Sub-Zero SZ-7505T User manual

Touch Sensitive Temperature Indicator
TM
SZ-7505T
Operating Instructions
Introduction :
ThenewSZ-7505TisthenextgenerationofSubzeroindicators.Their
IP ratings are greatly improved and have an excellent iconic display.
The touch feature whilst increasing reliability also gives a great user
experience.
Their operation is very user friendly and is easily understood with the
examplesintheinstructionsbelow.
The SZ-7505T can be used as a indicator with a temperature range
O O
from-45 Cto99 C.Calibrationviakeypadispossible.
CAUTION
WIRING: The probe and its corresponding wires should never be
installed in a conduit next to control or power supply lines. The
electrical wiring should be done as shown in the diagram. The power
supply circuit should be connected to a protection switch. The
terminalsadmitwiresofupto2.5sqmm.
WARNING: Improper wiring may cause irreparable damage and
personal injury. Kindly ensure that wiring is done by qualified
personnelonly.
Maintenance: Cleaning: Clean the surface of the indicator with a
soft moist cloth. Do not use abrasive detergents, petrol, alcohol or
solvents.
Notice: The information in this document is subject to change in
ordertoimprove reliability ,designor function withoutpriornotice and
does not represent a commitment on the part of the company. In no
eventwill thecompany beliablefor direct,indirect, special,incidental
or consequential damage arising out of the use or inability to use the
product or documentation, even if advised of the possibility of such
damages.Nopart of thismanualmay be reproducedortransmittedin
any form or by any means without the prior written permission of the
company.
Installation:Fixinganddimensionsofpanelmodels:
To fix the unit, slide the fastener 1 through the guides 2 as per the
position shown in the figure. Move the fastener in the direction of the
arrow, pressing tab 3 it permits to move the fastener in the opposite
directionofthearrow.
Indicator : Indicator should be installed in a place protected by
vibration, water and corrosive gasses and where ambient
temperature does not exceed the values specified in the technical
data.
Probe :To give a correct reading, the probe must be installed in a
place protected from thermal influences, which may affect the
temperaturetobecontrolled.
71mm
29mm
Panel cutout
3
1
2
36mm
59 mm
10mm
MAX
Panel Cutout and Dimensions :
Side Lock
Indicator
GASKET
PANEL
Suggested Wiring SZ-7505T (230Vac)
Caution:Wiring for 230Vac load only
TO SZ-T75
5
3
21
PROBE 230VAC
50-60Hz
PHASE
NEUTRAL
USER INTERFACE
UP In Program mode:
Scroll through parameters &
parameter value. Increases
Touch and hold for 2sec to enter into
program mode.
In program mode:
Decreases parameter value
PRG
Down/
Program
In program mode :
set/save the changed value of parameter.
SET
Set
dd Parameter
2Function : This parameter is used to delay
the display of temperature by the set in
this parameter.
Each value corresponds to 5 seconds, if the value is set to 1, it
corresponds to 5 seconds, if it is set to 2, it corresponds to 10
secondsandsoon.
Min Max
Fac.
036 0
For example, if this parameter is set to 1, temperature on the
displaywillbe updatedafter5seconds.Thesamevaluewillbe
consideredforcalculationandlogging.
Display delay parameter is applicable only when temperature
isincreasing(rising).Whentemperatureisdecreasing(falling)
thisparameterwillnotbeapplicable.
Ifthisparameterissetto0,thisfeaturewillbedisabled.
PA Parameter
3
Min Max
Fac.
-99 99 0
Function : To change Password.
User cannot enter into program mode, if correct password is
notentered.
If the password is kept other then 0, user need to enter correct
passwordtoenterintoprogrammode.
Ifpasswordis0,usercandirectlyaccessprogrammode.
Min Max
Fac.
10OC
-10OC0OC
Function: To set probe calibration.
In time it may be possible that the display may be offset by a
degreeorso.
To compensate for this error, user may need to add or minus
thedegreesrequiredtoachievethecorrecttemperature.
O
Example : The temperature on the display is 28 C, whereas
O
the actual temperature is 30 C. User will have to set the P5
parameterto2,whichmeansthatonceoutoftheprogramming
O O O
mode,thetemperatureondisplaywillbe 30 C(28 C+2 C).
Touch & hold key for 2 seconds.
P5 Parameter
1
Parameter List :
Suggested Wiring SZ-7505T (12Vac/dc)
TO SZ-T75
5
3
21
PROBE 12V
ac/dc
Suggested Wiring SZ-7505T (24V )ac/dc
TO SZ-T75
5
3
21
PROBE 24V
ac/dc
Housing : Black ABS Plastic, Auto-extinguish
Front Cover : Polycarbonate Plastic
Dimensions : Frontal : 78 X 36mm,
Depth : 59mm
Panel Cutout : 29 X 71mm
Mounting : Flush panel mounting with fasteners
Protection : IP65 Front (with gasket)
Connections : Screw terminal blocks.
< 2.5sq mm terminal only.
Display : 2 X 17mm 7 segment display &
1 LED for Indication
Data storage : Non-volatile EEPROM memory
Power input : 230 Vac ±15 %, 50-60Hz.
Others on request.
Operating temp. : 0°C to 60°C (non-condensing)
Operating humidity : 20% to 85% (non-condensing)
Storage temp : -25°C to 60°C (non-condensing)
Measuring Range : -45°C to 99°C
Input : NTC probe, SZ-T75
Resolution : +/- 1°C
Accuracy : +/- 1°C
TECHNICAL DATA
Change Password
Keypad Lock
Restore factory defaults
End Programming
Delay the display of temperature.
INDEX
Sr.
No. Para. Description
1
3
4
5
6
7
P5 Probe calibration.
dd
PA
LP
FS
EP
Password Function
Temperature Logging
2
Operating Messages
Suggested Wiring SZ-7505T (48VDC)
TO SZ-T75
5
3
21
PROBE 48VDC
Suggested Wiring SZ-7505T (110Vac)
TO SZ-T75
5
3
21
PROBE 110Vac

LLLast low temperature
Last low temperature
logged.
LH Last high temperature
Last high temperature
logged.
Probe fail
PPProbe short circuit, circuit
open or without probe, or
0
temperature is > 99 C or
0
<-45 C
OPERATING MESSAGES
EP Parameter
6Function: To end programming.
Oncethe keyispressed,theindicatorgoes
into the normal mode and displays the
temperatureandallsettingsarerecorded.
T
programming
press "SET" key
o end
FS Parameter
5Function : To restore default settings of the
indicator.
When set to 1 all parameters are programmed to factory set
values.
Useful to debug setting related Problems.
When set to,
0 = FS is disable.
1 = FS as per default value. Min Max
Fac.
010
Passwordfunction
In Program mode:
Touch & hold " " key for 2sec. Display will flash "P5"
parameter if "PA" value is kept "0". If other than "0", then "PA'
and "0" will flash. Use " "and " " keys to enter the
password. On entering correct value, display will flash the first
parameter User can scroll through parameters using
" " or " " keys.
"P5".
HighandLow temperature loggingfunction
How to reset the Logged values
While the display is showing the logged values, if user
touches & holds the "SET" key for 1sec, the logged values will
be cleared and "rS" will be displayed.
Log Values will get reset after Power ON/OFF.
How to see the logged values:
LL : Last Low temperature
LH : Last High temperature
Touch and hold " " key for 1sec. display will flash “LL” and
the corresponding temperature for 4 seconds. After this,
display will flash “LH” and the corresponding temperature for
4 seconds and come out of Log mode and will display
Control probe temperature.
Disclaimer:This manual&itscontentsremain the solepropertyof
PVR CONTROLS . India and shall not be reproduced or
distributed without authorization. Although great care has been
taken in the preparation of this document, the company or its
vendors in no event will be liable for direct, indirect, special,
incidental or consequential damage arising out of the use or
inabilitytousethe product ordocumentation,evenif advised ofthe
possibility of such damages. No part of this manual may be
reproducedor transmittedin any form or by anymeans without the
prior written permission of the company. PVR CONTROLS.,
reserves the right to make and changes or improvements without
priornotice.
Warranty: This product is warranted against defects in materials
and workmanship for a period of one year from the date of
purchase.During thewarranty period, product determined by usto
be defective in form or function will be repaired or, at our option,
replaced at no charge. This warranty does not apply if the product
hasbeendamagedbyaccident,abuse,andmisuseorasaresultof
serviceor modification other than bythe company.This warrantyis
inlieu ofany other warranty expressed or implied.In no event shall
the company be held liable for incidental or consequential
damages, including lost revenue or lost business opportunity
arisingfromthepurchaseofthisproduct.
Ball Valves
Globe Valves
Hand Valves
Flow Switches
Solenoid Valves
R
Cold Room Controller
Chiller Controller
Two Compressor Controller
Heating Controller
Humidity Controller
Pressure Controller
TM
OUR OTHER PRODUCTS
00 / 19.05.17
7
LP Parameter
4
Min Max
Fac.
010
Function: To lock keypad.
Flashing
When locked all parameters can only be viewed, but not
modified.
Note: If LPparameterissetto1 andifusertriesto changeany
parametervalue,"LP"willflashonthedisplay.
This parameter is used to lock the keypad so that tampering is
notpossiblebyby-standers.
0=keypadunlocked
1=keypadlocked
Other Sub-Zero Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Kistler
Kistler dmf-p a300 instruction manual

Interscan Corporation
Interscan Corporation GF1900 manual

Lincoln
Lincoln 961 Installation and maintenance guide

Agilent Technologies
Agilent Technologies U1731C quick start guide

M-system
M-system MXAB instruction manual

Fieldpiece
Fieldpiece SC54 Operator's manual