ESR-Auto Plus®RAM Reset Procedure
For Model 504-V9,V10 and Model 505-V3
Sep 2014, 880004-4
Page 1 of 4
Note: Performing a RAM reset will clear all existing patient and QC data. If possible, print the
patient log file and QC data before proceeding with a RAM reset.
1. Unplug the instrument (Model 504) or power off the instrument using the ON/OFF switch (Model 505)
and wait at least 5 seconds. While holding down the red ‘X’button on the instrument keypad, plug in
or power on the instrument. The following instrument display readings may differ slightly, depending
on the software version.
2. Instrument reads: “Setup, Reset System?”
Release the red ‘X’button and press the green ‘’button.
3. Instrument reads: “Reset All to Default! Sure?”
Press the green ‘’button.
4. Instrument reads: “RAM Cleared Restart System.”
Unplug or turn off the instrument and wait at least 5 seconds.
Plug in or turn on the instrument without holding the red ‘X’button.
5. Instrument reads: The software version will display, followed by “New Software, [OK] for Setup?”
Press the green ‘’button.
6. Instrument reads: “Motor Setup, Please Wait...”
The instrument will cycle and print out the system settings.
This step may take a couple of minutes.
7. Instrument reads: “IR Auto Set, Remove Tubes [OK]”
Remove all tubes/test rack and press the green ‘’button.
8. Instrument reads: “Calibrating…”
The instrument will print some settings.
9. Instrument reads: “RAM Now Reset to Default. [OK]”
Press the green ‘’button.
10. Instrument reads: “Setup User Config?”
Press the green ‘’button.
11. Instrument reads: “Set Date & Time?”
Press the green ‘’button. Default date format is YY-MM-DD.
Press the green ‘’button to accept the date information on the display.
Press the red ‘X’button to change the date.
Press the green ‘’button to advance to the time menu.
Press the green ‘’button to accept the time information on the display.
Press the red ‘X’button to change the time.
Press the green ‘’button to advance to the next prompt.
12. Instrument reads: “Set ESR Method?”
Press the green ‘’button.
13. Instrument reads: “Set Tube Type?”
Press the green ‘’button.
ESR-Auto Plus®RAM Reset Procedure
For Model 504-V9,V10 and Model 505-V3
Sep 2014, 880004-4
Page 2 of 4
14. Instrument reads: “1.2ml (PAED)?”
The available tube settings are “1.2ml” only, “2.0ml” only or “Both 1.2ml & 2.0ml”.
Press the red ‘X’button to scroll through the available options.
Press the green ‘’button to accept the desired tube type.
15. Instrument reads: “Set Calibr.?”
Press the green ‘’button.
If the instrument reads “+0%”, press the green ‘’button.
If not, press the red ‘X’button, enter “0” and press the green ‘’button.
16. Instrument reads: “Set Meas. Time?”
Press the red ‘X’button.
17. Instrument reads: “Audible Alarm” (Model 505 only)
Press the green ‘’button to access this feature.
Press the green ‘’button to enable the audible alarm.
18. Instrument reads: “Set Printouts?”
Press the green ‘’button.
19. Instrument reads: “Auto Print?”
Press the green ‘’button.
20. Instrument reads: “Ticket Style?”
Patient results can print in either ticket style or log style.
For ticket style, press the green ‘’button.
For log style, press the red ‘X’button.
NOTE: If the instrument is interfaced with an LIS, the print style selected
should match the data output format written in the LIS interface.
21. Instrument reads: “Config Ports?”
Press the green ‘’button.
22. Instrument reads: “Int Prn On?”
Press the green ‘’button.
23. Instrument reads: “Com 1 On?”
Press the green ‘’button.
24. Instrument reads: “Com 2 On?”
Press the green ‘’button.
25. Instrument reads: “Main Printer, Int. Prn?”
Press the green ‘’button.
26. Instrument reads: “Print Setup?”
Press the red ‘X’button.
27. Instrument reads: “Set Access Code?”
Press the red ‘X’button.
ESR-Auto Plus®RAM Reset Procedure
For Model 504-V9,V10 and Model 505-V3
Sep 2014, 880004-4
Page 3 of 4
28. Instrument reads: “See Sys Info?”
Press the red ‘X’button.
29. Instrument reads: “Service Menu?”
Press the green ‘’button.
30. Instrument reads: “Restricted Entry, Give Code.”
Enter “48” and press the green ‘’button.
31. Instrument reads: “Set Language?”
Press the red ‘X’button.
32. Instrument reads: “Date Style?”
Press the green ‘’button.
Press the red ‘X’button to scroll through the time and date formats.
Press the green ‘’button to accept the preferred format.
33. Instrument reads: “Code USR Config?”
Press the red ‘X’button.
34. Instrument reads: “Set IR Power?”
Press the red ‘X’button.
35. Instrument reads: “LCD Contrast?”
Press the red ‘X’button.
36. Instrument reads: “Set Alarms?”
Press the green ‘’button.
37. Instrument reads: “Over Fill”
If the instrument reads “690”, press the green ‘’button.
If not, press the red ‘X’button, enter “690” and press the green ‘’button.
38. Instrument reads: “1.2 ml Low Fill”
If the instrument reads “510”, press the green ‘’button.
If not, press the red ‘X’button, enter “510” and press the green ‘’button.
39. Instrument reads: “No Fill”
If the instrument reads “200”, press the green ‘’button.
If not, press the red ‘X’button, enter “200” and press the green ‘’button.
40. Instrument reads: “Report Range Max”
If the instrument reads “120,” press the green ‘’button.
If not, press the red ‘X’button, enter “120” and press the green ‘’button.
41. Instrument reads: “Motor Settings?”
Press the red ‘X’button.
ESR-Auto Plus®RAM Reset Procedure
For Model 504-V9,V10 and Model 505-V3
Sep 2014, 880004-4
Page 4 of 4
42. Instrument reads: “Printouts?”
If instrument is not connected to the LIS, press the red ‘X’button.
When transmitting data to an LIS, press the green ‘’button. (The most
common LIS printer settings: “Int Prn Type” = 1; “Main Port” = 3; “Enabled Ports”
= 7.)
43. Instrument reads: “Set Prediction?”
Press the red ‘X’button.
44. Instrument reads: “Reg Test Rack?”
Press the green ‘’button.
45. Instrument reads: “New Test Rack, Sure?”
Press the green ‘’button.
46. Instrument reads: “Insert Rack [OK]”
Insert the test rack and press the green ‘’button.
The instrument will scan the rack twice and print the results.
47. Instrument reads: “Set Thermometer?”
Press the green ‘’button.
48. Instrument reads: “Compensation On?”
Press the green ‘’button.
49. Instrument reads: “Celsius:___”
Press the green ‘’button.
NOTE: This setting is the internal working temperature of the instrument.
The temperature sensor reading will flash on the screen temporarily.
50. Instrument reads: “Test Mode On?”
Press the red ‘X’button.
51. Instrument reads: “Reading Test?”
Press the red ‘X’button.
52. Instrument reads: “Power Test?”
Press the red ‘X’button.
53. Instrument reads: “Initiating Motor…”
This step may take a couple of minutes.
54. Instrument reads: “Next Sample.”
The instrument is now ready for sample analysis.

This manual suits for next models

2

Other Streck Laboratory Equipment manuals

Streck ESR-Auto Plus 506 User manual

Streck

Streck ESR-Auto Plus 506 User manual

Streck ESR-Auto Plus 505 User manual

Streck

Streck ESR-Auto Plus 505 User manual

Streck ESR-Auto Plus Instruction Manual

Streck

Streck ESR-Auto Plus Instruction Manual

Streck Cell-Free DNA BCT CE User manual

Streck

Streck Cell-Free DNA BCT CE User manual

Streck ESR-10 User manual

Streck

Streck ESR-10 User manual

Streck A1c-Cellular User manual

Streck

Streck A1c-Cellular User manual

Popular Laboratory Equipment manuals by other brands

SynOptics SYNBIOSIS Protos 3 quick guide

SynOptics

SynOptics SYNBIOSIS Protos 3 quick guide

Metrohm 917 Coulometer Tutorial

Metrohm

Metrohm 917 Coulometer Tutorial

Cole Parmer GenoLyte HG-200 quick start guide

Cole Parmer

Cole Parmer GenoLyte HG-200 quick start guide

Thermo Scientific Evolution 200 Series user guide

Thermo Scientific

Thermo Scientific Evolution 200 Series user guide

3M 9900 Series instructions

3M

3M 9900 Series instructions

Apex Instruments XC-5000 Operator's manual

Apex Instruments

Apex Instruments XC-5000 Operator's manual

uv guard T Series Installation and operation manual

uv guard

uv guard T Series Installation and operation manual

Molecular Devices StakMax user guide

Molecular Devices

Molecular Devices StakMax user guide

STAVIOR UVC-TMS 360 user manual

STAVIOR

STAVIOR UVC-TMS 360 user manual

Labtech MultiVap 54 user manual

Labtech

Labtech MultiVap 54 user manual

DINKO Instruments 1.9745.00 manual

DINKO Instruments

DINKO Instruments 1.9745.00 manual

Klein Tools TI220 instruction manual

Klein Tools

Klein Tools TI220 instruction manual

Philips AVENT SVF271/42 Instruction leaflet

Philips

Philips AVENT SVF271/42 Instruction leaflet

OKW Datec-Control XS Series Assembly instruction

OKW

OKW Datec-Control XS Series Assembly instruction

Olympus U-APT instructions

Olympus

Olympus U-APT instructions

EG Xcellerex XDUO 2500 operating instructions

EG

EG Xcellerex XDUO 2500 operating instructions

Cognionics CGX Quick-20m manual

Cognionics

Cognionics CGX Quick-20m manual

NUVE BM 15 instruction manual

NUVE

NUVE BM 15 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.