Taik T210A User manual

T210A
Instruction Manual
07/2020
TAIK
Senflow T210A
OK
PROG

1. Hardware Structure:
1.1Push Buttons On Front Panel:
R2
COM
AO2
AO1
P%
√
OKPROG
TAIK
Senflow T210A
8
7
6
5
3
2
1
9ABCD E F
4
1
8
7
6
5
3
2
4
9
A
B
C
D
E
F
Flow input indication flashes interactive
6 digits value display
Parameters display symbol
Tens digit accumulative value display
Password setting
Square root calculation
Unit display
5 push buttons
If setting first analog output will display
If setting second analog output will display
If setting second relay contact will display :Relay contact activated
If setting pulse output will display
RS-485 activation
Analog output value percentage
Analog output value display in bar
1.2 Terminals Configuration:
NL
24V
COM
5V 12V
A
Ext CIN
B C D
4
2
3
5
7
6
1
1
7
6
5
3
2
4
Power supply
Signal input
Aux. power option
RS-485 module
AOP1 Analog output module
Pulse and relay contact output module
AOP2 Analog output module

91+0.5
-0.0
91 +0.5
-0.0
96
17 46 18 20
81
101
2. Installation/ Wiring:
2.1 Case Dimension and Panel Cut-out:Unit:mm
Cut-out size
Side view
96
96
2. Connection Diagrams2
NL
24V
COM
5V 12V
A
Ext CIN
B C D
Aux. Source
(+)(-)
Analog Input
(+)(-)
Rear view Meter connection
Module
A+B-O
RS-485
RS-485
1+1-
Analog output
1G 1A 2G
Pulse and
relay contact
2A
Relay Contact
(+)(-)
Analog Output
(+) (-)
MD01 MD07MD06
Pulse Output

T210A Display Mode:
Instantaneous flow
Resettable integrated flow
Non-resettable
integrated flow
Ao1 Output display
%
Ao2 Output display
%
瞬間功率與
不可歸零能量累計值
&
Non-resettable integrated flow
Instantaneous flow
AO2
AO1
PROG
Keys Operation:
Advanced to the previous display
Advanced to the next display
+
Clear integrated values at T1
Hold down to relieve error code
OK
+
Access to the set-up page and enter password

3.4 Error Code3.2 Formula Calculation
Mass flow rate calculation: W = Q * DS
W:Mass flow rate(kg/s)
Q:Volume flow rate(m³/s)
DS:Density of fluid (kg / m³)
3.3 Unit Conversion
1L = 1000mL = 0.001m³ = 0.2642GAL
1kg = 2.205Lb = 0.001TON
1day = 24hour = 1440min = 86400s
ERR10
NOINIT :Ex-factory without calibration
ERR1 OVERFD :Instantaneous flow exceed set value
ERR2 OVER P
:Pulse output exceed available output range
ERR4 AO1ERR :AOP1 output exceed 20mA
ERR3 AIERR :21mAInput exceed
ERR7 ROMERR:EEPROM malfunction
ERR5 AO2ERR :AOP2 output exceed 20mA
**After error exclusion, hold down can back to display
OK

E
Buttons:
Values set-up
Vary value
Vary value
Vary digits position
Confirm and access to next setting
4 mA set as min.
20 mA set as max
Default code is 1000
Basic parameters set-up
Energineering units set-up
Analog output set-up
E
4mA Corresponding value
Correction factor set-up
Decimal point set-up
Density set-up
0.0001
99999
000000
999999
0.000
9999
00
99
Low flow cutoff set-up
Set-up page password
LCD backlight set-up
Square root activates
OFF
ON
00.0001
99.9999
0000 01
9999 99
.
.
L/F c ut off
dec imal po int set -up
L/F cutoff value s et- up
ON
dEFuLT
OFF
Integrated flow decimal point
0000
.0000
Instantaneous flow units
Volume units
Time units
L
ML
TON
Lb
KG
SFT3
/S
/d
/H
/M
Integrated flow units
AO1
Ao1 Output option
Instantaneous flow/ power
decimal point
0000
.0000
AO1
Ao1 4mA
Corresponding value
AO1
Ao1 20mA
Corresponding value
AO2
Ao2 Output option
AO2
AO2
Ao2 20mA
Corresponding value
OFF
FLOW
000000
999999
000000
999999
000000
999999
000000
999999
Main menus
Advance to the last page
Advance to the next page
Access to Sub-menus
Confirm and return to the Display
Sub-menus
Advance to the last page
Advance to the next page
Access to Values set-up
Return to the Main menus
E
PROG
E
OK
E
E
E
OK
PROG PROG
OK OK
E
PROG OK PROG OKPROG OK
OK
OK
OK
OK
OK
OK
4. Set-up Page Flowchart:
4.1 Menus Flowchart
PROG
PROG
PROG
20mA Correspond ing va lue
OK
NM3
NL
FT3
GAL
M3
L
ML
TON
Lb
KG
SFT3
NM3
NL
FT3
GAL
M3 OFF
FLOW
Average value set-up
(00 represents10)
(Scr ee n ba cklight w il l be turn off after 1 0 se c. )
Ao2 4mA
Corresponding value

RELAY,Pulse set-up
Communication set-up
COM
Address
COM
Parity setting
COM
Baud rate
N.8.1
O.8.1
E.8.1
N.8.2
19200
9600
4800
2400
001
199
PROG OK PROG OK
E
E
R1
R1 Output set-up
OFF
PuLSE
Set to OFF
P
Pulse corresponding options
L
M3
P
Pulse corre spon ding value
decimal point posi tion
00.0.0 0.1.
P
Pulse corresponding value
000000
999999
Set to PuLse
R2
Relay activation value
R2
Relay activation set point
R2 Output option set-up
R2
OFF
EALAM
RELAY
R2
Hysteresis value
0000
9999
R2
Relay delay time
When set to OFF or EALAM
00
99
000000
999999
When set to RELAY
OK OK OK
OK
OKOK
OK
OK OK
OK
OK
P
Pulse width
0.001
9.999
OK

Low flow cutoff set-up
Backlight set-up
4.2 Menu Legends:
Correction factor set-up
Set-up page password
Integrated flow decimal point
Instantaneous flow unit
Instantaneous flow
decimal point
Average value set-up
Density set-up
4mA and 20mA corresponding value
Range: 0.0001~99999
Correction factor set-up : A coefficient can correct deviation
Range: 00.001~99.999
When calculating fluid weight need to set-up liquid density
0000.01~9999.99Range:
Change password
Range: 000000~999999
Square root of input value
OFF、ON
Set-up for average sampling times
Range: 00~99
(00 means 10 times)
LCD backlight set-up
LCD backlight
ON:LCD backlight stays on
dEFuLT:LCD backlight shut down after 10 secs.when not in use
OFF:Close
Intantaneous flow decimal point
Range: .0000~0000
Intantaneous flow unit
Volume or weight units: L,ML,TON,Lb
KG、SFT3,NM3
NL,FT3,GAL,M3
Time units:S:sec., d:day, H:hour, M:month
If input is lower than set value will be 0
Range: 0.000~9999
Integrated flow decimal point
Range: .0000~0000
Enter corresponding
value
Square root activates
Integrated flow unit
Integrated flow unit set-up
L,ML,TON,Lb
KG,SFT3,NM3
NL,FT3,GAL,M3

R1 P
Pulse corresponding value
R1 P
Pulse corre spon ding value
decimal point posi tion
R1 P
Pulse corresponding options
AO1
Analog output set-up
AO1
Ao1 4mA
Corresponding value
AO2
Ao2 20mA
Corresponding value
AO1
Ao1 20mA
Corresponding value
AO2
Analog output set-up
Ao2 4mA
Corresponding value
AO2
Ao2 output corresponding value
OFF:No output, FLOW:flow
AO1 4mA corresponding value
Range: 000000~999999
Ao1 20mA corresponding value
Range: 000000~999999
Ao1
OFF:No output, FLOW:flow
output corresponding value
Ao2 4mA corresponding value
Range: 000000~999999
Ao2 20mA corresponding value
Range: 000000~999999
R1
R1 Output set-up
Setting the corresponding to pulse
Range: 000000~999999
accumulative value that
Width of pulse output signal
0.001~9.999
Pulse corresponding value decimal point position
Range: 000001~00.0001
Pulse corresponding options
KWH:power, L:flow
R1 output option
OFF:output disable, PuLSE:pulse output
P
Pulse width

R2
Relay activation set point
R2
Relay activation value
R2
Hysteresis value
R2
Relay delay time
COM
Rs485 address set-up
COM
Baud rate
COM
Parity setting
R2 activation set-up point value
Range: 000000~999999
Activation set-up point
Range:
hysteresis setting
0000~9999
RS-485 p
N.8.1, O.8.1, E.8.1,N.8.2
arity setting
RS-485 communication address set-up
Range:001~199
R2 activation set-up point
≧:more than or equal to set value
≦:less than or equal to set value
RS-485 baud rate set-up
2400,4800,9600,19200
R2 delay time set-up
Range:00~99(sec.)
R2 Output option set-up
R2
R2 output option
OFF:output disable, EALAM:Will activate when error happens
RELAY:Set point activation

5. Specifications:
5.1 Specifications
Analog input:
DC 4-20mA
Input range
±0.05% F.S. ±2 digit
Deviation range
Display screen:
Format LCD white backlit
5 digits
Instantaneous flow digits
10 digits
Integrated flow digits
Relay contact status, RS-485 communication status
Pulse output,Analog output percentage
Input signal status, Unit display, Error message
Status symbol
Analog output:
Output range DC4-20mA isolated
Corresponding value Instantaneous flow
Maximum load ≦350Ω
Deviation range ±0.1%R.O.
Communication:
Interface RS485
Protocol MODBUS , RTU format
Baud rate
2400 ~ 19200 programmable
Address
1 ~ 199 programmable
Data formal N,8,1/N,8,2 O,8,1 E,8,1
/ /
Parallel connection 32 meters
programmable
Pulse output:
Output genre
Open Collector (O.C.)
Corresponding value Integrated flow
Setting range 0.001-999999
Pulse width 0.001-9.999 sec.
Relay contacts output:
Output genre A contact (N.O.)
Corresponding value
Instantaneous flow high/low alarm or error alarm
Setting range 0-999999
Contact load AC240V 5A
DC24V 5A
Maximum load DC30V 30mA
Auxiliary power supply
Output voltage
DC5V, DC12V, DC24V(option)
Max. current 30mA
Power supply:
Auxiliary power supply AC/DC85-265V
Frequency Range 45-70Hz
Power consumption ≦15VA(All modules)
Permissible module quantity:At most 4 sets
DC analog output module 2 sets(B、C)
Rs485 module quantity 1 set(A)
Pulse and relay contacts module 1 set(D)

Application environment:
Operating temperature 0-60°C
5-95%RH,non condensing
Operating humidity
-10-70°C
Storage temperature
Characteristic and terms:
Sampling time 0.3 secs.
Input/Output/Power AC2KV,1min
Dielectric strength
Terminals/Case AC3KV,1min
Ip54 front,IP20 rear
Ingress protection
Input/Output/Power/Case
DC500V ≧100MΩ
Surge immunity
Conducted emission
EN 55011
Radiated emmission
EN 55011
Harmonic current emissions
EN 61000-3-2
Voltage changes, voltage fluctuations,and flicker
En 61000-3-3
Electrostatic discharge
IEC61000-4-2
Electromagnrtic field immunity
IEC61000-4-3
Electrical fast transient/burst immunity
IEC61000-4-4
Surge immunity
IEC61000-4-5
immunity to conducted disturbances
IEC61000-4-6
Power frequency magnetic field immunity
IEC61000-4-8
Short interruptions and voltage variations immunity
IEC61000-4-11
CE certification
EMC Testing

6.1 Communication protocol:
6. Communications:
Adopting MODBUS communication shall use a repeater as the meters
are in parallel connection more than 30 pcs.
6.2 Transmission mode:
RTU MODE
6.3 Communication method:
Rs485 (Half-Duplex)
6.4 MODBUS Frame:
6.4.1 Basic Comm and Framing: Hexadecimal Code
Start of
frame
Address
Field
Function
Code
Data
Field
Error
Check
End of
Frame
Start of frame :The data is not transmitted by a silent period of at least
4 characters
Address field :The valid MODBUS addresses are in the range of 1-255, the address 0 for
broadcast command is only valid for Function Code ð 06H
Function code :03HðRead data
06HðWrite data
Data field :The start address of a register. Reading N WORDS and Writing N values
Error check :16bit CRC。
End of frame :The date is not transmitted by a silent period of at least 4 characters
6.4.2 Bit Per Byte: Access to sub-menus 485àFrAE to set-up
Start Bit Data Bit Parity Stop Frame
1 8 None 2 N,8,2
1 8 Odd 1 O,8,1
1 8 Even 1 E,8,1
1 8 None 1 N,8,1
6.5 Reading Register Command:
Query:
Start of
Frame
Address
Field
Functio
n
Code
Start
Address
Hi
Start
Address
Lo
Number
of
Word
Hi
Number
of
Word
Lo
Error
Check
End of
Frame
01H~FFH 03H 0~nnH 0 ~ nnH 0H 1~nnH CRC L o CRC Hi
1 Byte 1 Byte 2Byte 2 Byte 2 Byte
Response:(Com mand is correct)
Start of
Frame
Address
Field
Function
Code
Number of
Data Byte
Count
D0、D1.. Dn
(Hi,Lo,Hi,Lo....)
Error
Check
End of Frame
01H~FFH 03H CRC Lo CRC Hi
1 Byte 1 Byte 1Byte 2 Byte
6.6 Writing Register Command: A single writing WORD command
Query:
Start of
Frame
Address
Field
Functio
n
Code
Start
Address
Hi
Start
Address
Lo
Value
Hi..
Value
..Lo
Erro r
Check
End of
Fra me
01H~FFH 06H 0~nnH 0 ~ nnH Setting Value CRC Lo CRC Hi
1 Byte 1 Byte 2Byte 2 or 4 Byte 2 Byt e
Response:(Response to the writing data if the command is correct)
Start of
Frame
Address
Field
Functio
n
Code
Start
Address
Hi
Start
Address
Lo
Value
Hi..
Value
.. Lo
Erro r
Check
End of
Fra me
01H~FFH 06H 0~nnH 0 ~ nnH Setting Value CRC Lo CRC Hi
1 Byte 1 Byte 2Byte 2 or 4 Byte 2 Byt e
6.7 Message Error:(A command is error)
Start of
Frame
Address
Field
Function
Code
Error
Code
Error
Check
End of
Frame
01H~FFH 83H or 86H CRC Lo CRC Hi
1 Byte 1 Byte 1 Byte 2 Byte
l Function Code:Response to the received Function Code but MSB is set to 1, it
functions like 03H 83H
l Error Code:
01:Error Function。
02:Error Data Address。
03:Error Data Value。

6.8 The CRC Calculation:
The CRC is calculated on all byte of a message from the Address Filed to the
last data byte ended (Data Field); Furthermore, it means that the data
received is in error if the CRC calculation performed on host does not match
the received data.
The CRC Calculation Performed:
1. Load a CRC register with 0 x FFFF
2 Exclusive- OR the first 8 bits of the message with the low-order byte of the
CRC register. Put the result in the CRC register.
3.Shift the CRC register one bit to the right then fill the message in high-order
of the CRC register with a zero, and compare the bit shifted out (SLSB).
4. Repeat step 3 if the SLSB=0; Exclusive- OR the CRC register with the
value A001(Hex), then put the result in the CRC register if the SLSB=1.
5. Repeat steps 3 and 4 uutil the 8 bits have been performed and tested.
6. Repeat steps 2 to 5 until all bytes have been performed.
7. Swap a message with low and high order bytes of the CRC register.
CRC Checkup Routine:
The CRC register returned is as unsigned short int.
The starting address and the field of the data are transmitted, and the
high and low order bytes of the returned CRC register have been
swapped.
/*CRC Generation Function with ‘C’ language*/
/* Msg:*message to calculate CRC upon*/
/* usDatalen: number of bytes in message*/
unsigned int CRC16(char *Msg,unsigned char usDatalen)
{
unsigned char uchCRCHi=0xFF;/*CRC high byte*/
unsigned char uchCRCLo=0xFF; /*CRC low byte*/
unsigned char uIndex;
while(usDatalen--)/*pass through message buffer*/
{
uIndex=uchCRCHi^*Msg++;/*calculate the CRC*/
uchCRCHi=uchCRCLO^auchCRCHi[uIndex];
uchCRCLo=auchCRClo[uIndex];
}
}
return (uchCRCHi<<8|uchCRCLo);
static unsigned char auchCRCHi[]={
0x00,0xc1,0x81,0x40,0x01,0xc0,0x80,0x41,0x01,0xc0,
0x80,0x41,0x00,0xc1,0x81,0x40,0x01,0xc0,0x80,0x41,
0x00,0xc1,0x81,0x40,0x00,0xc1,0x81,0x40,0x01,0xc0,
0x80,0x41,0x01,0xc0,0x80,0x41,0x00,0xc1,0x81,0x40,
0x00,0xc1,0x81,0x40,0x01,0xc0,0x80,0x41,0x00,0xc1,
0x81,0x40,0x01,0xc0,0x80,0x41,0x01,0xc0,0x80,0x41,
0x00,0xc1,0x81,0x40,0x01,0xc0,0x80,0x41,0x00,0xc1,
0x81,0x40,0x00,0xc1,0x81,0x40,0x01,0xc0,0x80,0x41,
0x00,0xc1,0x81,0x40,0x01,0xc0,0x80,0x41,0x01,0xc0,
0x80,0x41,0x00,0xc1,0x81,0x40,0x00,0xc1,0x81,0x40,
0x01,0xc0,0x80,0x41,0x01,0xc0,0x80,0x41,0x00,0xc1,
0x81,0x40,0x01,0xc0,0x80,0x41,0x00,0xc1,0x81,0x40,
0x00,0xc1,0x81,0x40,0x01,0xc0,0x80,0x41,0x01,0xc0,
0x80,0x41,0x00,0xc1,0x81,0x40,0x00,0xc1,0x81,0x40,
0x01,0xc0,0x80,0x41,0x00,0xc1,0x81,0x40,0x01,0xc0,
0x80,0x41,0x01,0xc0,0x80,0x41,0x00,0xc1,0x81,0x40,
0x00,0xc1,0x81,0x40,0x01,0xc0,0x80,0x41,0x01,0xc0,
0x80,0x41,0x00,0xc1,0x81,0x40,0x01,0xc0,0x80,0x41,
0x00,0xc1,0x81,0x40,0x00,0xc1,0x81,0x40,0x01,0xc0,
0x80,0x41,0x00,0xc1,0x81,0x40,0x01,0xc0,0x80,0x41,
0x01,0xc0,0x80,0x41,0x00,0xc1,0x81,0x40,0x01,0xc0,
0x80,0x41,0x00,0xc1,0x81,0x40,0x00,0xc1,0x81,0x40,
0x01,0xc0,0x80,0x41,0x01,0xc0,0x80,0x41,0x00,0xc1,
0x81,0x40,0x00,0xc1,0x81,0x40,0x01,0xc0,0x80,0x41,
0x00,0xc1,0x81,0x40,0x01,0xc0,0x80,0x41,0x01,0xc0,
0x80,0x41,0x00,0xc1,0x81,0x40};
static unsigned char auchCRCLo[]={
0x00,0xc0,0xc1,0x01,0xc3,0x03,0x02,0xc2,0xc6,0x06,
0x07,0xc7,0x05,0xc5,0xc4,0x04,0xcc,0x0c,0x0d,0xcd,
0x0f,0xcf,0xce,0x0e,0x0a,0xca,0xcb,0x0b,0xc9,0x09,
0x08,0xc8,0xd8,0x18,0x19,0xd9,0x1b,0xdb,0xda,0x1a,
0x1e,0xde,0xdf,0x1f,0xdd,0x1d,0x1c,0xdc,0x14,0xd4,
0xd5,0x15,0xd7,0x17,0x16,0xd6,0xd2,0x12,0x13,0xd3,
0x11,0xd1,0xd0,0x10,0xf0,0x30,0x31,0xf1,0x33,0xf3,
0xf2,0x32,0x36,0xf6,0xf7,0x37,0xf5,0x35,0x34,0xf4,
0x3c,0xfc,0xfd,0x3d,0xff,0x3f,0x3e,0xfe,0xfa,0x3a,
0x3b,0xfb,0x39,0xf9,0xf8,0x38,0x28,0xe8,0xe9,0x29,
0xeb,0x2b,0x2a,0xea,0xee,0x2e,0x2f,0xef,0x2d,0xed,
0xec,0x2c,0xe4,0x24,0x25,0xe5,0x27,0xe7,0xe6,0x26,
0x22,0xe2,0xe3,0x23,0xe1,0x21,0x20,0xe0,0xa0,0x60,
0x61,0xa1,0x63,0xa3,0xa2,0x62,0x66,0xa6,0xa7,0x67,
0xa5,0x65,0x64,0xa4,0x6c,0xac,0xad,0x6d,0xaf,0x6f,
0x6e,0xae,0xaa,0x6a,0x6b,0xab,0x69,0xa9,0xa8,0x68,
0x78,0xb8,0xb9,0x79,0xbb,0x7b,0x7a,0xba,0xbe,0x7e,
0x7f,0xbf,0x7d,0xbd,0xbc,0x7c,0xb4,0x74,0x75,0xb5,
0x77,0xb7,0xb6,0x76,0x72,0xb2,0xb3,0x73,0xb1,0x71,
0x70,0xb0,0x50,0x90,0x91,0x51,0x93,0x53,0x52,0x92,
0x96,0x56,0x57,0x97,0x55,0x95,0x94,0x54,0x9c,0x5c,
0x5d,0x9d,0x5f,0x9f,0x9e,0x5e,0x5a,0x9a,0x9b,0x5b,
0x99,0x59,0x58,0x98,0x88,0x48,0x49,0x89,0x4b,0x8b,
0x8a,0x4a,0x4e,0x8e,0x8f,0x4f,0x8d,0x4d,0x4c,0x8c,
0x44,0x84,0x85,0x45,0x87,0x47,0x46,0x86,0x82,0x42,
0x43,0x83,0x41,0x81,0x80,0x40};

6.9 Setting Data Address
6.9.1 Communication address
Address (Hex) Contents Format Codename Access Range & Unit
0032 0020H
0045 002DH
0044 002CH
0043 002BH
0042 002AH
0036 0024H
0035 0023H
0034 0022H
0033 0021H
0047 002FH
0041 0029H
0040 0028H
0039 0027H
0038 0026H
0037 0025H
0046 002EH
0048 0030H
0049 0031H Relay1 status set-up
0050 0032H Relay2 status set-up
0051 0033H
Pulse corresponding value /Low byte
0052 0034H
Pulse corresponding value /High byte
0053 0035H
R/W
0054 0036H
Instantaneous flow time
Instantaneous flow units
Instantaneous flow decimal point
Pulse decimal point Integer
0058 003AH
Relay2 alarm hysteresis value
0057 0039H
0056 0038H
Relay2 alarm value/High byte
0055 0037H
Relay2 alarm value/Low byte
Integrated flow decimal point
0063 003FH
0062 003EH
0061 003DH
0060 003CH
0059 003BH
integrated flow units
0~999 999
0~99
R1V
R2V
R1_P.L
R1_P.H
R2_FW.H
R2_HS
R2T
PDP
DPF
UTF
TMF
DPT
UTT
R/W
R/W
R/W
R/W
R/W
R/W
0~999 999
0~999 9
Integer
Integer
Long
Long
Long
Long
Integer
Integer
Address (Hex) Contents Format Codename Access Range & Unit
0000 0000H T2.A
0013 000DH
0012 000CH
0011 000BH
Resettable integrated flow
0010 000AH
Resettable integrated flow
0004 0004H
Non-resettable integrated flow
0003 0003H
0002 0002H
0001 0001H
0015 000FH
Instantaneous flow/High byte
0009 0009H
0008 0008H
0007 0007H
0006 0006H
0005 0005H
Resettable integrated flow
Resettable integrated flow
0014 000EH
Instantaneous flow Low byte/
0016 0010H
0017 0011H
0018 0012H
0019 0013H
4mA corresponding flow/High byte
0020 0014H
0021 0015H
Correction factor / Low byte
R
R/W
0022 0016H
Correction factor / High byte
0026 001AH
0025 0019H
0024 0018H Density / High byte
0023 0017H Density / Low byte
Average value set-up
0031 001FH
0030 001EH
0029 001DH
0028 001CH
0027 001BH
Low flow cutoff decimal point
Corresponding flow decimal point
0~4(Note 3)
Low flow cutoff
0~9999999999
0~99999
1~999999
0~99
0~9999
T2.B
T2.C
T2.D
T1.A
T1.B
T1.C
T1.D
FW.L
FW.H
04.L
04.H
CF.L
CF.H
DS.L
DS.H
AG
LC
MAK
LCD
R
R
R
R
R
R
R
R
R
R/W
R/W
R/W
R/W
R/W
R/W
R/W
R/W
R/W
0~9999999999
0~9999999999
0~9999999999
0~9999999999
0~9999999999
0~9999999999
0~9999999999
0~999 99
0~999 99
0~99999
1~999999
1~999999
1~999999
0~3(Note 3)
Integer
Integer
Integer
Integer
Long
Long
Long
Long
Long
Long
Long
Long
(Note 1)
(Note 1)
(Note 1)
(Note 1)
(Note 1)
(Note 1)
(Note 1)
(Note 1)
Integer
Integer
Integer
Integer
Integer
R/W
R/W
R/W
R/W
R/W
0~3(Note 3)
0~3Note 3( )
0~6(Note 4)
0~3(Note 5)
0~3(Note 3)
0~6(Note 4)
Relay2 alarm delay
R2_FW.L
R/W
R/W
0~1(Note 6)
0~2(Note 7)
0~999 999
0~999 999
Non-resettable integrated flow
Non-resettable integrated flow
Non-resettable integrated flow
4mA corresponding flow/Low byte
20mA corresponding flow/High byte
R/W 0~99999
20.L
20.H R/W 0~99999
Long
Long
20mA corresponding flow/Low byte
Input square root 0~1(Note 2)
RTR/W
Integer
Ao1 /4mA corresponding Low byte
Ao1 4mA corresponding /High byte
Ao1 20mA corresponding /Low byte
Ao1 20mA corresponding /High byte
Ao2 4mA corresponding /Low byte
Ao2 4mA corresponding /High byte
Ao2 20mA corresponding /Low byte
Ao2 20mA corresponding /High byte
Long
Long
Long
Long
Long
Long
Long
Long
A2.20.H
A2.20.L
A2.40.H
A2.40.L
A1.20.H
A1.20.L
A1.40.H
A1.40.L
R/W
R/W
R/W
R/W
R/W
R/W
R/W
R/W 0~9 99 999
0~999 999
0~999 999
0~999 999
0~999 999
0~999 999
0~999 999
0~999 999
Pulse units R1_P.U
Integer R/W
0~1(Note 8)

6.9.2 Notes
Address (Hex) Contents Format Codename Access Range & Unit
0064 0040H
0068 0044H
0067 0043H
0066 0042H
0065 0041H
0072 0048H
0071 0047H
0070 0046H
0069 0045H
RS-485 parity
RS-485 baud rate
RS-485 address ADD
BRD
PAR
1~199
0~3(Note 9)
0~6(Note 10)
Integer
Integer
Integer R/W
R/W
R/W
Note Description
1Integrated value byte order from high to low is
ntegrated value:
DCBA, if using tens digit calculation,
the i D * 2^48 + C * 2^32 + B * 2^16 + A
5
4
3
2
10
9
8
7
6
Decimal position: 0:No decimals , 1:First decimal place, 2:Second decimal place
3:Third 4:Fourth decimal place decimal places,
Flow units: 0:ml, 1:L, 2:m³, 3:Gallon
4:in³, 5: 、6:7:
8:kg, 9:lb, 10:ton
ml m³, in³
Time units: 0:Sec., 1:Mins, 2:Hours, 3:Days
Relay1 status: 0:Off, 1:Pulse
Relay2 status: 0:Off, 1:Alarm set value, 2:Alarm error
Pulse units: 0:L, 1:m³
RS-485 Baud rate: 0:2400、1:4800、2:9600、3:19200
RS-485 Parity: 0:n.8.1, 1:n.8.2, 2:e.8.1, 3:o.8.1
Input square root: 0:Off, 1:On
Zero resettable integrated energy
Zero resettable integrated flow
RST.E
RST.E
R/W
R/W
Integer
Integer
1: Reset
1: Reset
Table of contents
Other Taik Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

VOLTCRAFT
VOLTCRAFT VC-524 AC/DC operating instructions

Advanced Energy
Advanced Energy MONROE 282H quick start guide

Kingfisher
Kingfisher KI 6700 user manual

sopra
sopra PoolLab 1.0 user manual

Precision Digital Corporation
Precision Digital Corporation ProtEX-MAX PD8-6000 instruction manual

ABB
ABB PFEA 111 user manual

Camlab
Camlab CW6120 instruction manual

Quantum Data
Quantum Data 980 user guide

SPARLING
SPARLING Tigermag EP FM-626 Installation, operation & maintenance manual

Weidmuller
Weidmuller Transclinic 16i+ user guide

NDT Systems
NDT Systems Nova Eclipse TG930 user manual

Chauvin Arnoux
Chauvin Arnoux 1030 user manual