
Operating Instructions-7Al1
5. Switch the oscilloscope Calibrator to 4 V and note
that the upper excursions of the square wave are no longer
visible. If
it
is
desired to display the upper portions of the
square wave, pull out on the oscilloscope BEAM FINDER
knob, causingthe entiredisplay toappear on the CRT.
6.
Depress the OFFSET ON button and adjust the OFF-
SET control
as
necessary to place the top of the square
waves
at
graticule center.
7. Return the BEAM FINDER to normal by depressing
and releasing the button. Note that the upper portion of
the square wave now appears near graticule center. Refine
the OFFSET adjustment (if necessary) to align the tops of
the square waves with the graticuleverticalcenter.
8. Calculate the OFFSET voltage being applied.
It
is
equal to theOFFSET decimal value (indicatedby the OFF-
SET knob) multiplied by the OFFSET RANGE (indicated
by the VOLTS/DIV switch). Inthis case, thecalculation
is
-.2
X
20 V
=
-4 V. (-4 volts is needed to return
a
+4 V
signal to graticulecenter.)
9.
Using a DC voltmeter and the Cable Assembly, check
the voltage available
at
the OFFSET OUT jack.
It
is
equal
to the value indicated by the OFFSET knob, and can never
exceed approximately
*I
V. Notethat only 0.2 V
is
present
there, even though the offsetvoltage beingapplied to offset
the incomingsignal
is
equivalentto
4
V.
10. Turn the OFFSET button OFF and disconnect the
coaxial cable from the oscilloscope and the STORED
PROBE INPUTAdapter.
Probe Operation
1. Pull out on the 7Al
I
Locking Latch and remove the
7Al
I
from the oscilloscope plug-inunit compartment.
2. Lay the 7A11 on
its
left side and remove the Probe as
explained in the beginning of this section. Replace the
7A11 in the oscilloscope plug-in compartment. Allow
approximately 1 minute to elapse before continuing with
the procedure.
3. Depress the IDENTIFY button near the rear of the
Probe body and note that the trace moves approximately
114 division just as
it
did when the front panel IDENTIFY
button was depressed. If the oscilloscope
is
equipped with
readout circuitry, also notice that the word "IDENTIFY"
appears inthe 7A11 readoutarea of the CRT.
4. Switch the oscilloscope Calibrator to 0.4 V. Now
touch the Probe tip to the Calibrator output jack center
terminal and note that approximately 4 divisionsof square
wave appears on the upper half of the CRT.
5.
Disconnect the Probe from the jack. Attach the AC
Coupler accessory to the Probe tip. Connect the Coupler tip
to the Calibrator jack and note that the 4 division square-
wave display now straddles the vertical center of the grati-
cule, since the capacitor in the AC Coupler blocks the DC
component.
6.
It
is
suggested that the Probe's other accessories be
experimented with at this time for familiarizationpurposes.
Note that when the Retractable Hook Tip
is
installed on
either the Probe or the AC Coupler tip, the hook can be
exposed by pulling back on the retracting tab. Also note
that either the 3-inch or 12-inch screw-in grounding lead
can be used with the AC Coupler tip. An alligator clip
should be attached to the opposite end of the grounding
leads for attaching to convenientgrounded surfaces.
7. Remove the 7Al
I
from the plug-in compartment.
Depress the tab on the side of the STORED PROBE INPUT
Adapter and remove the Adapter through the front of the
7A11. Retract the cable into the 7Al
I
and wind
it
on the
cable spool as necessary to draw the Probe through the
7A11 front panel. Snap the STORED PROBE INPUT
Adapter intoplace and insert the Probe intothe back of the
Adapter. The 7A11
is
again ready for use in the stored
probemode.
Unscaled Deflection Factors
On occasion,
it
may be convenient to establish vertical
deflection factors other than those which are provided by
specific positions of the VOLTSIDIV switch. Forexample,
assume that 0.8 voltsldivisiondeflectionfactor
is
desired. If
a
4-volt square wave calibrator signal
is
appliedto
a
vertical
unit having 0.8 volts deflection factor, the displayed signal
amplitude
is
equal to the applied signal (4volts) divided by
the deflection factor, (0.8 volts) which
is
equal to 5 divi-
sions of display. With this in mind, set the VOLTSIDIV to
the next lower deflection factor (in this case,
.5
volts) and
apply a 4-volt signal from the oscilloscopeCalibratorto the
7Al
I
Probe input. Then depress and release the VARI-
ABLE knob and adjust
it
as necessary to obtain
5
divisions
of deflection. A 0.8 volt/division deflectionfactor now
is
in
effect.
It
should be noted that the same ratio of actual deflec-
tion factor to VOLTSIDIV switch value exists in all other
switch positions. Thus if the VARIABLE control
is
in the
position established in the preceding paragraph and the
VOLTSIDIV switch is placed in the 2 VOLTSIDIV posi-
tion, a 3.2 voltsldivisiondeflection factor will be in effect
(0.8
is
to 0.5 as 3.2
is
to 2).