
10
k.Q.
The signal
is
not
affected by the Oscilloscope Verti-
cal
Mode switch. The signal
is
changed
by
the Units/Div
switch, the DC OFFSET controls,
and
the
NORMAL/
SMOOTH switches. The
INVERT
switch
does
not
invert
the signal.
Output
voltage swing
is
limited
to
2.5 volts peak
or about 6 divisions above
and
below the graticule center
depending on the actual position. The Units/Div switch
is
active
but
the
output
voltage
is
always proportional
to
the
CRT display (0.2
V/Div)
up
to
12 divisions maximum peak
to
peak. Since the signal at the Vertical
Output
jack
is
an
equivalent time signal,
it
is
useful
for
pen
recorder applica-
tions when the sampling
sweep
unit
external
input
is
driven
by the recorder.
The OFFSET
OUT
monitor
jack permits accurate
measurement
of
the DC voltage set
by
the DC OFFSET
controls. The
output
voltage
has
a
range
of
+10
to
-10
volts. Unless otherwise specified on the sampling
head
front
panel, the
monitor
jack
output
voltage
is
ten times the
internal DC offset voltage. The
output
resistance
is
10
k.Q.
BASIC
APPLICATIONS
Vertical Deflection Measurements
Vertical displacement
of
the trace on the CRT
is
directly
proportional
to
the signal at the sampling
head
input
con-
nector installed in the 7S11. The amount
of
displacement
for
a given signal
can
be
selected
with
the Units/Div switch.
To
provide sufficient deflection
for
best resolution, set the
Units/Div switch
so
the display
spans
a large
portion
of
the
graticule. Also, when measuring between points on the dis-
play,
be
sure
to
measure consistently from either the
bottom,
middle,
or
top
of
the trace. This prevents the
width
of
the trace from affecting the measurements.
To
make a vertical deflection difference measurement
between
two
points on the display, proceed
as
follows:
1.
Note the vertical deflection, in graticule divisions,
between the
two
points on the display. Make
sure
the
VARIABLE
control
is
in the
CAL
position.
NOTE
The
vertical deflection factor is determined
by
the
Units/Div switch and the value
of
the units
of
measure
as
stated on the sampling head
front
panel.
The
ratio
of
the
input
signal
to
the resultant deflec-
tion
is called the deflection factor.
For
example,
with
the sampling head
unit
values
stated
as
mV/Div, and
the Units/Div switch set
to
20, the vertical deflection
factor
will
be
20
mV/division.
2.
Multiply
the divisions
of
vertical deflection
by
the
deflection factor, and the external attenuator
or
probe
®
Operating I
nstructions-7S
11
attenuation factor
(if
any). The product
is
the voltage
difference between the
two
points measured.
For
example, suppose you measure 4.4 divisions
of
de-
flection between
two
points on the display
and
the
Units/
Div switch
is
set
for
20
mV
/Div.
Multiply
20
millivolts/
division
by
4.4 divisions, the
product
is
88 millivolts. This
is
the voltage difference at the
input
connector between the
two
points on the display.
Now
assume
there
is
a 10X
external attenuator (probe) between the
input
connector
and the signal source.
To
determine the actual signal voltage
at the source,
multiply
10 (the attenuation factor
of
the
probe)
by
88 millivolts; this
product
(880
mi"ivolts
or 0.88
volts)
is
the actual voltage
at
the signal source.
It
is
also possible
to
measure
an
instantaneous (or DC)
voltage
to
ground
from
the display. This measurement
is
accomplished in the
same
manner, except
that
with
no
sig-
nal applied, you must
first
establish a ground-reference
point
on the CRT.
NOTE
To
establish the ground reference
point
be sure the
input
connector is either terminated
by
a
50.Q
resis-
tor
or
coaxialcable connected to the input.
To
do
this, trigger the sampling
sweep
unit
with
the
signal source
and
display a trace. Then, position the trace
so
it
is
exactly aligned
with
one
of
the graticule lines. The
actual graticule line you select
will
be
largely determined
by
the polarity and amplitude
of
the applied signal.
After
establishing the ground reference, make no
further
adjust-
ments
with
the DC OFFSET control.
Apply
the signal and measure the voltage in the manner
previously described. Make all measurements 'from the
established ground reference
point.
Accuracy
of
this
meas-
urement
is
within
±3%.
If
the applied signal
has
a relatively high DC level, the
ground-reference
point
and the actual signal may
be
so
far
apart
that
neither
will
appear on the CRT. In this
case,
refer
to
the following discussion on "Voltage Measurements
Using the DC Offset
Control."
Voltage Measurements Using the DC OFFSET
Control. Unless otherwise stated on the sampling
head
front
panel, the DC offset voltage cancels the effect
of
an
applied DC voltage
of
up
to
±1
volt
at
the sampling
head
input. Also, accurate slideback amplitude measurements
of
the applied signal
can
be
obtained
by
positioning the dis-
play at various points and measuring the amount
of
voltage
change at the OFFSET
OUT
monitor
jack.
2-9
Scans by ArtekMedia => 2010