
maximum operaling time and useful battery life is provided
when the batteries are charged and discharged at about
+25'C.
During normal usage or storage, the individual battery
cells in this instrument attain slightly different charge
characteristics. To provide the best overall operation and
-..\maximum operating life, the charge on the battery cells
'should be equali.,ed periodically. This can be done without
damage to the battery cells by charging the batteries at the
full,charge rate tot 24 hours. This should be done after
ever occurs first.
Operating I nstructions-Type 321A
o, boric acid in water. A 2% solution of boric acid can be
obtained by dissolving 1 1/4 teaspoons o{ boric acid powder
in one cup of water. After the residue has been cleaned
from the batteries and the battery companment, dry the
wetted area thoroughly with a soft cloth.
ll the total potential across the battery terminals (see
Fig- 2-38) does not reach 13-14 volts even after the recom-
mendd charge time. one or more of the battery cells is
probably de{ective. To locate the defective cell, first be sure
the batteries have been charged for the recommended
charge period. Then, while operating the instrument on the
internal batteries, measure the potential across each individ-
ual cell. Use a meter which has a long, thin probe tip to
reach the contacts between the batteries. lf an individual
battery cell measures below about 1.15 volts, it is probably
defective; above this voltage, the battery cell is probably
operating correctly. lf the total potential across the battery
terminals is 13-14 volts after the recommended charge
period, but the LOW BATTERIES lighl comes on too soon
when operating on internal batteries, the charge retaining
capacity of one or more of the battery cells has probably
decreased below rhe acceptable limits. (NOTE: Batteries
may not be reaching full charge due to a defective charger
circuit.) This can be checked in a manner similar to that
described above. First be sure that the batteries have been
charged for the recommended charge period. Then, operate
the instrument on the internal batleries until the LOW
BATTEBIES light comes on. Now measure the potential
across each cell. lf an individual cell measures below about
1.15 volts, it is probably defective; above this voltage, it is
probably acceptable. lf the voltage of all the cells is above
1.2 volts and the charger circuit is operating correctly. the
LOW BATTERIES circuit may be at fault.
When replacing battery cells in this instrument, charge
the entire set of batteries at the recommended charge rate
and charge period before operating the instrument from the
internal batteries. This equalizes the charqe on all the bat-
tery cells and protects the new cells from polarity reversal
damage. The batteries should be replaced only with cells
which have the same charge rate and charge capacity as
those in the instrument (see Battery Charging for more
information).
Battery Storage. The battery cells used in this instru,
ment may be stored in a charged or partially charged condi,
tion. For best shelf-life when storing individual cells, fully
recharge the battery cells at three to six month intervals.
Although replacement cells obtained from Tektronix, lnc.
are fully charged before shipment, recharge the complete
set of batteries before operating the instrument.
Charge retention characteristics of nickel-cadmium cells
vary with the storage temperature. They may be stored at
any temperature between -4O' C and +60'C without
damage, either in the instrument or as individual cells. How
ever, the self-discharge rate increases with ambient tempera-
ture and humidity. For example, cells stored at +20'C will
lose about 50% of their stored charge in thre€ monlhs but
Nickel-Cadmium Battery I nformation
General, The nickel-cadmium battery cells available
from Tektronix, lnc. for use with this instrument have been
selected as a result of extensive evaluation. Each battery
cell has received an ampere hour test, has met or exceeded
the minimum ampere-hour storage requirement and has
been rigidly inspected. The battery cells used in this instru,
ment should provide a useful operating life extending over
several hundred charge-discharge cycles.
Precautions To extend the useful operating life of the
nickel-cadmium battery cells used in this instrument, ob-
serve the following precaulions.
l. Do not exceed the recommended charge rate and
/.,\ period (see Battery ChargingJ.
2. Be sure the battery charger is operated correctly as
described previously in this section.
3. Be sure to observe the cell polarity when inserting the
batteries (see Fig. 2-38).
4. Excessive discharge of the batteries after the LOW
BATTEBIES light comes on may cause one or more of the
cells to reverse polarity. Although the cells are protected
against immediale damage, repeated polarity reversal will
shonen the useful life of the batteries.
E2-5
5. Observe the temperature limits given in this manual
for battery charging, operation and storage-
Maintsnance, When the battery cells are overcharged or
when discharged to the point of polarity reversal, gas is
formed within the nickel cadmium cells. This gas produces
pressure within the cells. The nickel cadmium cells supplied
by Tektronix, lnc. (Tektronix Part No. '146-0010,00)
incorporate a vent so this internal pressure does not damage
^ the battery. However, as the internal pressure is relieved, a
\small amount of the electrolyte may be expelled with the
gas. Although the cell will probably not be damaged, this
loss oI electrolyte may result in shorter overall battery li{e.
The batteries and battery compartment should be inspected
occasionally for any electrolyte residue on the batteries
themselves or in the battery compartment. Any residue
which is found should be cleaned away with a 2% solution
Scan by vintageTEK - Your donations help support the museum - vintagetek.org