manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Synthes
  6. •
  7. Medical Equipment
  8. •
  9. Synthes Small Battery Drive II User manual

Synthes Small Battery Drive II User manual

Manual and Mechanical Cleaning.
Instructions for the Small Battery Drive II.
Cleaning Chart
© 2013 Synthes, Inc. or its affiliates. All rights reserved. Synthes is a trademark of Synthes, Inc. or its affiliates. Printed in U.S.A. 2/13 J12030-A
www.synthes.com
Synthes USA
1302 Wrights Lane East
West Chester, PA 19380
Telephone: (610) 719-5000
To order: (800) 327-6887
Fax: (800) 327-6661
E-mail: [email protected]
Synthes (Canada) Ltd.
2566 Meadowpine Boulevard
Mississauga, Ontario L5N 6P9
Telephone: (905) 567-0440
To order: (800) 668-1119
Fax: (905) 567-3185
9 Inspect the device.
Remove all the devices from the washing basket. Inspect
the cannulations, coupling sleeves, etc. for visible soil. If
necessary, repeat the manual pre-clean/automated cleaning
cycle. Confirm that all parts are completely dry.
Mechanical cleaning/disinfection is an additional stress
for power equipment, especially for seals and bearings.
Therefore, systems must be properly lubricated and
regularly sent to be serviced (at least once per year).
Step Duration Cleaning instructions
(minimum)
Rinse 2 minutes Cold tap water
Pre-wash 1 minute Warm water (≥ 40 °C); use detergent
Cleaning 2 minutes Warm water (≥ 45 °C); use detergent
Rinse 5 minutes Rinse with de-ionized (DI) or purified water (PURW)
Thermal disinfection 5 minutes Hot DI water, ≥ 93 °C
Dry 40 minutes ≥ 90 °C
8 Automated cleaning cycle parameters
Note: The washer/disinfector should fulfill the requirements as
specified in ISO 15883.
Manual and Mechanical Cleaning.
Instructions for the Synthes Small Battery Drive II.
1 Remove debris.
Rinse the device under running cold tap water for a
minimum of 2 minutes. Use a sponge, soft lint-free cloth
or soft-bristled brush to assist in removing gross soil.
For cannulations of the handpiece and attachments, the
cleaning brush (519.40) should be used.
Note: Do not use pointed objects for cleaning.
2 Manipulate moving parts.
Manipulate all moving parts such as the triggers, sleeves
and switches under running tap water to loosen and
remove gross debris.
Manual Cleaning Instructions
Mechanical/Automated Cleaning Instruction with Manual Pre-cleaning
Important:
– Manual pre-cleaning prior to mechanical/automated
cleaning/disinfection is important to ensure that
cannulations and other difficult to access areas are
clean.
– Alternative cleaning/disinfection procedures other
than in the procedure described below (including
manual precleaning) have not been validated by
Synthes.
3 Spray and wipe.
Spray and wipe the device using a neutral pH enzymatic
solution for a minimum of 2 minutes.
Follow the enzymatic detergent manufacturer’s
directions for correct temperature, water quality
(i.e. pH, hardness) and concentration/dilution.
4 Clean with detergent.
Clean the device manually under running warm water
using an enzymatic cleaner or detergent for a minimum
of 5 minutes. Manipulate all moving parts under running
water. Use a soft-bristled brush and/or soft lint-free cloth
to remove all visible soil and debris.
Follow the enzymatic cleaner or detergent
manufacturer’s instructions for use for correct
temperature, water quality and concentration/dilution.
5 Rinse with tap water.
Rinse the device thoroughly using cool to lukewarm
running water for a minimum of 2 minutes. Use a
syringe, pipette or water jet to flush lumens and
channels. Actuate joints, handles and other movable
device features in order to rinse thoroughly under
running water.
6 Visually inspect device.
Inspect the cannulations, coupling sleeves, etc. for visible
soil. Repeat steps 1 – 6 until no visible soil remains.
7 Final rinse with de-ionized/
purified water.
Final rinse with de-ionized or purified water for a
minimum of 2 minutes.
8 Dry.
Dry device using a soft lint-free cloth or clean
compressed air.
1 Remove debris.
Rinse the device under running cold tap water for a
minimum of 2 minutes. Use a sponge, soft lint-free
cloth or soft-bristled brush to assist in removing gross
soil. For cannulations of the handpiece and attachments,
the cleaning brush (519.40) should be used.
Note: Do not use pointed objects for cleaning.
2 Manipulate moving parts.
Manipulate all moving parts such as the triggers, sleeves
and switches under running tap water to loosen and
remove gross debris.
3 Spray and wipe.
Spray and wipe the device using a neutral pH enzymatic
solution for a minimum of 2 minutes. Follow the
enzymatic detergent manufacturer’s directions for
correct temperature, water quality (i.e. pH, hardness)
and concentration/dilution.
4 Clean with detergent.
Clean the device manually under running warm water
using an enzymatic cleaner or detergent for a minimum
of 5 minutes. Manipulate all moving parts under running
water. Use a soft-bristled brush and/or soft lint-free cloth
to remove all visible soil and debris.
Follow the enzymatic cleaner or detergent manu-
facturer’s instructions for use for correct temperature,
water quality and concentration/dilution.
5 Rinse with tap water.
Rinse the device thoroughly using cool to lukewarm
running water for a minimum of 2 minutes. Use a
syringe, pipette or water jet to flush lumens and
channels. Actuate joints, handles and other movable
device features in order to rinse thoroughly under
running water.
6 Visually inspect device.
Inspect the cannulations, coupling sleeves, etc. for visible
soil. Repeat steps 1 – 6 until no visible soil remains.
7 Load washing basket.
Place devices in the specially designed tray for machine
washing supplied by Synthes (68.001.610). Ensure
that all cannulations (handpiece and attachments), if
applicable, are positioned vertically, i.e. in an upright
position as shown.
Note: Specific lid (68.001.602) is available for the
washing basket.
Manual and Mechanical Cleaning.
Instructions for the Small Battery Drive II.
Cleaning Chart
© 2013 Synthes, Inc. or its affiliates. All rights reserved. Synthes is a trademark of Synthes, Inc. or its affiliates. Printed in U.S.A. 2/13 J12030-A
www.synthes.com
Synthes USA
1302 Wrights Lane East
West Chester, PA 19380
Telephone: (610) 719-5000
To order: (800) 327-6887
Fax: (800) 327-6661
E-mail: [email protected]
Synthes (Canada) Ltd.
2566 Meadowpine Boulevard
Mississauga, Ontario L5N 6P9
Telephone: (905) 567-0440
To order: (800) 668-1119
Fax: (905) 567-3185
9 Inspect the device.
Remove all the devices from the washing basket. Inspect
the cannulations, coupling sleeves, etc. for visible soil. If
necessary, repeat the manual pre-clean/automated cleaning
cycle. Confirm that all parts are completely dry.
Mechanical cleaning/disinfection is an additional stress
for power equipment, especially for seals and bearings.
Therefore, systems must be properly lubricated and
regularly sent to be serviced (at least once per year).
Step Duration Cleaning instructions
(minimum)
Rinse 2 minutes Cold tap water
Pre-wash 1 minute Warm water (≥ 40 °C); use detergent
Cleaning 2 minutes Warm water (≥ 45 °C); use detergent
Rinse 5 minutes Rinse with de-ionized (DI) or purified water (PURW)
Thermal disinfection 5 minutes Hot DI water, ≥ 93 °C
Dry 40 minutes ≥ 90 °C
8 Automated cleaning cycle parameters
Note: The washer/disinfector should fulfill the requirements as
specified in ISO 15883.
Manual and Mechanical Cleaning.
Instructions for the Small Battery Drive II.
Cleaning Chart
© 2013 Synthes, Inc. or its affiliates. All rights reserved. Synthes is a trademark of Synthes, Inc. or its affiliates. Printed in U.S.A. 2/13 J12030-A
www.synthes.com
Synthes USA
1302 Wrights Lane East
West Chester, PA 19380
Telephone: (610) 719-5000
To order: (800) 327-6887
Fax: (800) 327-6661
E-mail: [email protected]
Synthes (Canada) Ltd.
2566 Meadowpine Boulevard
Mississauga, Ontario L5N 6P9
Telephone: (905) 567-0440
To order: (800) 668-1119
Fax: (905) 567-3185
9 Inspect the device.
Remove all the devices from the washing basket. Inspect
the cannulations, coupling sleeves, etc. for visible soil. If
necessary, repeat the manual pre-clean/automated cleaning
cycle. Confirm that all parts are completely dry.
Mechanical cleaning/disinfection is an additional stress
for power equipment, especially for seals and bearings.
Therefore, systems must be properly lubricated and
regularly sent to be serviced (at least once per year).
Step Duration Cleaning instructions
(minimum)
Rinse 2 minutes Cold tap water
Pre-wash 1 minute Warm water (≥ 40 °C); use detergent
Cleaning 2 minutes Warm water (≥ 45 °C); use detergent
Rinse 5 minutes Rinse with de-ionized (DI) or purified water (PURW)
Thermal disinfection 5 minutes Hot DI water, ≥ 93 °C
Dry 40 minutes ≥ 90 °C
8 Automated cleaning cycle parameters
Note: The washer/disinfector should fulfill the requirements as
specified in ISO 15883.

Other Synthes Medical Equipment manuals

Synthes MAXFRAME AUTOSTRUT User manual

Synthes

Synthes MAXFRAME AUTOSTRUT User manual

Synthes DHS User manual

Synthes

Synthes DHS User manual

Synthes MatrixRIB User manual

Synthes

Synthes MatrixRIB User manual

Synthes Hybrid Fixator Proximal Tibia Frame User manual

Synthes

Synthes Hybrid Fixator Proximal Tibia Frame User manual

Synthes ComPact Air Drive II User manual

Synthes

Synthes ComPact Air Drive II User manual

Synthes TRS Configuration guide

Synthes

Synthes TRS Configuration guide

Synthes Dorsal User manual

Synthes

Synthes Dorsal User manual

Synthes Pelvic C-Clamp User manual

Synthes

Synthes Pelvic C-Clamp User manual

Synthes Codman Neuro DirectLink User manual

Synthes

Synthes Codman Neuro DirectLink User manual

Synthes USS Universal User manual

Synthes

Synthes USS Universal User manual

Synthes SynFrame Light-transmitting Rod 387.362 User manual

Synthes

Synthes SynFrame Light-transmitting Rod 387.362 User manual

Synthes USS II User manual

Synthes

Synthes USS II User manual

Synthes Electric pen drive Configuration guide

Synthes

Synthes Electric pen drive Configuration guide

Synthes Synex System User manual

Synthes

Synthes Synex System User manual

Synthes USS II User manual

Synthes

Synthes USS II User manual

Synthes TFN User manual

Synthes

Synthes TFN User manual

Synthes Air Power Line II User manual

Synthes

Synthes Air Power Line II User manual

Synthes SYNMESH User manual

Synthes

Synthes SYNMESH User manual

Synthes TRS User manual

Synthes

Synthes TRS User manual

Synthes Pelvic C-Clamp II User manual

Synthes

Synthes Pelvic C-Clamp II User manual

Synthes LCP User manual

Synthes

Synthes LCP User manual

Synthes CODMAN NEURO CODMAN CERTAS Plus User manual

Synthes

Synthes CODMAN NEURO CODMAN CERTAS Plus User manual

Synthes The Titanium Femoral Nail System User manual

Synthes

Synthes The Titanium Femoral Nail System User manual

Synthes ACTIS DUOFIX User manual

Synthes

Synthes ACTIS DUOFIX User manual

Popular Medical Equipment manuals by other brands

Itamar Medical WatchPAT Operation manual

Itamar Medical

Itamar Medical WatchPAT Operation manual

PROTEOR 1P650 KNEE Assembling and adjustment instructions

PROTEOR

PROTEOR 1P650 KNEE Assembling and adjustment instructions

Liberate Medical VentFree user manual

Liberate Medical

Liberate Medical VentFree user manual

Dicarre DA20 user guide

Dicarre

Dicarre DA20 user guide

Stryker Medical M Series Operation manual

Stryker Medical

Stryker Medical M Series Operation manual

Invacare ScanBed 750 user manual

Invacare

Invacare ScanBed 750 user manual

Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Getinge

Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Mettler Electronics Sonicator 730 Maintenance manual

Mettler Electronics

Mettler Electronics Sonicator 730 Maintenance manual

Pressalit Care R1100 Mounting instruction

Pressalit Care

Pressalit Care R1100 Mounting instruction

Denas MS DENAS-T operating manual

Denas MS

Denas MS DENAS-T operating manual

bort medical ActiveColor quick guide

bort medical

bort medical ActiveColor quick guide

AccuVein AV400 user manual

AccuVein

AccuVein AV400 user manual

agiliti NP Trio user manual

agiliti

agiliti NP Trio user manual

Rocket Medical Rocket Digital Oocyte Aspiration Pump Service manual

Rocket Medical

Rocket Medical Rocket Digital Oocyte Aspiration Pump Service manual

Monteris Medical Neuroblate Instructions for use

Monteris Medical

Monteris Medical Neuroblate Instructions for use

Otto Block 8E33 7 manual

Otto Block

Otto Block 8E33 7 manual

ADVANCED HOME CARE Helios Guide

ADVANCED HOME CARE

ADVANCED HOME CARE Helios Guide

NeuLog NUL-208 Guide

NeuLog

NeuLog NUL-208 Guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.