
Maintenance-7
A
15A/7
A15AN
C m
E C E C E
B B B
Plastic Case
Transistors
o
S G
Dual-FET
a
~\-
common
anode/
\
cathode
clip
off
Dual-Diode
Fig. 4-1. Electrode configuration
for
semiconductors
in
the
7A
15AI
7A15AN.
Troubleshooting Equipment
The following
equipment
is
useful for troubleshooting
the
7A15A/7A15AN.
1. Semiconductor
Tester-Some
means
of
testing
the
transistors, diodes, and FET's used
in
this instrument
is
helpful. A transistor-curve tracer such as
the
Tektronix
Type
576
will give
the
most
complete information.
2.
DC
Voltmeter
and
Ohmmeter-A.
voltmeter
is
required for checking voltages within the !Circuits, and an
ohmmeter
for checking resistors and diodes.
3.
Test Oscilloscope-A.
test
oscilloscop1e
is
required
to
view waveforms
at
different points
in
the
circuit. A
Tektronix
7000-series Oscilloscope equipped with a
readout
system, 7D13 Digital Multimeter unit, 7B-series Time-Base
unit, and a 7A-series Amplifier
unit
with a
lOX
probe will
meet
the
needs
of
both
items 2 and
3.
4.
Plug-in
Extender-A
fixture
that
permits operation of
the
unit
outside of
the
plug-in
compartment
for
better
accessibility during troubleshooting. Order
Tektronix
part
number 067-0589-00.
4-2
Troubleshooting Procedure
This troubleshooting procedure
is
arranged
in
an order
which checks
the
simple
trouble
possibilities before
proceeding with extensive troubleshooting.
1. Check Control Settings. An incorrect setting
of
the
7A
15A/7
A15AN controls can ind icate a trouble
that
does
not
exist.
If
there
is
any question
about
the
correct
func-
tion
or
operation of a
control
or front-panel
connector,
see
the
Operating Instructions section.
2. Check Associated Equipment. Before proceeding
with troubleshooting
of
the
7A15A/7A15AN, check
that
the
equipment
used with this
instrument
is
operating
correctly.
If
possible,
substitute
an amplifier
unit
known
to
be operating correctly
into
the
indicator
unit
and see
if
the
problem persists_ Check
that
the
input
signals are properly
connected
and
that
the
interconnecting cables are
not
defective.
3.
Visual Check. Visually check
the
portion
of
the
instrument
in
which
the
trouble
is
suspected. Many troubles
can be located by visual indications, such
as
unsoldered
connections, broken wires, damaged circuit boards,
damaged
components,
etc.
4.
Check
Instrument
Performance. Check
the
calibra-
tion
of
the
unit
or
the
affected circuit, by performing
Performance Check
of
Section 5.
The
apparent
trouble may
only be a result
of
mis-adjustment, and may be corrected
by calibration. Complete calibration instructions are given
in
Part
II
of
Section 5.
5.
Check Voltages.
Often
the
defective
component
or
stage can be located by checking for
the
correct
voltage
in
the
circuit. Typical voltages are given on the diagrams; how-
ever, these are
not
absolute and may vary slightly between
instruments.
To
obtain operating conditions similar
to
those used
to
take
these readings, see
the
instructions
in
the
Diagrams section.
6.
Check
Individual Components.
The
following
methods are provided for checking
the
individual com-
ponents
in
the
7A
15A/7
A15AN.
Components
which are
soldered
in
place are best checked by disconnecting one end
to
isolate
the
measurement
from
the
effects
of
surrounding
circuitry.
NOTE
To locate
intermittent
or
temperature sensitive com-
ponents mounted
on
the attenuator board, Quik
Freeze (Miller Stephenson, MS-240, Tektronix Part
Number 006-0173-01)
is
recommended.
Dry
ice
or
dichlorodi-fluorremethane (Freon 12,
Dupont
or
Can-
a-Gas) may also be
used.
Other types
of
circuit
coolant
may
damage
the polyphenylene oxide boards.
®
I
.'
I
I
I
I
I
I
I
I
I
I
I
I
I
I
.,
I