0986-001-011 Rev-R EN
www.stryker.com 7
To Disconnect the CastVac from the Power
Source and the Cast Cutter
WARNINGS:
▪ ALWAYS set the CastVac function switch to the off position before
disconnecting the CastVac from the facility power outlet.
▪ The CastVac function switch does not remove electrical power from the
CastVac. To remove the electrical power, you must disconnect the power
cord plug from the facility power outlet.
CAUTION: To reduce the risk of damage to the power cord, ALWAYS grasp the
power cord plug, not the power cord, when disconnecting the CastVac.
1. Set the CastVac function switch to the off position.
2. Disconnect the CastVac power cord plug from the facility power outlet.
3. Disconnect the Cast Cutter power cord plug from the CastVac power
outlet.
4. Remove the Cast Cutter power cord from the Cast Cutter power cord
straps.
5. Depress the button on the hose cuff and pull the hose from the Cast
Cutter.
6. Install the vacuum line cap onto the Cast Cutter.
Cleaning
WARNINGS:
▪ ALWAYS disconnect the power cord plug from the facility power outlet
before cleaning the equipment. Failure to comply may cause fire, electric
shock, or injury.
▪ DO NOT immerse or soak any component of the equipment in liquid. DO
NOT allow moisture or liquid to soak into electrical plugs, receptacles, or
openings. Moisture or liquid inside the equipment may create an electric
shock hazard.
▪ DO NOT use solvents, lubricants, or other chemicals, unless otherwise
specified.
1. Disconnect the power cord plug from the facility power outlet.
2. Wipe the external surfaces of the equipment with a soft cloth moistened
with a non-abrasive, mild detergent and water.
3. Thoroughly dry the equipment with a soft, absorbent cloth.
Inspection, Testing, and Maintenance
WARNINGS:
▪ Only individuals trained and experienced in the maintenance of reusable
medical devices should inspect, test, and maintain this equipment.
▪ DO NOT disassemble or service this equipment, unless otherwise
specified.
▪ ALWAYS disconnect the power cord plug from the facility power outlet
before performing maintenance procedures on the equipment. Failure to
comply may cause fire, electric shock, or injury.
NOTES:
▪ Maintenance documentation for this equipment is available upon request
to Stryker-authorized service personnel only.
▪ For service, contact your Stryker sales representative or call Stryker
customer service. Outside the US, contact your nearest Stryker subsidiary.
INTERVAL ACTION INSPECTION CRITERIA
Upon initial
receipt and
before each
use.
Inspect the
equipment.
If damage or excessive wear is
apparent, contact Stryker for repair.
If any components are loose or
missing, contact Stryker for repair.
Inspect the
power cord
and/or power
source cord.
If cracks, cuts, or damage is apparent,
contact Stryker for repair.
Operate the
equipment.
If unusual sound, excessive noise, or
vibration is apparent, contact Stryker
for repair.
As required. Inspect the
filter cartridge.
If the filter cartridge is full or damaged,
replace the filter cartridge. See the To
Replace the High Performance Filter
Cartridge section.