Installation
WARNING
When installing a Swagelok®self-venting regulator, position the vent connection or line away
from operating personnel. Operating personnel must protect themselves from exposure to system
fluids.
CAUTION
Do not use the regulators as a shutoff device.
The preferred mounting position of the regulator is horizontal with the dome facing upwards.
If grounding is required, connect a ground wire under a dome bolt.
Filling the dome
The dome can be filled using manual or electronic loading.
Manual loading:
This can be done by taking the gas pressure from the system and feeding this through a spring
loaded pilot regulator into the dome. This is shown in sketch A.
In liquid systems the gas pressure for manual dome loading can be taken from a cylinder or mains.
This is shown in sketch B.
Substituting the spring loaded pilot regulator for a proportional control valve and a pressure sensor,
allows you to control the pressure electronically.
Electronic loading:
This can be done by taking the gas pressure from the system and feeding this through a proportional
control valve into the dome. This is shown in sketch C.
This method can be used for low and medium-pressure systems.
This can be done by taking the gas pressure from the system and feeding this through a proportional
controlled ratio regulator into the dome. This is shown in sketch D.
This method can be used for high-pressure systems.
The best results will be achieved by allowing a small flow to continuously pass through the pilot regulator.
This flow can either be vented or, in gas systems, fed back through an orifice into the downstream piping.
This is usually referred to as “dynamic regulation”.
It is not recommended to place a gauge on the dome to set or check the outlet pressure.
Because of forces in the regulator, the dome pressure will always be higher than the outlet
pressure. Place a gauge in the outlet line to set or check the outlet pressure.