Synetica enLink Zone User manual

Quick Start Guide
enLink Zone
Wireless environmental sensor
•LoRa long range wireless
•Battery powered
•Built in sensors for:
•Temperature (°C)
•Humidity (%RH)
•Carbon Dioxide (CO2 )*
•Motion (PIR)*
•Barometric Pressure (Pa)*
•Light level (lux)*
•Volatile Organic Compounds (VOC’s)*
Temperature
Humidity
Light level
VOC’s
CO2
Pressure
Motion
enLink Zone accurately measures multiple environmental
parameters including room temperature, humidity, light level,
VOC’s, CO2, barometric pressure and human presence (motion)*.
Volatile Organic Compounds (VOC’s) from paints (such as
formaldehyde), lacquers, paint strippers, cleaning supplies,
furnishings, office equipment, glues, adhesives and alcohol can be
detected.
Readings are transmitted to the cloud using long range LoRa
wireless, where the data can be displayed and analysed.
The unit is battery powered with long life of 3+ years#.
A built in USB port allows all parameters including sensor, air
quality data, wireless signal strength and wireless network
configuration to be viewed and set using simple menus via any
USB enabled host such as a PC or Mac.
Features
•
Multiple sensor options*
•
LoRa long range wireless
•
Frequency Range 863-870MHz*
•
Frequency Range 902-928MHz*
•
Up to +16dBm Tx Power
•
Built in USB port for power and
configuration
•
Battery or mains powered
•
CE compliant
•
RoHS compliant
•
Made in the UK
*Model dependent, see Selection Guide section
#Dependent on factors such as transmission interval

SYNETICA enLink Zone
Quick Start Guide Rev 1.4
Introduction
LoRaWAN devices are configured using OTAA (Over-the-Air-Activation).
OTAA is the most secure way to connect a device to the LoRaWAN network. In OTAA, the device performs a Join-procedure with
the network, during which a dynamic DevAddr (device address) is assigned and security keys are negotiated with the device.
The OTAA configuration requires the following parameters to be correctly set:
•DevEUI: End-device Identifier. It is unique for every device and is set at device manufacture.
•AppEUI / JoinEUI*: Application Identifier. Used to identify the end application.
•AppKey: Application key. Used to create the session keys.
*Note: In LoRaWAN 1.1, AppEUI was renamed to JoinEUI.
For many applications Synetica can supply enLink Zone units with the AppEUI and AppKey pre-configured, so providing the LoRa
gateway has the matching values, the join process will happen automatically once the Zone unit is in wireless range and switched
on.
The DevEUI is always set at device manufacture and is unique. The device AppEUI and AppKey can easily be set via the USB
connection if required and the process is detailed later in this document.

SYNETICA enLink Zone
Quick Start Guide Rev 1.4
Battery Installation / Replacement
enLink devices use SAFT LS14500 AA size 3.6 Volt Lithium Thionyl Chloride (Li-SOCl2) batteries (non-rechargeable) or direct
equivalent.
No other batteries are approved for use in the device.
Lithium Thionyl Chloride batteries have very high energy capacity and must be used and handled with care observing the guidance
below.
WARNING
Risk of death or serious injury from explosion or fire.
•Keep out of sight and reach of children.
•Fire, explosion and burn hazard - do not recharge, short circuit, crush, disassemble, incinerate
•Due to the high terminal voltage (3.6V), they are not suitable as direct replacements for other battery technologies in the same can sizes.
•When not in use the Batteries must be stored in a non-Hazardous Area.
•Do not change batteries in an explosive gas atmosphere.
•When installing batteries, do not snag the battery terminal on the clip or the battery may be damaged. Do not apply excessive force.
•Do not drop. Dropping the battery may cause damage. If a battery is dropped, do not install the dropped battery into the unit. Dispose of dropped battery
promptly per local regulations or per the battery manufacturer’s recommendations.
Guidance
•Always install the batteries correctly as per instructions taking great care to observe the battery polarity.
•Ensure that the contact points are clean and conductive.
•All batteries must be the same model from the same manufacturer.
•Do not mix old and new batteries or batteries from different manufacturers.
•Do not heat or attempt to recharge the battery.
•Do not dispose of in a fire.
•Only install approved batteries: SAFT LS14500 Lithium Thionyl Chloride AA Battery 3.6 Volt, or direct equivalent.
Safe disposal
•Please recycle responsibly, a wide range of schemes are available.
•Do not dispose of in normal waste or in a fire.

SYNETICA enLink Zone
Quick Start Guide Rev 1.4
To get started with enLink Zone wireless sensor, follow these steps:
1. Setup using Multitech Conduit® AEP
MultiTech’s Conduit is a leading configurable, manageable, and scalable LoRa gateway and can be quickly and easily configured to
create a LoRaWAN network with enLink LoRa devices.
To enable LoRa devices to connect to the Conduit the AppEUI and AppKey must match those of the enLink devices.
For Multitech Conduit units running V1.6.4/1.7.0, follow the process below. Detailed instructions can be found on the Multitech
Web Site (www.multitech.net).
Logon to the Conduit and select LoRaWAN -> Network Settings
Set the Mode to Network Server, then set the Channel Plan required and set the Mask to 00FF as shown below.
Next, select the Key Management menu and set the AppEUI and AppKey. Ensure that these match the enLink devices you
want to join the LoRa network.

SYNETICA enLink Zone
Quick Start Guide Rev 1.4
Once the keys are set, select Submit. The Conduit will prompt to Save and Restart. The settings will not take effect until
the Conduit has restarted. Once the unit has restarted you are ready to join enLink devices to the network.
The Conduit gateway is now ready to receive devices onto the network. On the Conduit, select LoRaWAN -> Devices as
shown below. This screen shows details of devices which have joined the network, when they joined and when the last
message was received from them. As you add devices to the network, this is where they will appear.

SYNETICA enLink Zone
Quick Start Guide Rev 1.4

SYNETICA enLink Zone
Quick Start Guide Rev 1.4
2. Join enLink devices to the LoRaWAN network
enLink devices in wireless range and with the correct AppEUI and AppKey settings set will automatically join the
LoRaWAN network when they are first powered up.
enLink Zone Unit and Label
The unique DevEUI is printed on all enLink devices and is also present in the QR code. The DevEUI can be used to identify
the device once joined to the network.
To power the device ON, remove the rear cover of the enLink Zone by inserting a small screwdriver blade into the slots on
the bottom of the unit as shown below.
Cover Removal

SYNETICA enLink Zone
Quick Start Guide Rev 1.4
Locate the power switch and using small screwdriver gently slide the power switch towards the ON position marked on
the unit as shown.
Switch to the ON position
Once powered ON, the enLink device will send a Join request message to the Conduit. The Status LED will blink RED as
shown below whilst the Join process is taking place. Depending on factors such as signal strength, RF interference etc. the
join process may take several seconds to complete.
Blinking Red LED –Attempting Join
When the device has successfully joined the network the Status LED will blink GREEN five times to show that the join has
been completed. The LED’s will then switch off to conserve the batteries.
Blinking Green LED –Device Successfully Joined
Devices which have Joined the network appear in the Conduit LoRaWAN - > Devices menu as shown previously.
ON
OFF

SYNETICA enLink Zone
Quick Start Guide Rev 1.4
3. Setting / changing the enLink LoRaWAN keys
For many applications, Synetica can supply enLink Zone units with the LoRaWAN AppEUI and AppKey parameters pre-configured
to your requirements, whereby if the LoRa gateway has matching keys the join process will happen automatically once the Zone
unit is in wireless range and switched on.
The DevEUI is always set at device manufacture and is unique. The device AppEUI and AppKey can easily be set via the USB
connection as detailed below.
Remove the enLink Zone board from its enclosure as shown previously. Take care when removing the board from the
front part of the enclosure and only use the slots provided at the bottom of the board to gently lift the board up and out
of the enclosure.
Connect a micro USB cable to the enLink unit. The device will attach to a COM port on your PC.
Using a terminal program (e.g. Tera Term https://ttssh2.osdn.jp/) connect to the COM port used by the enLink device.
To verify which COM port is being used, check the Windows™ Device Manager (In Windows - Click the Start button, type
device manager into the search box and tap Device Manager on the menu.) Expand the Ports (Com & LPT) menu as
shown below.
In your terminal program press the Enter key. An enLink summary screen will appear as shown below. The default
password is the last four digits of the displayed DevEUI, in the screen below this is 668b.
enLink logon screen
The screen below will show with the enLink Main Menu options. Enter Qto enter the Quick Start Menu.
enLink Main Menu
The Quick Start Menu contains only the parameters that normally need to be configured to setup the device and join the
LoRa network.

SYNETICA enLink Zone
Quick Start Guide Rev 1.4
From the Quick Start Menu you can change the AppEUI and AppKey.
Quick Start Settings Menu
From the Quick Start Settings Menu, access the AppEUI setting by entering e. Enter the 16 character AppEUI using
numbers and letters a to f. Do not include spaces or any other characters. Pressing Swill enter the default AppEUI which
you can then edit. Press Enter when the key is correctly entered to return to the Quick Start Settings Menu.
AppEUI setting
From the Quick Start Settings Menu access the AppKey setting by entering k. Enter the 32 character Appkey using
numbers and letters a to f. Do not include spaces or any other characters. Pressing Swill enter the default AppKey which
you can then edit. Press Enter when the key is correctly entered to return to the Quick Start Settings Menu.
AppKey setting
Press Xfrom the Quick Start Settings Menu to return to the enLink Main menu.
The header will show ** Reboot Required ** as shown below. The new key settings will not take effect until the enLink
device is restarted. Enter Rto reboot followed by OK. The device will restart with the entered AppEUI and AppKey and
attempt to join the LoRa network.
Reboot Required notification
Check the Conduit LoRaWAN -> Devices menu detailed previously to verify that the enLink device has joined successfully.

SYNETICA enLink Zone
Quick Start Guide Rev 1.4
Live Menu
enLink Zone incorporates a live data screen which shows all readings and device status for easy data validation. To enter
the Live status screen, from the Main Menu enter c for Configure Device followed by dfor Live readings display. A screen
similar to the one below will show. The sensors will vary according to the enLink Zone model and the installed sensors.
Live Display
Configuration Menu
The enLink Zone configuration menu allows you to view current sensor readings and also to change various functions of
their behaviour such as calibration data. To enter the Configure Device menu press cfrom the main menu. A screen
similar to the one below will show. The exact parameters shown will vary according to the Zone model and sensors fitted.
CO2Sensor Auto Calibration Configuration
To view and set CO2sensor calibration information, enter c and the screen below will show.

SYNETICA enLink Zone
Quick Start Guide Rev 1.4
Please see the table below for information on each menu item.
Menu Item
Description / details
Last/Minimum Reading
Shows the last CO2value read and the minimum CO2value read since the last auto calibration.
Next Auto-Cal due
Shows when the next autocalibration routine will occur
Last Auto-Cal result
Shows the value of the last auto calibration result. Used internally by the sensor.
Calibration Success
This shows the total number of successful auto calibrations since the device was powered up.
Out-of-bounds Ignored
Shows the number of times that auto calibration did not run due to the Out Of Bounds setting.
E - Enable/Disable Auto-Cal
Enables or disables the auto calibration routine.
T - Set Target CO2 Level
This is the known CO2corresponding to the minimum value the sensor has read since power-up or
last calibration. It is normally 'fresh air' or the lowest level when the building is unoccupied
overnight or at weekends. Typically this is 400 ~ 450 ppm.
K - Set to Known CO2 Level
This will re-calibrate the zero point of the sensor to a known gas concentration. The sensor should
be placed in this gas concentration and allowed to stabilise.
This command runs in the background and will take 8 to 10 seconds to complete.
As an example, fresh air is typically around 400 ~ 450 ppm.
O - Out-of-bounds check
The Out-of-bounds value is used to ignore the calibration if the minimum value the sensor has read
is not within a sensible range of the target concentration level.
So, if the target concentration level is 400, the Out-of-bounds value is ±50 and the minimum
reading is 451 (or more), the calibration routine is ignored.
I - Initial Interval
It is possible for the first auto-calibration to take place more quickly than the regular auto-
calibration events. This can be useful to stabilise the readings quickly after installation.
R - Regular Interval
This is the standard calibration interval, it is set to 8 days by default to accommodate a week long
period where the minimum sensed CO2level should have fallen to background levels.
The CO2sensor needs to be exposed to fresh, clean air periodically for the auto calibration to be successful. Most
occupied areas are unoccupied for some time during a week-long period, typically at night, or at the weekend and
therefore the auto calibration runs every 8 days by default. Background CO2levels are typically around 400-450 ppm, if
the background CO2level is known to be a different value then this can be set in the “Set Target CO2Level” parameter
If a unit is placed in an area where the CO2level may not fall below a certain level, e.g. 450ppm, during the calibration
interval then the “Out-of-bounds check” parameter can be set so that the auto calibration routine does not run. As an
example, if an area is continuously occupied for a long period and the minimum CO2reading does not fall below, say
450ppm, then it is undesirable to run the autocalibration routine based on a target of 400ppm. In this case, if the “Set
Target CO2 Level” is set to 400ppm and the “Out-of-bounds check” value is set to +/-50 ppm then the autocalibration
routine will not run unless the minimum read value falls below 451ppm in the interval.
Technical Support
For technical assistance, please visit the downloads section of our web site at www.synetica.net or email us at
support@synetica.net
Table of contents
Other Synetica Accessories manuals
Popular Accessories manuals by other brands

Burnham
Burnham Aquastat IQ Option Card instruction sheet

Gemini
Gemini 7709 Installation and maintenance manual

STEINEL
STEINEL SensorSchalter IR 180 UP operating instructions

Clas Ohlson
Clas Ohlson WA PUSH WHITE manual

Black & Decker
Black & Decker Slice Right EK700 Series Use and care book

ITW
ITW VORTEC PANEL GUARD 701 Operation & Safety Instructions

STEINEL PROFESSIONAL
STEINEL PROFESSIONAL IR 2180 UP ECO Information manual

Anker
Anker eufy Security T8740 user manual

B&C Speakers
B&C Speakers HF Drivers DE1050 Specification sheet

Red Arrow Trading
Red Arrow Trading PIR-B1 instructions

Forma Scientific
Forma Scientific 3920 Operating and maintenance manual

Arenti
Arenti VBELL 1 quick guide