESCRIPTION
2.3.1 Intended Use:
Essential Performance: This bed is designed for positioning and transporting
patients, as an aid to diagnosing, monitoring, treating, and alleviating illnesses
or compensating for injuries or disabilities. For detailed instructions for use, see
Chapter 9.6.
This bed is suitable for accommodating adult patients only (= persons with a
height of at least 146 cm).
The bed itself is not life sustaining or life supporting.
The bed has no medical indication.
2.3.2 Designated Use
This bed is only for use in hospitals and comparable medical facilities with
trained specialist staff and must only be used inside enclosed spaces.
Specialist staff must be skilled in handling the beds through being thoroughly
conversant with the operating instructions.
If the hospital bed changes owners, the instruction manual must be handed
over with the bed.
The bed with its stable, smooth-running castors is designed for being moved
within the room as well as frequently in corridors and across thresholds with a
height of up to 2 cm.
This bed may be operated without restrictions and with a permanent
maximum load of 260 kg (= safe operating load, consisting of
occupant and accessories) Symbol:
The permitted weight of the patient depends on the total weight of
the accessories attached at the same time (e.g. respirators,
infusions,…) Symbol:
The information applicable for your bed is given on a sticker with the above
symbol which is located on the chassis of the bed.
Example:
Weight of accessories
(with mattress)
Max. permitted weight
of patient
10 kg
250 kg
40 kg
220 kg
This bed is designed for re-use. When re-using the bed, observe the following
requirements:
-Cleaning and Disinfection (see Chapter 5)
-Maintenance / Repeat Inspections (see Chapter 6).

Instruction Manual for Sicuro pesa Hospital Bed
Page 13
This bed may not be used in explosive environments caused, for example, by
cleaning agents or anaesthetics.
This bed may not be used in combination with high frequency surgical
equipment.
Take suitable measures to prevent pressure sores in long-stay patients.
Before using the safety sides, assess and take into consideration the clinical
condition and particular physical build of the patient:
-For example, if the patient is extremely confused or very restless, avoid
using safety sides as much as possible and make use of alternative safety
measures such as posey belts, restraint sheets, etc.
-For especially small, slim patients, additional protective measures for
reducing the distance between the bars on the safety sides may be
necessary. In these cases, use protective covers (accessory), posey belts,
etc.
2.3.3 Contraindications
This bed is only suitable for occupants who do not fall
below the following minimum body size/weight:
oHeight: 146 cm,
oWeight: 40 kg
oBody Mass Index1“BMI”: 17.
Owing to the smaller limbs of occupants with lower body
measurements/weight, there is an increased risk of entrapment between the
open spaces of the safety sides when safety side systems are used.
2.3.4 Side Effects
Take suitable measures to prevent pressure sores in long-stay patients.
The Sicuro pesa hospital bed may only be used under the operating conditions
described in this instruction manual.
Its use for any other type of application is deemed to be contrary to the intended
purpose!
1 Calculation of BMI =
; example: a)
OK!; b)
Not OK
Sticker on
bed chassis

Instruction Manual for Sicurao pesa Hospital Bed
Page 14
2.3.5 Special Features
-Fully motorised version (optionally with ± 25° lateral tilting)
-Integrated Multiflex safety sides (up to a protection height of 428 mm)
-Safe working load: 260 kg
-Integrated linen holder
-Control box for medical personnel in linen holder
-Bed extension, at the foot end
-Trendelenburg position up to 15° / reverse-Trendelenburg position up to 18°
-Castors (diameter: 15 cm), with central locking; steering lock for moving in a
straight line
-Mattress base in four sections, mattress base area 200 x 87 cm; external
dimensions approx. 99 x 215 cm
-Rail for universal holders, on safety sides on both sides
-Mattress base height adjustment range from approx. 45 to 80 cm
-Adjustment of the backrest from 0° to approx. 70°
-Adjustment of the thigh rest from 0° to approx. 80°
2.3.6 Materials Used
-The bed is made predominantly of steel sections. Their surface has a polyester
powder coating or a metallic coating made of zinc or chrome.
-The mattress base is made of high-quality, PVC-free plastics (PP).
-The chassis is made of steel sections / cast metal parts.
-The headboards and footboards consist of moulded laminate boards or high-
quality PVC-free PE plastics.
-The bed does not contain any mutagenic substances or phthalates.
-All surfaces within reach of patients are recognised as being safe for contact
with skin.

Instruction Manual for Sicuro pesa Hospital Bed
Page 15
2.3.7 Structural Design
Mattress base
The mattress base is divided into a backrest, a seat section, a thigh rest and a
lower leg rest. The mattress base height can be adjusted horizontally, and can be
tilted into a Trendelenburg or anti-Trendelenburg position. Optionally, the mattress
base can be tilted sideways along its longitudinal axis (lateral tilting).
All adjusting functions are electrically powered. The components for the electric
drive system are mounted in the mattress base.
Chassis
Under the mattress base is a crank chassis that has an electric adjustment facility
for adjusting the height and inclination. The bed has four castors that can be
centrally locked. One of the castors is equipped with a steering lock which
enables the bed to be moved in a straight line.
Safety sides
The bed is equipped with an integrated "Multiflex" safety side system to protect
the occupant from accidentally falling out of bed.
Electric drive system
The electric drive system for this bed consists of:
-A central control unit, installed underneath the mattress base, with a
transformer that generates a 24 volt protective low voltage, which is harmless
for patients and users. The drives, the handset, and the control box are
connected to the control unit. These components operate on the 24 volt
protective low voltage generated and have dust and water protected plug
connections.
-Rechargeable batteries for emergency operation, independent of the mains
supply;
-A handset for use by the patient to adjust the bed;
-Two drives for mattress base height adjustments and tilting to a Trendelenburg
and reverse-Trendelenburg position
-A drive for the backrest
-A drive for the thigh rest
-An electric motor for lateral tilting of the mattress base (optional feature).
-An integrated control box
Nursing staff can use this to disable adjusting functions individually, if necessary,
and also carry out a variety of special functions and emergency adjustments.

Instruction Manual for Sicurao pesa Hospital Bed
Page 16
3Putting into Service
No electrical measurements are necessary prior to putting this bed into service for
the first time, since the bed was tested by the manufacturer for electrical safety
and functionality and left our factory in perfect condition.
Before putting the bed into service for the first time:
Remove all transport securing devices and packaging film.
Clean and disinfect the hospital bed (see Chapter 5).
Use only authorised accessories (e.g. mattresses, external safety sides), to
minimize endangering patients through becoming trapped or falling out of
bed. For further details see Chapters 4.5.1; 8.
Check that
-The bed is connected to a properly earthed mains socket only;
-The mains voltage from the socket is the same as the rating shown on the
electrical type-plate on the bed. (The type plate is located under the
backrest beside the mains cable entry);
-The mains cable is connected and routed in such a way that it cannot be
damaged and can easily be reached at any time to unplug it from the
mains socket;
-The mains cable, drive cables and handset cable cannot be damaged by
moving parts of the bed;
-If the bed is equipped with rechargeable batteries, it is connected to the
mains supply for at least 24 hours. This is the only way to ensure that the
batteries are fully charged and are available in case of an emergency (see
Chapter 4.2.5).
Each time before putting the bed into service, the user must check that:
-The bed has been cleaned and disinfected;
-The castor brakes are applied (see Chapter 4.1);
-No obstacles such as bedside cabinets, supply rails or chairs will impede bed
adjustments;
-All adjustment facilities work properly and have been checked (see Chapter
3.1).
The bed may be put into operation only after carrying out these checks!

Instruction Manual for Sicuro pesa Hospital Bed
Page 17
3.1 CHECKLIST:INSPECTION BY THE USER
Inspection
Ok
Not
Ok
Description of
Defect
Visual Inspection of the Electrical Components (if installed)
Handset
Damage, foil
Handset cable
Damage, routing
Mains cable
Damage, routing
Control box
Damage, foil
Visual Inspection of the Mechanical Components
Patient lifting pole,
location sleeves
Damage, deformation
Grab handle with strap
Damage
Chassis
Damage,
deformations
Mattress base, covers
Damage, deformation
Performance Check of the Electrical Components (if installed)
Handset
Functional test
Foot pedal
Functional test
Control box
Function test, locking
functions
Performance Check of the Mechanical Components (if installed)
Bed extension
Functional test
Castors
Locking, moving
Emergency lowering of
backrest
Functional test
according to
instruction manual
Safety sides
Functional test,
securely engaged
Lower leg rest, lower leg
rest extension
Functional test
Accessories (e.g.
patient lifting pole, grab
handle with strap,
safety sides)
Fixings, damage,
suitability
Inspector’s signature:
Inspection results:
Date:

Instruction Manual for Sicurao pesa Hospital Bed
Page 18
3.2 LOCATION REQUIREMENTS
There must be sufficient room available to accommodate the bed’s entire
range of adjustments. Obstacles such as bedside cabinets, supply rails,
furniture, or windowsills must not impede adjustments.
Before using the bed on parquet flooring, check whether the castors can
leave marks on the parquet varnish. We accept no liability for such wear. The
bed can be used on tiles, carpet, linoleum or laminate flooring without causing
any damage.
A properly installed and earthed mains socket must be available close to the
bed, at the head end.
The supply voltage of this mains socket must comply with the information on
the electrical rating plate on the bed (located at the head end, facing
inwards towards the bed).
If other equipment (e.g. compressors for positioning systems, etc.) is attached,
ensure that all items of equipment are fixed securely and function properly.
Pay special attention here to the safe routing of all loose connector cables,
tubing, etc. If you have any queries or concerns, consult the manufacturer of
the equipment or Stiegelmeyer.
Minimise, as far as possible, the risk of fire due to external influences.
Instruct users on these points:
Use only flame-retardant mattresses and bedding if possible.
Avoid smoking in bed, since the mattress and bedding used may
not be resistant to smokers' accessories.
Only use additional equipment (e.g. electric blankets) and other
electrical devices (e.g. lamps, radios) that are in perfect working
order!
-Ensure that this equipment is used only for the purpose intended.
-Ensure that this equipment is not unintentionally placed on or
under the bedding (danger of overheating)!
Avoid using extension cables or multiple socket bars under the bed
(risk of fire due to penetrating fluids).
Extension cables and/or multiple socket outlets should not be used
at all.

Instruction Manual for Sicuro pesa Hospital Bed
Page 19
4Operation
4.1 MOVING AND BRAKING THE BED
The bed is equipped with four lockable castors and central locking at the foot
end. A steering lock can be activated for one castor at the foot end of the bed,
which makes it easier to move the bed in a straight line and to manoeuvre the
bed.
As a general rule, always apply the brakes when the bed is not being
moved or when a patient is left unattended in the bed.
Each time before moving the bed, ensure that:
-The mains cable cannot be stretched, driven over or damaged in any
other way.
-The mains cable has been hung in the mains cable holder (see Chapter
4.4.6).
-The mains cable does not trail on the floor.
-The cables, tubes or wires of any attached equipment are adequately
secured and cannot be damaged.
Otherwise, the mains cable could sustain damage as a result of being
torn off, crushed or driven over. Such damage could lead to electrical
hazards and malfunctions.
When the bed is moved, ensure that objects hanging in the sliding rails
[10] do not cause any damage (e.g. to door frames).
Risk of crushing! When transporting a patient, ensure that the patient
hands and feet do not protrude over the edge of the bed and that they
cannot get in the way of the wall deflection rollers.
Braking
Press down on a foot lever [9].
Check to ensure that the bed is
properly braked.
Moving without the steering lock
Foot lever [9] in horizontal middle position.
Moving with the steering lock
Raise a foot lever [9] with the back
of the foot.
If you push the bed from the foot end now,
it will be easier to steer.
Brake alarm (optional feature)
A continuous acoustic signal sounds (for up to 5 minutes) if the castor brake is
released and the mains cable is plugged in.
Advantage: Prevents damage to mains cables and sockets
Prevents endangering patients due to the bed rolling away.
Moving without
the steering lock
Similar to
illustration
Braking
Head section
Moving with the
steering lock
Direction of travel

Instruction Manual for Sicurao pesa Hospital Bed
Page 20
4.2 ELECTRICAL DRIVE SYSTEM
This instruction manual describes all the available options. It is possible
that certain functions or features are described which are not
incorporated in your bed.
4.2.1 Special Safety Information on the Electrical Drive System
This hospital bed may not be used in combination with high-
frequency surgical equipment or in explosive environments!
When using accessories:
Make sure that the attachment of accessories does not produce
any crush or shearing zones when the bed sections are adjusted. If
this cannot be ensured, the user must lock the handset using the
control box or the locking box!
When making any adjustments, always ensure that there are no
limbs belonging to patients, users, other persons, and especially
playing children, that could be trapped underneath the rests or
the mattress base.
When this bed is connected to the mains supply:
oIt must be connected to the hospital’s potential equalisation
system for intravascular and intracardial applications.
oIt must only be used in medical rooms which meet the electrical
requirements of the VDE 0100 Part 710 standard.
Please also refer to the Chapter: Specifications for the use of
Additional Electromedical Equipment, on page 61.
Ensure that no obstacles such as bedside cabinets, supply rails or
chairs could impede adjustments to the bed.
Ensure that there are no objects on the chassis or under the bed.
Ensure that no collisions can occur if equipment such as patient
lifts or cleaning equipment is used under the chassis of the bed.
Otherwise, the bed and/or lifting frame could be damaged.
When routing the handset cable, ensure that it cannot be
damaged by any moving parts of the bed.
Ensure that the mains cable and handset cable cannot be driven
over or otherwise crushed when the bed is moved.
Continuous operation must not exceed two minutes! After this
time, a rest period of at least 18 minutes must be observed.
Alternatively: one minute continuous operation followed by a nine
minute rest period.
If the electric drive is operated for a much longer period, e.g. due
to the patient "playing" continually with the handset, the thermal
protection device integrated in the transformer of the control unit
will deactivate the drive permanently.
In this extremely rare case, the control unit must be replaced.
Table of contents
Other Stiegelmeyer Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Leftside Industries
Leftside Industries Crossover G3 Infinity Knee manual

Baylis Medical
Baylis Medical SureFlex Instructions for use

Amico
Amico Exergen 134201 Quick Installation and Maintenance Guide

Graham Field
Graham Field Lumex FR587W Series Assembly and operating instructions

idi
idi ilex installation guide

Heraeus
Heraeus PALAMIX Instructions for use