TaiDoc Technology Corporation TD-1242 User manual

TD-1242
Forehead Thermometer
Operation Instructions
WARNING AND PRECAUTIONS:
• As with any thermometer, proper technique is crucial to getting accurate
temperature readings. Please read this manual thoroughly and carefully
before use.
• Always operate the thermometer in an operating temperature range 10°C to
40°C (50°F to 104°F), and relative 30% to 85% relative humidity.
• Always store the thermometer in a cool and dry place: temperatures between
-20°C to 60°C (-4°F to 140°F); 30% to 85% relative humidity. Avoid direct
sunlight.
• Avoid dropping the thermometer.
• Basic safety precautions should always be observed, especially when the
thermometer is used on or near children and disabled persons.
• This thermometer is not intended to be a substitution for a consultation with
your physician.
LCD SCREEN
311-1242000-001
Version 1.0 2020/04
INTRODUCTION
Thank you for choosing TD-1242 Forehead Thermometer. Please read this
instruction manual first so you can use this thermometer safely and correctly.
Please keep this instruction manual for future reference. This innovative
medical device uses advanced infrared (IR) technology to measure
temperature instantly and accurately on the forehead / surface. TD-1242
Forehead Thermometer delivers a body temperature reading from the thermal
radiation emitted from the forehead without contact to the body.
INTENDED USE
TD-1242 Forehead Thermometer is intended for the intermittent measurement
and monitoring of human body temperature from the forehead. The device is
intended for use of all ages for home and professional use by one with a good
understanding of the operation instruction, where the patient may be an
operator.
IMPORTANT SAFETY INSTRUCTIONS
READ THIS BEFORE USING AND KEEP THESE INSTRUCTIONS IN A
SAFE PLACE
1. Close supervision is necessary when the thermometer is used by, on, or
near children, handicapped persons or invalids.
2. Use the thermometer only for the intended use described in this manual.
3. Do not use the thermometer if it is not working properly, or if it has
sustained any damage.
4. Keep the sensor end clean and free of debris. See Maintenance section for
instructions.
5. Do not use ethylene oxide gas, heat, autoclave, or any other harsh
methods to sterilize the device.
6. Do not use the device shortly after exercise, bathing or coming indoors.
7. If coming from an environment of warmer or cooler temperature or after a
period of exertion, allow the user and the thermometer to acclimate to room
temperature for 20 minutes prior to taking a measurement.
8. As the forehead temperature may be affected by sweat, oil and the
surrounding temperature, the reading shall be taken as a reference only.
9. Do not use in presence of flammable anesthetic mixtures.
10. Do not use accessories which are not supplied or recommended by the
manufacturer.
11. Proper maintenance is essential to the longevity of your device. If you are
concerned about the accuracy of measurement, please contact the local
customer service or place of purchase for help.
APPEARANCE AND KEY FUNCTIONS OF
THE THERMOMETER
1Display Screen
Power / Scan Button
Probe
Mode Button
Memory Button
Battery Cover
2
3
4
5
6
MEMORY MORE
1
2
4
5
6
3
USING THE DEVICE
Selecting the measuring unit
When the thermometer is off, press MODE key for 2 seconds to enter selection
mode. Press MODE key again to switch between °C and °F. Press POWER
key.
Install Battery
1. Remove battery cover by pressing down at the arrow mark and slide in the
direction of the arrow as shown in the figures below (Figure 1).
2. Install (2) AAA alkaline batteries and close the battery cover (Figure 2).
3. If thermometer will be stored without use, remove batteries.
Warning:
Make sure the battery cover is properly closed and secured.
MEASURING TEMPERATURES-GENERALS
(older than 36 months)
1. The forehead should be clear of hair and perspiration.
2. Aim at the center forehead area 3cm to 7cm (1.18” to 2.75”) away from
skin surface. Be sure the thermometer is perpendicular to the skin
surface.
1Temperature display
Error message
Memory mode
Hold symbol
Temperature unit
Battery indicator
General body temperature indicator
Child body temperature indicator
Object surface temperature indicator
2
3
4
5
6
7
8
9
6789
3 4
2
1
5
( Figure 1 ) ( Figure 2 )
3. Press and release the SCAN button to take a measurement. A double
“beep” sound indicates a reading has been taken and displayed on the
LCD screen.
Manufactured by
TaiDoc Technology Corporation
B1-7F, No. 127, Wugong 2nd Rd., Wugu Dist.,
24888 New Taipei City, Taiwan
www.taidoc.com
MedNet EC-REP GmbH
Borkstraβe 10, 48163 Münster - Germany
0123
4. To take another measurement, follow step 2 and 3.
5. The thermometer turns off automatically after 30 seconds.
WARNING AND PRECAUTIONS:
• If the reading is < 31.9°C (89.5°F), the display will show “Lo”.
• If the reading is ≥ 32.0°C (89.6°F) and ≤ 37.9 °C (100.3°F), the display will
show the reading with backlight.
• If the reading is ≥ 38°C (100.4 °F) and ≤ 43°C (109.4 °F), the display will
show the reading with flash backlight.
• If the reading is ≥ 43.1°C (109.5 °F), the display will show “Hi”.
MEASURING TEMPERATURES-CHILDREN
(younger than 36 months)
1. Press the SCAN button to turn on the thermometer.
2. Press and hold the MODE button for1 second before releasing. An adult
facial profile should be flashing,indicating that the preset mode is for adults.
3. Press the MODE key to switch to Child mode, indicated by a smiley face
icon.
4. Aim the scanner at the center of the child’s forehead or temple 3cm to 7cm
(1.18” to 2.75”) away from skin surface. Be sure the thermometer is
perpendicular to the skin surface. Press and release the SCAN button to
take a measurement.
5. The thermometer automatically switches off when left idle for 30 seconds.
NOTE:
• If the reading is ≤ 31.9°C (89.5°F), the display will show “Lo”.
• If the reading is ≥ 32°C (89.6°F) and ≤ 37.6 °C (99.8°F), the display will show
the reading with backlight.
• If the reading is ≥ 37.7 °C (99.9°F) and ≤ 43°C (109.4°F), the display will
show the reading with flash backlight.
• If the reading is ≥ 43.1°C (109.5°F), the display will show “Hi”.
• Parents should not rely only on temperature readings. If you have concerns,
please seek medical advice.
MEASURING TEMPERATURES-SURFACE
1. Press the SCAN button to turn on the thermometer.
2. Press and hold the MODE button for 1 second before releasing. A general
facial profile should be flashing, indicating that the preset mode is for
generals.
3. Press the MODE key to switch to Surface mode, indicated by a
thermometer icon.
When a reading is greater than 38°C (100.4°F), a warning symbol and
backlight will start flashing.
Flash backlight with a warning symbol indicates a reading greater than
37.6 °C (99.8°F).
4. Make sure the probe is flat and close to the object surface, not at an angle.
Perform an object measurement with a distance within 5 cm (1.9 in). Press
and hold the SCAN button as you move the meter along the surface. The
HOLD symbol will be flashing.
3. Press and release the MEMORY button to cycle through older readings.
4. When left idle for 30 seconds, the thermometer automatically switches off.
NOTE:
When the memory is full, the oldest result is deleted as the neones are added.
While recalling previous readings, you may take a measurement by pressing
the SCAN button.
ABOUT NORMAL BODY TEMPERATURE &
FEVER
Body temperature can vary from one individual / person to the next. It also
varies by location on the body and time of day. Fever indicates that the body
temperature is higher than normal. This symptom may be caused by infection,
overdressing or immunization. Some people may not experience fever even
when they are ill. These include, but are not limited to, infants younger than 3
months old, persons with compromised immune systems, persons taking
antibiotics, steroids or antipyretics (aspirin, ibuprofen, acetaminophen), or
persons with certain chronic illnesses. Please consult your physician if you
are concerned about your body temperature readings.
MAINTENANCE
• TD-1242 has no user serviceable internal parts except battery replacement.
• Store in a dry location free of dust and away from direct sunlight.
CARE AND CLEANING
• Use a soft dry cloth to clean the plastic casing or a cloth dampened with 70%
isopropanol.
• For the measuring probe you can use only soft dry cloth or Cotton swabs for
cleaning.
5. Release the button and read the result. If the reading is ≥ 0°C (32°F) and
≤ 100.1°C (212.2°F), the display will show the reading with backlight.
NOTE:
• If the reading is ≤ 0°C (32°F), the display will show “Lo”.
• If the reading is ≥ 100.1°C (212.2°F), the display will show “Hi”.
RECALLING PAST READINGS
TD-1242 stores 30 most recent readings.
1. Press and release the SCAN button to turn on the thermometer.
2. Press and hold the MEMORY button for 1 second to enter the memory
mode indicated by a flashing " " symbol. The most recent reading stored
will be displayed.
Read instructions before use.
Caution, consult accompanying documents.
TROUBLESHOOTING SPECIFICATIONS
Model No.: TD-1242
Dimension & Weight:
155.46 (L) x 40.14 (W) x 39.45 (H) mm, 61.8g (without battery)
Power Source: 2 x 1.5V AAA alkaline batteries
Battery Life: With new batteries, approx. 3,000 measurements.
Displayed Temperature range:
• Forehead: 32°C to 43°C (89.6°F to 109.4°F)
• Surface: 0°C to 100°C (32°F to 212°F)
Display Resolution: 0.1°C / 0.1°F
Accuracy: Meet the accuracy requirement specified in ASTM E1965-98
• Forehead: ±0.2°C (±0.4°F) for the range of 35.0°C to 42.0°C (95°F to
107.6°F) / ±0.3°C (±0.5°F) for the range of < 35.0°C (95°F) or > 42.0°C
(107.6°F)
• Surface: ±1°C (±2°F)
Device Standard: Device Corresponds to the requirements of the standard for
infrared thermometers ASTM E1965-98, EN ISO 80601-2-56, EN 60601-1-2,
IEC 60601-1, EN 60601-1, EN 60601-1-6
Electromagnetic Compatibility:
Device fulfills the stipulations of the standard IEC 60601-1-2, EN 60601-1-2.
The stipulations of EU-Directive 93/42/EEC for Medical Devices Class Ila have
been fulfilled.
Temperature unit: °C (Default) or °F
Operating temperature range: 10°C to 40°C (50°F to 104°F)
Operating humidity: 30% to 85% relative humidity (non-condensing)
Operating Atmosphere Pressure: 700hPa to 1060hPa
Storage / transportation temperature range: -20°C to 60°C (-4°F to 140°F)
Storage / transportation humidity: 30% to 85% relative humidity (non-condens-
ing)
Memory capacity: 30 measurements
Water-resistance: IP22
Type BF Applied part
The specifications may be changed without prior notice.
Electromagnetic environment-guidance
(for home and professional healthcare environment)
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations / flicker emissions IEC 61000-3-3
Group 1
Class B
Not applicable
Not applicable
Compliance
Manufacturer’s declaration-electromagnetic emissions
The TD-1242 is intended for use in the electromagnetic environment (for home and professional healthcare) specified below.
The customer or the user of the TD-1242 should assure that it is used in such an environment.
The TD-1242 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The TD-1242 is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes.
Electromagnetic environment-guidance
(for home and professional healthcare environment)
Electrostatic
discharge(ESD) IEC
61000-4-2
Compliance levelIEC 60601 test levelImmunity test
Manufacturer’s declaration-electromagnetic immunity
The TD-1242 is intended for use in the electromagnetic environment (for home and professional healthcare) specified below.
The customer or the user of the TD-1242 should assure that it is used in such an environment.
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30%
Mains power quality should be that of a typical home and
professional healthcare environment.
Mains power quality should be that of a typical home and
professional healthcare environment.
The TD-1242 power frequency magnetic fields should be at
levels characteristic of a typical location in a typical home
healthcare environment.
Mains power quality should be that of a typical home and
professional healthcare environment. If the user of the
TD-1242 requires continued operation during power mains
interruptions, it is recommended that the TD-1242 be
powered from an uninterruptible power supply or a battery.
Electrical fast
transient/burst IEC
61000-4-4
Surge IEC
61000-4-5
Power frequency(50,
60 Hz) magnetic field
IEC 61000-4-8
Voltage Dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
Contact:±8 kV
Air±2 kV,±4 kV,
±8 kV,±15 kV
Not applicable
Not applicable
Not applicable
Not applicable
30 A/m
50 Hz and 60 Hz
Voltage dips:
Not applicable
Not applicable
Not applicable
Voltage interruptions:
Not applicable
Contact:±8 kV
Air±2 kV,±4 kV,
±8 kV,±15 kV
+ 2kV for power supply lines
+ 1kV for input/output lines
+ 0.5kV, +1kV line(s) to line(s)
+ 0.5kV, +1kV,+ 2kV line(s) to
earth
30 A/m
50 Hz or 60 Hz
Voltage dips:
0 % UT; 0,5 cycle
0 % UT; 1 cycle
70 % UT; 25/30 cycles
Voltage interruptions:
0 % UT; 250/300 cycle
NOTE UT is the a.c. mains voltage prior to application of the test level.
Separation distance according to frequency of transmitter
m
0,01
0,1
1
10
100
N/A
N/A
N/A
N/A
N/A
0,12
0,38
1,2
3,8
12
0,23
0,73
2,3
7,3
23
Rated maximum output power
of transmitter
W150 kHz to 80 MHz
d =1,2√P
80 MHz to 800 MHz
d =1,2√P
800 MHz to 2,7 GHz
d =2,3√P
Recommended separation distance between
portable and mobile RF communications equipment and the TD-1242
The TD-1242 is intended for use in an electromagnetic environment (for home and professional healthcare) in which radiated RF
disturbances are controlled. The customer or the user of the TD-1242 can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the TD-1242 as recommended below,
according to the maximum output power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where p is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
385 380 –390 TETRA 400
Pulse
modulation b)
18 Hz
1,8 0,3 27 27
710
745
780
704 – 787 LTE Band 13,
17
Pulse
modulation b)
217 Hz
0,2 0,3 9 9
5 240
5 500
5 785
5 100 –
5 800
WLAN 802.11
a/n
Pulse
modulation b)
217 Hz
0,2 0,3 9 9
810
870
930
800 – 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation b)
18 Hz
2 0,3 28 28
2 450 2 400 –
2 570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation b)
217 Hz 2 0,3 28 28
1 720
1 845
1 970
1 700 –
1 990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse
modulation b)
217 Hz
2 0,3 28 28
450 430 – 470 GMRS 460,
FRS 460
FM c)
±5 kHz deviation
1 kHz sine
2 0,3 28 28
Test
frequency
(MHz)
Maximum
power
(W)
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
Compliance LEVEL
(V/m) (for home and
professional healthcare)
Band a)
(MHz)
Service a) Modulation b)
Manufacturer’s declaration-electromagnetic immunity
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
The TD-1242 is intended for use in the electromagnetic environment (for home and professional healthcare) specified below.
The customer or the user of the TD-1242 should assure that it is used in such an environment.
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or
ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual
modulation, it would be worst case.
Electromagnetic environment-guidance
(for home and professional healthcare environment)
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Compliance levelIEC 60601 test levelImmunity test
Manufacturer’s declaration-electromagnetic immunity
The TD-1242 is intended for use in the electromagnetic environment (for home and professional healthcare) specified below.
The customer or the user of the TD-1242 should assure that it is used in such and environment.
Portable and mobile RF communications
equipment should be used no closer to any part of the
TD-1242 including cables, than the recommended
separation distance calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance:
d = 1,2 √P
d = 1,2 √P 80MHz to 800 MHz
d = 2,3 √P 800MHz to 2,7 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Interference may occur in the vicinity of equipment marked
with the following symbol:
Not applicable
Not applicable
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms:
in ISM and amateur
radio bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
SYMBOL INFORMATION
Symbol Referent
Consult instructions for use
This device does not belong to household waste and must be
returned to a collection point for recycling electric and electronic
devices according to local laws. If it contains batteries, the
batteries should be removed and disposed in accordance with local
regulations for separate collection of spent batteries.
Manufacturer
Serial number
Caution
Authorized representative in the European Community
CE mark
Symbol Referent
Type BF applied part
Temperature limit
Humidity limitation
Resistant to liquid ingress
RoHS Compliance
MESSAGE WHAT IT MEANS WHAT TO DO
Appear when ambient
temperature is outside of
the operating temperature
range.
Only operate the thermometer
in ambient temperature range of
10°C to 40°C (50°F to 104°F).
Surface temperature
measured outside of the
measurement range 0°C to
100°C/32°F to 212°F
Review the instructions and
re-start the measurement
procedure.
Low or no power. Replace with new batteries.
Measured temperature
below the measurement
range (below 32°C /
89.6°F).
Review instructions and repeat
measurement. Make sure
distance from forehead is not
greater than 7cm (2"), and that
the forehead is clear.
Measured temperature
above the measurement
range (above 43°C /
109.4°F).
Review instructions and repeat
measurement. Consult with a
physician if the problem
persists.
IP22
0123

Other TaiDoc Technology Corporation Thermometer manuals

TaiDoc Technology Corporation Riester ri-thermo sensioPRO User manual

TaiDoc Technology Corporation

TaiDoc Technology Corporation Riester ri-thermo sensioPRO User manual

TaiDoc Technology Corporation Riester ri-thermo tymPRO User manual

TaiDoc Technology Corporation

TaiDoc Technology Corporation Riester ri-thermo tymPRO User manual

Popular Thermometer manuals by other brands

SIKA DiTemp LCK operating manual

SIKA

SIKA DiTemp LCK operating manual

Maverick Industries XR-40 instruction manual

Maverick Industries

Maverick Industries XR-40 instruction manual

GREISINGER GMH 3710 operating manual

GREISINGER

GREISINGER GMH 3710 operating manual

TFA 30.3048 instruction manual

TFA

TFA 30.3048 instruction manual

Haushalt International 70176 instructions

Haushalt International

Haushalt International 70176 instructions

ALCO AL-82955 Directions for use

ALCO

ALCO AL-82955 Directions for use

GTC LTX10 user guide

GTC

GTC LTX10 user guide

JOYTECH DMT-427 Instructions and warranty

JOYTECH

JOYTECH DMT-427 Instructions and warranty

Sealey VS900.V4 instructions

Sealey

Sealey VS900.V4 instructions

TFA 30.3034 instruction manual

TFA

TFA 30.3034 instruction manual

Oregon Scientific EMR801 user manual

Oregon Scientific

Oregon Scientific EMR801 user manual

Ascon tecnologic e30V operating instructions

Ascon tecnologic

Ascon tecnologic e30V operating instructions

Microlife IR 1DE1 instruction manual

Microlife

Microlife IR 1DE1 instruction manual

Viproud FC-IR109 user manual

Viproud

Viproud FC-IR109 user manual

Fluke 62 MAX manual

Fluke

Fluke 62 MAX manual

BIOS DT300 instruction manual

BIOS

BIOS DT300 instruction manual

Prestige medical DT-30 instructions

Prestige medical

Prestige medical DT-30 instructions

TrueLife CARE T3 instruction manual

TrueLife

TrueLife CARE T3 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.