Instruction Manual
Flue Gas Analysis Computer
Type: MULTILYZER®STe

2 MULTILYZER®ST/STe

MULTILYZER®ST/STe 3
Contents
1About this instruction manual ..................................................................................5
1.1 Warning Terms.............................................................................................5
1.2 Explanation of symbols and typeface...........................................................5
2Safety.......................................................................................................................6
2.1 Intended use.................................................................................................6
2.2 Incorrect use.................................................................................................6
2.3 Safe handling ...............................................................................................6
2.4 Qualification of personnel.............................................................................7
2.5 Calibration ....................................................................................................7
2.6 Modifications to the product .........................................................................7
2.7 Usage of spare parts and accessories.........................................................7
2.8 Liability information.......................................................................................7
3Product description..................................................................................................8
3.1 Control panel (Key- and scroll function).......................................................8
3.2 Package contents.........................................................................................9
3.3 Measurement and calculation parameters...................................................9
3.4 Measuring methods....................................................................................10
3.5 Technical specifications .............................................................................12
3.6 Calculation formulae (extract) ....................................................................16
3.7 Approvals, tests and conformities..............................................................17
4Transportation and storage ...................................................................................18
5Commissioning......................................................................................................19
5.1 Connection diagram...................................................................................19
5.2 Use of the IR-printer...................................................................................20
5.3 Leakage test of the flue gas probe (Option)...............................................21
6Operation...............................................................................................................22
6.1 Measuring mode.........................................................................................22
6.2 Data Logger function (Option)....................................................................27
6.3 Fluegas measurement program.................................................................30
6.4 CO-Measurement program ........................................................................37
6.5 Temperature Measurement program .........................................................40
6.6 "Pressure / Draft measurement" program..................................................43
6.7 Pressure Loss / Let-By / Tightness Test program.....................................47
6.8 Pitot Measurement program (Option).........................................................49
7Settings configuration menu..................................................................................51
7.1 Set Time / Date ..........................................................................................52
7.2 Set Display.................................................................................................52
7.3 Set Sound Levels.......................................................................................55

4 MULTILYZER®ST/STe
7.4 Show device information ............................................................................56
8Memory mode & memory structure .......................................................................57
8.1 Create a customer database......................................................................57
8.2 Use of the memory.....................................................................................62
8.3 Enter the user's address.............................................................................65
9Battery management .............................................................................................66
9.1 Battery mode/charging mode .....................................................................66
9.2 Charging the batteries ................................................................................66
10 Maintenance ..........................................................................................................68
11 Troubleshooting.....................................................................................................69
12 Shutting down and disposal...................................................................................70
13 Spare parts and accessories .................................................................................71
14 Warranty ................................................................................................................72
15 Copyright................................................................................................................72
16 Customer satisfaction ............................................................................................72
17 Addresses..............................................................................................................72
18 Certification............................................................................................................73
18.1 DIN EN 50379-Certificate...........................................................................73
18.2 Option: „Dust measurement“ (Emission-measurement).............................74

MULTILYZER®ST/STe 5
1About this instruction manual
This instruction manual is part of the product.
Read this manual before using the product.
Keep this manual during the entire service life of the product
and always have it readily available for reference.
Always hand this manual over to future owners or users of the
product.
1.1 Warning Terms
WARNING TERM
The type and source of danger is shown here.
Precautions to take in order to avoid the danger are shown
here.
There are three different levels of warning:
Warning term
Meaning
DANGER
Imminent danger!
Failure to observe the information will result in
death or serious injuries.
WARNING
Possible imminent danger!
Failure to observe the information may result in
death or serious injuries.
CAUTION
Dangerous situation!
Failure to observe the information may result in
minor or serious injuries as well as damage to
property.
1.2 Explanation of symbols and typeface
Symbol
Meaning
m
Prerequisite for an activity
Activity consisting of a single step
1.
Activity consisting of several steps
Result of an activity
Bulleted list
Text
Indication on a display
Highlighting
Highlighting

6 MULTILYZER®ST/STe
2Safety
2.1 Intended use
The flue gas analysis computer MULTILYZER®STe is exclusively
suitable for:
Professional settings and control measurements at all small
combustion systems (low temperature and burner value boilers
and thermal systems) for gas and oil systems.
Any use other than the application explicitly permitted in this instruc-
tion manual is not permitted.
2.2 Incorrect use
The MULTILYZER®STe flue gas analysis computer must never be
used in the following cases:
Hazardous area (Ex)
If the device is operated in hazardous areas, sparks may cause
deflagrations, fires or explosions
Use as a safety (alarm) unit or continuous measuring device
Ambient air monitoring
Use in humans and animals
2.3 Safe handling
This product represents state-of-the-art technology and is made ac-
cording to the pertinent safety regulations. Each device is subjected
to a function and safety test prior to shipping.
Operate this product when it is in perfect condition. Always ob-
serve the operating instructions, all pertinent local and national
directives and guidelines as well as the applicable safety regula-
tions and directives concerning the prevention of accidents.
Perform an overall visual inspection of the measuring device
(including any accessories) prior to each operation of the
MULTILYZER®STe in order to ensure proper operation of the
device.
Protect the product against impact
Use the product only indoor
Insulate the product from moisture
WARNING
Severe burns or death due to live parts.
Do not touch live parts with the instrument or sensors.

MULTILYZER®ST/STe 7
2.4 Qualification of personnel
The product may only be installed, commissioned, operated, main-
tained, shut down and disposed of by qualified, specially trained per-
sonnel.
Electrical work may only be carried out by qualified electricians in
accordance with local and national regulations.
2.5 Calibration
The flue gas analysis computer MULTILYZER®STe have to be cali-
brated annually by a recognized, relevant authority.
2.6 Modifications to the product
Changes or modifications made to the product by unauthorised per-
sons may lead to malfunctions and are prohibited for safety reasons.
2.7 Usage of spare parts and accessories
Usage of unsuitable spare parts and accessories may cause dam-
age to the product.
Use only the manufacturer’s genuine spare parts and accesso-
ries of the manufacturer.
2.8 Liability information
The manufacturer shall not be liable in any direct or consequential
damage resulting from failure to observe the technical instructions,
guidelines and recommendations.
The manufacturer or the sales company shall not be liable for costs
or damages incurred by the user or by third parties in the use or
application of this device, in particular in case of improper use of the
device, misuse or malfunction of the connection, malfunction of the
device or of connected devices. The manufacturer or the sales
company shall not be liable for damage resulting from any use other
than the use explicitly stated in this instruction manual.
The manufacturer shall not be liable for misprints.

8 MULTILYZER®ST/STe
3Product description
The MULTILYZER®STe flue gas analysis computer is a multiple-
function analyser with integrated calculating functions. Measure-
ments are in accordance with the general regulations set forth by the
German “BImSchV” at all kinds of combustion plants within the
framework of the monitoring of exhaust systems.
The MULTILYZER®STe flue gas analysis computer has an infrared
printer interface, a Bluetooth®interface (Bluetooth®low energy) and
can be fitted with a memory card (MicroSD).
User-friendly, color-coded menus support improved and intuitive op-
eration. The individual measuring programs, configuration menus,
etc. are assigned distinctive colors.
The innovative measuring device doesn’t have a traditional key-
board. A state of the art „Touchpad“ technology with an almost wear-
free operation provides perfect user interface. The sensibility and
speed of the control panel is individually adjustable.
3.1 Control panel (Key- and scroll function)
Taste
Funktion
Scroll panel (touchpad)
Adjustment/navigation functions to move up
and down in the menu section.
Cancel program (ESCAPE/CLEAR button).
„HOLD“-key / direct access
Confirm selection (ENTER button).
Switch on and off

MULTILYZER®ST/STe 9
3.2 Package contents
The Product contains:
Flue gas analysis computer
Protective sleeve with magnet
Flue gas probe with gas treatment and condensate cartridge
Air temperature sensor
Replace filter
USB power-supply
Calibration certificate
Instruction manual
3.3 Measurement and calculation parameters
Table 1: Measured Values
Value
Measured medium
Unit
Tgas
Flue gas temperature
°C, °F
Tair
Air temperature
°C, °F
O2
Oxygen concentration
Vol.-%
CO
Carbon monoxide concentration
ppm, mg/m³,
mg/kWh, mg/MJ,
Vol.-%
NO
Nitrogen monoxide concentration
(Option)
ppm, mg/m³,
mg/kWh, mg/MJ
SO2
Sulfur dioxide concentration
(Option)
ppm, mg/m³,
mg/kWh, mg/MJ
NO2
Nitrogen dioxide concentration
(Option)
ppm, mg/m³,
mg/kWh, mg/MJ
CO+
Carbon monoxide high (Option)
Vol.-%
Draft
Draft
Pa, hPa, kPa,
mbar, bar,
mmWs, mmHg,
inHg, Psi
Barometer
Barometric pressure
hPa

10 MULTILYZER®ST/STe
Table 2: Calculated values
Value
Calculated medium
Unit
CO2
Carbon dioxide
Vol.-%
COref
Carbon monoxide referenced to an O2
reference value
ppm
Eta
Combustion efficiency value
%
Lambda
Excess air value
qA
Flue gas loss
%
Dewpnt
Fuel-specific dew point
°C, °F
T.Diff
Differential temperature (TG - TL)
°C, °F
NOx
Nitrogen oxide (Option)
ppm, mg/m³,
mg/kWh, mg/MJ
NOref.
Nitrogen oxide referenced to an O2
reference value
ppm
NOxref.
Nitrogen oxides referenced to an O2
reference value (Option)
ppm
SO2ref
Sulfur dioxide referenced to an O2
reference value (Option)
ppm
NO2 ref
Nitrogen dioxide referenced to an O2
reference value (Option)
ppm
3.4 Measuring methods
Table 3: Measuring procedure
Function
Explanation
Temperature meas-
urement
Thermocouple NiCr-Ni (type K)
O2measurement
Electrochemical measuring cell
CO measurement
Electrochemical measuring cell
CO2measurement
(Option)
NDIR measuring cell
CO+ measurement
(Option)
Electrochemical measuring cell

MULTILYZER®ST/STe 11
Function
Explanation
NO measurement
(Option)
Electrochemical measuring cell
SO2measurement
(Option)
Electrochemical measuring cell
NO2measurement
(Option)
Electrochemical measuring cell
Pressure/draft
Piezo-resistive sensor with internal tem-
perature compensation
Measuring duration
Short-term, stable measurements of max.
60 minutes are possible, followed by a
new calibration phase with ambient air.
Flue gas measurement
Via an external water separator and filter,
the flue gas is supplied to the sensors by
means of a gas pump.
Sensor calibration
After switching on the instrument, there is
a calibration phase that takes 30 seconds
after a cold start.
CO Sensor protection
The standard CO sensor with dynamic H2
compensation is protected automatically
by separate flushing pump. When the
permissible concentration of CO is
reached (>4000 ppm), the additional CO
pump turns on and flushes the sensor
with fresh air. The measurement starts
again automatically when the CO concen-
tration drop below 1600 ppm.
Flue gas sampling
Flue gas sampling is done by means of a
probe which enables either a “one-point
measurement” (combi probe) or a “multi-
point measurement” (multi-hole probe).
WARNING
The life of the sensors depends essentially on utilization and
use of the instrument. The expected life of the gas sensors is
about 24 months.

12 MULTILYZER®ST/STe
3.5 Technical specifications
Table 2: Device description
Parameter
Value
General Specifications
Dimensions housing
including protective
sleeve (W xHxD)
90 x 53 x 220 mm
(3,6 x 2,1 x 8,7 inch)
Weight (including pro-
tective sleeve)
Ca. 625 g - 685 g (22.05 oz –24.20 oz)
(depends on count of sensors)
Material Protective
sleeve
Polyamid (PA)
Display
High-resolution graphical 3,5“ TFT-
display (240 x 320).
Data communication
Infrared printer interface, Bluetooth®
interface (Bluetooth®low energy).
Printer
External infrared thermal printer (EURO-
PRINTER)
Memory
Micro-SD memory card with folder/file
structure
Temperature range
Ambient
0 °C to +40 °C (+32 °F to +104 °F)
Medium
0 °C to +40 °C (+32 °F to +104 °F)
Storage
-20 °C to +50 °C (-4 °F to +122 °F)
Air pressure range
Ambient
750 hPa to +1100 hPa
Humidity range
Ambient
20 % rH to 80 % rH
Power supply
Rechargeable battery
Lithium-Ion-Battery 3,6 V / 2900 mAh
Power adapter
USB power supply
Electrical guard
Protection type
IP 42 EN 60529

MULTILYZER®ST/STe 13
Electromagnetic compatibility (EMC)
Interference
DIN EN 55022 (VDE 0878-22)
Noise immunity
DIN EN 61000-4-3 (VDE 0847-4-3)
ESD
DIN EN 61000-4-2 (VDE 0847-4-2)
Table 3: Device specifications
Parameter
Value
Flue gas temperature measurement
Measuring range
0 °C to +1000 °C
Max. deviation
±1°C (0 °C to +300 °C)
±1.0 % of measured value (above +300 °C)
Resolution
1 °C
Sensor
Thermocouple NiCr-Ni (type K)
Combustion air temperature
Measuring range
-20 °C to +200 °C
Max. deviation
± 3 °C + 1 digit (-20 °C to 0 °C)
±1°C + 1 digit (0 °C to +200 °C)
Resolution
0,1 °C
Sensor
Thermocouple NiCr-Ni (type K)
Draft / pressure measurement
Measuring range
± 70 hPa (Draft) / ± 150 hPa (Diff.-Pressure)
Max. deviation
± 2 Pa + 1 Digit (<2,00 hPa)
± 1 % reading (>2,00 hPa)
Resolution
± 0,01 hPa bzw. 1 Pa
Sensor
Semiconductor sensor

14 MULTILYZER®ST/STe
Pitot measurement (option)
Measuring range
0.5 - 70 m/s
Max. deviation
± 0.8 m/s
Resolution
0.1 m/s
Sensor
Semiconductor sensor
O2-measurement
Measuring range
0,0 Vol. % to 21,0 Vol. %
Max. deviation
± 0,2 Vol. % by volume of measured value
Resolution
0,1 Vol. %
Sensor
Electrochemical measuring cell
Response time
(T90)
30 seconds
CO2- determination
Range
0 –CO2max
Max. deviation
±0.2 Vol.%
Resolution
0.1 Vol.%
Sensor
calculation from O2 value
Response time
(T90)
30 seconds
CO- measurement (with H2compensation)
Measuring range
0 ppm to 4000 ppm (nominal)
Accuracy
5 ppm (to 50 ppm)
5 % of measured value (above 50 ppm)
Resolution
1 ppm
Sensor
Electrochemical measuring cell
Response time
(T90)
60 seconds

MULTILYZER®ST/STe 15
Table 4: Device specifications - options
Parameter
Value
NO- measurement
Measuring range
0-2000 ppm
Accuracy
5 ppm (to 50 ppm)
5 % of measured value
Resolution
1 ppm
Sensor
Electrochemical measuring cell
Response time (T90)
60 seconds
COhigh- measurement (without H2compensation)
Measuring range
0-2,0 Vol.-% (20.000 ppm)
Accuracy
5 % of measured value (± 1 Digit)
Resolution
0,01 Vol.-%
Sensor
Electrochemical measuring cell
Response time (T90)
60 seconds
CO2- measurement
Range
0-20 Vol.%
Max. deviation
±0.3 Vol.% or 5% of measured value
Resolution
0.1 Vol.%
Sensor
NDIR measuring cell
Response time (T90)
60 seconds
SO2- measurement
Measuring range
0-2000 ppm
Accuracy
10 ppm (to 200 ppm)
5 % of measured value (ab 200 ppm)
Resolution
1 ppm
Sensor
Electrochemical measuring cell
Response time (T90)
60 seconds

16 MULTILYZER®ST/STe
NO2- measurement
Measuring range
0-200 ppm
Accuracy
10 ppm (to 50 ppm)
10 % of measured value (at 50 ppm)
Resolution
1 ppm
Sensor
Electrochemical measuring cell
Response time (T90)
60 seconds
3.6 Calculation formulae (extract)
Calculation of the CO2value
CO2= CO2 max * (1 -
21
O2
) in %
CO2
Calculated carbon dioxide value in %
CO2max
Max. CO2value (fuel-specific) in % by volume
O2
Measured oxygen concentration in %
21
Oxygen concentration of the air in % by volume
Calculation of the flue gas loss
qA = (TG - TA) * (
2
2
O21A
+ B) in %
qA
Flue gas loss in %
TG
Flue gas temperature in °F or °C
TA
Combustion air temperature in °F or °C
A2, B
Fuel-specific factors
O2
Measured O2value

MULTILYZER®ST/STe 17
Calculation of the excess air value Lambda
Lambda =
2
max2
CO
CO
=
2
21
21
O
Lambda
Excess air value
Calculation of the combustion efficiency value (Eta)
Effic. = 100 - qA in %
Effic.
Combustion efficiency value in %
Calculation of CO reference (CO ref)
COref. = CO *
2
2
21
21
O
Oref
COref.
Carbon monoxide reference
CO
Measured CO value
O2ref
O2reference value
3.7 Approvals, tests and conformities
The MULTILYZER®STe flue gas analysis computer is approved in
accordance with the German "1. BundesImmissionsSchutzVer-
ordnung" (1. BImSchV) and EN 50379-2, is TÜV-tested (VDI 4206)
and meets the applicable directives of 2004/108/EC.

18 MULTILYZER®ST/STe
4Transportation and storage
CAUTION
Damage to the device due to improper transportation.
Do not throw or drop the device.
Transporting the device only in the device-specific case.
CAUTION
Damage to the device due to improper storage.
Protect the device from shock when storing it.
Store the device in a clean and dry environment.
Only store the device within the permissible temperature
range.
Store the device away from paint, solvent and glue.

MULTILYZER®ST/STe 19
5Commissioning
WARNING
Before using the MULTILYZER®STe you have to do a visual
inspection of the entire measurement equipment (Device and
accessories).For a correct operation of the device.
5.1 Connection diagram
1
Flue gas probe
with draft for
measurements
2
Plug for flue gas
temperature (yel-
low)
3
Hose for meas-
urement gas (yel-
low)
4
Draft hose
5
(Combustion) air
temperature sen-
sor blue
6
(Combustion) air
temperature sen-
sor with 2.5 m line
and magnet re-
tainer (Option)
7
Outside wall air
sensor
8
Measurement gas
treatment (see
extra sheet)
9
Adjustable cone
Fig. 1: Connection diagram (Sensor side)

20 MULTILYZER®ST/STe
1
USB-Host
connection
2
MicroSD-card
holder
3
USB power sup-
ply unit (100-
240 V / 50-60
Hz)
4
Speaker
5
IR-Printer-
interface (not
visible from in-
strument surface)
Fig. 2: Connection diagram (Interface side)
5.2 Use of the IR-printer
For printing, the IR interface of the MULTILYZER®STE unit must
point towards the printer as shown in the following illustration, keep-
ing a minimal distance of ca. 25 cm! (Max. ca. 70 cm).
Fig. 3: Positioning of the MULTILYZER®STe towards the printer
WARNING
Transmission errors if positioning is not correct
The optical transmission zone must be kept straight and free
of obstacles!
Ensure Bluetooth printing option is switched off in –Settings-
BluetoothSMART menu for correct IR printer operation.