Supelco ATIS User manual

ATIS™ OPERATION MANUAL
Adsorbent Tube Injector System
T702019A
1

Introduction
The Adsorbent Tube Injector System (ATIS) is a sample preparation device that provides a quick
and efficient way to transfer calibration standards or samples onto an adsorbent tube. The ATIS
employs the technique of flash vaporization to vaporize the sample in a continuous flow of inert
gas. The inert gas carries the sample to the tube. After enough time has elapsed, typically less
than 5 minutes, the tube is removed from the ATIS and analyzed using the appropriate technique
for that adsorbent tube.
The ATIS performs these tasks:
• Injects calibration standards onto adsorbent tubes to calibrate your analytical system.
• Injects surrogates and system monitoring compounds onto an adsorbent tube before or after
sampling.
• Removes moisture from the tube prior to analysis (dry purging).
• Vaporizes neat compounds into a gas-sampling bag to create a calibration standard.
The ATIS can be expanded:
• Thermally extract solid samples and collect the volatiles onto an adsorbent tube. The
temperature range is adjustable from ambient to 150°C. Extraction glassware is available
(ordered separately) that will accommodate solid samples up to 76mm (3”) length, and
13mm (1/2”) in diameter.
• Purge volatiles from an aqueous sample onto an adsorbent tube at ambient temperatures. A
purge & trap / humidifier module is available (ordered separately) that interfaces with the
ATIS. This module can also be used to generate a dynamic humidified stream of an inert
gas for spiking calibration standards. The purge & trap module includes purge and trap
glassware, and a separate flow controller that allows the user to set a separate purge (wet)
flow rate independently of the (dry) flow rate.
How the ATIS Works
The ATIS is comprised of the following major components:
• Injection Glassware
• Adsorbent Injector Heating Block
• Carrier Gas Flow Control
Injection Glassware
The sample pathway of the ATIS is constructed of glass and stainless steel. The calibration
standard is injected by a syringe through a replaceable septum in the center of the injection
glassware. The injection glassware has an internal volume of 10 milliliters. This provides
adequate volume for the vaporized liquid to expand.
Injector Heater Block
The injector heater block transfers heat produced by the block heater to the injection glassware.
The heated glassware vaporizes the sample. A thermometer port is located in the top of the
injector block to monitor the temperature during use.
Carrier Gas Flow Control
A supply of clean gas (nitrogen, helium, or air) is required to flow through the injection glassware
during sample introduction. The gas carries the calibration standards / sample to the adsorbent
tube during sample preparation.
The ATIS includes an adjustable flow controller and an on/off control valve. The flow controller
will maintain a constant flow of gas, regardless of changing downstream pressures i.e., an
adsorbent tube. The flow range is adjustable from 0 to 110mL/min. This flow controller provides
better accuracy and control of the flow rate than a rotameter or needle valve. The on/off valve
allows the gas supply to be turned off when the unit is not in use. A gas purifier is included that
2

will remove trace levels of hydrocarbon contaminants from your gas stream. However, additional
purification may be required if your gas supply has high contamination levels.
Safety Information
The TALBOYS®Dry Block Heater included with this product is the heat source for the ATIS. This
product operates at temperatures high enough to cause serious burns to the skin. A heat shield
is provided to reduce the risk of getting burned while the ATIS is in use. The temperature of the
heat shield will be approximately half of the set block temperature.
Danger: Do not use the ATIS in the presence of flammable or combustible materials or explosive
gases. Fire or explosion may result, causing death or injury. This device contains electrical
components that may ignite such materials.
Danger: Do not connect the ATIS to any flammable gases i.e., hydrogen. This will create a fire
and/or explosion hazard.
Warning: The ATIS should be placed in a fume hood or have its own ventilation duct when in
use.
Warning: Be careful when working around the ATIS when the block heater temperature is turned
on.
Warning: Do not block the exit port of the pressure relief valve. Doing so could cause pressure
to build in the system that may cause the septum to blow out of the septum holder, or worse
case, cause the glassware to explode. The pressure relief valve will open to relieve the gas
pressure if it exceeds 6psig.
Warning: Before performing any maintenance, turn off the block heater, unplug the electrical
cord, and allow sufficient time to let the block heater to cool down.
In addition to reading this manual, please read the operating manual included with the TALBOYS
Dry Block Heater (follow the instructions for the Standard Dry Block Heater) before using the
ATIS.
Specifications
Dimensions
Requires a 61mm x 61mm (2’ x 2’) space to house the assembled unit.
Power Requirements for the TALBOYS Block Heater
Cat# 28520-U ATIS 120 VAC +/-10% 50/60 Hz, 110 Watts, 0.9 Amps
Cat# 28521-U ATIS 240 VAC +/-10% 50/60 Hz, 110 Watts, 0.5 Amps
Certifications for the TALBOYS Block Heaters
Safety Standards:
IEC 61010-1 Safety requirements for electrical equipment for measurement, control and
laboratory use. Part I: General Requirements.
IEC 61010-2-010 Part II: Particular requirements for laboratory equipment for the heating of
materials.
UL Std.No. 61010-1
EMC standards: EN6100-3-2 EN61000-3-3, EN61000-4-5 FCC-B, EN61000-4-4 EN55022-A,
EN61000-4-2, EN61000-4-3, EN61000-4-11, EN61000-4-6, EN45501
Associated EU guidelines:
EMC directive 89/336/EEC, LVD directive 73/23/EEC
Temperature Range with all the ATIS parts attached:
Slightly above ambient to 150°C (302°F)
Pneumatic Specifications:
Maximum operating pressure 60psig (Suggested operating pressure 25-30psig)
Gas Flow Range:
0-110 mL/min at 25-30 psig
3

The ATIS Components
• Operation Manual
1. Block Heater
2. Injector Heater Block
3. Gas Purifier
4. Copper Tubing 1/8” OD
5. Flow Controller and On/Off Valve & Pressure Relief
6. Polypropylene Tubing 1/8” OD
7. 1/4” to 1/8” Reducing Union
8. Injection Glassware
9. 1/4” Union with Thumbwheel
10. Heat Shield
11. Thermometer
12. Luer/Hose-Barb Adapter
Spare Parts
SP1. 6mm Ferrules “Black”
SP2. 1/4” Ferrules “Orange”
SP3. Half-Hole Septa
SP4. Spare Relief Valve and 3/16” Ferrules
SP5. 1/4” PTFE Ferrules “White”
4

Installation
Site Preparation
Place the ATIS in a fume hood and away from any combustibles. The ATIS requires a regulated
supply (25–30psig) of nitrogen, helium, or air. Nitrogen is your best choice because it’s dry,
clean, and relatively inexpensive. The ATIS also requires an electrical supply 120 VAC or 240
VAC (depending on the unit you purchased). The Block Heater must be plugged into a grounded
outlet. See the operation manual included with TALBOYS Dry Block Heater for additional
information.
Tools Required for Assembly (not included)
• Small tubing cutter
• Knife or scissors
• 7/16” open-end wrench
• 1/2” open-end wrench
• 9/16” open-end wrench
• Flow meter (to set the gas flow rate)
Assembling the ATIS
To assemble the ATIS requires you to cut the supplied tubing to fit your workspace and to attach
the tubing to the appropriate fittings. We have installed the appropriate ferrules in each of the
fittings for your convenience. The fittings that connect to the ATIS glassware include a VESPEL®
ferrule. Use caution when tightening the fittings to the glassware – very little force is needed to
create a seal to the glassware. PTFE ferrules are also included if you prefer to use them instead
of the “orange” VESPEL type ferrules.
For additional help in plumbing the ATIS see the pneumatic diagram in Appendix 1 of this
manual.
The picture to the right illustrates what the
ATIS should look like AFTER you have
assembled it.
Place the injector heater block (with the
open slot facing to your right) in the
opening of the dry block heater.
5

Using a tubing cutter, cut a piece of
copper tubing to connect the gas
purifier to a regulated gas supply (user
supplied) of nitrogen, helium, or
compressed air. Pressure should be set
to 25 to 30psig after installation is
complete.
Using a wrench, tighten the nut 3/4 of a
turn, past fingertight.
Cut a second piece of copper tubing to
connect the outlet of the gas purifier to
the inlet of the on/off valve connected to
the flow controller.
Using a wrench, tighten both of these nuts
3/4 of a turn, past fingertight.
NOTE: Position the flow controller so you
have access to it during use.
Connect the 1/4” to 1/8” reducing union
to the injection glassware. The
SUPELCO Logo on the glassware should
face you.
Using a wrench, gently tighten the larger
nut 1/8 of a turn, past fingertight.
NOTE: Over-tightening will cause the
glassware to break.
Cut a piece of the white polypropylene
tubing to connect the outlet of the flow
controller to the union you attached to the
Injection Glassware.
Allow extra slack in the tubing so the tubing will absorb
the shock if the tubing is accidentally bumped. This helps
to prevent the glassware from breaking during use.
Using a wrench, tighten both of these nuts
3/4 of a turn, past fingertight.
Danger: Do not plumb the ATIS to an
y
flammable gases i.e., hydrogen. This wil
l
create a fire and/or explosion hazard.
6

Slide the injection glassware into the
injector heater block until the stem of the
injection glassware protrudes through the
recess of the injector block.
Install the thumbwheel to the 1/4” union.
See the next section to choose the correct size
ferrule for the adsorbent tube you plan to use.
Install the 1/4” union over the stem of the
glassware that protrudes through the recess.
Using a wrench, tighten the CENTER hex
portion of the union 1/8 of a turn while holding
the injection glassware by the septum holder.
The recess secures the union so you can attach your adsorbent tube by using only the thumbwheel. The
recess also promotes heat to transfer to the union. This prevents the calibration standard/sample from
condensing inside the union.
Place the heat shield over the injector
heater block so the septum holder
protrudes through the hole of the heat
shield. The base of the heat shield will
rest on the gray cover of the dry block
heater.
Insert the thermometer through the hole
of the heat shield. Note: The thermometer
port is dual tapered and will accept both a
normal thermometer, or a smaller diameter
digital thermometer.
The ATIS is now completely assembled.
Your set-up should look like the picture to
the right.
7

How to Attach your Adsorbent Tubes
The thumbwheel will accept tubes with either 6mm or 1/4” by placing the appropriate ferrule in the
thumbwheel. For odd size tubes use the Luer/Hose-Barb Adapter included. The pictures below
illustrate some examples.
NOTE: While using the ATIS, the temperature of the adsorbent tube will be close to the ambient
temperature of your lab. However if the adsorbent tube is attached to the thumbwheel for a long
period, the heat from the injector block will transfer to the attached adsorbent tube. The rate at
which this occurs is dependent on both the block heater temperature and whether the adsorbent
tube is made of glass or stainless steel. The transfer of heat will be greater with stainless steel
than glass tubes.
Installing Thermal Desorption Tubes
For 6mm OD tubes, install a 6mm ferrule
“black” in the thumbwheel.
For 1/4” OD tubes, install a 1/4” ferrule
“orange” in the thumbwheel.
Installing Solvent Desorption Tube
Use the Luer/Hose-Barb Adapter and a
piece of flexible tubing (not supplied) to
connect odd size adsorbent tubes. To
attach the Adapter, install a 1/4” ferrule
“orange” in the thumbwheel. The smooth
section of the adapter goes into the
thumbwheel.
Caution: Make sure the flexible tubing you
choose will withstand the temperatures
you plan on using. The wrong tubing
could melt or off-gas by-products if left
attached for a long period of time.
Installing Adsorbent Tubes with Male
Luer fittings
The inner diameter of the Luer/Hose Barb
Adapter will accept adsorbent tubes that
have a male luer fitting.
8

Installing Gas-Sampling Bags
Gas-sampling bags can be attached using
the Luer/Hose-Barb Adapter and a piece
of flexible tubing. If the stem of the bag is
3/16” in diameter you can use 3/16”
ferrule (included SP4) in the thumbwheel
in place of the adapter.
Setting the ATIS Parameters
There are two parameters you need set; one is the flow rate, the other is the block temperature.
Setting the Flow Rate (Carrier Gas)
Insert the type of adsorbent tube you plan to use in the thumbwheel of the ATIS, and gently tighten
the thumbwheel. With the gas supply on, attach a flow meter to the opposite end of the adsorbent
tube. Turn the on/off valve attached to the flow controller to the ON position. (The arrow on the
knob should be pointing towards the flow controller when it is ON.) Adjust the gas flow rate using
the black knob of the flow controller.
• To increase the flow rate -- turn the knob counter-clockwise.
• To decrease the flow rate -- turn the knob clockwise.
The flow controller has a 13-turn adjustment knob; each complete turn of the knob is
approximately 8 to 10mL/min change in the flow rate. The flow range of the flow controller is 0 to
110mL/min at 25-30psig.
Caution: Do not over-tighten the flow controller knob in either direction, or you could damage the
flow controller.
Setting the Temperature
The power switch is located on the right side of the block heater. When the switch is in the center
position, the unit is OFF. Depressing the switch to the left engages the Low Heat range control.
Depressing the switch to the right engages the High Heat range control.
The temperature range of the block heater with all the ATIS components attached:
• Low Heat Range is slightly above ambient to approx. 100°C (212°F)High Heat Range is approx. 75°C
to 150°C (167°F to 302°F)
9

Maintenance
Suggested maintenance for the ATIS includes the following:
• Check the temperature setting daily.
• Check the flow rate daily.
• Replace the septum after 25 injections.
• Replace the injection glassware after the first sign of any deposits.
• Replace the gas purifier annually.
Replacing the Pressure Relief Valve
In the event that you need to replace the pressure relief valve, a spare valve is included with the
ATIS.
To replace the valve:
1) Remove the nut that holds the relief valve.
Note: You may have to break the old relief valve to remove it from the nut.
2) Place a new 3/16” ferrule in the fitting.
3) Start the nut on the threads of the fitting.
4) Insert the relief valve through the nut and into the ferrule.
5) Gently push the relief valve so it is flush with nut.
6) Tighten the nut 1/8 to 1/4 of a turn. Do not over-tighten.
Guidelines for Using the ATIS
To accurately quantify the concentration of a sample acquired with an adsorbent tube, the
analytical system must first be calibrated. The ATIS simplifies this task. Below are a few starting
parameters, some experimentation may be needed to optimize the conditions for your application.
Allow enough time for the block heater to equilibrate to the desired temperature before injecting
your standards. You will also want to make sure you have gas flowing through your adsorbent
tube, prior to injecting your standard so the compounds will be carried to the tube.
Injecting Gas Phase Standards
Gas phase standards are available in compressed gas cylinders, made up in gas sampling bags,
or gas sampling bulbs.
• Set the block temperature between 60° to 75°C.
• Set the flow rate of the carrier gas to 50mL/min.
• Allow the adsorbent tube to remain attached for 2 to 5 minutes after injecting the gas mix with
the carrier gas on.
• The delivery rate of the gas syringe should be less than 10mL/min.
Large syringe volumes of gas standards can be injected through the septum of the injection
glassware, but the injection rate of the syringe plunger must be slow. The rate that you inject the
gas mix will have an affect on the total flow rate passing through the adsorbent tube. For
example: If you injected a 100mL syringe volume of a gas mix into the injection glassware, and it
takes you 10 seconds to inject the entire 100mL volume, the flow rate going through the adsorbent
tube for those 10 seconds is 600mL/min. That is too fast. There is a good chance that the
compounds in the gas mix could breakthrough the adsorbent and not be retained. In general, the
delivery rate of the syringe should be less than 10mL/min.
10

Injecting Liquid Phase Standards
Liquid standards are typically made up in a solvent such as: methanol, hexane, or carbon
disulfide.
• Set the block temperature 10°C above the boiling point of the solvent.
• Set the flow rate of the carrier gas to 50mL/min
• Inject the liquid standard, then allow the tube to remain attached for 2 to 5 minutes with the
carrier gas flowing through the tube.
Note: If the compounds in your liquid standard are larger than C12, set the block heater
temperature to 100°C
Unlike gas standards there is a limit to how much
volume you can inject. If too much liquid is
injected, the liquid could flashback into the
plumbing of the system. Injection volumes of
liquid standards should be less than 25 microliters.
The injection rate for liquid standards can be
quicker than that of gas standards, since the
displacement of microliter volumes will have
minimum impact on the flow rate going through
the adsorbent tube. It is important to allow
enough time for the contents of the liquid standard
to be transferred to the tube before removing the
adsorbent tube. For example; if the flow rate of the carrier gas is set to 50mL/min, after two
minutes the injection glassware will be purged 10 times. Ten exchanges are typically enough to
completely purge all of the compounds from the injection glassware.
Dry Purging
When samples are taken in humid atmospheres, the adsorbent tube may retain some moisture.
Moisture retention is dependent on the type of adsorbent used. In general, it is best to remove this
moisture prior to analysis. The ATIS simplifies this task. Make sure the injection glassware is
clean prior to performing this task.
• Set the block temperature to 50°C
• Set the flow rate of the carrier gas to 50mL/min.
• The total time/volume to leave the adsorbent tube attached will need to be determined
experimentally.
The key to dry purging is to allow enough (dry) gas to pass through the tube and carry the
moisture away. Too much dry purge can actually start to push the compounds of interest through
the adsorbent (Breakthrough). Dry purge is usually referenced as a volume instead of time.
Recommended Dry Purge Volumes for a few adsorbents
Tenax®and Carbotrap™adsorbents usually need only 0.25 liters of dry purge (five minutes at
50mL/min).
Carbosieve®and Carboxen™adsorbents may need anywhere from 0.5 liters to 3 liters, depending
on how much moisture was in the air during sampling.
Gas-Sampling Bags
The ATIS can be used to vaporize neat compound(s) prior to injecting them into the bag. Gas
sampling bags are often used to make calibration standards. Simply install the gas-sampling bag
in the thumbwheel, and the process of filling the bag and injecting the compound can happen at
the same time. By volatizing the compounds with the ATIS, this allows you to use the contents of
the bag immediately without having to let it equilibrate overnight.
11

Using the Extraction Glassware (ordered separately)
Two different styles of extraction glassware are
available, one with a micro connector, the other with a
ground joint. The style you choose is your preference;
both have their advantages. Both styles include a glass
frit at the outlet and a replaceable septum. The
replaceable septum allows a surrogate or internal
standard to be injected during the extraction process.
The glass frit helps to prevent solids from passing
through to the adsorbent tube during the extraction. Solid samples up to 13mm (1/2”) OD and
76mm (3”) in length can be inserted into both styles.
To use the extraction glassware, simply remove the injection glassware from the injector heater
block and reinstall the fittings as described in the assembly instruction of injection glassware. See
Appendix 2: ATIS Glassware Configurations for additional information.
This picture illustrates an example of a flower being placed
into the extraction glassware.
In most extraction applications, the block
temperature is set to the desired temperature,
and the sample is placed in the glassware
using a probe or tweezers. Caution: Be
extremely careful when working with high
temperatures. Whether you choose to have the
adsorbent tube attached to the thumbwheel
before, or after you place the sample in the
extraction glassware is up to you. By attaching
the adsorbent after the sample is placed in the
glassware will allow some of the volatiles to exit
before being collected by the adsorbent tube.
In some applications this may be desirable.
Replacement Parts
Consumable Parts:
28526-U Injection Glassware
20668 Half-Hole Septa (6mm OD 9mm Long) pk/100
22445-U Gas Purifier
22393 6mm Ferrules M-2A “Black” pk/10
22320-U 1/4” Ferrules M-2 “Orange” pk/10
29024-U 1/4” Ferrules Teflon “White” pk/10
See the SUPELCO catalog for additional ferrules
Replacement Parts:
33315 Block Heater 110 VAC
33318-U Block Heater 220 VAC
Z423394-1EA Thermometer
28529-U ATIS Thumbwheel (Does not include the union)
28525-U Luer/Hose-Barb Adapter
12

Accessories:
28522-U Purge & Trap/Humidifier Module
28523-U Extraction Glassware with Micro Connector
64700-U Replacement 20/20 Micro Connectors pk/6
28524-U Extraction Glassware with 19/22 Ground Joint
64763 Replacement Clip for 19/22 Ground Joint pk/5
64762 Teflon Sleeves for 19/22 Ground Joint (used to create a tighter seal)
22754-U Digital Timer
22435-U Stainless Steel Tweezers
Troubleshooting
Below are some general troubleshooting tips. For expert answers to your questions about the
ATIS, contact our Technical Service Department. (Phone 800-359-3041 or 814-359-3041, Fax
800-359-3044 or 814-359-5468), E-mail techservice@sial.com.
No flow is observed at the outlet of the adsorbent tube.
• Check to see if there is a blockage in the adsorbent tube.
• Check that gas delivery pressure is set to 25-30psig
• Make sure the on/off valve is in the ON position.
• Check for leaks; Check all of the gas connections for leaks.
• Replace the septum.
• Make sure the correct size ferrule is installed in the thumbwheel for the size of tube you are
using.
• The pressure relief valve is open. If the backpressure of the adsorbent tube is greater than 6
psig the pressure relief valve will open, thus all or most of flow will exit the relief valve.
The flow controller will not maintain the set flow rate:
• Check that the gas delivery pressure is set at 25-30psig.
• There is an extreme difference in the backpressure of the adsorbent tube you are using than
with the one you used to calibrate the flow rate. You should check the flow rate when using
different types of tubes.
• The backpressure of the adsorbent tube exceeds 6psig and the pressure relief valve has
opened.
I do not see any peaks after injecting my calibration mix.
• The gas flow may have been OFF during the injection.
• The temperature of the block heater is too low or it was off.
• There may be leak. Replace the septum; Check the injection glassware for breakage.
• Check the pressure relief valve.
I am seeing extraneous peaks in my analyses.
• Check the glassware, it may be contaminated.
• Gas purifier may need to be changed.
• May need to install additional gas purification.
• The calibration standard or sample may have flashed back into the pneumatics of the ATIS. This may
require you to purchase a new flow controller, on/off valve, and to replace the tubing. Contact our
Technical Service for assistance in ordering replacement parts.
13

Questions & Answers
Q: Can I leave the unit turned ON for an extended period?
A: Yes, leaving the temperature of the block heater on, and a flow of gas going through the
injection glassware will help to keep the glassware clean and ready for use. To save electrical
energy: Utilize the feature of the two temperature ranges of the block heater: Set the low
range temperature to around 50°C, and use the high range setting for the temperature you are
using to inject your standards. When you are not using the ATIS for an extended period of
time, just simply set the power switch to the low range.
Q: Can I use PTFE ferrules instead of the VESPEL ferrules to connect the ATIS glassware?
A: Yes, You can use PTFE (TEFLON®) ferrules instead of VESPEL ferrules. The VESPEL
ferrules were chosen for their inert properties and high reusability. All of the ATIS glassware
uses 1/4” OD connections. However if you prefer PTFE ferrules we have included a pack of 3
with the unit.
Q: How can I be sure I completely transferred my calibration standard?
A: This may require you to do some experimentation by making several analytical runs of your
spiked adsorbent tubes. Change the amount of time you leave the tube attached after making
an injection from 2 to 5 minutes. Also, experiment by raising the temperature of the injection
block. Compare the results. In most cases you will not need to change the flow rate from
50mL/min.
Q: How can I determine if my adsorbent tube is dry? (Dry Purge)
A: The best way is to weigh the adsorbent tube and record its tare weight prior to sampling.
Then after sampling, reweigh the tube. If the difference is greater than 2 milligrams you may
want to remove the moisture prior to analysis. Attach the tube to the ATIS and allow 0.5 to 1
Liter of the carrier gas to pass through the tube. Reweigh the tube. The adsorbent tube can
be considered dry when it is within 1.0mg of the weight of the tube prior to sampling.
Q: Why does the pressure relief valve open at 6psig?
A: The pressure relief valve is designed to relieve the pressure if an adsorbent tube with a
complete blockage was accidentally installed. The relief valve prevents pressure from building
up in ATIS glassware. The pressure relief valve will reseal itself, once the pressure falls below
6psig.
Q: Does the white polypropylene tubing off-gas volatiles?
A: No, this tubing does not off-gas any volatiles at ambient temperatures. However if the
tubing is heated, some volatiles should be expected. The maximum temperature of this tubing
is 93°C (200°F).
Trademarks:
ATIS – Sigma-Aldrich
Carbotrap – Sigma-Aldrich
Carboxen – Sigma-Aldrich
TEFLON, VESPEL – E.I du Pont de Nemours & Co., Inc.
Tenax – Enka Research Institute Arhem
TALBOYS – Henry Troemner LLC
14

Appendix 1: ATIS Pneumatic Diagram
Copper Tubing
Copper
Tubing
White Polypropylene Tubing
Flow Element
Flow
Controller
IN
Out
Gas Purifier
Orange 1/4" M-2 Vespel Ferrule
Suggested operating pressure to 25-30psig
Maximum operating pressure 60psig
(Regulator not included)
Half-Hole
Septa
SS Reducing
Union
1/4" to 1/8"
Regulator
SS Union
1/4" to 1/4" with Thumbwheel
Relief Valve
15

Appendix 2: ATIS Glassware Configurations
White Polypropylene
Tubing
White Polypropylene
Tubing
White Polypropylene
Tubing
SS Reducing Union
1/4" to 1/8"
SS Reducing
Union
1/4" to 1/8"
SS Reducing
Union
1/4" to 1/8"
3
2
1
Orange 1/4" M-2 Vespel Ferrule
SS Union
1/4" to 1/4" with Thumbwheel
SS Union
1/4" to 1/4" with Thumbwheel
SS Union
1/4" to 1/4" with Thumbwheel
1. Standard Injection Glassware Cat# 28526-U
2. Extraction Glassware with Ground Joint Cat# 28524-U
3. Extraction Glassware with Micro Connector Cat# 28523-U
Orange 1/4" M-2 Vespel Ferrule
Orange 1/4" M-2 Vespel Ferrule
16
Table of contents
Other Supelco Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

SciCan
SciCan statim 2000 G4 Operator's manual

Thermo Scientific
Thermo Scientific thermoscientific Vanquish UHPLC System operating manual

IKA
IKA ULTRA -TURRAX Tube Drive P control operating instructions

HACH LANGE
HACH LANGE BUHLER 3010 Short manual

VOLTCRAFT
VOLTCRAFT LSP-1165 operating instructions

Reverberi
Reverberi STERILIX 2 PLUS Operation and maintenance manual

Claind
Claind Hydro200 user manual

SPEX SamplePrep
SPEX SamplePrep Genomax 2050 Operation manual

Hakko Electronics
Hakko Electronics FT-8003 instruction manual

Thermo Scientific
Thermo Scientific 5030i SHARP instruction manual

EMP-Centauri
EMP-Centauri P.123-A instruction manual

Safe Fleet
Safe Fleet DH6 Quick installation guide