Sub-Zero VAF433-T User manual

Instruction Manual
VAF433-T / VAF433-K
Introduction
The VAF is a panel mounted 96 x 96mm Digital Panel Meter
for the measurement of important electrical parameters like
AC Voltage, AC Current, RPM, Frequency and Power Factor
VAF can be configured and Programmed On site for the
following:
PT Primary, PT Secondary, CT Primary, CT Secondary (5A
or 1A) and System Type 3 phase 3W or 4W or single-phase
system.
The front panel has Four keys using which the user can scroll
through different screens and configure the product.
Introduction
WIRING: The probe and its corresponding wires should
never be installed in a conduit next to control or power supply
lines. The electrical wiring should be done as shown in the
diagram. The power supply circuit should be connected to a
protection switch. The terminals admit wires of upto 2.5sq
mm.
WARNING: Improper wiring may cause irreparable damage
and personal injury. Kindly ensure that wiring is done by
qualified personnel only.
Maintenance: Cleaning: Clean the surface of the controller
with a soft moist cloth. Do not use abrasive detergents, petrol,
alcohol or solvents.
Notice: The information in this document is subject to change
in order to improve reliability, design or function without prior
notice and does not represent a commitment on the part of
the company. In no event will the company be liable for direct,
indirect, special, incidental or consequential damage arising
out of the use or inability to use the product or documentation,
even if advised of the possibility of such damages. No part of
this manual may be reproduced or transmitted in any form or
by any means without the prior written permission of the
company.
Controller:
Controller should be installed in a place
protected by vibration, water and corrosive gasses and where
ambient temperature does not exceed the values specified in
the technical data.
Probe: To give a correct reading, the probe must be installed
in a place protected from thermal influences, which may
affect the temperature to be controlled.
Dimensions & Panel Cutout
Connection Diagram
Index
Sr.
No. Parameter
Description
Technical Specification.
User Interface.
Home Page Screen Description
1
Parameter setting mode.
2
Network selection 3P-4W, 3P-3W&1P-2W.
3
Select CT primary ratio.
4
Select PT primary ratio.
5
Select PT secondary ratio.
6
Select number of poles.
7
Clear run hours.
8
Restore factory defaults
Error Messages
Technical Specification
Housing : Front Cover: Polycarbonate Plastic
Back Cover: ABS Plastic
Dimensions : Frontal: 96 X 96mm, Depth: 61mm
Panel Cutout : 92 X 92mm
Mounting : Flush panel mounting with Fasteners
Protection : IP65 Front
Connections : Terminal connectors.
< 2.5sq mm terminal only with
U-type lugs.
Display : 3 X 17mm 7 segment WHITE display,
15 Iconic LEDs for Indication
3 X 17mm 7 segment RED display,
8 Iconic LEDs for Indication
Data storage : Non-volatile flash memory
Operating temp. : 0°C to 60°C (non-condensing)
Operating humidity : 20% to 85% (non-condensing)
Storage temp. : -25°C to 60°C (non-condensing)
Supply Voltage : 230 Vac ± 10% , 50/60Hz Standard.
Wiring Input : 3ø-4W, 3ø-3W system.
Measuring Range
Input Rated Voltage : 11 to 375 V AC (L-N)
: 19 to 649 V AC (L-L)
Input Rated Current : Nominal 5A AC
CT Primary : 5 A to 10 kA
CT Secondary : 5 A
PT Primary : 100 V to 500 kV
PT Secondary : 100 to 375 V AC (L-L)
Electrical Connection : 3ø-4W, 3ø-3W, 2ø-3W, 1ø-2W.
Resolution
Current / Voltage : Resolution depends upon CT,
PT Primary settings
RPM : 0.1
Run Hour : 0.1 Hr
Accuracy
Voltage : ± 0.5% of F.S. ± 2 digits.
Current & Average Current : ± 1% of F.S. ± 2 digits
Frequency : ± 0.1 Hz ± 1 digit
Run Hour : ± 1%
RPM : ± 0.5%
Display Scrolling : Automatic or Manual
(Programmable)
Power Consumption : 1.5VA Max
RPM : 1350 to 1950
Pole: 0
(Range: 0-98, selectable in
steps of 2)
Frequency : 40-70 Hz
Run Hours : 0 to 99999.9 Hrs
Burden : 0.5VA @5A per phase
User Interface
Sr.
No.
Description
1
Electrical parameters like AC Voltage, AC Current,
RPM, Frequency and Power Factor will be seen as
Per screen no and auto or manual mode. Kindly refer
Home Page Descriptor.
2
3
4
Unit Kilo Volts
5
Unit Volts (Only for T model)
6
Kilo Ampere
7
Ampere
8
Unit for Run Hours
9
Frequency units Hertz
10 Line to Line value of corresponding
Parameter (only for T model)
11
Line to Neutral value of First Phase
12
Average value of corresponding parameter
13
Line to Neutral value of Second Phase
14
Line to Neutral value of Third Phase
15
Line to line voltage
16
Line to neutral voltage
17 Next key:
In Manual Mode: To scroll for Next Screen.
In Program Mode: To scroll to Next
Parameter.
Note: Kindly refer ‘Home Page Screen
Description’ for further details.
18
Down key/ I/ h:
In Normal Mode (Auto/Manual):
Press this key for 1 second, the First Screen
Displays phase current of three phases. Line
to neutral and line to line voltage of three
phases are displayed one by one.
Press this key for 3 seconds,
to display
Run Hours (for non-zero poles).
In Program M
ode: To increase Parameter
value.
Note:
To return back to current page, press
for 3 seconds.
19
Up key:
In Normal Mode (Auto/Manual):
Press this key for 1 second.
The First Screen displays line to neutral
voltage of three phases.
The
Second Screen displays line to line
voltage of three phases.
Press this key for 3 seconds.
Displays rpm for non-zero poles.
In Program Mode: To increase Parameter
value.
Note:
To return back to current page, press
for 3 seconds.
20 Exit key:
In Normal Mode:
Press this key for 4 seconds to toggle
between Automatic/Manual Mode.
In Program Mode:
To save the changed Parameter and exit to
Normal mode.
Home Page Screen Description
Automatic Mode / Manual Mode Settings:
Press A/M mode for 4
seconds to toggle between
Automatic/Manual Mode.
Note: By Default, unit operates in automatic mode. In
automatic mode online pages scroll automatically at the rate
of 5 sec per page. In automatic mode if any key is pressed,
unit temporarily switched to man
ual mode and the
appropriate page is displayed.
In Automatic Mode:
1. Display Descriptions in 3Ø – 4W:
1st Screen Displays Line to Neutral voltage of all 3 Phase
2nd Screen Displays Line to Line voltage of all 3 Phase
3rd Screen RPM
4th Screen Displays current of all 3 Phase
5th Screen Displays Run Hours
6th Screen
Displays Line to Neutral voltage, current of first
Phase
7th Screen
Displays line to neutral voltage, current of
second Phase
8th Screen
Displays line to neutral voltage, current of third
Phase
9th Screen
Displays average line to neutral voltage,
average current of three phase and frequency.
10th Screen
Displays average line to line voltage, average
current of three phases and frequency of
present phase.
2. Display Descriptions in 3Ø – 3W:
1st Screen Displays Line to line voltage of all 3 Phase
2nd Screen RPM
3rd Screen Displays Line to line current of all 3 Phase
4th Screen Displays run hours
5th Screen Displays Line to line voltage, current and
frequency of first and second Phase
6th Screen Displays line to line voltage, current and
frequency of second and third Phase
7th Screen Displays line to line voltage, current and
frequency of third and first Phase
8th Screen Displays average line to line voltage, average
current of three phase and frequency
3. Display Descriptions in 1Ø – 2W:
1st Screen
Displays line to neutral voltage, current of
phase and frequency.
2
nd
Screen
Displays Run Hours
3
rd
Screen
RPM
In Manual Mode:
To scroll through next screen press key.
1.Display Descriptions in 3Ø – 4W:
1st Screen Displays Line to Neutral voltage of all 3 Phase
2nd Screen Displays Line to Line voltage of all 3 Phase
3rd Screen RPM
4th Screen Displays current of all 3 Phase
I h
V r
A M
5th Screen Displays Run Hours
6th Screen
Displays Line to Neutral voltage, current of
first Phase
7th Screen
Displays line to neutral voltage, current of
second Phase
8th Screen
Displays line to neutral voltage, current of
third Phase
9th Screen Displays average line to neutral voltage,
average current of three phase and frequency.
10th Screen
Displays average line to line voltage, average
current of three phases and frequency of
present phase.
2. Display Descriptions in 3Ø – 3W:
1st Screen Displays Line to line voltage of all 3 Phase
2nd Screen RPM
3rd Screen Displays Line to line current of all 3 Phase
4th Screen Displays run hours
5th Screen Displays Line to line voltage, current and
frequency of first and second Phase
6th Screen Displays line to line voltage, current and
frequency of second and third Phase
7th Screen Displays line to line voltage, current and
frequency of third and first Phase
8th Screen Displays average line to line voltage, average
current of three phase and frequency
3. Display Descriptions in 1Ø – 2W:
1st Screen
Displays line to neutral voltage, current of
phase and frequency.
2nd Screen Displays Run Hours
3rd Screen RPM
Parameter Setting Mode
1. Program
Mode
Function: To enter into Program Mode.
Press & hold key for 4 seconds to enter into Program
Mode. When release the key, second display will flash .
Press UP/DOWN keys to modify the set value and to go to
the next parameter by pressing key.
Press the key to save the set value and to come out of
parameter setting after changing the set value.
2.
Function: Network selection.
Network selection, 3 Phase 4 wire, 3 Phase 3 wire and 1
Phase 2 wire
Min Max Fac.
3P4
1P2
3P4
3.
Function: To select CT Primary ratio.
To select CT Primary ratio.
Min Max Fac.
5A 10kA 5A
4.
Function: To select PT Primary ratio.
PT Primary: To select PT Primary ratio.
Min Max Fac.
100V 500kV 350V
5.
Function: To select PT Secondary ratio.
PT Secondary: To select PT Secondary ratio.
Min Max Fac.
100V 375V 350V
6.
Function: To select no. of Poles.
To select number of poles.
Min Max Fac.
0
98
0
7.
Function: To clear run hours.
If selected yes run hours will be cleared.
Min Max Fac.
8.
Function: To restore default settings of the
controller.
When set to YES all parameters are programmed to factory
values.
Useful to debug setting related problems.
Min Max Fac.
Error Messages
Message
Description
It indicates corresponding parameter reach
above specified measuring range.
Typical Wiring Diagram
NOTE: # All fuse types: 0.5A class CC UL type;
0.5A fast acting 600V
3 Phase - 4 Wire
3 ø- 4 Wire, 3 CT's
3 Phase - 4 Wire
3 ø- 4 Wire, 3 CT's and 3 PT's
3 Phase - 3 Wire
3 ø- 3 Wire, 2 CT's and 2 PT's
Note: Do not use 7,8, 9 & 10 connector for 3P 3W.
P2 Phase - 3 Wire
2 ø- 3 Wire, 2 CT's and 2 PT's
1 Phase - 2 Wire
1 ø- 2 Wire, 1 CT's
Disclaimer: These manual & its contents remain the sole
property of PVR CONTROLS, India and shall not be
reproduced or distributed without authorization. Although
great care has been taken in the preparation of this
document, the firm or its vendors in no event will be liable for
direct, indirect, special, incidental or consequential damage
arising out of the use or inability to use the product or
documentation, even if advised of the possibility of such
damages.
Prospective purchasers should not confine
themselves to the contents but should make their own
enquiries to satisfy themselves in all respects. No part of this
manual may be reproduced or transmitted in any form or by
any means without the prior written permission of the firm.
PVR CONTROLS, reserves the right to make and changes
or improvements without prior notice. PVR CONTROLS will
not accept any responsibility should any details prove to be
incomplete or incorrect. PVR CONTROLS assumes no
responsibility for any misleading content, or incorrectly listed
information due to inaccuracies in content or data supplied by
any source to the information available.
Warranty: This product is warranted against defects in
materials and workmanship for a period of one year from the
date of purchase. During the warranty pe
riod, product
determined by us to be defective in form or function will be
repaired or, at our option, replaced at no charge. Such
rectification shall be provided / carried out only upon
submitting a valid purchase receipt. Any claim raised after
warranty period shall not be entertained. This warranty does
not apply if the product has been damaged by accident,
abuse, willful default on part of the user, negligent use and
misuse or as a result of service or modification other than by
the firm. (De)mounting and/or (de)installation, and labor costs
are excluded from warranty. In no event shall the firm be held
liable for incidental or consequential damages, including loss
of revenue or loss of business opportunity arising from the
purchase of this product nor compensate you for any reason
whatsoever.
OUR OTHER PRODUCTS
Digital Panel Meter
Power Analyzer | Timer
PLC | HMI | Data Logger
REV2 – 11.03.2022

This manual suits for next models

1

Other Sub-Zero Measuring Instrument manuals

Sub-Zero MFM443-TX User manual

Sub-Zero

Sub-Zero MFM443-TX User manual

Sub-Zero MFM443-T User manual

Sub-Zero

Sub-Zero MFM443-T User manual

Sub-Zero VAF433-T User manual

Sub-Zero

Sub-Zero VAF433-T User manual

Sub-Zero SZ-7505T User manual

Sub-Zero

Sub-Zero SZ-7505T User manual

Popular Measuring Instrument manuals by other brands

Omron F3UV Series datasheet

Omron

Omron F3UV Series datasheet

LumaSense SmartDGA Hardware manual

LumaSense

LumaSense SmartDGA Hardware manual

CAISSON PM-20 Operation manual

CAISSON

CAISSON PM-20 Operation manual

EnerSys Wi-iQ3 user manual

EnerSys

EnerSys Wi-iQ3 user manual

Starrett 2900 Series user manual

Starrett

Starrett 2900 Series user manual

Epiq Solutions Flying Fox ES012-518-A Getting started guide

Epiq Solutions

Epiq Solutions Flying Fox ES012-518-A Getting started guide

PDK PDR250 user manual

PDK

PDK PDR250 user manual

socomec COUNTIS E25 quick start

socomec

socomec COUNTIS E25 quick start

MONARCH INSTRUMENT EXAMINER 1000 instruction manual

MONARCH INSTRUMENT

MONARCH INSTRUMENT EXAMINER 1000 instruction manual

AerobTec Altis v4 user manual

AerobTec

AerobTec Altis v4 user manual

QUARTA RD1503+ quick start guide

QUARTA

QUARTA RD1503+ quick start guide

Fluke FOM Series Quick reference card

Fluke

Fluke FOM Series Quick reference card

Leica DISTO D810 touch manual

Leica

Leica DISTO D810 touch manual

Tecnosystemi Toolsplit TSTI1 user manual

Tecnosystemi

Tecnosystemi Toolsplit TSTI1 user manual

EUTECH INSTRUMENTS WATERPROOF CYBERSCAN PD 300 PHDISSOLVED OXYGEN... instruction manual

EUTECH INSTRUMENTS

EUTECH INSTRUMENTS WATERPROOF CYBERSCAN PD 300 PHDISSOLVED OXYGEN... instruction manual

PCB Piezotronics IMI Sensors M623C00 Installation and operating manual

PCB Piezotronics

PCB Piezotronics IMI Sensors M623C00 Installation and operating manual

Tyrepal Solar manual

Tyrepal

Tyrepal Solar manual

Hach TOC-X5 user manual

Hach

Hach TOC-X5 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.